K Number
K130565

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-11-27

(268 days)

Product Code
Regulation Number
886.1850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nidek Specular Microscope CEM-530 is a non-contact ophthalmic microscope, optical pachymeter, and camera intended for examination of the corneal endothelium and for measurement of the thickness of the cornea.

Device Description

The Nidek Specular Microscope CEM-530 provides non-contact, high magnification image capture of the endothelium enabling observation of the size and shape of cells. Information such as the number of endothelial cells, cell density, and cell area is analyzed through the captured images. The captured images and analysis results of the endothelium are used in intraocular or corneal surgery, postoperative follow-up, and corneal observation such as for endothelial disorders or the corneal state of patients who wear extended-wear contact lenses. Observation is possible in the central area (visual angle: 5°) and peripheral area (visual angle: 27°) using a periphery capture function as well as in the center of the captured images and analysis results can be printed on the built-in printer or optional video printer, or output to an external device over LAN connection. In addition to the specular microscopy, the corneal thickness can be optically measured in a non-contact method. The CEM-530 has auto-tracking and auto-shooting functions. Results can be printed using the the built-in thermal printer or captured images can be transferred to a filing system via LAN connection.

AI/ML Overview

The Nidek Specular Microscope CEM-530 performed a clinical study to demonstrate substantial equivalence to the predicate device Konan Medical, Inc. Cellchek Plus (K120264).

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria were not explicitly stated as numerical thresholds in the provided text. Instead, the study aimed to demonstrate "agreement, accuracy, and precision" between the Nidek CEM-530 and the predicate device, with results being "acceptable" and "substantially equivalent." The key metrics evaluated were:

VariablePerformance of Nidek CEM-530 (Mean)Performance of Konan CELLCHEK XL (PLUS) (Mean)Mean Difference (CEM-530 - Predicate)Mean Difference as a % of Predicate ReadingCorrelation (R²) vs. Predicate95% Limits of Agreement (LOA)Repeatability Ratio (CEM-530/Predicate)Reproducibility Ratio (CEM-530/Predicate)
Endothelial Cell Density (CD)2482.6 cells/mm²2553.1 cells/mm²-70.5 cells/mm²-0.42%0.9654(-406.3, 265.2)1.21021.1887
Coefficient of Variation Endothelial Cell Area (CV)29.9%31.2%-1.3%0.77%-0.0088(-18.5, 15.9)0.87461.0016
% Hexagonality69.0%59.3%9.7%18.19%0.2036(-7.2, 26.5)0.75860.7466
Central Corneal Thickness (CCT)551.5 microns565.2 microns-13.8 microns-2.38%0.8856(-53.0, 25.5)0.2634N/A (Not provided for CCT)

Summary of Device Performance against Implicit Acceptance Criteria:

  • Agreement: The study reported that "Agreement of the measurements with the CEM-530 and the predicate device was found to be acceptable." Bland-Altman plots generally showed LOAs including 0, indicating no overwhelming systematic bias.
  • Accuracy: Differences were generally small (3-5% for CD, CV, CCT) except for % Hexagonality (15%). Deming regression plots showed good agreement for CD, CCT, and CV (especially after outlier removal for CV).
  • Precision: "Overall, within eye/subject variability was acceptable, and similar for both machines." Repeatability was notably better for Nidek CEM-530 for CCT, and comparable for other endpoints.
  • Substantial Equivalence: The overall conclusion was that the Nidek CEM-530 is "substantially equivalent" to the predicate device based on these performance metrics.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Agreement Assessment: 74 eyes
    • 24 non-pathologic young eyes (18-28 years of age)
    • 25 non-pathologic adult eyes (29-80 years of age)
    • 25 pathologic adult eyes (29-80 years of age)
  • Sample Size for Precision Assessment: 62 evaluable eyes
    • 20 non-pathologic young eyes
    • 22 non-pathologic adult eyes
    • 20 pathologic adult eyes
  • Data Provenance: The study was a prospective clinical study. The country of origin of the data is not explicitly stated, but the sponsor and manufacturer are NIDEK Co., Ltd. in Japan, and the contact person is in Andover, MA, USA. It's common for such studies to be conducted in multiple locations, but this information is not provided.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not specify the number of experts used or their qualifications for establishing the ground truth. The "ground truth" for this study was established by comparing the Nidek CEM-530's automatic measurements against the predicate device (Konan Cellchek Plus)'s manual measurements. It is implied that the manual measurements from the predicate device serve as the reference standard.

