K Number
K130775

Validate with FDA (Live)

Device Name
DUET SYSTEM
Manufacturer
Date Cleared
2014-05-09

(415 days)

Product Code
Regulation Number
866.4700
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Duet™ System is an automated scanning microscope and image analysis system. It is intended for in-vitro diagnostic use as an aid to the pathologist in the detection, classification and counting of cells of interest based on color, intensity, size, pattern and shape.

The Duet™ System is intended to:

  1. Detect Hematopoietic cells stained by Giemsa stain, Immunohistochemistry or ISH (with brightfield and fluorescent) prepared from cell suspension.

  2. Detect Amniotic cells stained by FISH (using direct labeled DNA probes for chromosomes X. Y. 13, 18 and 21).

  3. Detect Aneuploidy for chromosomes 3,7. 17 and loss of the 9p21 locus via FISH in Urine specimens from subjects with transitional cell carcinoma of the bladder, probed by the Vysis Urovysion Bladder Cancer Kit.

  4. Detect and quantify chromosome 17 and the HER-2/neu gene via fluorescence in situ hybridization (FISH) in interphase nuclei from formalin-fixed, paraffin embedded human breast cancer tissue specimens, probed by the Vysis® Path Vysion™ HER-2 DNA Probe Kit. The Duet™ is to be used as an adjunctive automated enumeration tool, in conjunction with manual review of the digital image, to assist in determining HER-2/neu gene to chromosome 17 signal ratio.

  5. Qualitatively detect rearrangements involving the ALK gene via fluorescence in situ hybridization (FISH) in formalinfixed paraffin-embedded (FFPE) non-small cell lung cancer (NSCLC) tissue specimens, probed with the Vysis ® ALK Break Apart FISH Probe Kit. The Duer™ is to be used as an adjunctive autonated enumeration with manual review of the digital image. Note: The pathologist should verify the image analysis software application score.

Device Description

The Duet™ System is a fully integrated imaging and scanning platform that automates time-consuming and difficult laboratory tasks of slide scanning.

The Duet™ System workstation integrates a microscope, CCD camera, motorized stage / slide-loader, computer, keyboard, mouse, joystick, monitor and a dedicated software program.

The Duet™ System is software controlled and includes features such as: acquisition of images, views, editing, relocation, enhancement capabilities, automatic/manual counting and classification, printing, export of images and backups.

The Duet™ System scans in high resolution cell samples at high speed both in bright light illumination and in fluorescent illumination.

The Duet™ System suggests classification of the cells according to their morphological features, their staining (Giemsa, IHC) and fluorescent signals, and allows the user to quickly examine the results, correct them as needed and generate a report summarizing the sample's data. The Duet™ system allows combined presentation of morphological and specific staining information of the same cell, for all the cells of the sample.

AI/ML Overview

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Coefficient of Variation, CV)Reported Device Performance (Mean %CV Range for Individuals Slides)Reported Device Performance (Overall %CV derived from Random Model)
Positive samples: CV < 25%Within-run: 0.0% - 19.0% (for positive slides with mean >10%)Below specified goals
Negative samples, with mean percentage below 4%: CV ≤ 180%Within-run: 0.0% - 86.6% (for negative slides with mean <10%)Below specified goals
Negative samples, with mean percentage above 4%: CV < 70%Within-run: 0.0% - 86.6% (for negative slides with mean <10%)Below specified goals
Additional Criteria (Binary Outcome: Positive/Negative)Reported Performance
100% Repeatability100% Repeatability
100% Reproducibility100% Reproducibility
Method Comparison (Accuracy)Reported Performance
Concordance with manual countingPercent Agreement with Manual Negative: 100% (95% CI: 95.5%-100%)
Percent Agreement with Manual Positive: 96.9% (95% CI: 84.3%-99.5%)
Percent Overall Agreement: 99.1% (95% CI: 95.2%-99.8%)

