K Number
K130809

Validate with FDA (Live)

Date Cleared
2014-02-18

(330 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder Free Viny! Patient Examination Gloved) is a non-sterile disposable device intended for medical purposs that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) that meets all of the requirements of ASTM standard D 5250-06(Reaffirmation 2011).

-- How the device functions:

PVC films form a barrier to body fluids and bloodborne Pathogens

-- Scientific concepts that form the basis for the device

The PVC rubber is water tight under normal conditions of use. Its tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure.

-- Physical and performance characteristics such as design, materials and physical properties: PVC gloves are known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151 requirements.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Feature & DescriptionAcceptance Criteria (from ASTM/FDA Guidance)Reported Device Performance (Subject Device)Result of Comparison
Dimensions - LengthASTM D5250: 230mm min for all sizes230mm min for all sizesSubstantially equivalent
Dimensions - WidthASTM D5250: Small 80-90 mm, Medium 90-100mm, Large 100-110mm, X large 110-120 mmSmall 80-85 mm, Medium 95-97 mm, Large 102-108mm, X large 114-118 mmSubstantially equivalent
Dimensions - ThicknessASTM D5250: Finger 0.05mm min, Palm 0.08mm minFinger 0.05mm min, Palm 0.08mm minSubstantially equivalent
Physical Properties (Aging)ASTM D5250: Before aging/after aging: Elongation ≥300%, Tensile Strength >14MPaBefore aging/after aging: Elongation ≥300%, Tensile Strength ≥ 14MPaSubstantially equivalent
Freedom from Pinholes21 CFR 800.20, ASTM D5250, ASTM D 5151; AQL 2.5 (Inspection Level I)Meets ASTM D5151-06 (Reapproved 2011), Holes Inspection Level I, AQL 2.5Substantially equivalent
Residual PowderASTM D 6124 (values below 2mg of residual powder)Meets ASTM D 6124-06 (Reapproved 2011), Results generated values below 2mg of residual powderSubstantially equivalent
MaterialsPolymer/other type of material identification (If vinyl: PVC)PVCSubstantially equivalent
Donning Powder/LubricantComposition and biocompatibility data if lubricant usedPU (Polyurethane) / Surface Coating AgentSubstantially equivalent
Sterility statusNon sterileNon sterileSubstantially equivalent
Single Patient UseSingle Patient UseSingle Patient UseSubstantially equivalent
BiocompatibilitySKIN IRRITATION DERMAL and SENSITIZATION STUDIES (ISO 10993-10)SKIN IRRITATION DERMAL and SENSITIZATION STUDIES Meets ISO 10993-10:2002/Amd.1:2006Substantially equivalent
Labeling-Powder Free, -Patient Examination Glove, -Single Use Only, - Manufactured For:, -Lot-Powder Free, -Patient Examination Glove, -Single Use Only, - Manufactured For:, -LotSubstantially equivalent

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample sizes used for testing each of the acceptance criteria (e.g., number of gloves tested for dimensions, pinholes, tensile strength). Instead, it refers to compliance with ASTM standards (D5250, D5151, D6124) which inherently define the sampling plans and test procedures for these performance characteristics. The provenance of the data is implied to be from the manufacturer (Jewel Plastic Co., Ltd.) as they are reporting that their device "Meets" these standards. No specific country of origin for the test data is mentioned, nor is it specified whether the studies were retrospective or prospective, though testing against an ASTM standard is typically a prospective and controlled laboratory evaluation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not applicable as the acceptance criteria are based on objective, quantifiable physical and performance characteristics defined by recognized consensus standards (ASTM D5250, D5151, D6124) and regulatory requirements (21 CFR 800.20, ISO 10993-10). Expert clinical or image-based "ground truth" is not established for these types of device characteristics. The "ground truth" is the standard itself.

4. Adjudication Method for the Test Set

This information is not applicable for the same reasons as point 3. The testing involves standardized measurements and procedures, not subjective interpretations requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the Effect Size of how much Human Readers improve with AI vs without AI assistance

This information is not applicable. The device is a patient examination glove, not an AI-assisted diagnostic tool or an imaging device. Therefore, MRMC studies and AI assistance are irrelevant to its evaluation.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a physical product (gloves), not a software algorithm.