4. Adjudication Method for the Test Set

The document does not explicitly describe an adjudication method for the test set. The comparison was primarily between the automated measurements of the Nidek CEM-530 and the (presumably expert-performed) manual measurements from the predicate device.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed to assess the improvement of human readers with AI vs. without AI assistance. This study focused on the performance of a new automated device (Nidek CEM-530) compared to an existing predicate device (Konan Cellchek Plus), which primarily involved a comparison of device measurements rather than human reader performance with or without AI assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Yes, a standalone performance study was done. The Nidek Specular Microscope CEM-530 is described as having "auto-tracking and auto-shooting functions" and "automatic image analysis." The clinical study reported on the "agreement, accuracy and precision of the CEM-530 by comparing results across three machines/operators to those obtained with the predicate device," and specifically "evaluates the precision and accuracy of the automatic analyses performed by the CEM-530 compared to manual measurements performed with the predicate device." This indicates that the CEM-530's automatic analysis was tested in a standalone capacity. An additional analysis was performed on 24 images comparing the automated method to a manual method on the same image.

7. Type of Ground Truth Used

The primary "ground truth" used for comparison was the manual measurements performed with the predicate device, the Konan Cellchek Plus. For the additional analysis on 24 images, the "manual method of analysis on the same image" served as the comparison for the automated method.

8. Sample Size for the Training Set

The document does not provide information regarding a separate training set or its sample size. The clinical study was conducted to assess the performance of the already developed CEM-530 device.

9. How the Ground Truth for the Training Set Was Established

Since information about a specific training set or its sample size is not provided, the method for establishing ground truth for a training set is not available in the document.

{0}------------------------------------------------

K130565

510(k) Summary

This summary of the 510(k) premarket notification for the NIDEK Specular Microscope CEM-530 is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR S 807.92.

Date Prepared: November 27, 2013

SPONSER/ 510(k) OWNER/ MANUFACTURER

NIDEK Co., Ltd. 34-14 Maehama, Hiroishi-cho, Gamagori, Aichi, 443-0038 Japan Telephone: +81-533-67-8901 Facsimile: +81-533-67-6628 E mail: yoneji mizuno@nidek.co.jp Establishment Registration Number: 8030392

NOV 27 2013

CONTACT PERSON

Aron Shapiro Ora, Inc. 300 Brickstone Square Andover, MA 01810 Telephone: (978) 332-9443 Facsimile: (978) 689-0020 E-mail: ashapiro@oraclinical.com

NAME OF DEVICE

Trade Name: CEM-530 Common Name: Specular Microscope

DEVICE CLASSIFICATION/FDA REVIEWING BRANCH

The Ophthalmic Branch has classified AC Powered Slit Lamp Biomicroscopes as Class II devices pursuant to 21 C.F.R. §886.1850.

PRODUCT CODE: CLASSIFICATION / CFR TITLE NQE, 21 CFR 886.1850

PREDICATE DEVICE

Konan Medical, Inc. Cellchek Plus (K120264)

{1}------------------------------------------------

INDICATIONS FOR USE

The Nidek Specular Microscope CEM-530 is a non-contact ophthalmic microscope, optical pachymeter, and camera intended for examination of the corneal endothelium and for measurement of the thickness of the cornea.

PRODUCT DESCRIPTION

The Nidek Specular Microscope CEM-530 provides non-contact, high magnification image capture of the endothelium enabling observation of the size and shape of cells. Information such as the number of endothelial cells, cell density, and cell area is analyzed through the captured images. The captured images and analysis results of the endothelium are used in intraocular or corneal surgery, postoperative follow-up, and corneal observation such as for endothelial disorders or the corneal state of patients who wear extended-wear contact lenses. Observation is possible in the central area (visual angle: 5°) and peripheral area (visual angle: 27°) using a periphery capture function as well as in the center of the captured images and analysis results can be printed on the built-in printer or optional video printer, or output to an external device over LAN connection. In addition to the specular microscopy, the corneal thickness can be optically measured in a non-contact method. The CEM-530 has auto-tracking and auto-shooting functions. Results can be printed using the the built-in thermal printer or captured images can be transferred to a filing system via LAN connection.

