K Number
K130664

Validate with FDA (Live)

Device Name
HYRDROMAX HMEF
Manufacturer
Date Cleared
2013-06-21

(101 days)

Product Code
Regulation Number
868.5260
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and / or expired gases is desired and to add and retain moisture in the circuit as required.

It is single patient use, disposable for patients with Tidal Volumes > 250 ml.

Duration of use < 24 hours.

Device Description

The HydroMax™ HMEF is standard configuration housing with a CO2 sampling port.

The common features are:

  • All have standard conical 15 mm / 22 mm fittings for connections ●
  • Female luer lock port for gas sampling for end-tidal CO2 .

The principle of operation:

  • The filter media is an electrostatic type of media and filters via electrostatic attraction .
  • . The HME media a porous foam that has hygroscopic properties to retain and release moisture from the patient
AI/ML Overview

The A-M Systems, Inc. HydroMax™ HMEF (Heat and Moisture Exchanger Filter) is a bacterial/viral filter intended for use with ventilators, anesthesia machines, and open flow systems to filter inspired and/or expired gases and retain moisture in the circuit. It is a single-patient use, disposable device for patients with tidal volumes greater than 250 ml and a duration of use of less than 24 hours. The device's performance was evaluated through bench testing.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the HydroMax™ HMEF are generally established by comparing its performance to a legally marketed predicate device, the GE (Engstrom) HMEF 1000 (K964204), and by meeting relevant international standards for medical devices (e.g., ISO 9360-1, ASTM F316-80). The "acceptance criteria" are implied by demonstrating substantial equivalence to the predicate and meeting the performance characteristics of the relevant standards.

Performance CharacteristicPredicate Device (GE (Engstrom) HMEF 1000) PerformanceAcceptance Criteria (Implied)Reported Device Performance (HydroMax™ HMEF)
Filtration Efficiency
Bacterial Filtration Efficiency (BFE)>99.9999%≥ 99.9999% (based on predicate)>99.99969% @ 30 lpm
Viral Filtration Efficiency (VFE)>99.99%≥ 99.99% (based on predicate)>99.9946% @ 30 lpm
Humidification Performance
Moisture Output (Vt 250 ml)N/A(Comparable to predicate/industry standard for HME)36.5 mg/L
Moisture Output (Vt 500 ml)33 mg/L≥ 33 mg/L (based on predicate)35.13 mg/L
Moisture Output (Vt 750 ml)32 mg/L≥ 32 mg/L (based on predicate)33.2 mg/L
Moisture Output (Vt 1000 ml)30 mg/L≥ 30 mg/L (based on predicate)32.53 mg/L
Moisture Loss (Vt 250 ml)N/A(Comparable to industry standard for HME)7.63 mg/L
Moisture Loss (Vt 500 ml)4.5 mg/L(Comparable to predicate/industry standard for HME)8.97 mg/L
Moisture Loss (Vt 750 ml)5.5 mg/L(Comparable to predicate/industry standard for HME)10.9 mg/L
Moisture Loss (Vt 1000 ml)7.5 mg/L(Comparable to predicate/industry standard for HME)11.57 mg/L
Respiratory Mechanics
Pressure Drop @ 30 lpm1.0 cmH2O≤ 1.0 cmH2O (based on predicate)0.7 cmH2O
Pressure Drop @ 60 lpm2.3 cmH2O≤ 2.3 cmH2O (based on predicate)2.0 cmH2O
Pressure Drop @ 90 lpmN/A(Comparable to industry standard for HME)3.5 cmH2O
Dead Space (Internal Volume)77 mlNot explicitly defined as 'criteria', but for comparison.56 ml
Structural Integrity & Safety
Filter Integrity (Bubble Point)N/A(Meet ASTM F316-80 requirements)244.43 microns
Housing Burst StrengthN/A(Meet safety requirements/industry standard)>10 psi < 20 psi
Leakage per ISO 9360-1 6.4N/A0 ml/min @ 70 cm H2O (per ISO standard)0 ml/min @ 70 cm H2O
Compliance per ISO 9360-1 6.5N/A(Meet ISO standard requirements)1.1 ml/kPa
Other
Standard 22/15 mm connectionsYesYes (per ISO 5356-1)Yes
Luer port for gas sampling (ISO 594-2)YesYes (per ISO 594-2)Yes
Material BiocompatibilityStandard materials to ISO 10993-1Standard materials to ISO 10993-1Standard materials to ISO 10993-1

