K Number
K130801

Validate with FDA (Live)

Date Cleared
2013-06-14

(84 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SonoScape S11 device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), OB/Gyn and Urology.

Device Description

The SonoScape S11 Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications. The all digital architecture with progressive dynamic receive focusing allows the system to maximize the utility of all imaging transducers to enhance the diagnostic utility and confidence provided by the system. The exam dependent default setting allows the user to have minimum adjustment for imaging the patient, while the in-depth soft-menu control allows the advanced user to set the system for different situations. The architecture allows cost-effective system integration to a variety of upgrade-able options and features. This SonoScape system is a general purpose, software controlled, diagnostic ultrasound system. Its basic function is to acquire ultrasound data and display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Doppler and Power Doppler, or a combination of these modes, 4D.

AI/ML Overview

The provided 510(k) summary is for a diagnostic ultrasound system, which is a hardware device. These types of submissions typically focus on demonstrating substantial equivalence to a predicate device based on technical specifications and adherence to recognized safety standards, rather than clinical performance metrics often associated with AI/ML-enabled devices.

Therefore, many of the requested criteria (acceptance criteria related to clinical performance, specific study types like MRMC, sample sizes for test/training sets, ground truth establishment, or expert qualifications) are not applicable or detailed in this document. The submission asserts substantial equivalence based on comparisons of design, function, and safety characteristics with a predicate device.

Here's a breakdown of what can be extracted and what is not provided based on the input:

1. A table of acceptance criteria and the reported device performance

The document does not specify quantitative acceptance criteria for device performance in terms of diagnostic accuracy, sensitivity, or specificity, as it's an ultrasound hardware device submission focused on substantial equivalence rather than a new clinical claim. Its "performance" is implicitly deemed acceptable if it meets the technical specifications and safety standards, and is substantially equivalent to the predicate.

The table below summarizes the comparison items where "SE" (Substantially Equivalent) is indicated, implying the proposed device meets the predicate's performance in those aspects.

Comparison ItemAcceptance Criteria (Predicate)Reported Device Performance (Proposed Device)Remark
Classification NameUltrasonic Pulsed Doppler Imaging System, Ultrasonic Pulsed Echo Imaging System, Diagnostic Ultrasound TransducerUltrasonic Pulsed Doppler Imaging System, Ultrasonic Pulsed Echo Imaging System, Diagnostic Ultrasound TransducerSE
Product Code90-IYN/90-IYO/90-ITX90-IYN/90-IYO/90-ITXSE
Regulation Number892.1550/892.1560/892.1570892.1550/892.1560/892.1570SE
PanelRadiologyRadiologySE
ClassIIIISE
Intended UseGeneral-purpose ultrasonic imaging for Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), OB/Gyn and Urology.The same intended uses, with the addition of one more probe (C354) that does not introduce new intended uses, maintaining substantial equivalence.SE
Probe Type & ConnectorsL741, L742, L743 Linear Array; C344, C362 Curved Array; C611, 6V1, 6V3, EC9-5 Micro-curved Array; 2P1, 5P1 Phased Array. Multi-port connector connects 2 transducers.Includes all predicate probes plus C354 (Curved Array, 2.0-5.0 MHz). Multi-port connector connects 3 transducers.SE (Analysis 1, 2)
Acoustic TrackTRACK 3TRACK 3SE
Design (Functions Comparison - Excerpt)Based on embedded Linux OS, 64 channel full digital beam former, Autocorrelation for color/FFT for pulse/CW Doppler, supports Linear, Curve linear and Phase array probes from 2 to 15 MHz, Cine play back, Image file archive, Software upgrades with USB, Digital Scan Converter 800x600, TGC 8 slider, Depth Range 3 to 32 cm, Image sector size 32 lines to full B (256 lines), Image Sector position Steering.Same as predicate.SE
Operation ModesB, M, PW, CW, CFM, DPI, TDI, Tissue Harmonic Image, 3D/4D Mode, Color M Mode.B, M, PW, CW, CFM, DPI, TDI, Tissue Harmonic Image, 4D Mode, Color M Mode. (Note: 3D function is present in predicate but not explicitly listed for proposed device, addressed in SE Analysis 3).SE (Analysis 3 specifies absence of 3D function but deems SE)
Display ModesDual B, Quad Display, B and M, B and Doppler, B + Color, Dual B (Flow), Triplex mode, Compound Imaging, Panoramic Imaging, Trapezoid Imaging.Same as predicate.SE
Measurement ItemsDistance; area; circumference; calipers; volume, velocity, HR, PI, RI. Cardiac. OB/GYN, Urology, Vascular and small part packageSame as predicate.SE
Cine LoopAutomatic/manual review. Review speed can be adjusted.Automatic/manual review. Review speed can't be adjusted.SE
Power SupplyVoltage: 110-240 VAC, Frequency: 50/60 Hz, Power Consumption: 2.7-1.2AVoltage: 100-240 VAC, Frequency: 50/60 Hz, Power Consumption: 2.7-1.2ASE (Analysis 4)
Operating ConditionTemperature: 10-40°C, Relative humidity: 30-85%, Air pressure: 700-1060hPaSame as predicate.SE
Storage ConditionTemperature: -20-55°C, Relative humidity: 20-90%, Air pressure: 700-1060hPaSame as predicate.SE
Screen Size15 inch Widescreen LCD monitorSame as predicate.SE
Acoustic OutputTrack 3: MI,TIS,TIC,TIB, Derated ispta: 720Mw/cm² maximum, TIS/TIB/TIC: 6.0 Maximum, Mechanical Index: 1.9 Maximum, or Derated Isppa: 190W/cm² maxSame as predicate.SE
Electrical SafetyIEC 60601-1Same as predicate.SE
EMCIEC 60601-1-2Same as predicate.SE
PerformanceIEC 60601-2-37Same as predicate.SE
BiocompatibilityISO 10993-5, ISO 10993-10Same as predicate.SE
Level of Concern of SoftwareModerate level of concern systemSame as predicate.SE