SUBSTANTIAL EQUIVALENCE

The Specular Microscope CEM-530 is substantially equivalent to the Konan Medical, Inc. Cellchek Plus (K120264). The Specular Microscope CEM-530 has the same intended use and indications for use, technological characteristics, and principles of operation as the previously cleared predicate device. Clinical performance data is provided which demonstrates that the CEM-530 is substantially equivalent to the Konan predicate device.

The CEM-530 and the predicate device are both non-contact ophthalmic microscopes, optical pachymeters, and cameras intended for examination of the corneal endothelium and for measurement of the thickness of the cornea. Both the CEM-530 and the predicate device offer automatic capture features and manual capture modes. Both the CEM-530 and the predicate device have a built-in CCD camera. Slight differences in flash, illumination for focusing and fixation lamps were evaluated in terms of light safety and found to meet the requirements of ISO 15004-2.

Both the CEM-530 and the predicate device include an optical pachymeter with an accuracy of ± l 0 microns. 、

Regarding image analysis, both the CEM-530 and the predicate device offer automatic image analysis while the predicate device also offers manual analysis of images. Clinical performance data is provided which evaluates the precision and accuracy of the automatic analyses performed

{2}------------------------------------------------

by the CEM-530 compared to manual measurements performed with the predicate device. The clinical performance data demonstrates the substantial equivalence of the CEM-530 automatic measurement mode to the predicate device's manual mode.

Both the CEM-S30 and the predicate device comply with applicable electrical safety and light safety standards.

NON-CLINICAL PERFORMANCE SUMMARY

The performance testing conducted using the NIDEK Specular Microscope CEM-530 verified that the device operates as intended. The specifications to which the CEM-530 was verified to are substantially equivalent to the predicate devices and therefore, support a determination of substantial equivalence. The pachymetry functionality was evaluated in model eyes and the measurement accuracy of ± 10 microns was confirmed.

Additionally, the CEM-530 was subjected to electrical safety testing in accordance with IEC 60601-1, electromagnetic compatibility (EMC) testing in accordance with IEC 60601-1-2, and optical radiation safety testing in accordance with ISO 15004-1 and ISO 15004-2.

CLINICAL PERFORMANCE SUMMARY

A prospective clinical study was conducted to assess the agreement, accuracy and precision of the CEM-530 by comparing results across three machines/operators to those obtained with the predicate device, the Cellchek Plus. Three populations were studied: young (18-28 years of age) and adult (29-80 years of age) healthy subjects and pathologic adult eves (29-80 years of age).

A total of 74 eyes were evaluated (24 non-pathologic young eyes, 25 non-pathologic adult eves, and 25 pathologic adult eyes) for the assessment of device agreement and 62 evaluable eyes (20 non-pathologic young eyes, 22 non-pathologic adult eyes, and 20 pathologic adults eyes) for the assessment of device precision.

All evaluable study eyes (74 eyes) were included in the assessment of device agreement. The differences were on the order of 3-5% of the overall mean value for endothelial cell density, coefficient of variation of endothelial cell area, and central corneal thickness measurements and approximately 15% of the overall mean for % hexagonality. All of the 95% Limits of Agreement (LOAs) included 0 and the majority was well centered around 0. However, for % hexagonality, the measurements for the CEM-530 device were generally higher than those from the Cellchek Plus machine. The correlation coefficients were generally high for the endothelial cell density and central corneal thickness measurements, indicative of strong linear relationships; but they were low for coefficient of variation of endothelial cell area and % hexagonality.

The mean differences for endothelial cell density are illustrated on the Bland Altman plot (Figure i. The greatest absolute differences between the two machines were seen at the extremes of the measurement range (Figure 2). The Deming regression line (Figure 3) showed good agreement between the devices. Table I provides a summary of the agreement data for all subjects.

{3}------------------------------------------------

Endothelial Cell Density Figure 1: Bland-Altman Plot- Observed Data-Endothelial Cell Density- All Subjects, Effectiveness Population

Image /page/3/Figure/1 description: This image is a scatter plot with the title "Difference" on the y-axis and "Mean" on the x-axis. The x-axis ranges from 400 to 3400, while the y-axis ranges from -700 to 700. There are three horizontal lines across the plot at approximately y = -400, y = -50, and y = 250. The data points are clustered between x = 2200 and x = 2800.