Note: "N/A" for the predicate indicates the information was not provided in this specific comparison table, not necessarily that the predicate does not meet these parameters. The "Acceptance Criteria (Implied)" are derived from the direct comparison to the predicate device and the referenced ISO/ASTM standards.

2. Sample size used for the test set and the data provenance

The document does not explicitly state the sample sizes for each specific test (e.g., how many filters were tested for BFE, VFE, pressure drop, etc.). It refers to "a number of bench tests."

The data provenance is from bench testing performed by Nelson Laboratories (for Bacterial/Viral filtration efficiency) and potentially other internal or external labs for other physical characteristics.
The data is retrospective in the sense that the device was manufactured, and then tested. There is no information on the geographical origin of the data beyond "Nelson Laboratories" and "bench testing."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This section is not applicable as the study described is a non-clinical bench testing study of a medical device, not a human reader study or a diagnostic accuracy study involving human expert interpretation. The "ground truth" for these tests is established by standardized testing methodologies and instrument readings (e.g., filtration efficiency is determined by particle count, pressure drop by manometry).

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This section is not applicable as the study is a non-clinical bench testing study. Adjudication methods like 2+1 or 3+1 are used in clinical studies or reader studies where human interpreters are involved in making decisions or interpretations that need to be reconciled.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable as the study described is a non-clinical bench testing study for a physical medical device (a filter), not an AI-powered diagnostic tool. Therefore, there is no human reader component or AI assistance to evaluate.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This section is not applicable as the device is a physical medical device (filter), not an algorithm or AI product. The "standalone performance" refers to the device's inherent physical characteristics as measured by the bench tests.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc.)

For this non-clinical bench testing study, the "ground truth" for the device's performance characteristics is established by:

  • Standardized laboratory measurements: Using calibrated equipment and established protocols (e.g., ISO 9360-1, ASTM F316-80) to measure physical properties like filtration efficiency, pressure drop, moisture output, etc.
  • Predicate device performance: The performance of the legally marketed predicate device serves as a benchmark for substantial equivalence.

8. The sample size for the training set

This section is not applicable as this is a non-clinical bench testing study of a physical medical device, not a machine learning or AI model. Therefore, there is no "training set."

9. How the ground truth for the training set was established

This section is not applicable for the same reasons as #8.

{0}------------------------------------------------

510(k) Summary Page 1 of 5 18-Jun-13

JUN 2 1 2013

A-M Systems, Inc.131 Business Park LoopSequim, WA 98382Tel - 360-683-8300Fax - 360-683-3525
Official Contact:David Green, Product Manager
Proprietary or Trade Name:HydroMax™ HMEF
Common/Usual Name:Bacterial / Viral Filter
Classification Name:Filter, Bacterial, Breathing Circuit,CAH - 21 CFR 868.5260Class 2
Predicate Devices:GE (Engstrom) HMEF 1000 - K964204

Device Description

The HydroMax™ HMEF is standard configuration housing with a CO2 sampling port.

The common features are:

  • All have standard conical 15 mm / 22 mm fittings for connections ●
  • Female luer lock port for gas sampling for end-tidal CO2 .

The principle of operation:

  • The filter media is an electrostatic type of media and filters via electrostatic attraction .
  • . The HME media a porous foam that has hygroscopic properties to retain and release moisture from the patient

Indications for Use

For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and / or expired gases is desired and to add and retain moisture in the circuit as required.