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states "No clinical testing was required." Therefore, there is no information on sample size for a test set or data provenance related to clinical data. The testing mentioned in Section 7 refers to "Laboratory testing" to verify design specifications and compliance with safety standards, which is not a clinical test set.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as no clinical testing or test set requiring ground truth was conducted for this submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no clinical testing or test set requiring adjudication was conducted for this submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a diagnostic ultrasound system, not an AI-enabled device designed to assist human readers. No MRMC study was mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a hardware device; it does not perform as a standalone algorithm in the sense of AI/ML.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable, as no clinical testing or test set requiring ground truth was conducted for this submission.

8. The sample size for the training set

Not applicable. This is a hardware device; there is no "training set" in the context of AI/ML algorithms.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this hardware device.

{0}------------------------------------------------

510(k) Summary

JUN 1 4 2013

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number: __________________________________________________________________________________________________________________________________________________

    1. Date of submission: March 13, 2013

2. Submitter

SonoScape Company Limited

Address: Yizhe Building, Yuquan Road, Nanshan, Shenzhen 518051, P.R. China

Tel: (86) 755-26722890

Fax: (86) 755-26722850

Contact Person: Toki Wu

E-mail:Faith@sonoscape.net/wusq@sonoscape.net

3. Proposed Device Identification

Trade/Proprietary Name: S11 Digital Color Doppler Ultrasound System Common Name: Diagnostic Ultrasound System and Transducers Classification:

21 FR 892.1550 Ultrasonic Pulsed Doppler Imaging System (90-IYN)

21 FR 892.1560 Ultrasonic Pulsed Echo Imaging System (90-IYO)

21 CFR 892.1570 Diagnostic Ultrasound Transducer (90-ITX)

Classification Panel: Radiology

Device Class: II

4. Legally Marketed Predicate Device

SonoScape Company Limited, Diagnostic Ultrasound System, Model S6 has been cleared by FDA through 510(k) No.K112602 (Decision Date - November 07, 2011).

