K Number
K130535

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-04-17

(47 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The iBGStar™ Blood Glucose Monitoring System is intended for the quantitative measurement of blood glucose levels from fresh capillary whole blood samples drawn from the fingertip, palms (at the base of the thumb), or forearms. It is intended to be used by a single patient and should not be shared. The iBGStar™ Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The iBGStar™ Blood Glucose Monitoring System is not for the diagnosis of, or screening for diabetes, and is not intended for use with neonates.

The iBGStar Diabetes Manager Application is intended for use in the home with the capability of sending glucose readings through email to an individual's healthcare professional in the review, analysis and evaluation of glucose test results to support an effective diabetes management program. It is an optional data management software accessory for use with the iBGStar Blood Glucose Monitoring System.

Device Description

The iBGStar Blood Glucose Monitoring System (BGMS) consists of:

  • iBGStar Blood Glucose Meter .
  • . BGStar Test Strips
  • BGStar Control Solution .

The iBGStar. Diabetes Manager Application (DMA) is an optional software accessory for the iBGStar Blood Glucose Monitoring System. It is a digital logbook and diabetes tool designed to operate using an iPhone or iPod touch. An individual can manually enter blood glucose readings or can download readings directly to the DMA installed on an iPhone or iPod touch from the iBGStar meter.

AI/ML Overview

Here's an analysis of the provided text regarding the AgaMatrix iBGStar Blood Glucose Monitoring System and iBGStar Diabetes Manager Application, focusing on acceptance criteria and supporting studies.

Based on the provided 510(k) summary, the device under review (K130535) is an update to an existing device (predicate K103544) to ensure compatibility with newer Apple devices (iPhone 5, iPod touch 5th generation) using a Lightning adapter. Crucially, the document states that there were no changes to the fundamental scientific technology of the Blood Glucose Monitoring System (meter, strips, controls) or the Diabetes Manager Application itself. The improvements are primarily related to connectivity and physical compatibility with new phone models.

Therefore, the performance data presented is focused on demonstrating that these new connectivity components (Lightning adapters) do not negatively impact the previously established performance, and that the updated system as a whole remains substantially equivalent to the predicate.


1. Table of Acceptance Criteria and Reported Device Performance

Given the nature of the submission (compatibility update, not a new sensing technology), the acceptance criteria mentioned are related to:

  • Disinfection and Robustness of the new adapter hardware and associated Apple devices.
  • Verification and Validation that the new adapters facilitate data transfer as intended.
  • Human Factors to ensure users can correctly use the new adapters with the system.

The original performance characteristics for the blood glucose monitoring system (e.g., accuracy of glucose measurement) are assumed to be met as they were established for the predicate device, which remains unchanged in its core technology.

Acceptance CriteriaReported Device Performance (K130535)
DisinfectionCriterion: Ensure the new components (iPhone 5, iPod touch 5, Apple Lightning to 30-pin Adapter, and Apple Lightning to 30-pin Adapter (0.2m) cable) can be effectively disinfected without performance degradation. Performance: Disinfection testing with PDI® Super Sani-Cloth® Germicidal Disposable Wipes (EPA number 9480-4) demonstrated complete inactivation of live virus.
RobustnessCriterion: Ensure the new components maintain performance and material integrity after repeated cleaning/disinfection cycles. Performance: Robustness testing involved 260 cleaning/disinfection cycles (simulating 5 years of use) with the PDI® Super Sani-Cloth® Germicidal Disposable Wipes. Results showed no change in performance or material of the tested devices.
Verification and Validation (of new adapters/connectivity)Criterion: Demonstrate that the Apple Lightning to 30-pin Adapter and the Apple Lightning to 30-pin Adapter (0.2m) cable correctly enable the iBGStar DMA to function as intended (i.e., transfer glucose readings). Performance: Verification and validation protocols were successfully executed, and results demonstrated "substantial equivalence to the predicate system" regarding the use of these adapters.
Human FactorsCriterion: Ensure users can successfully connect and transfer readings via the new Apple Lightning to 30-pin Adapter and cable. Performance: A human factors study was conducted, and results demonstrated "substantial equivalence to the predicate system" in relation to users' ability to connect and transfer readings using the new adapters.
(Implied) Blood Glucose Measurement Accuracy/PerformanceCriterion: (Inherited from predicate) The iBGStar BGMS performs quantitative measurements of blood glucose levels in fresh capillary whole blood with acceptable accuracy and precision. Performance: The document explicitly states: "There have been no modifications to the iBGStar BGMS to allow it to operate with the iPhone 5 and iPod touch 5th generation. The Meter, Test Strips and Control Solution are identical to the previously cleared system. The iBGStar BGMS has all the same technological and performance characteristics as the predicate." Therefore, the previous performance data for K103544 is considered applicable and sufficient. No new studies on the core blood glucose measurement accuracy are reported here.