2. Sample Size and Data Provenance for the Test Set

  • Reproducibility and Repeatability Study (Study 1):
    • Sample Size: A panel of 16 archived clinical specimen slides. These were categorized into four groups based on positive cell percentage: <10%, 10-25%, 25-50%, and ≥ 50%. There were 4 samples in each of these four value ranges.
    • Data Provenance: The document states "archived clinical specimen slides." No specific country of origin is mentioned, but the manufacturer is based in Israel. The data is retrospective.
  • Analytical Performance/Methods Comparison Study (Study 2):
    • Sample Size: 113 specimen slides. Out of these, 32 were positive and 11 were equivocal.
    • Data Provenance: The slides were prepared and probed using the ALK Kit. No specific country of origin is mentioned. The data appears to be retrospective clinical samples.
  • Configurations Method Comparison Study (Study 3):
    • Sample Size: Not explicitly stated as a single number. It included "patient slides from each of the four previously cleared indications" and "samples were selected to cover the intended use of both Normal, Abnormal and near the medical decision/cut-off (as applicable)."
    • Data Provenance: Clinical specimens. No specific country of origin is mentioned. The data appears to be retrospective.

3. Number of Experts and Qualifications for Ground Truth

  • For the Reproducibility and Repeatability Study (Study 1) and Analytical Performance/Methods Comparison Study (Study 2): The studies compare the Duet™ System's performance to "manual system" or "manual scoring method." The document implies that the manual scoring is performed by pathologists in the intended use. While not explicitly stated, it's typical for the ground truth in such studies to be established by qualified pathologists or clinical laboratory scientists with expertise in FISH analysis. The number of experts is not specified.
  • For the Configurations Method Comparison Study (Study 3): The comparison was made between the new and cleared configurations, with the "final interpretation of results" being identical. This suggests the ground truth was established based on the existing standard of interpretation (likely by pathologists) for the previously cleared indications. The number and qualifications of experts are not explicitly detailed.

4. Adjudication Method for the Test Set

  • The document does not explicitly describe an adjudication method like 2+1 or 3+1 for resolving discrepancies in the test set.
  • For the Analytical Performance/Methods Comparison Study (Study 2), it's a direct comparison against a "manual scoring method." The outcome suggests a high level of agreement, implying that the manual scoring itself served as the reference.
  • For the Reproducibility and Repeatability Study (Study 1), variability between runs, days, and sites was measured, not necessarily involving adjudication for a single ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No specific MRMC comparative effectiveness study is explicitly mentioned to evaluate how much human readers improve with AI vs. without AI assistance. The studies presented focus on the device's accuracy and precision compared to a manual method (Study 2) or its own internal consistency (Study 1). The device is noted as an "adjunctive automated enumeration tool, in conjunction with manual review of the digital image," implying a human-in-the-loop workflow, but a formal MRMC study demonstrating improvement in human performance with the device is not detailed.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

  • Yes, a standalone performance ("algorithm only without human-in-the-loop performance") was done. The Analytical Performance/Methods Comparison Study (Study 2) directly compares the "Duet™ System method" (which is automated imaging and analysis, i.e., the algorithm) to the "manual scoring method." The results presented, such as "Percent Agreement with Manual Negative" and "Percent Overall Agreement," are measures of the algorithm's performance against the human read. The device is intended as an "adjunctive automated enumeration tool," meaning the pathologist verifies the score, but the study assesses the machine's initial analytical performance.

7. Type of Ground Truth Used

  • Expert Consensus / Manual Scoring: For the Analytical Performance/Methods Comparison Study (Study 2), the ground truth was established by "manual scoring method." This implies that trained personnel (likely pathologists or clinical laboratory scientists) manually interpreted the slides to establish the reference standard.
  • Clinical Specimen Characteristics: For the Reproducibility and Repeatability Study (Study 1), the ground truth for "positive" and "negative" categorizations, as well as the percentage of positive cells, was based on the characteristics of the "archived clinical specimen slides."
  • Existing Standards: For the Configurations Method Comparison Study (Study 3), the ground truth was based on the "final interpretation of results" for the previously cleared indications, implying adherence to established clinical criteria and interpretations.

8. Sample Size for the Training Set

  • The document does not explicitly state the sample size used for training the Duet™ System's algorithms. The studies described are performance evaluation studies, which would typically use a separate test set, distinct from any training data.