510(k) Summary

{1}------------------------------------------------

5. Device Description

The SonoScape S11 Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.

The all digital architecture with progressive dynamic receive focusing allows the system to maximize the utility of all imaging transducers to enhance the diagnostic utility and confidence provided by the system. The exam dependent default setting allows the user to have minimum adjustment for imaging the patient, while the in-depth soft-menu control allows the advanced user to set the system for different situations. The architecture allows cost-effective system integration to a variety of upgrade-able options and features.

This SonoScape system is a general purpose, software controlled, diagnostic ultrasound system. Its basic function is to acquire ultrasound data and display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Doppler and Power Doppler, or a combination of these modes, 4D.

6. Intended Use Statement

The SonoScape S11 device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), OB/Gyn and Urology.

7. Testing

Laboratory testing was conducted to verify that the S11 system with added transducer met all design specification and was substantially equivalent to the Predicate Device. The device has been found to conform to applicable medical device safety standards in regards to thermal, mechanical and electrical safety as well as biocompatibility. The acoustic output is measured and calculated per "NEMA UID 2: 2004 Acoustic Output

510(k) Summary

{2}------------------------------------------------

Measurement Standard for Diagnostic Ultrasound Equipment" and "NEMA UD3: 2004 Standards for Real-time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment".

IEC 60601-1: 1988+A1:1991+A2:1995 Medical Electrical Equipment - Part 1: General Requirements for Safety

IEC 60601-1-2: 2007 Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests. IEC 60601-2-37: 2008 Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment.

NEMA UD 2-2004, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment Version 3.

NEMA UD3: 2004 Standards for Real-time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment

8. Clinical Test:

No clinical testing was required.

9. Comparison Table

The differences between the S11 and the predicate device S6 in almost every part are listed in the tables below.

IDComparisonItemsProposed DeviceSonoscape S11Predicate DeviceSonoscape S6Remark
2ClassificationNameUltrasonic Pulsed DopplerImaging SystemUltrasonic Pulsed EchoImaging SystemDiagnostic UltrasoundTransducerUltrasonic Pulsed DopplerImaging SystemUltrasonic Pulsed EchoImaging SystemDiagnostic UltrasoundTransducerSE
3ProductCode90-IYN/90-IYO/90-ITX90-IYN/90-IYO/90-ITXSE
IDComparisonItemsProposed DeviceSonoscape S11Predicate DeviceSonoscape S6Remark
4RegulationNumber892.1550/892.1560/892.1570892.1550/892.1560/892.1570SE
5PanelRadiologyRadiologySE
6ClassIIIISE
1IntendedUseThe device is ageneral-purpose ultrasonicimaging instrument intended foruse by a qualified physician forevaluation of Fetal, Abdominal,Pediatric, Small Organ (breast,testes, thyroid), Cephalic(neonatal and adult),Trans-rectal, Trans-vaginal,Peripheral Vascular,Musculo-skeletal (Conventionaland Superficial), Cardiac(neonatal and adult), Urologyand OB/Gyn.The device is ageneral-purpose ultrasonicimaging instrument intended foruse by a qualified physician forevaluation of Fetal, Abdominal,Pediatric, Small Organ (breast,testes, thyroid),Cephalic(neonatal and adult),Trans-rectal, Trans-vaginal,Peripheral Vascular,Musculo-skeletal (Conventionaland Superficial), Cardiac(neonatal and adult), OB/Gynand Urology.SE
7Probe Type&ConnectorsL741 Linear Array, 5.0-10.0MHzL742 Linear Array, 5.0-12.0MHzL743 Linear Array, 5.0-10.0MHzC344 Curved Array, 2.0-5.0MHzC362 Curved Array, 2.0-6.0MHzC354 Curved Array, 2.0-5.0MHzC611 Micro-curved Array,4.0-8.0 MHzVC6-2 Curved Array, 2.0-6.0MHz6V1 Micro-curved Array,4.0-8.0 MHz6V3 Micro-curved Array,5.0-9.0 MHzEC9-5 Micro-curved Array,5.0-9.0 MHzL741 Linear Array, 5.0-10.0MHzL742 Linear Array, 5.0-12.0MHzL743 Linear Array, 5.0-10.0MHzC344 Curved Array, 2.0-5.0MHzC362 Curved Array, 2.0-6.0MHzC611 Micro-curved Array,4.0-8.0 MHzVC6-2 Curved Array, 2.0-6.0MHz6V1 Micro-curved Array,4.0-8.0 MHz6V3 Micro-curved Array,5.0-9.0 MHzEC9-5 Micro-curved Array,5.0-9.0 MHzSEAnalysis1SESESE
2P1 Phased Array, 2.0-4.0 MHz2P1 Phased Array, 2.0-4.0 MHzSE
IDComparison ItemsProposed DeviceSonoscape S11Predicate DeviceSonoscape S6Remark
5P1 Phased Array, 4.0-7.0 MHz5P1 Phased Array, 4.0-7.0 MHz
Multi-port connector connects 3 transducersMulti-port connector connects 2 transducersSEAnalysis2
8Acoustic TrackTRACK 3TRACK 3SE