7. The Type of Ground Truth Used

The ground truth used for evaluating the device's performance against its acceptance criteria is consensus standards and regulatory requirements. Specifically, this includes:

  • ASTM D5250-06 (Reaffirmation 2011): Standard Specification for Poly(vinyl chloride) Gloves for Medical Application
  • ASTM D5151-06 (Reapproved 2011): Standard Test Method for Detection of Holes in Medical Gloves
  • ASTM D6124-06 (Reapproved 2011): Standard Test Method for Residual Powder on Medical Gloves
  • 21 CFR 800.20: Subpart B - Performance Standards, specific to medical devices (likely related to leakage testing).
  • ISO 10993-10: 2002/Amd. 1:2006(E): Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization (for biocompatibility).

These standards define the measurable properties and the acceptable limits for those properties.

8. The Sample Size for the Training Set

This information is not applicable. The device is a physical product, not a machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

Section C

FEB 1 8 2014

510(k) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: K130809 - " (applicant leave blank)

Premarket Notification [510(k)] Summary

[(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :

Submitter's name :Jewel Plastic Co., Ltd.
Submitter's address :No. 11 Jian She St., Shijiazhuang, 050000, China
Phone number :(86) 311-67699858
Fax number :(86) 311-67699906
Name of contact person:Mr. WANG Hai
Date the summary was prepared:2014-02-07

[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

Device Name:Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored)
Proprietary/Trade name:Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored)
Common Name:Patient examination glove
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LYZ

[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .

Class I* Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) that meets all of the requirements of ASTM standard D 5250-06 (Reaffirmation 2011).

Predicate device : Powder free Vinyl Patient Examination Gloves, Clear(Non-colored), Shijiazhuang Fuguan Plastic Products Co., Ltd. K032908 .

{1}------------------------------------------------

[(a)(4)] A description of the device

Device Description : Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) that meets all of the requirements of ASTM standard D 5250-06(Reaffirmation 2011).

-- How the device functions:

PVC films form a barrier to body fluids and bloodborne Pathogens

-- Scientific concepts that form the basis for the device

The PVC rubber is water tight under normal conditions of use. Its tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure.

-- Physical and performance characteristics such as design, materials and physical properties: PVC gloves are known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151 requirements.

{{a)(5)| The summary describes the intended use of the device

Device Intended Use: Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.

The Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

Features &DescriptionPredicate DeviceMedicalGloveGuidanceManual(1661)Subject DeviceResult ofComparison
CompanyShijiazhuang FuguanPlastic Products Co., Ltd.Jewel Plastic Co., Ltd.-
510(K) NumberK032908-
Product namePowder Free Vinyl PatientExamination Gloves,Clear(Non-colored)Powder Free VinylPatient ExaminationGloves, Clear(Non-colored)Same
Product CodeLYZLYZLYZSame
SizeSmall/Medium/Large/X largeSmall/Medium/Large/X largeSubstantiallyequivalent
Intend for usePowder free Vinyl PatientExamination Gloves,Clear(Non-colored)is adisposable deviceintended for medicalpurposes that is worn onthe examiner's hand orfinger to preventcontamination betweenpatient and examiner.Powder FreeExaminationGloves:A powder-freepatient examinationglove is adisposable deviceintended formedical purposesthat is worn on theexaminer's hand orfinger to preventcontaminationbetween patient andexaminer.Powder free VinylPatient ExaminationGloves, Clear(Non-colored) is adisposable deviceintended for medicalpurposes that is worn onthe examiner's hand orfinger to preventcontamination betweenpatient and examiner.Substantiallyequivalent
DeviceDescription andSpecificationsMeets ASTM D5250-06 (Reapproved 2011)If vinyl:Do the vinylexamination glovesmeet all the currentspecifications listedunder ASTMSpecification D5250Meets ASTM D5250-06 (Reapproved 2011)Substantiallyequivalent

