K Number
K130617

Validate with FDA (Live)

Device Name
RAYSTATION
Date Cleared
2013-05-10

(64 days)

Product Code
Regulation Number
892.5050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plaus provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.

The system functionality can be configured based on user needs.

The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.

Device Description

RayStation 3.5 is a treatment planning system. i.e. a software program for planning and analysis of radiation therapy plans. Typically, a treatment plan is created by importing patient from a CT scanner, defining regions of interest either manually or semi-automatically, deciding on a realment setup and objectives, optimizing the treatment parameters, comparing rival plans to find the best compromise, computing the clinical dose distribution, approving the plan and exporting it.

AI/ML Overview

This 510(k) submission for RayStation 3.5 focuses on demonstrating substantial equivalence to predicate devices (RayStation 2.5, DCM 2.0, and Proton Vision 7.0) rather than establishing specific acceptance criteria and proving performance against them in a traditional sense for a novel AI device. The document primarily describes the device's functionality and its verification through testing against its own internal specifications, which were developed from the predicate devices' specifications.

Here's an analysis based on the provided text, addressing the requested information to the best extent possible given the nature of the submission:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not outline explicit, quantitative acceptance criteria for device performance in a table format that would typically be seen for an AI diagnostic device. Instead, the "acceptance criteria" are implied by the verification and validation (V&V) activities ensuring the device performs as well as or substantially equivalent to its predicate devices.

The reported device performance is described qualitatively as "successful verification and validation" and that "RayStation 3.5 is as safe and effective, and performs as well as the predicate devices."

Feature/Functionality VerifiedAcceptance Criteria (Implicit)Reported Device Performance
General Technology (RayStation 3.5 vs. RayStation 2.5)Performs as safely, effectively, and with the same basic design and requirements as RayStation 2.5."Successful verification and validation of RayStation 3.5 therefore support the substantial equivalence of the two RayStation versions." "RayStation 3.5 is as safe and effective, and performs as well as the predicate devices."
Electron Planning (RayStation 3.5 vs. DCM 2.0)Performs calculations using Voxel Monte Carlo (VMC++) algorithm with electron beam model, supporting treatment planning with electron energies from 4 to 25 MeV, and calculation of asymmetric and non-coplanar fields, equivalent to DCM 2.0."The verification performed for electron planning verifies the functionality for... Treatment planning with electron energies from 4 to 25 MeV... Dose calculation for 3D radiotherapy treatment with combined modality plans... Calculation of asymmetric and non-coplanar fields. This is the same functionality as included in the predicate device DCM 2.0. The test results thereby support a determination of substantial equivalence."
Proton Planning (RayStation 3.5 vs. Proton Vision 7.0)Calculates and displays prospective or verification treatment plans for proton therapy, provides tools for proton energy range estimation, proton dose calculation, and dosimetric plan review, equivalent to Proton Vision 7.0."The verification performed for proton planning verifies the functionality for... Proton planning treatment plan calculation... Proton energy range estimation... Proton dose calculation. This is the same functionality as included in the predicate device Proton Vision 7.0. The test results thereby support a determination of substantial equivalence."

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size (e.g., number of patient cases) used for the test set. It refers generally to "verification" and "test results" and that "The test specification of RayStation 3.5 is a further developed version of the test specification of RayStation 2.5."

There is no information provided regarding data provenance (e.g., country of origin, retrospective or prospective nature).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not mention the use of experts to establish ground truth for a test set. The verification seems to be against the expected output of the algorithms and comparison to the predicate devices' established performance, rather than against an expert-derived ground truth.

4. Adjudication Method for the Test Set

As there's no mention of experts establishing ground truth, there is no adjudication method (e.g., 2+1, 3+1) described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No MRMC comparative effectiveness study was mentioned in the provided text. The submission focuses on the technical equivalence of the software's functionality, not on human reader performance with or without the device.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the studies described are standalone performance evaluations of the algorithms and functionalities within RayStation 3.5. The verification reports on the system's ability to perform calculations (dose, energy range, plan generation) and display plans correctly. The overall device is a "software program for planning and analysis of radiation therapy," implying an algorithm-only evaluation for its core functions.

7. The Type of Ground Truth Used

The ground truth implicitly used for the verification tests is the expected output of the algorithms based on established physics models (Voxel Monte Carlo for electrons, and general proton dose calculation models) and the functionality of the predicate devices. It's not expert consensus, pathology, or outcomes data, but rather adherence to computational accuracy and functional specifications derived from existing, cleared devices.

8. The Sample Size for the Training Set

The document does not mention any training set or machine learning components in the traditional sense that would require a training set. The device is described as performing calculations based on algorithms (e.g., VMC++) and models, not learning from data in the way a modern AI algorithm might. Therefore, a "training set sample size" is not applicable based on the provided information.

9. How the Ground Truth for the Training Set Was Established

Since there is no mention of a training set, the establishment of ground truth for it is not applicable. The algorithms used (like VMC++) are based on physics principles, not learned from labeled data.

{0}------------------------------------------------

K130617

MAY 1 0 2013

Title:Document ID:Version:
510(k) Submission RayStation 3.5RSL-D-61-041.0

5. 510(k) Summary

5.1 510(k) owner

RaySearch Laboratories AB Sveavägen 25. plan 9 111 34 Stockholm Sweden

liel: +46 (8) 54506130 +46 (8) 54506139 Fax:

5.2 Contact person

David Hedfors, Quality and Regulatory Affairs Manager

5.3 Preparation date

February 22, 2013

5.4 Trade name

RayStation

Trade name and version number are written together. i.e. "RavStation 3.5" to easily distinguish the submitted device from the predicate device RayStation 2.5.