Table 1 General Comparison

510(k) Summary

{3}------------------------------------------------

SonoScape Company Limited

510(k) Submission

510(k) Summary

.

.

{4}------------------------------------------------

SonoScape Company Limited

510(k) Submission

Table 2 Functions Comparison

IDComparisonItemsProposed DeviceSonoScape S11Predicate DeviceSonoScape S6Remark
9DesignBased on an embedded Linuxoperating system.Based on a 64 channel fulldigital beam former.Autocorrelation for colorprocessing and FFT for pulseand CW Doppler processing.Supporting Linear, Curve linearand Phase array probes from 2to 15 MHz.Cine play back capabilityImage file archiveSoftware upgrades with USBflash drive.Digital Scan Converter800×600TGC 8 sliderDepth Range: 3 to 32 cmImage sector size: 32 lines tofull B (256 lines)Image Sector position: Steeringwithin full maximumBased on an embedded Linuxoperating system.Based on a 64 channel fulldigital beam former.Autocorrelation for colorprocessing and FFT for pulseand CW Doppler processing.Supporting Linear, Curve linearand Phase array probes from 2to 15 MHz.Cine play back capabilityImage file archiveSoftware upgrades with USBflash drive.Digital Scan Converter800×600TGC 8 sliderDepth Range: 3 to 32 cmImage sector size: 32 lines tofull B (256 lines)Image Sector position: Steeringwithin full maximumSE
10OperationControlsB orientation flip :L/R key withmarking on the screenB Dynamic range control:preset 14 curves over 140 dBGray Scale Control: 7 SettingsFocal Number: 12 focal zonesettingB persistence: 0-95%Image Processing: SmoothingB orientation flip :L/R key withmarking on the screenB Dynamic range control:preset 14 curves over 140 dBGray Scale Control: 7 SettingsFocal Number: 12 focal zonesettingB persistence: 0-95%Image Processing: SmoothingSE
ComparisonProposed DevicePredicate Device
IDItemsSonoScape S11SonoScape S6Remark
edge enhancementedge enhancement
PW sweeping speed 2,4,6,8PW sweeping speed 2,4,6,8SE
sec over displaysec over display
PW Wall filter setting: 16PW Wall filter setting: 16SE
settings,25 to 750 HZsettings,25 to 750 HZ
PW sample volume:0.5 toPW sample volume:0.5 toSE
20mm20mm
PW/B update: with UPDATEPW/B update: with UPDATESE
keykey
PW cursor steering: Steer softPW cursor steering: Steer softSE
keykey
PW angle correction:0 to 80PW angle correction:0 to 80SE
degree user controldegree user control
PW spectrum dynamicPW spectrum dynamicSE
range: 10 preset curve overrange:10 preset curve over
15-48 dB15-48 dB
Spectrum baseline shift andSpectrum baseline shift andSE
invertinvert
Color ROI setting: trackball andColor ROI setting: trackball andSE
set key to control size andset key to control size and
positionposition
Color steering on flat probe:±20Color steering on flat probe:±20SE
±160±160
Color Wall Filter: Color wallColor Wall Filter: Color wallSE
filter with 16 selection, 25-750filter with 16 selection, 25-750
of PRFof PRF
Color priority-B priority softColor priority-B priority softSE
menumenu
Color Packet size: preset perColor Packet size: preset perSE
Exam, horizontal, vertical, offExam, horizontal, vertical, off
Zoom adjustableZoom adjustableSE
Freeze control: Toggling freezeFreeze control: Toggling freezeSE
keykey
Cine control: step, playCine control: step, playSE
backward, play continuouslybackward, play continuously
11OperationB, M, PW, CW, CFM, DPI, TDI,B, M, PW, CW, CFM, DPI, TDI,SE
ModeTissue Harmonic ImageTissue Harmonic ImageAnalysis
4D Mode3D/4D Mode3
Color M ModeColor M Mode
12DisplayDual B, Quad Display,Dual B, Quad Display,SE
ModesB and M, B and DopplerB and M, B and Doppler
IDComparisonItemsProposed DeviceSonoScape S11Predicate DeviceSonoScape S6Remark
B + Color, Dual B(Flow)B + Color, Dual B(Flow)
Triplex mode: B,CFM, andTriplex mode: B,CFM, and
PW/CW; B,DPI, andPW/CW; B,DPI, and
PW/CW;B,THI and Color M,steer MPW/CW;B,THI and Color M,steer M
Dual B and Color in real timeDual B and Color in real time
Compound Imaging,Compound Imaging,
Panoramic Imaging,Panoramic Imaging,
Trapezoid Imaging.Trapezoid Imaging.
13Measurement ItemsDistance; area; circumference;Distance; area; circumference;SE
calipers; volume, velocity, HR,calipers; volume, velocity, HR,
PI, RI. Cardiac. OB/GYN,PI, RI. Cardiac. OB/GYN,
Urology, Vascular and smallUrology, Vascular and small
part packagepart package
14Cine LoopAutomatic review/ manualreviewAutomatic review/ manualreviewSE
Review speed can't be adjustReview speed can be adjustSE

