(136 days)
The NuVasive Expandable Lumbar Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The system is designed for use with autogenous bone graft to facilitate fusion.
The Expandable Lumbar Interbody System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the Expandable Lumbar Interbody System. The Expandable Lumbar Interbody System is intended to be used with supplemental internal spinal fixation systems (e.g. pedicle screw/rod systems) that are cleared by the FDA for use in the lumbar spine.
The NuVasive Expandable Lumbar Interbody System is manufactured from Ti-6Al-4V ELI conforming to ASTM F136 and ISO 5832-3, Ti-6A1-4V conforming to ASTM 1472, and Nitinol SE508 conforming to ASTM F2063. The implants are available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient.
The provided document describes the NuVasive Expandable Lumbar Interbody System, a Class II medical device intended for intervertebral body fusion of the spine. The evaluation criteria for this device are based on demonstrating substantial equivalence to predicate devices rather than meeting specific performance metrics for an AI system. Therefore, many of the requested categories related to AI model evaluation are not applicable.
Here's an analysis based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Demonstrated through testing) | Reported Device Performance (Summary of Test Results) |
|---|---|
| Static and Dynamic Axial Compression per ASTM F2077 | No new worst-case for performance testing. |
| Compression Shear per ASTM F2077 | No new worst-case for performance testing. |
| Torsion Testing per ASTM F2077 | No new worst-case for performance testing. |
| Range of Motion Testing | No new worst-case for performance testing. |
| Lateral Collapse Testing | No new worst-case for performance testing. |
| Cadaver Testing | No new worst-case for performance testing. |
| Corrosion Testing per ASTM F2129 | No new worst-case for performance testing. |
| Overall Goal: Substantial Equivalence to Predicate Devices (NuVasive CoRoent® System (K071795) and NLT Spine Ltd. Prow Fusion (K112359)) | The subject device was found to be substantially equivalent to the predicate devices in terms of safety and effectiveness, having similar indications for use, technological characteristics, and principles of operation. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not specified for individual tests. The document refers to various "testing" and "cadaver testing" without detailing the number of units or cadavers used.
- Data Provenance: Not explicitly stated. Nonclinical testing was performed, but the location or specific origin of materials/samples is not mentioned. This is a medical device, so the tests would be performed on engineered samples or cadaveric specimens, not retrospective or prospective patient data in the sense of an AI study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Not Applicable. This is a physical medical device clearance, not an AI or diagnostic tool where expert ground truth is established for a test set. The "ground truth" here is compliance with established ASTM standards and comparable performance to predicate devices.
4. Adjudication Method for the Test Set
- Not Applicable. See point 3. Testing involves objective measurements against established technical standards, not subjective expert adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance
- Not Applicable. This is not an AI-assisted diagnostic or imaging device. Therefore, no MRMC study or AI performance evaluation was conducted.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Not Applicable. This device does not involve an algorithm or AI.
7. The Type of Ground Truth Used
- Technical Standards and Predicate Device Performance: The "ground truth" for this device's performance evaluation is established by:
- ASTM International Standards: F2077 (for static and dynamic mechanical testing), F2129 (for corrosion testing).
- Performance of Legally Marketed Predicate Devices: The "substantial equivalence" claim is based on demonstrating that the new device performs comparably to the NuVasive CoRoent® System (K071795) and NLT Spine Ltd. Prow Fusion (K112359) without introducing new worst-case scenarios.
8. The Sample Size for the Training Set
- Not Applicable. This device does not involve an AI model requiring a training set.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. This device does not involve an AI model requiring ground truth for a training set.