9. How the Ground Truth for the Training Set was Established

  • Since the document does not explicitly mention the training set size, it also does not detail how the ground truth for the training set was established. However, for medical imaging analysis systems, training data ground truth is generally established by expert annotation (e.g., pathologists marking regions of interest, classifying cells) or by using slides with known clinical outcomes, often confirmed by pathology or other definitive diagnostic methods.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for "bioview". The logo consists of the letters "b", "i", "o", and "view" in a stylized font. The "b" and "i" are contained within a solid black circle, while the "o" is contained within a circle that is connected to two other circles. The word "view" is written in a simple, sans-serif font.

MAY 0 9 2014

Pre Market Notification Submission - 510(k)

510(K) SUMMARY Duet™ System 510(k) Number K130775

5.1 Company Name

Bioview Ltd. 3 Pekeris Street Rehovot 7670203, Israel Tel: +972-8-9366868 Fax: + 972-8-9366869

Contact Person 5.2

Orly Maor 25 A Sirkin Street Kfar-Saba 4442156, Israel Tel: 972-9-7453607 Fax: 972-153-9-7453607 Mail: oram.ma@gmail.com

5.3 Trade/Proprietary Name

Duet™ System

Classification Name 5.4

  • Automated cell-location device and: .
  • Automated Fluorescence in situ Hybridization (FISH) Enumeration Systems. 트

Product Code/Regulation No. 5.5

  • Automated cell-locating devices, product code: JOY, Regulation No. 864.5260. ■
  • Automated Fluorescence in situ Hybridization (FISH) Enumeration Systems, 트 product code: NTH, Regulation No. 866.4700.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Bioview. The logo consists of the word "bioview" in a sans-serif font. To the left of the word is a graphic of three circles connected by lines. The top circle is black, the middle circle is black with a white circle inside, and the bottom circle is gray.

Device Classification 5.6

Class II

5.7 Panel

Hematology and Immunology

Predicate Devices 5.8

  • Duet™ System, manufactured by BioView Ltd., cleared under K030192, I K040591, K050840, and K061602.
  • Human manual visualization of formalin-fixed paraffin-embedded human lung ■ cancer tissue specimens, probed by Vysis ALK Break Apart FISH Probe Kit (Hereinafter, ALK).

5.9 Intended Use

The Duet™ System is an automated scanning microscope and image analysis system. It is intended for in-vitro diagnostic use as an aid to the pathologist in the detection, classification and counting of cells of interest based on color, intensity, size, pattern and shape.

The Duet™ System is intended to:

  • Detect Hematopoietic cells stained by Giemsa stain, Immunohistochemistry or 1. ISH (with brightfield and fluorescent) prepared from cell suspension.
  • Detect Amniotic cells stained by FISH (using direct labeled DNA probes for 2. chromosomes X,Y,13, 18 and 21).
  • Detect Aneuploidy for chromosomes 3,7, 17 and loss of the 9p21 locus via 3. FISH in Urine specimens from subjects with transitional cell carcinoma of the bladder, probed by the Vysis Urovysion Bladder Cancer Kit.
  • Detect and quantify chromosome 17 and the HER-2/neu gene via fluorescence 4. in situ hybridization (FISH) in interphase nuclei from formalin-fixed, paraffin embedded human breast cancer tissue specimens, probed by the Vysis® Path Vysion™ HER-2 DNA Probe Kit. The Duet™ is to be used as an adjunctive automated enumeration tool, in conjunction with manual review of the digital image, to assist in determining HER-2/neu gene to chromosome 17 signal ratio.
  • Qualitatively detect rearrangements involving the ALK gene via fluorescence in న. situ hybridization (FISH) in formalin-fixed paraffin-embedded (FFPE) non-

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for "bioview". The logo consists of the word "bioview" with the "bio" part stylized with three circles connected in a triangular shape. The "b" is inside a solid black circle, the "i" is inside a white circle with a black outline, and the third circle is below the other two and is white with a gray pattern inside.

small cell lung cancer (NSCLC) tissue specimens, probed with the Vysis ® ALK Break Apart FISH Probe Kit. The Duet™ is to be used as an adjunctive automated enumeration tool, in conjunction with manual review of the digital image.

Note: The pathologist should verify the image analysis software application score.

5.10 Device Description

The Duet™ System is a fully integrated imaging and scanning platform that automates time-consuming and difficult laboratory tasks of slide scanning.

The Duet™ System workstation integrates a microscope, CCD camera, motorized stage / slide-loader, computer, keyboard, mouse, joystick, monitor and a dedicated software program.