ි බ

{5}------------------------------------------------

SonoScape Company Limited

510(k) Submission

:

510(k) Summary

{6}------------------------------------------------

SonoScape Company Limited

Table 3 Specifications Comparison

IDComparisonItemsProposed DeviceSonoScape S11Predicate DeviceSonoScape S6Remark
15PowerSupplyVoltage: 100-240 VACFrequency: 50/60 HzPower Consumption:100-240V AC, 2.7-1.2AVoltage: 110-240 VACFrequency: 50/60 HzPower Consumption: 110-240VAC, 2.7-1.2ASEAnalysis4
16OperatingConditionTemperature: 1040°CRelative humidity: 3085%Air pressure: 700hPa~1060hPaTemperature: 1040°CRelative humidity: 3085%Air pressure: 700hPa~1060hPaSE
17StorageConditionTemperature: -2055°CRelative humidity: 2090%Air pressure: 700hPa~1060hPaTemperature: -2055°CRelative humidity: 2090%Air pressure: 700hPa~1060hPaSE
18Screen Size15 inch Widescreen LCDmonitor15 inch Widescreen LCDmonitorSE
19AcousticOutputTrack 3:MI,TIS,TIC,TIBDerated ispta: 720Mw/cm²maximum.Track 3: MI,TIS,TIC,TIBDerated ispta: 720Mw/cm²maximum.SE

{7}------------------------------------------------

SonoScape Company Limited

IDComparisonItemsProposed DeviceSonoScape S11Predicate DeviceSonoScape S6Remark
TIS/TIB/TIC: 6.0 Maximum,Mechanical Index: 1.9Maximum, or Derated Isppa:190W/cm² maxTIS/TIB/TIC: 6.0 Maximum,Mechanical Index: 1.9Maximum, or Derated Isppa:190W/cm² max