Figure 2: Device Difference by Cellchek Plus Value- Endothelial Cell Density- All Subjects, Effectiveness Population

Image /page/3/Figure/3 description: The image is a scatter plot with the title "Device Difference" on the y-axis and "CELLCHEY EL Value" on the x-axis. The x-axis ranges from 500 to 3500, while the y-axis ranges from -700 to 700. The data points are scattered, showing a negative correlation between the two variables. As the CELLCHEY EL Value increases, the Device Difference tends to decrease.

{4}------------------------------------------------

Image /page/4/Figure/0 description: The image is a scatter plot with two trend lines. The y-axis is labeled "CEM-520 Value" and ranges from 400 to 3600. There are two trend lines on the plot, one labeled "Deming Regression Line" and the other labeled "45 Degree Line."

400

うこと

8 3 ວ

Figure 3: Deming Regression Plot- CEM-530 by Cellchek Plus- Endothelial Cell Density- All Subjects, Effectiveness Population

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------1こぐり !さむ とここは 16な 1695 ::この こちらのと 2600 3202 CELLCHECK XL Vslue

{5}------------------------------------------------

Table 1: Four Corneal Specular Microscopic Variables Assessed with the Two Devices, All Configurations, All Subjects, Effectiveness Population

Endothelial CellDensityCoefficient ofVariationEndothelial CellArea (CV)Coefficient ofVariationEndothelial CellAreaWITHOUTSubject 001-3-008%HexagonalityCentralCornealThickness
Nidek CEM-530
N7474737474
Mean2482.629.929.469.0551.5
(SD)(436.55)(7.21)(5.86)(5.19)(40.7)
Median2574.028.028.069.0551.0
Min-Max731 - 309319 - 6619 - 5256 - 82411 - 640
Deming RegressionIntercept (95%Confidence Interval)452.83278.7-42.255.3-4.2
(193.1, 712.5)(-120246.9,126804.3)(-90.2, 5.8)(38.0, 72.5)(-120.5,112.2)
Deming RegressionSlope(95% ConfidenceInterval)0.8-104.32.30.21.0
(0.7, 0.9)(-4072.4, 3863.8)(0.7, 3.8)(-0.1, 0.5)(0.8, 1.2)
Konan CELLCHEK XL (PLUS)
N7474737474
Mean (SD)2553.131.231.459.3565.2
(544.85)(4.63)(4.00)(7.80)(41.32)
Median2649.531.031.059.5565.0
Min- Max515 - 347211 - 4022 - 4040 - 75474 - 685
Device Comparisons
Mean Difference (SD)-70.5-1.3-2.19.7-13.8
(167.89)(8.60)(5.52)(8.44)(19.63)
Mean Difference (SD)as a % of theCELLCHEK reading-0.42%0.77%-6.07%18.19%-2.38%
(15.416%)(61.409%)(17.73%)(17.464%)(3.558%)
95% Limits ofAgreement (LOA)(-406.3, 265.2)(-18.5, 15.9)(-13.1, 9.0)(-7.2, 26.5)(-53.0,25.5)
Correlation (R2)0.9654-0.00880.42470.20360.8856

{6}------------------------------------------------

Coefficient of Variation of Endothelial Cell Area

For the total population, the mean differences illustrated on the Bland Altman plot (Figure 4) were generally small. The flatness of the device difference plot (Figure 5) throughout the measurement range also illustrates good agreement between the devices, although an outlier is visible that had a low Cellchek Plus value and a very high Nidek CEM-530 value. This outlier caused a very poor fit on the Deming regression of Figure 6, but this greatly improved in Figure 7 when the regression lines were drawn with this subject excluded.

Image /page/6/Figure/2 description: This image is titled "Coefficient of Variation of Endothelial Cell Area". The image is labeled as "Figure 4: Bland-Altman Plot Observed Data- Coefficient of Variation Endothelial Cell Area- All Subjects, Effectiveness Population". The figure is likely a Bland-Altman plot, which is a scatter plot that is used to compare two methods of measurement.

Image /page/6/Figure/3 description: This image is a scatter plot with the x-axis labeled as "Mean" and the y-axis labeled as "Difference". The data points are scattered around the zero line, with some points above and some below. There are also two horizontal lines, one above and one below the zero line, which may represent the upper and lower limits of agreement.