Summary of the Study:
The study described is a nonclinical performance testing regimen designed to demonstrate the mechanical and material characteristics of the NuVasive Expandable Lumbar Interbody System. The methodologies employed are based on established ASTM standards for intervertebral body fusion devices. The primary objective was to show that the subject device performs similarly to or better than two specific predicate devices already on the market (NuVasive CoRoent® System (K071795) and NLT Spine Ltd. Prow Fusion (K112359)) across various mechanical and environmental tests. The conclusion states that the device presents "no new worst-case for performance testing," thereby establishing substantial equivalence to the predicate devices in terms of safety and effectiveness. This type of clearance (510(k)) relies on demonstrating similarity to existing legally marketed devices, rather than establishing de novo safety and effectiveness through extensive clinical trials for software performance.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a stylized graphic to the left of the word "NUVASIVE" in a sans-serif font. Below the company name is the tagline ":Speed of Innovation" in a smaller font. The logo is simple and modern, and the tagline suggests that the company is focused on innovation and speed.
510(k) Summary, K130820
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
A. Submitted by:
AUG 8 2013
Jeremy Markovich Senior Specialist, Regulatory Affairs NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 309-3783
Date Prepared: July 15, 2013
B. Device Name
| Trade or Proprietary Name: | Expandable Lumbar Interbody System |
|---|---|
| Common or Usual Name: | Intervertebral Body Fusion Device |
| Classification Name: | Spinal Intervertebral Body Fixation orthosis |
| Device Class: | Class II |
|---|---|
| Classification: | 21 CFR § 888.3080 |
| Product Code: | MAX |
C. Predicate Devices
The subject Expandable Lumbar Interbody System is substantially equivalent to the following predicate devices: NuVasive CoRoent® System (K071795) and NLT Spine Ltd. Prow Fusion (K112359).
D. Device Description
The NuVasive Expandable Lumbar Interbody System is manufactured from Ti-6Al-4V ELI conforming to ASTM F136 and ISO 5832-3, Ti-6A1-4V conforming to ASTM 1472, and Nitinol SE508 conforming to ASTM F2063. The implants are available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a stylized graphic to the left of the company name. Below the company name is the tagline "Speed of Innovation".
E. Intended Use
The NuVasive Expandable Lumbar Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The system is designed for use with autogenous bone graft to facilitate fusion.
The Expandable Lumbar Interbody System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the Expandable Lumbar Interbody System. The Expandable Lumbar Interbody System is intended to be used with supplemental internal spinal fixation systems (e.g. pedicle screw/rod systems) that are cleared by the FDA for use in the lumbar spine.
F. Technological Characteristics
As was established in this submission, the subject Expandable Lumbar Interbody System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, and function.
G. Performance Data
Nonclinical testing was performed to demonstrate that the subject Expandable Lumbar Interbody System is substantially equivalent to other predicate devices. The following testing was performed:
- . Static and dynamic axial compression, compression shear, and torsion testing per ASTM F2077
- Range of Motion Testing .
- Lateral Collapse testing
- . Cadaver testing
- . Corrosion testing per ASTM F2129
The results demonstrate that the subject Expandable Lumbar Interbody System presents no new worst-case for performance testing, and the subject device was therefore found to be substantially equivalent to the predicate.
H. Conclusions
The subject Expandable Lumbar Interbody System has been shown to be substantially equivalent to legally marketed predicate devices in terms of safety and effectiveness, having similar indications for use, technological characteristics, and principles of operation as the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure embracing or supporting a globe-like shape.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 8, 2013
NuVasive, Incorporated Mr. Jeremy Markovich Senior Specialist. Regulatory Affairs 7475 Lusk Boulevard San Diego, California 92121
Re: K130820
Trade/Device Name: NuVasive® Expandable Lumbar Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: July 15, 2013 Received: July 16, 2013
Dear Mr. Markovich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Jeremy Markovich
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erin Keith
For
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name: NuVasive® Expandable Lumbar Interbody System
Indications For Use:
The NuVasive Expandable Lumbar Interbody System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The system is designed for use with autogenous bone graft to facilitate fusion.
The Expandable Lumbar Interbody System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the Expandable Lumbar Interbody System. The Expandable Lumbar Interbody System is intended to be used with supplemental internal spinal fixation systems (e.g. pedicle screw/rod systems) that are cleared by the FDA for use in the lumbar spine.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anton E. Dmitriev, PhD Division of Orthopedic Devices
N/A