It is single patient use, disposable for patients with Tidal Volumes > 250 ml.

Duration of use < 24 hours.

Environment of Use

Hospital, sub-acute care, and home settings

{1}------------------------------------------------

510(k) Summary Page 2 of 5 18-Jun-13

Predicate Device Comparison:

The proposed HydroMAX™ HMEF is similar in all respects to the predicate device, GE (Engstrom) HMEF 1000-K964204.

Comparative Table 5.1 discusses the major features of the proposed device and the legally marketed predicate device.

The HydroMAX™ HMEF is viewed as substantially equivalent to the predicate device based upon the following:

Indications -

  • Intended to be placed in circuit for filtration and humidification of inspired and expired . gases - GE (Engstrom) HMEF 1000 - K964204.
    Discussion - The indications for use are identical for the proposed and predicate device.

Technology -

  • The filter and HME media are standard materials and the fundamental principle of ● operation are identical to the predicate.
    Discussion - There are no differences in technology between the HydroMAX™ HMEF and the predicate GE (Engstrom) HMEF 1000 - K964204

Environment of Use -

  • Home, Hospitals, Sub-acute institutions .
    Discussion - The environment of use are identical between the HydroMAX™ HMEF and the predicate GE (Engstrom) HMEF 1000 - K964204

{2}------------------------------------------------

510(k) Summary

.

Table 1 - Comparison to the Predicate

FeaturesPredicateGE (Engstrom) HMEF 1000K964204ProposedHydroMAX™ HMEF
Indications for useFor use with ventilators, anesthesia machines and open flow systems where filtration of inspired and / or expired gases is required.For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and / or expired gases is required.
Patient Use / Duration if useSingle patient use, disposable, < 24 hoursSingle patient use, disposable, < 24 hours
Environment of UseHospital, Sub-acute Institutions, HomeHospital, Sub-acute Institutions, Home
Patient PopulationTidal volume > 300 mlTidal volume > 250 ml
ContraindicationsNoneNone
Standard 22 / 15 mm connectionsYesYes
Per ISO 5356-1YesYes
Luer port for gas samplingPer ISO 594-2YesYes
Various configurationsStraight, angledStraight
Filtration methodElectrostaticElectrostatic
Placement within the circuitPatient-sidePatient-side
Weight (gm)24 gm21 gm
Internal volume/ Dead space per ISO 9360-177ml56 ml
Bacterial / Viral filtration efficiency (Nelson Labs)BFE->99.9999%VFE->99.99%BFE->99.99969%VFE->99.9946%
Moisture Output (mg/L)Vt 250 - N/AVt 500 - 33Vt 750 - 32Vt 1000 - 30Vt 250 - 36.5Vt 500 - 35.13Vt 750 - 33.2Vt 1000 - 32.53.
Moisture Loss (mg/L)Vt 250 - N/AVt 500 - 4.5Vt 750 - 5.5Vt 1000 - 7.5Vt 250 - 7.63Vt 500 - 8.97Vt 750 - 10.9Vt 1000 - 11.57
FeaturesPredicateGE (Engstrom) HMEF 1000K964204ProposedHydroMAXTM HME/F
Pressure Drop1.0 cmH2O @ 30 lpm2.3 cmH2O @ 60 lpmN/A0.7 cmH2O @ 30 lpm2.0 cmH2O @ 60 lpm3.5 cmH2O @ 90 lpm
Filter integrityper ASTM F316-80N/A244.43 microns
Housing burst strengthN/A>10 psi < 20 psi
Leakage per ISO 9360-1 6.4N/A0 ml/min @ 70 cm H2O
Compliance per ISO 9360-1 6.5N/A1.1 ml/kPa
MaterialsStandard materials to ISO 10993-1Standard materials to ISO 10993-1
StandardsISO 9360-1 HMEISO 9360-1 HME

Page 3

{3}------------------------------------------------

510(k) Summary Page 4 of 5 18-Jun-13

Non-clinical Testing Summary -

Performance -- Bench Testing

We performed a number of bench tests pre- and post-conditioning these included the results are summarized in Table 2 below.