5.5 Common name

Radiation treatment planning system

5.6 Classification name

Medical charged-particle radiation therapy system (21 CFR 892.5050, Product Code MUJ)

5.7 Predicate devices

RayStation 2.5K120387
DCM 2.0K020971
Proton Vision 7.0K000922

5.8 Device description

RayStation 3.5 is a treatment planning system. i.e. a software program for planning and analysis of radiation therapy plans. Typically, a treatment plan is created by importing patient from a CT scanner, defining regions of interest either manually or semi-automatically, deciding on a realment setup and objectives, optimizing the treatment parameters, comparing rival plans to find the best compromise, computing the clinical dose distribution, approving the plan and exporting it.

The main workflow, creating a treatment plan from imported patient image data, is described below;

Flow of Events
UserSystem
1. The user launches RayStation 3.5
2. The user imports a patient with CT images through DICOM3. The system imports the data and checks data consistency
4. The user enters the Structure Definition module and creates ROIs using the contouring tools5. The system adds the ROIs to the patient structure set
6. The user enters the Treatment Specification module and creates a plan and a treatment setup with specified

{1}------------------------------------------------

Title:Document ID:Version
510(k) Submission RayStation 3.5RSL-D-61-041.0
7.machine, treatment energy and delivery typeThe user specifies beam configuration including isocenter, dose grid and fluence grid resolution8.The system adds the plan and treatment setup to the patient
9.The user enters the Plan Optimization module and creates an optimization problem
10.The user defines the algorithm and delivery settings and starts the optimization
11.The system generates a deliverable treatment plan
12.The system displays the plan in a number of user configurable displays
13.The user reviews the plan
14.The user enters the Plan Approval module and evaluates the plan
15.The user approves and exports the plan together with dose, structure sets and images
16.The system exports the plan and patient data in DICOM format

5.9 Intended use

.

RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.

The system functionality can be configured based on user needs.

The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.

The intended use for RayStation 3.5 is the same as for the predicate device RayStation 2.5.

5.10 Technological characteristics summary

5.10.1 General technology

The technological characteristics are the same for RayStation 3.5 as for the predicate device RayStation 2.5.

Comparing RayStation 3.5 with RayStation 2.5, the newer version includes usability, performance and connectivity improvements. Both versions are built on the same software platform and share basic design. Both versions have been developed under the same quality system meeting the same requirements for safety and effectiveness.

5.10.2 Electron planning

The RayStation 3.5 electron planning technology is substantially equivalent to the predicate device DCM 2.0 electron planning.

Comparing RayStation 3.5 with DCM 2.0. both systems perform calculations using the Voxel Monte Carlo (VMC++) algorithm, supported by an electron beam model.

5.10.3 Proton planning

The RayStation 3.5 proton planning technology is substantially equivalent to the predicate device Proton Vision 7.0 proton planning.

{2}------------------------------------------------

Title:Document ID:Version:
510(k) Submission RayStation 3.5RSL-D-61-041.0

Comparing RayStation 3.5 and Proton Vision 7.0, both systems calculate and display prospective or verification treatment plans for patients undergoing a course of proton therapy. Both systems provide tools for proton energy range estimation, proton dose calculation and dosimetric plan review.

5.11 Assessment of non-clinical performance data

5.11.1 General technology

The test specification of RayStation 3.5 is a further developed version of the test specification of RayStation 2.5. This is supported by the requirements specification, for which the same is true. The successful verification and validation of RayStation 3.5 therefore support the substantial equivalence of the two RayStation versions.

5.11.2 Electron planning

The verification performed for electron planning verifies the functionality for

  • Treatment planning with electron energies from 4 to 25 MeV. . This is a small extension of the energy range of the predicate device. which ranges from 6 to 25 McV. The extension does not require a new or modified algorithm or technology.
  • Dose calculation for 3D radiotherapy treatment with combined modality plans. .
  • Calculation of asymmetric and non-coplanar fields.

This is the same functionality as included in the predicate device DCM 2.0. The test results thereby support a determination of substantial equivalence.

5.11.3 Proton planning

The verification performed for proton planning verifies the functionality for

  • Proton planning treatment plan calculation .
  • Proton energy range estimation .
  • Proton dose calculation .

This is the same functionality as included in the predicate device Proton Vision 7.0. The test results thereby support a determination of substantial equivalence.

5.12 Test conclusion

The summary of the performed non-clinical tests shows that RayStation 3.5 is as safe and effective, and performs as well as the predicate devices.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 10, 2013

RaySearch Laboratories AB % Mr. David Hedfors Quality and Regulatory Affairs Manager Sveavägen 25, Plan 9 111 34 STOCKHOLM SWEDEN

· Re: K130617

Trade/Device Name: RayStation Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ Dated: March 4, 2013 Received: April 4, 2013

Dear Mr. Hedfors:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Hedfors

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm1.

Sincerely yours,

Michael D. O'Hara

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

for

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K130617

Device Name: RayStation

Indications for Use:

RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plaus provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.

The system functionality can be configured based on user needs.

The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Machiel D. O'Hara

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k) K130617

Page 1 of 1

N/A