{7}------------------------------------------------

Image /page/7/Figure/0 description: The image is a scatter plot titled "Endothelial Cell Area- All Subjects, Effectiveness Population". The y-axis is labeled "Device Difference" and ranges from -60 to 60. The x-axis is labeled "CELLCHEK XL Value" and ranges from 10 to 40. The data points are scattered across the plot, with a higher concentration between 25 and 35 on the x-axis and between -10 and 10 on the y-axis.

Figure 5: Device Difference by Cellchek Plus Value- Coefficient of Variation

Figure 6: Deming Regression Plot- CEM-530 by Cellchek Plus - Coefficient of Variation Endothelial Cell Area- All Subjects, Effectiveness Population, with Outlier ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Image /page/7/Figure/3 description: The image is a scatter plot with a regression line. The x-axis is labeled "CELLCHECK XL Value" and ranges from approximately 5 to 40. The y-axis is labeled "CEN-530 Value" and ranges from 0 to 100. The data points are scattered around the regression line, which has a positive slope.

{8}------------------------------------------------

Figure 7: Deming Regression Plot- CEM-530 by Cellchek Plus - Coefficient of Variation Endothelial Cell Area- All Subjects, Effectiveness Population, without Outlier

Image /page/8/Figure/1 description: This image is a scatter plot with two trend lines. The y-axis is labeled "CEM-536 Value" and ranges from -10 to 70. The x-axis is labeled "CELLCHECK XL Value" and ranges from 20 to 50. One trend line is labeled "Deming Regression Line" and the other is labeled "45 Degree Line".

Percent Hexagonality

While the LOAs included 0, the mean differences were on the order of approximately 15% of the mean for % hexagonality (Figure 8), more than for the other 3 variables. The lack of flatness of the device difference plot, Figure 9, also shows that the largest device differences are seen at the lower end of the scale. The Deming regression lines (Figure 10) show some agreement between the two devices.

{9}------------------------------------------------

Image /page/9/Figure/0 description: The image is a scatter plot titled "Effectiveness Population". The x-axis is labeled "Mean" and ranges from 52 to 80. The y-axis is labeled "Difference" and ranges from -40 to 40. There are three horizontal lines at approximately -15, 10, and 30, and the data points are scattered around the center of the plot.

Percent Hexagonality Figure 8: Bland-Altman Plot- Observed Data- % Hexagonality- All Subjects, Effectiveness Population

Figure 9: Device Difference by CELLCHEK XL (PLUS) Value- % Hexagonality- All Subjects, Effectiveness Population

Image /page/9/Figure/3 description: The image is a scatter plot with the title "Device Difference" on the y-axis and "CELLTHEK YL Value" on the x-axis. The x-axis ranges from 40 to 75, while the y-axis ranges from -40 to 40. The scatter plot shows a negative correlation between the two variables, with the points generally decreasing as the x-axis increases.

{10}------------------------------------------------

Image /page/10/Figure/0 description: This image is a scatter plot with two lines. The x-axis is labeled "CELLCHECK XL Value" and ranges from 40 to 80. The y-axis is labeled "CEM-530 Value" and ranges from 40 to 90. The scatter plot shows the relationship between the two variables, and the two lines represent the Deming Regression Line and the 45 Degree Line.

Figure 10: Deming Regression Plot- CEM-530 by CELLCHEK XL (PLUS)- % Hexagonality- All Subjects, Effectiveness Population

Central Corneal Thickness

The Bland Altman plot of Figure 11 shows that the LOAs include 0, indicating a lack of an overwhelmingly large systematic bias. The device differences plot (Figure 12) shows no notable effect of scale on agreement. The Deming regression lines (Figure 13) show an excellent fit.

{11}------------------------------------------------

Image /page/11/Figure/0 description: The image shows a Bland-Altman plot of observed data for central corneal thickness in an effectiveness population. The plot displays the difference between two measurements on the y-axis, ranging from approximately -80 to 120. There are three horizontal lines at approximately -50, -10, and 30, and data points are scattered around the middle line. The title of the plot is "Figure 11: Bland-Altman Plot- Observed Data- Central Corneal Thickness- All Subjects, Effectiveness Population".