{2}------------------------------------------------

or an equivalentconsensus standard?
Dimensions- LengthMeets ASTM D5250-06 (Reapproved 2011)ASTM D5250230mm min for all sizesSubstantiallyequivalent
Dimensions-- Width≥230mm min.Meets ASTM D5250-06 (Reapproved 2011)Substantiallyequivalent
Small 80-90 mmMedium 90-100mmLarge 100-110mmX large 110-120 mmSmall 80-85 mmMedium 95-97 mmLarge 102-108mmX large 114-118 mm
Dimensions-- ThicknessMeets ASTM D5250-06 (Reapproved 2011)
Finger 0.05mm min.Palm 0.08mm min.Finger 0.05mm min.Palm 0.08mm min.
PhysicalPropertiesMeets ASTM D5250-06 (Reapproved 2011)ASTM D5250Before aging/after agingSubstantiallyequivalent
Before aging/after agingElongation ≥300%Tensile Strength>14MPaElongation ≥300%Tensile Strength≥ 14MPa
Freedom fromPinholesMeets• 21 CFR 800.20• ASTM D5250-06(Reapproved 2011)• ASTM D 5151-06(Reapproved 2011)21 CFR 800.20ASTM D5250ASTM D 5151Meets ASTM D5151-06(Reapproved 2011)HolesInspection Level IAQL2.5Substantiallyequivalent
ResidualPowderMeets ASTM D 6124-06(Reapproved 2011)ASTM D 6124Meets ASTM D 6124-06(Reapproved 2011)Results generatedvalues below 2mg ofresidual powderSubstantiallyequivalent
Compare allmaterials usedto fabricate thedevicesPVCIf the glove is madeof a polymer orother type ofmaterial. identifythe material.PVCSubstantiallyequivalent
Dusting orDonningPowder:PUIf a donninglubricant is used,state thecompositionand includebiocompatibilitydata for thelubricant in anidentifiedattachment; alsostate the name,manufacturer, andaddress belowPUSubstantiallyequivalent
Dusting orDonningPowder: namePULubricantGeneric Name/LubricantBrand NameSurface Coating AgentSubstantiallyequivalent
Compareperformancedata supportingsubstantialsubstantialequivalenceMeetsASTM D5151-06(Reapproved 2011)ASTM D5250-06ASTM D5250-06(Reapproved 2011)ASTM D6124-06At this time FDArecognizes thefollowing standards:Patient ExaminationPatient ExaminationGloves(PVC)ASTMD5151(Detection ofHoles in MedicalGloves)ASTMMeetsASTM D5151-06(Reapproved 2011)ASTM D5250-06ASTM D5250-06(Reapproved 2011)ASTM D6124-06Substantiallyequivalent

{3}------------------------------------------------

D6124(ResidualPowder on MedicalGloves)ASTMD5250(Poly(vinylchloride) Gloves)
Sterility statusNon sterile-Non sterileSubstantiallyequivalent
SinglePatientUseSingle Patient UseSingle Patient UseSingle Patient UseSubstantiallyequivalent
BiocompatibilitySKIN IRRITATIONDERMAL andSENSITIZATIONSTUDIES Meets ISO10993-10 :2002/Amd.1:2006SKIN IRRITATIONDERMAL andSENSITIZATIONSTUDIESISO 10993-10SKIN IRRITATIONDERMAL andSENSITIZATIONSTUDIES Meets ISO10993-10 :2002/Amd.1:2006Substantiallyequivalent
Labeling for thelegallymarketeddevice to whichsubstantialequivalence isclaimed.-Powder Free-Patient ExaminationGlove-Single Use Only- Manufactured For:- LotChapter 4-Powder Free-Patient ExaminationGlove-Single Use Only- Manufactured For:- LotSubstantiallyequivalent

The manufacturing process has been described on the appendix 3.0.

[(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored)meet requirements per ASTM D5250-06(Reaffirmation 2011), per ASTM D6124-06(Reaffirmation 2011), per 21 CFR 800.20 and ISO 10993-10: 2002/Amd. 1:2006(E).

[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

[(b)(3)] The conclusions drawn from the nonclinical tests that demonstrate that the device is as safe,as effective, and performed as well or better than the legally marketed device identified in (a)(3).

It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims and the Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) is as safe, as effective, and performs as well as the predicate device, Powder free Vinyl Patient Examination Gloves, Clear(Non-colored), Shijiazhuang Fuguan Plastic Products Co., Ltd. K032908 .

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in bold, uppercase letters. The text is black and appears to be centered. The font is sans-serif and the letters are closely spaced.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 18, 2014

Jewel Plastic Co. Ltd. C/O Mr. Chu Xiaoan Official Correspondent Rm. 1606 Bldg.1 Jianxiang Yuan No.209 Bei Si Huan Zhong Road Haidian District Beijing, 100083 CHINA

Re: K130809

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: January 2, 2014 Received: January 8, 2014

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-rclated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Image /page/5/Picture/6 description: The image shows a signature block with the text "Sincerely yours,". Below that is the name "Tejashri Purohit-Sheth, M.D." and the title "Clinical Deputy Director". The acronym "DAGRID" is also present, and the word "FOR" is at the bottom right.

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known) K130809

Device Name

Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored)

Indications for Use (Describe)

Powder Free Viny! Patient Examination Gloved) is a non-sterile disposable device intended for medical purposs that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

[] Prescription Use (Part 21 CFR 801 Subpart D)

[x] Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FOR FDA USE ONLY . :
ాలు . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/6/Picture/14 description: The image shows the name Elizabeth F. Claver. Below the name is a date and time stamp. The date is 2014.02.14 and the time is 21:16:05.

{7}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.