The Duet™ System is software controlled and includes features such as: acquisition of images, views, editing, relocation, enhancement capabilities, automatic/manual counting and classification, printing, export of images and backups.

The Duet™ System scans in high resolution cell samples at high speed both in bright light illumination and in fluorescent illumination.

The Duet™ System suggests classification of the cells according to their morphological features, their staining (Giemsa, IHC) and fluorescent signals, and allows the user to quickly examine the results, correct them as needed and generate a report summarizing the sample's data. The Duet™ system allows combined presentation of morphological and specific staining information of the same cell, for all the cells of the sample.

5.11 Performance Standards

No performance standards have been established for such device under Section 514 of the Federal Food, Drug, and Cosmetic Act. However, the Duet™ System complies with the following voluntary standards:

  • IEC 60601-1-4 I
  • . ISO 14971:2012

5.12 Guidance

  • Guidance for Industry and FDA Staff Class II Special Controls Guidance ■ Document: Automated Fluorescence in situ Hybridization (FISH) Enumeration Systems, issued March 23, 2005.
  • Guidance for the Content of Premarket Submissions for Software Contained in 미 Medical Devices, issued May 11, 2005.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Bioview. The logo consists of the word "bioview" in black font, with the "bio" portion of the word connected to a three-lobed shape. The "b" is lowercase, while the rest of the word is in lowercase. The three-lobed shape is made up of three circles, with the top circle filled in black, the middle circle containing three concentric circles, and the bottom circle filled with a dotted pattern.

5.13 Substantial Equivalence

The Duet™ System with detection of the Vysis® ALK Break Apart FISH Probe Kit is as safe and effective as the previously cleared Duet™ System with detection of the Vysis® HER-2 DNA Probe Kit. The systems have similar intended use, technological characteristics, and principles of operation as its predicate device. The technological differences between the current Duet™ System and its predicate device which are listed in the table below raise no new issues of safety or effectiveness.

Performance data demonstrate that the current Duet™ System is as safe and effective as previously cleared Duet™ System. Thus, the Duet™ System is substantially equivalent.

A substantial equivalence table, which summarizes the differences between the current Duet™ system, and its predicate device, is provided below.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for "bioview". The logo consists of the word "bioview" in black font, with the "bio" portion of the word connected to a three-circle design. The three circles are arranged in a triangular shape, with the top circle filled with a black pattern, the middle circle containing a smaller circle inside, and the bottom circle filled with a light gray pattern.

CharacteristicBioView Ltd.Duet™ System (Current)BioView Ltd.Duet™ System (Predicate)
510k NumberK130775K061602
Product CodeJOY 864.5260, NTH 866.4700JOY 864.5260, NTH 866.4700
Intended useThe Duet™ System is an automated scanning microscope and image analysis system. It is intended for in vitro diagnostic use as an aiding tool to the pathologist in the detection, classification and counting of cells of interest based on color, intensity, size, pattern, and shape. The Duet™ System is intended to detect:• Hematopoietic cells stained by Giemsa stain, Immunohistochemistry or ISH (with bright field and fluorescent) prepared from cell suspension.• Amniotic cells stained by FISH (using direct labeled DNA probes for chromosomes X, Y, 13, 18 and 21).• Cells in urine specimens, stained by FISH (using the Vysis UroVysion™ Bladder Cancer Recurrence Kit for chromosomes 3, 7, 17 and 9p21 locus), from subjects with transitional cell carcinoma of the bladder.• Amplification of the HER-2/neu gene via fluorescence in situ hybridization (FISH) in human breast cancer tissue specimens, probed by the Vysis® PathVysion™ HER-2 DNA Probe Kit.• Detect qualitatively rearrangements involving the ALK gene via fluorescence in situ hybridization (FISH) in formalin-fixed paraffin-embedded (FFPE) non-small cell lung cancer (NSCLC) tissue specimens, probed with the Vysis ALK Break Apart FISH Probe Kit. The Duet™ is to be used as an adjunctive automated enumeration tool, in conjunction with manual visualization.The Duet™ System is an automated scanning microscope and image analysis system. It is intended for in vitro diagnostic use as an aiding tool to the pathologist in the detection, classification and counting of cells of interest based on color, intensity, size, pattern, and shape. The Duet™ System is intended to detect:• Hematopoietic cells stained by Giemsa stain, Immunohistochemistry or ISH (with bright field and fluorescent) prepared from cell suspension.• Amniotic cells stained by FISH (using direct labeled DNA probes for chromosomes X, Y, 13, 18 and 21).• Cells in urine specimens, stained by FISH (using the Vysis UroVysion™ Bladder Cancer Recurrence Kit for chromosomes 3, 7, 17 and 9p21 locus), from subjects with transitional cell carcinoma of the bladder.• Amplification of the HER-2/neu gene via fluorescence in situ hybridization (FISH) in human breast cancer tissue specimens, probed by the Vysis® PathVysion™ HER-2 DNA Probe Kit.
Software VersionDUET SW 3.5DUET SW 2.5
Specimen Typeformalin-fixed paraffin-embedded (FFPE) non-small cell Lung Cancer (NSCLC) tissue specimensformalin-fixed paraffin-embedded (FFPE) Breast cancer tissue specimens
CharacteristicBioView Ltd.Duet™ System (Current)BioView Ltd.Duet™ System (Predicate)
Probe kitVysis® ALK Break Apart FISH Probe KitVysis® PathVysion™ HER-2 DNA Probe Kit.
Slide CapacityUp to 200 SlidesUp to 8 slides
Device componentsAutomated microscopeAutomated microscope
Motorized stage with up to 200 slides using slide-loaderMotorized stage with up to 8-slides
Color or monochrome CCD cameraColor CCD camera
PCPC
Display: 22" high resolution' monitor , optional 22" touch-screen with Pen pointing deviceDisplay:17" high resolution LCD flat monitor