Central Corneal Thickness Figure 11: Bland-Altman Plot- Observed Data- Central Corneal Thickness- All

Figure 12: Device Difference by Cellchek Plus Value- Central Corneal Thickness-All Subjects, Effectiveness Population

Ke Kr

550

らぐつ

न में नौ

E S J

રેટર રે

・・・・ - : : : : :

41 J

Image /page/11/Figure/3 description: This image shows a scatter plot titled "All Subjects, Effectiveness Population". The y-axis is labeled "Device Difference" and ranges from -120 to 100. The x-axis is labeled "CELLCHEK XL Value" and ranges from 400 to 700. The data points are scattered across the plot, with a higher density between x values of 500 and 600 and y values between -40 and 20.

{12}------------------------------------------------

Figure 13: Deming Regression Plot- CEM-530 by Cellchek Plus - Central Corneal Thickness - All Subjects, Effectiveness Population

Image /page/12/Figure/1 description: The image is a scatter plot that compares two different values. The y-axis is labeled "DEM-520 Value" and ranges from 450 to 700. The x-axis is labeled "CELICHECK AL Value" and ranges from 450 to 570. There are two dotted lines on the plot, one labeled "Ceming Regression Line" and the other labeled "45 Degree Line".

Agreement of the measurements with the CEM-530 and the predicate device was found to be acceptable. Overall, within eye/subject variability was acceptable, and similar for both machines.

The precision of the two devices was assessed with repeatability and reproducibility measures: the first within a given subject and the second within and among configurations. Table 2 shows the repeatability and reproducibility data for each of the 4 variables in all subjects. Repeatability was notably better for the CEM-530 device for central corneal thickness. The two devices had comparable repeatability for the other endpoints.

VariableNidek CEM-530N=62KonanCELLCHEKPLUSN=62
Endothelial Cell Density
Repeatability SD75.562.4
Repeatability SD as a % of the Mean3.0%2.4%
Repeatability Limit211.5174.8
Repeatability Ratio (CEM-530/CELLCHEK PLUS)1.2102
Reproducibility SD113.295.2
Reproducibility SD as a % of the Mean4.5%3.7%
Reproducibility Limit317.0266.7
Reproducibility Ratio (CEM-530/CELLCHEK PLUS)1.1887

Table 2: Precision Analyses- All Subjects Effectiveness Population

{13}------------------------------------------------

Coefficient of Variation of Endothelial Cell Area (CV)
Repeatability SD2.32.7
Repeatability SD as a % of the Mean8.1%8.5%
Repeatability Limit6.67.5
Repeatability Ratio (CEM-530/CELLCHEK PLUS)0.8746
Reproducibility SD2.72.7
Reproducibility SD as a % of the Mean9.3%8.6%
Reproducibility Limit7.67.6
Reproducibility Ratio (CEM-530/CELLCHEK PLUS)1.0016
% Hexagonality
Repeatability SD4.15.4
Repeatability SD as a % of the Mean6.0%8.8%
Repeatability Limit11.415.0
Repeatability Ratio (CEM-530/CELLCHEK PLUS)0.7586
Reproducibility SD4.15.4
Reproducibility SD as a % of the Mean6.0%8.9%
Reproducibility Limit11.415.2
Reproducibility Ratio (CEM-530/CELLCHEK PLUS)0.7466
Central Corneal Thickness
Repeatability SD3.312.5
Repeatability SD as a % of the Mean0.6%2.2%
Repeatability Limit9.234.9
Repeatability Ratio (CEM-530/CELLCHEK XL (PLUS))0.2634
Reproducibility SD5.813.2
Reproducibility SD as a % of the Mean1.1%2.3%
Reproducibility Limit16.337.0

.

.