Table 4 Safety Comparison

IDComparisonItemsProposed DeviceSonoscape S11Predicate DeviceSonoscape S6Remark
20Electrical Safety-IEC 60601-1-IEC 60601-1SE
21EMC-IEC 60601-1-2-IEC 60601-1-2SE
22Performance-IEC 60601-2-37-IEC 60601-2-37SE
23Biocompatibility-ISO 10993-5,-ISO 10993-10-ISO 10993-5,-ISO 10993-10SE
24Level ofConcern OfSoftwareModerate level of concernsystemModerate level of concernsystemSE

SE Analysis 1

Compare to the predicate device, the proposed device is with one more probe C354. But the intended use of C354 is the same to the C344 and C362, no new intended use is added, therefore, the intended use could be considered Substantially Equivalent.

SE Analysis 2

The proposed device has 3 probe connection ports and the predicate device has 2 probe connection ports, but both of them comply with IEC 60601-1 and IEC 60601-1-2.

SE Analysis 3

Operation Mode, compare to the predicate device, the proposed device doesn't have 3D function. But both of them comply with AAMI UD2 and IEC 60601-2-37. Therefore, they are considered to be substantially equivalent, so the SE is not affected.

SE Analysis 4

The Voltage of the proposed device and the predicate device are 100-240 VAC and 110-240 VAC respectively, but both of them comply with IEC60601-1 and IEC 60601-1-2. Therefore, power supply can be considered Substantially Equivalent in safety and effectiveness.

:

{8}------------------------------------------------

K130801
Page 9 of 9

Discussion of Substantially Equivalent

The subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device. The differences above between the subject device and predicate device do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is raised regarding to effectiveness and safety.

10. Substantially Equivalent Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, SonoScape Company Limited concludes that S11 Digital Color Doppler Ultrasound System is substantially equivalent to predicate devices with regard to safety and effectiveness.

510(k) Summary

{9}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 14, 2013 .

SonoScape Company Limited % Ms. Toki Wu Regulatory Affairs Manager Yizhe Building, Yuquan Road, Nanshan Shenzhen, Guangdong 518051 CHINA

Re: K130801

Trade/Device Name: S11 Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: March 13, 2013 Received: March 22, 2013

Dear Ms. Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general-controls-provisions of the Act.include requirements.for annual_registration, listing_of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

This determination of substantial equivalence applies to the following transducers for use with the SII Digital Color Doppler Ultrasound System, as described in your premarket notification:

Transducer Model Number

2P1 Phase Array 5P1 Phase Array 6V1 Micro-curved Array 6V3 Micro-curved Array EC9-5 Micro-curved Array C611 Micro-curved Array C362 Curved Array C344 Curved Array . C354 Curved Array VC6-2 Curved Array

L743 Linear Array L741 Linear Arrav L742 Linear Array

{10}------------------------------------------------

Page 2 - Ms. Wu

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA `s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.hum.

-Sincerely-yours;-

Smh?)

for

Janine M. Morris Deputy Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

Indications for Use

510(k) Number: K130801

Device Name: S11 Digital Color Doppler Ultrasound System

Indications for Use:

The SonoScape S11 device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal, Transvaginal, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), OB/Gyn and Urology.

-Over-The-Counter Use. -AND/OR-Prescription-Use .. (21-GFR-801-Subpart-C)-(Part 21-CFR 801-Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Sm. 7)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

K130801 510(k)

{12}------------------------------------------------

System: SonoScape S11

Diagnostic Ultrasound Pulsed Echo System

Diagnostic Ultrasound Pulsed Doppler Imaging System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalNNNNNNote 1Notes 2,4
AbdominalNNNNNNote 1Notes 2,4
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricNNNNNNote 1Notes 2
Small Organ (specify)NNNNNNote 1Notes 2,5
Neonatal CephalicNNNNNNNote 1Notes 2,3
FetalAdult CephalicNNNNNNNote 1Notes 2,3
Imaging&Trans-rectalNNNNNNote 1Notes 2
OtherTrans-vaginalNNNNNNote 1Notes 2
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)NNNNNNote 1Notes 2
Musculo-skeletal(Superficial)NNNNNNote 1Notes 2
Intravascular
Other (Ob/GYN)NNNNNNote 1Notes 2,4
Cardiac AdultNNNNNNNote 1Notes 2,3
Cardiac PediatricNNNNNNNote 1Notes 2,3
CardiacIntravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vesselNNNNNNote 1Notes 2
VesselOther (specify)