.

:

:

:

.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Bioview. The logo consists of the word "bioview" with the "b" and "i" in black circles. The "o" is a circle with a smaller circle inside, and the "view" is in plain text. The "b", "i", and "o" are connected by lines that form a shape resembling a molecule.

5.14 Performance Characteristics of the Duet™ System

A performance evaluation study was performed in order to evaluate the performance of the Duet™ System method to detect qualitatively rearrangements involving the ALK gene via fluorescence in situ hybridization in formalin-fixed paraffin-embedded non-small cell lung cancer tissue specimens, probed with the Vysis ALK Break Apart FISH Probe Kit, in terms of reproducibility and repeatability and its accuracy in comparison to the manual system.

A method comparison study was performed to provide evidence that all four previous indications cleared for the Duet™ system, can be used safely and effectively on the current system configuration that include different camera, optional slide loader and optional pen display.

Note: The performance of the Vysis ALK Break Apart Fish Probe Kit has been established with the Duet Version 3.5 configuration. Duet Version 2.5 was not validated for the ALK Break-apart test and it is not intended to perform this test.

The performance evaluation study report is comprised of the following studies, which are summarized below:

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a logo for a company called "bioview". The logo consists of the word "bioview" with the "bio" part stylized. The "b" is inside a solid black circle, and the "io" is inside a circle with a white background and a black outline. To the left of the word is a graphic of three circles connected together.

Study 1 – Precision/Reproducibility Performance - an evaluation of the performance of the Duet™ System in terms of reproducibility and repeatability, within a system and across systems.

Study 2 - Analytical Performance/Methods Comparison - a comparison of the Duet™ System method to manual scoring method.

Study 3 - Configurations method comparison - to verify that the four indications previously cleared for use with Duet version 2.5 can be used safely and effectively on the current system configuration (3.5) that contains a different camera, slide loader, and pen display.

Repeatability and Reproducibility Study 5.14.1

The purpose of this study was to evaluate the precision of the Duet™ System, for its performance with the ALK™ Kit, in terms of reproducibility and repeatability. For this purpose, the Duet™ System was evaluated in three parts, as described below:

  • Within-system and within-day I.
  • Within-system and between-day II.
  • Between-systems and sites. III.

Each of these studies were done using four (4) slides in each of the following categories of positive cells percentage : <10%, 10-25%, 25-50% and ≥ 50%. The study was conducted at three sites, and slides were prepared according to the probe manufacturer instructions.

Acceptance Criteria

The following acceptance criteria were defined:

  • Positive samples: CV < 25% .
  • Negative samples, with mean percentage below 4%: CV ≤180% .
  • Negative samples, with mean percentage above 4%: CV <70% .