  • . Filtration efficiency
    • Bacterial / Viral filtration efficiency per Nelson Laboratories o
    • Pre - and Post - conditioning performance
    • Includes Pressure Drop per ISO 9360-1 o
    • Dead space per ISO 9360-1 ಂ
  • Leakage test per ISO 9360-1
  • . Filter integrity per ASTM F316-80
  • Housing burst strength .
  • . Environmental and Mechanical Testing

{4}------------------------------------------------

510(k) Summary Page 5 of 5 18-Jun-13

Table 2 - Summary of Non-clinical Testing

Performance CharacteristicsResults
Pressure Drop (cmH2O)0.7 cmH2O @ 30 lpm2.0 cmH2O @ 60 lpm3.5 cmH2O @ 90 lpm
Filter Integrity (Bubble Point) Test244.43 microns
Dead Space (Internal CompressibleVolume in ml)56 ml
Leakage per ISO 9360-1 6.40 ml/min @ 70 cm H2O
Compliance per ISO 9360-1 6.51 ml/kPa
Housing Burst Pressure (psig)>10 psi < 20 psi
Mechanical Vibration and ShockPressure Drop2.0 cm H2O @ 60 lpm
Storage - High and Low Temperatureand HumidityPressure Drop2.0 cm H2O @ 60 lpm
Bacterial Filtration Efficiency (BFE) %>99.99969% @ 30 lpm
Viral Filtration Efficiency (VFE) %>99.9946% @ 30 lpm

Material - Biocompatibility

  • The materials are identical to predicates that have the identical intended use. .
  • . This device would be considered -

.

  • External communicating / Tissue / mucosal / limited duration (< 24 hr) o
  • o However based on the potential of accumulative exposure the device would be consider permanent duration contact.

Discussion: We are utilizing identical materials and provide Material Certifications.

Substantial Equivalence Conclusion -

The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to be substantially equivalent.

Page 5

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 21, 2013

A-M Systems, Inc. c/o Mr. Paul Dryden President ProMedic, Inc. 24301 Woodsage Drive Bonita Springs, FL 34134

Re: K130664

Trade/Device Name: HydroMax™ HMEF Regulation Number: 21 CFR 868.5260 Regulation Name: Breathing circuit bacterial filter Regulatory Class: II Product Code: CAH Dated: May 14, 2013 Received: May 15, 2013

Dear Mr. Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warrantics. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 - Mr. Dryden

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOfTices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.lda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Image /page/6/Picture/6 description: The image shows a signature and contact information. The text "Sincerely yours," appears at the top of the image. Below the signature, the name "Tejashri Purohit-Sheth, M.D." is written, followed by "Clinical Deputy Director" and "DAGRID."

Kwame Ulmer, M.S. Acting Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use Statement

Page I of I

510(k) Number; K130664

Device Name: HydroMax™ HMEF

Indications for Use:

For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and / or expired gases is desired and to add and retain moisture in the circuit as required.

It is single patient use, disposable for patients with Tidal Volumes > 250 ml.

Duration of use < 24 hours.

Prescription Use XX (Part 21 CFR 801 Subpart D) or

Over-the-counter use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Digitally signed by James J. LeeON: c=US, o=U.S. Government, ou=HHS,ou=FDA-People, cn=James J. Lee,0.9.2342.19200300.100.1.1=2000954859Date: 2013.06.19 13:58:46 -04'00'ActingforDr. AHarry
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

(Division Sign-Off)

Division of Anesthesiology, General Hospital

Infection Control, Dental Devices

510(k) Number:K130664
-------------------------

ng BC Anya y

PDF PAGE 39 of 133

Page 04.2

§ 868.5260 Breathing circuit bacterial filter.

(a)
Identification. A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.(b)
Classification. Class II (performance standards).