{14}------------------------------------------------

Additional analysis was completed on 24 images with and without corneal pathology using both the manual method and automated method of analysis on the same image. The images were then analyzed and generated the following results for CD (Cell Density), CV(Coefficient of Variation) and HEX(%Hexagonality)

CD(Cell Density)CV(Coefficient ofVariation)HEX(Hexagonality)NUM(Number of Cells)
Mean2267.2 / 2165.3Auto / Manual28.9 / 36.3Auto / Manual68.7 / 57.4Auto / Manual173.3 / 153.3Auto / Manual
Mean Difference(SD)101.9 (53.76)-7.4(4.91)11.3(7.75)
Mean differenceas a % of theManual reading4.94%-19.41%20.98%13.47%
Correlation (R^2)0.99180.75230.22260.4306
Deming RegressionIntercept127.75.050.3-334.1
Deming RegressionSlope1.00.70.33.3

Table 3: Agreement Analysis on Same Image – Automated vs. Manual Method

Image /page/14/Figure/3 description: This image is labeled as "Figure 14: Deming Regression Plot - CEM-530 by Manual - Endothelial Cell Density (CD) - Same Image - All Subjects." The title indicates that the figure is a Deming Regression Plot. The plot is related to CEM-530 data obtained manually. The data is about endothelial cell density and includes all subjects from the same image.

Image /page/14/Figure/4 description: The image shows a scatter plot with two lines. The x-axis is labeled "Manual Value" and the y-axis is labeled "TEM-510 Value". There is a solid line labeled "Daring Regression Line" and a dashed line labeled "45 Degree Line". The scatter plot shows a positive correlation between the manual value and the TEM-510 value.

{15}------------------------------------------------

Figure 15: Deming Regression Plot - CEM-530 by Manual - Coefficient of Variation of Endothelial Cell Area (CV) – Same Image – All Subjects

Image /page/15/Figure/1 description: The image is a scatter plot with two lines. The x-axis is labeled "Input Value" and ranges from 20 to 55. The y-axis is labeled "Output Value" and ranges from 22 to 60. There is a solid line labeled "Dasing Regression Line" and a dashed line labeled "45 Degree Line."

Figure 16: Deming Regression Plot - CEM-530 by Manual - % Hexagonality (HEX). - Same Image - All Subjects

Image /page/15/Figure/3 description: The image is a scatter plot with two lines. The x-axis is labeled "Manual Value" and ranges from 10 to 80. The y-axis is labeled "CEM.430 Value" and ranges from 40 to 80. There are two lines on the plot, one solid and one dashed, and several data points scattered around the lines.

{16}------------------------------------------------

In summary, the agreement and precision of the Nidek CEM-530 was found to be substantially equivalent to the predicate device.

CONCLUSIONS

The Nidek Specular Microscope CEM-530 has the same intended use and indications for use, technological characteristics, and principles of operation as the previously cleared predicate. The minor differences between the subject device and the predicate device have been assessed in a human clinical trial which found agreement, accuracy and precision between the two devices. Therefore, the Nidek Specular Microscope CEM-530 is as safe and effective as its predicate device, and thus, substantially equivalent.

{17}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

November 27, 2013

Nidek Co., Ltd. c/o Mr. Aron Shapiro Vice President 300 Brickstone Square Andover, MA 01810

Re: K130565

Trade/Device Name: Nidek Specular Microscope CEM-530 Regulation Number: 21 CFR 886.1850 Regulation Name: AC-Powered Slitlamp Biomicroscope Regulatory Class: Class II Product Code: NQE Dated: October 17, 2013 Received: October 18, 2013

Dear Mr. Shapiro:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your boomed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to commerce proc to may 20, 1977, Liv accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Tournal controls provisions of the Act include requirements for annual registration, listing of I he general connous provisions and prohibitions against misbranding and ac vices, good manarataining prevates, as time, water information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), II your device is classified (500 above) a Existing major regulations affecting your device can be it may of Subject to adamonal conditions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{18}------------------------------------------------

Page 2 - Mr. Aron Shapiro

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/18/Picture/6 description: The image contains the text "Deborah L Falls -S". The text is written in a bold, sans-serif font. The word "Deborah" is on the left, followed by "L Falls" and then "-S" on the right. The text is slightly distorted, with some parts appearing blurred or pixelated.

for Malvina B. Evdelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{19}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name:_ Nidek Specular Microscope CEM-530

Indications For Use:

The Nidek Specular Microscope CEM-530 is a non-contact ophthalmic microscope, optical pachymeter, and camera intended for examination of the corneal endothelium and for measurement of the thickness of the cornea.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR .

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please Do NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Marsha L. Burke Nicholas -S 2013.12.03 15:14:38 -05'00'

Page 1 of _1

§ 886.1850 AC-powered slitlamp biomicroscope.

(a)
Identification. An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.(b)
Classification. Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.