Repeatability and Reproducibility on Binary Outcome Data (Positive/Negative final result) demonstrated 100% repeatability and reproducibility in all R&R studies.

Repeatability and Reproducibility results on coefficient of variation (CV), for the three studies performed are summarized in the following table. It includes %CV Range for Individuals Slides and %CV Overall, derived from the Random Model. Note that very high CV% is expected in FISH samples with low counts. For example, if one reading finds one or two positive cells while other two readings find zero positive cells, the resulting CV% is higher than 170%.

A panel of 16 archived clinical specimen slides (4 samples in each of 4 value ranges: <10%, 10-25%, 25-50% and >50%) were chosen to establish device within-run, between-day, and between-site variability

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for "bioview". The logo consists of the letters "b", "i", "o", and the word "view". The letters "b", "i", and "o" are inside of circles that are connected to each other. The circles are arranged in a triangular shape, with the "b" at the top left, the "i" at the top right, and the "o" at the bottom.

Within-run: Three runs for each of the panel members were performed on the same day.

CategorySlide IDMeanStandardDeviationCoefficient ofVariation (%)
<10%CYNK-402.0 %0.00.0
CYNK-633.3%1.234.6
CYNK-641.3 %1.286.6
CYNK-654.0 %2.050.0
10-25%CYNK-497.3%1.215.7
CYNK-5312.3%1.512.4
CYNK-556.7%1.217.3
CYNK-6716.0%1.06.3
25-50%BV Val 1348.7%9.219.0
BV Val 0750.7%8.116.1
BV Val 1134.0%3.08.8
CYNK-4130.3%4.013.3
>50%BV Val 0974.7%4.25.6
BV Val 1057.3%4.27.3
CYNK-3669.3%5.88.3
CYNK-5053.0%8.516.1

Between-day: Variability was assessed by assessing panel member performance of on three different days. The shortest between-day interval was five days.

CategorySlide IDMeanStandard DeviationCoefficient of Variation (%)
<10%CYNK-573.0 %5.2173.2
CYNK-632.0%2.0100.0
CYNK-640.7 %1.2173.2
CYNK-652.7 %1.243.3

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for "bioview". The logo consists of the letters "b", "i", "o", and the word "view". The letters "b", "i", and "o" are stylized and connected to form a unique design. The "b" is a filled circle, the "i" is a filled circle, and the "o" is a circle with two concentric circles inside.

CategorySlide IDMeanStandardDeviationCoefficient ofVariation (%)
10-25%CYNK-5314.0%0.00.0
CYNK-6715.7%1.27.4
CYNK-6910.3%3.534.0
CYNK-7010.7%2.119.5
25-50%BV Val 1356.7%3.15.4
BV Val 0751.0%5.610.9
BV Val 1140.0%10.827.0
CYNK-4729.3%3.110.4
>50%BV Val 0364.7%3.14.7
BV Val 0655.3%3.15.5
BV Val 0970.0%12.017.1
CYNK-5052.3%7.514.3

:

Site-to-Site: Reproducibility was validated by testing each slide three times, each at a different site and Duet system.

CategorySlide IDMeanStandard DeviationCoefficient of Variation (%)
<10%CYNK-574.0 %4.0100.0
CYNK-634.0%0.00.0
CYNK-643.7 %3.287.7
CYNK-653.3 %1.234.6
10-25%CYNK-5312.0%2.016.7
CYNK-5510.3%2.120.1
CYNK-6720.3%4.220.5
CYNK-699.7%0.66.0
25-50%BV Val 0744.0%1.73.9

5-11

.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for "bioview". The logo consists of the word "bioview" with the "b", "i", and "o" in black circles that are connected. The "view" part of the logo is in a standard font. The logo also has a design element that looks like a plus sign with circles at the end of each line.