E = added under this ppendix N = new indication; P = previously cleared by FDA; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color

Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 4D

Note 5: Small Organ: breast, thyroid, testes

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

510(k) K130801

Indications for Use

{13}------------------------------------------------

Transducer: 2P1 Phase Array

Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation Specific Color Power Other* Other* General CWD Combined (TRACKS 1 & 3) B M PWD Doppler (Amplitude) Specify (TRACK 1 Doppler ONLY) Ophthalmic Ophthalmic Fetal Fetal P P Notes 2 lmaging& Abdominal P P P Note 1 Other Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic P P p P Note 1 Notes 2,3 P P P b P Note 1 Notes 2,3 P P P Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Cardiac Adult P P P P P Note 1 Notes 2,3 Cardiac P P Cardiac Pediatric P P P P P Note 1 Notes 2,3 Intravascular(Cardiac) Trans-esoph. (Cardiac) Intra-cardiac Other (specify) Peripheral Peripheral vessel Vessel Other (specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M ; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 4D

Note 5: Small Organ: breast, thyroid, testes

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

510(k) K130801

Indications for Use

{14}------------------------------------------------

" Transducer: 5P1 Phase Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricPPPPPNote 1Notes 2
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2,3
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
CardiacCardiac Adult
Cardiac PediatricPPPPPPNote 1Notes 2,3
Intravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Other (specify)

E = added under this appendix N = new indication; P = previously cleared by FDA; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M ; B/Color Doppler; B/Color

Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 4D Note 3: TDI

Note 5: Small Organ; breast, thyroid, testes

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

510(k) K130801

Indications for Use

{15}------------------------------------------------

Transducer: 6V1 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Notes 2
Trans-vaginalPPPPPNote 1Notes 2
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal
(Conventional)
Musculo-skeletal
(Superficial)
Intravascular
Other (Ob/GYN)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph (Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselOther (specify)

N = new indication; · P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M ; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 4D

Note 5: Small Organ: breast, thyroid, testes

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

510(k) K130801

{16}------------------------------------------------

Transducer: 6V3 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Notes 2
Trans-vaginalPPPPPNote 1Notes 2
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Other (specify)

E = added under this appendix P = previously cleared by FDA; N = new indication;

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 4D

Note 5: Small Organ: breast, thyroid, testes

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

510(k) _______________________________________________________________________________________________________________________________________________________________________

Indications for Use

{17}------------------------------------------------

Transducer: EC9-5 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther *Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Notes 2
Trans-vaginalPPPPPNote 1Notes 2
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Other (specify)

P = previously cleared by FDA; E = added under this appendix N = new indication;

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M ; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 4D

Note 5: Small Organ: breast, thyroid, testes

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

510(k) K130801

{18}------------------------------------------------

Transducer: C611 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetal
AbdominalPPPPPNote 1Notes 2
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
PediatricPPPPPNote 1Notes 2
Small Organ (specify)
Neonatal CephalicPPPPPPNote 1Notes 2,3
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)
CardiacCardiac Adult
Cardiac PediatricPPPPPPNote 1Notes 2,3
Intravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Other (specify)

E = added under this appendix N = new indication; P = previously cleared by FDA; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M ; B/Color Doppler; B/Color

Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 4D Note 3: TDI

Note 5: Small Organ: breast, thyroid, testes

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

K130801 510(k) -

{19}------------------------------------------------

Transducer: C362 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalFetalPPPPPNote 1Notes 2
Imaging &AbdominalPPPPPNote1Notes 2
OtherIntra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal
(Conventional)
Musculo-skeletal
(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote1Notes 2
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph (Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselOther (specify)