CategorySlide IDMeanStandard DeviationCoefficient of Variation (%)
BV Val 1136.7%2.56.9
CYNK-4135.0%4.613.1
CYNK-4735.0%5.014.3
>50%BV Val 0658.7%3.15.2
BV Val 0976.7%8.310.9
BV Val 1055.3%2.34.2
BV Val 0279.3%12.916.2

Conclusions:

  • Overall %CV's derived from the Mixed Model were all well below the . performance goals specified
  • Of the 48 slides, each measured three (3) times, a single slide displayed a ● %CV above the 25% performance goal specified; %CV for the positive slide in the "Day" study was 27%, where the next highest %CV was 17.1%
  • For binary outcomes: .
    • Repeatability was 100% a.
    • b. Reproducibility was 100%

We thus conclude that the acceptance criteria were met and the Duet™ system is repeatable and reproducible.

Analytical Performance (Method Comparison) Study 5.14.2

The purpose of this study was to demonstrate the accuracy of the Duet™ System method for detection of rearrangements involving the ALK gene via fluorescence in situ hybridization (FISH) in formalin-fixed paraffin-embedded (FFPE) non-small cell lung cancer (NSCLC) tissue specimens, probed with the Vysis ALK Break Apart FISH Probe Kit, in comparison to the manual scoring method.

The study consisted of a total of 113 specimen slides. Out of these 113 slides, 32 slides were positive and 11 were equivocal.

The slides were prepared and probed using the ALK Kit, according to the manufacturer, while each slide was screened both manually and using the Duet™ system.

Acceptance Criteria

The objective of this study is to demonstrate that the Duet™ System is concordant with manual counting.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Bioview. The logo consists of the word "bioview" in a sans-serif font, with the "bio" portion of the word stylized with a connected set of three circles. The first circle, which contains the letter "b", is shaded with a white letter on a black background. The second circle, which contains the letter "i", is shaded with a black letter on a white background. The third circle, which contains the letter "o", is shaded with concentric circles.

The statistical analysis performed for the pooled results included levels of agreement with the manual. The statistical analysis presented in the following tables below demonstrates the accuracy of the Duet™ system as indicated by its high level of agreement in comparison to the manual method.

Manual scoring
Final Result-+Total
WithDuetMethod-81182
+03131
Total8132113

Method comparison - Summary of Pooled Results

Accuracy ParameterValueLower 95%CIUpper 95%CI
Percent Agreement with Manual Negative100%95.5%100%
Percent Agreement with Manual Positive96.9%84.3%99.5%
Percent Overall Agreement99.1%95.2%99.8%

Conclusion:

The test demonstrated that the acceptance criteria were met.

High correlation was found between the manual method and the results of the Duet™ System method, showing excellent accuracy in detecting positive and negative samples, as defined by the ALK Kit and the manual method. The overall agreement between both methods (the Duet™ and the manual) was 99.1% (95%, CI 95.2%-99.8%). No significant differences between the studies in the three sites were found, rendering the pooled analysis is valid.

Configurations Method Comparison Study 5.14.3

This study was intended to provide evidence that the changes in hardware configuration that includes different camera, optional slide loader and an optional pen display have not impacted performance with the indications cleared in previous 510(k) submissions.

The study was performed with the following guidelines:

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for "bioview". The logo consists of the word "bioview" with the "bio" part of the word stylized with circles. The "b" is inside a solid black circle, the "i" is inside a solid black circle, and the "o" is inside two concentric circles. The circles are connected by lines that form a three-way intersection, with the top circle being solid black and the bottom circle being gray.

  • The study included comparison of patient slides from each of the four ● previously cleared indications: Hematopoietic cells, Amniotic cells, Bladdercancer cells probed by Urovysion and Breast-cancer cells probed by Her-2 Path Vysion probe.
  • Same slides were evaluated on the cleared Duet™ configuration (version 2.5) . and by the current submission configuration (version 3.5).
  • Each of the previously cleared indications was tested using patient slides . samples. Samples were selected to cover the intended use of both Normal, Abnormal and near the medical decision/cut-off (as applicable). Clinical specimen were selected from each relevant sample type and from each scoring category.

Results summary:

  • The test sets included all required slides. .
  • Final interpretation of results was identical in all slides between new . configuration and cleared configuration.
  • The differences in numerical results between new configuration and cleared . configuration met the pre-defined acceptance criteria for each test, in all slides.

5.15 Final Conclusion

BioView Ltd. believes that the Duet™ System is substantially equivalent to the combination of its predicate devices in terms of Intended Use, Indications for Use, technological characteristics and mode of operation. Any differences between the Duet™ System and its predicate devices do not raise new safety or effectiveness issues, based on the performance results and the analysis of similarities and differences presented above.

{12}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 9, 2014

BIOVIEW LTD. ORLY MAOR 25A SIRKIN ST. KFAR SABA 4442156 ISRAEL

Re: K130775 Trade/Device Name: Duet™ System Regulation Number: 21 CFR 866.4700 Regulation Name: Automated fluorescence in situ hybridization (FISH) enumeration systems Regulatory Class: II Product Code: NTH Dated: May 3, 2014 Received: May 6, 2014

Dear Ms. Maor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 10 as batter in the May 28, 1976, the enactment date of the Medical Device Amendments, or to commerce provide thay 20, 2011 accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmoties for (1 corp too) the device, subject to the general controls provisions of the Act. The I ou may, morelors, manus of the Act include requirements for annual registration, listing of general controls pro revering practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability addition. Themse note. One of that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device is classified nal controls. Existing major regulations affecting your device can be fromd in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may round in the Code of I casiants concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please of advised hat I Dri 3 issuance or a cadesice complies with other requirements of the Act that FDA has made a determinations administered by other Federal agencies. You must of any rederal statutes and reguirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements inental device-related auvellas ovense (QS) regulation (21 CFR Part 820); and if applicable, the as set 1011 in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

{13}------------------------------------------------

Page 2-Ms. Maor

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Reena Romin -S

Reena Philip, PhD Director Division of Molecular Genetics and Pathology Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{14}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Expiration D
See PRA St

510(k) Number (if known) K130775

Device Name Duet™ System

Indications for Use (Describe)

The Duer™ System is an automated scanning microscope and imuge analysis system. It is intended for in-vitro diagnostic use as an aid to the pathologist in the detection, classification and counting of cells of interest based on color, intensity, size, pattern and shape.

The Duet™ System is intended to:

  1. Detect Hematopoietic cells stained by Giemsa stain, Immunohistochemistry or ISH (with brightfield and fluorescent) prepared from cell suspension.

  2. Detect Amniotic cells stained by FISH (using direct labeled DNA probes for chromosomes X. Y. 13, 18 and 21).

  3. Detect Aneuploidy for chromosomes 3,7. 17 and loss of the 9p21 locus via FISH in Urine specimens from subjects with transitional cell carcinoma of the bladder, probed by the Vysis Urovysion Bladder Cancer Kit.

  4. Detect and quantify chromosome 17 and the HER-2/neu gene via fluorescence in situ hybridization (FISH) in interphase nuclei from formalin-fixed, paraffin embedded human breast cancer tissue specimens, probed by the Vysis® Path Vysion™ HER-2 DNA Probe Kit. The Duet™ is to be used as an adjunctive automated enumeration tool, in conjunction with manual review of the digital image, to assist in determining HER-2/neu gene to chromosome 17 signal ratio.

  5. Qualitatively detect rearrangements involving the ALK gene via fluorescence in situ hybridization (FISH) in formalinfixed paraffin-embedded (FFPE) non-

ર-ર

small cell lung cancer (NSCLC) tissue specimens, probed with the Vysis ® ALK Break Apart FISH Probe Kit. The Duer™ is to be used as an adjunctive autonated enumeration with manual review of the digital image. Note: The pathologist should verify the image analysis software application score.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY . 1-15 Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/14/Picture/20 description: The image shows the text "Yun-fuHUAS" in a bold, sans-serif font. The letters "Yun-fu" are solid black, while the letters "HUAS" have a more intricate, patterned design. The text appears to be a title or heading, possibly related to a product or organization.

{15}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.2

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 866.4700 Automated fluorescence

in situ hybridization (FISH) enumeration systems.(a)
Identification. An automated FISH enumeration system is a device that consists of an automated scanning microscope, image analysis system, and customized software applications for FISH assays. This device is intended for in vitro diagnostic use with FISH assays as an aid in the detection, counting and classification of cells based on recognition of cellular color, size, and shape, and in the detection and enumeration of FISH signals in interphase nuclei of formalin-fixed, paraffin-embedded human tissue specimens.(b)
Classification. Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Automated Fluorescencein situ Hybridization (FISH) Enumeration Systems.” See § 866.1(e) for the availability of this guidance document.