P = previously cleared by FDA; E = added under this appendix N = new indication;

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M ; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 4D

Note 5: Small Organ: breast, thyroid, testes

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

K130801 510(k)

{20}------------------------------------------------

Transducer: C344 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalFetalPPPPPNote 1Notes 2
Imaging&AbdominalPPPPPNote 1Notes 2
OtherIntra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal
(Conventional)
Musculo-skeletal
(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph (Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselOther (specify)

E = added under this appendix N = new indication; P = previously cleared by FDA; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M ; B/Color Doppler; B/Color

Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging: The feature does not use contrast agents

Note 3: TDI Note 4: 4D

Note 5: Small Organ: breast, thyroid, testes

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

510(k)

Indications for Use

{21}------------------------------------------------

Transducer: C354 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalImaging&OtherFetalNNNNNNote 1Notes 2
AbdominalNNNNNNote 1Notes 2
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)NNNNNNote 1Notes 2
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vessel
Other (specify)

P = previously cleared by FDA; E = added under this appendix N = new indication;

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M ; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 4D

Note 5: Small Organ: breast, thyroid, testes

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

K130801 510(k)

Indications for Use

{22}------------------------------------------------

Transducer: VC6-2 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalFetalPPPPPNote 1Notes 2,4
Imaging &AbdominalPPPPPNote 1Notes 2.4
OtherIntra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph.(non-Card)
Musculo-skeletal(Conventional)
Musculo-skeletal(Superficial)
Intravascular
Other (Ob/GYN)PPPPPNote 1Notes 2,4
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph.(Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vessel
VesselOther (specify)

P = previously cleared by FDA; N = new indication;

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M ; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 4D

Note 5: Small Organ: breast, thyroid, testes

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

K130801 510(k)

Indications for Use

{23}------------------------------------------------

Transducer: L743 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalFetal
Imaging&Abdominal
OtherIntra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2
Intravascular
Other (Ob/GYN)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vesselPPPPPNote 1Notes 2
VesselOther (specify)

E = added under this appendix P = previously cleared by FDA; N = new indication;

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M ; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 4: 4D Note 3: TDI

Note 5: Small Organ: breast, thyroid, testes

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

510(k) _______________________________________________________________________________________________________________________________________________________________________ K130801

Indications for Use

{24}------------------------------------------------

Transducer: L741 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerOther*CombinedOther*Specify
OphthalmicOphthalmic
FetalFetal
Imaging&Abdominal
OtherIntra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2.5
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletalPPPPPNote 1Notes 2
(Conventional)
Musculo-skeletal
(Superficial)
Intravascular
Other (Ob/GYN)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)-
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralPeripheral vesselPPPPPNote 1Notes 2
VesselOther (specify)

P = previously cleared by FDA; E = added under this appendix N = new indication; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M ; B/Color Doppler; B/Color

Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 4D

Note 5: Small Organ: breast, thyroid, testes

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

K130801 510(k)

Indications for Use

{25}------------------------------------------------

Transducer: L742 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther*
General(TRACK 1ONLY)Specific(TRACKS 1 & 3)BMPWDCWDColorDopplerPower(Amplitude)DopplerCombinedSpecify
OphthalmicOphthalmic
FetalImaging&OtherFetal
Abdominal
Intra-operative Specify
Intra-operative Neuro
Laparoscopic
Pediatric
Small Organ (specify)PPPPPNote 1Notes 2,5
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card)
Musculo-skeletal(Conventional)PPPPPNote 1Notes 2
Musculo-skeletal(Superficial)PPPPPNote 1Notes 2
Intravascular
Other (Ob/GYN)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular(Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (specify)
PeripheralVesselPeripheral vesselOther (specify)PPPPPNote 1Notes 2

P = previously cleared by FDA; E = added under this appendix N = new indication; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color

Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 4D

Note 5: Small Organ: breast, thyroid, testes

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

510(k) K130801

Indications for Use

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.