2. Sample Size Used for the Test Set and Data Provenance

Due to the nature of this 510(k) submission being an update for compatibility with new accessories rather than a new core device, detailed sample sizes for the "test set" (in the typical sense of clinical performance) are not provided in this summary.

  • Disinfection and Robustness Test Set: This involved the physical components: iPhone 5, iPod touch 5th generation, Apple Lightning to 30-pin Adapter, and Apple Lightning to 30-pin Adapter (0.2m) cable. The "sample size" here refers to the number of individual devices/adapters tested, which is not specified but implicitly sufficient for the tests performed (e.g., "the tested devices").
  • Verification and Validation: The number of tests/data points for validating the data transfer via the new adapters is not specified.
  • Human Factors Study: The sample size for the human factors study is not specified in this summary.
  • Data Provenance: Not specified for these specific tests in this summary. The manufacturer is US-based. These tests are prospective, conducted for this submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This section is not applicable (N/A) for this submission.

The studies described (disinfection, robustness, verification/validation of connectivity, human factors for adapter connection) do not involve establishing a clinical "ground truth" using medical experts like radiologists. They are engineering and usability tests. The "ground truth" for the disinfection study was the complete inactivation of live virus, observed directly (laboratory result). For robustness, it was the maintenance of performance and material integrity. For verification/validation and human factors, it was the successful connection and data transfer by users, assessed against functional requirements.


4. Adjudication Method for the Test Set

This section is N/A for this submission.

The studies described do not involve a medical diagnosis or interpretation that would require an adjudication method (like 2+1 or 3+1 consensus). The outcomes are objective measurements or functional verifications.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

This section is N/A for this submission.

An MRMC study is typically for evaluating the diagnostic performance of a system (often AI-based) and comparing human readers with and without AI assistance. This submission is for an updated blood glucose monitoring system and its associated data management app's compatibility with new phone accessories. It does not involve a diagnostic AI component in the sense of image interpretation or similar tasks.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The "iBGStar Diabetes Manager Application" is a standalone software accessory in that it collects and displays data. This submission is about ensuring its connectivity with newer hardware accessories. The core 'standalone' performance (data logging, trend charting) of the application itself was likely established in the predicate 510(k) (K103544).

The described verification and validation protocols for K130535 specifically tested the standalone functionality of data transfer via the new adapters.


7. The Type of Ground Truth Used

  • Disinfection Study: Laboratory-confirmed complete inactivation of live virus.
  • Robustness Testing: Objective measurements of device performance and material integrity before and after simulated use cycles.
  • Verification and Validation: Successful data transfer and system functionality, verified against predefined technical specifications.
  • Human Factors Study: Observation and measurement of user success in connecting and transferring readings, likely against criteria for ease of use and error rates.

None of these studies utilized medical "gold standards" such as pathology, expert consensus on images, or long-term outcomes data, as they are not clinical performance studies for a new diagnostic capability.


8. The Sample Size for the Training Set

This section is N/A for this submission.

The device itself is a blood glucose monitor and a data management application. It is not described as an AI/Machine Learning device that requires a "training set" in the context of developing an algorithm for diagnostic or predictive tasks. The tests reported here are for hardware compatibility and usability, not algorithm development.


9. How the Ground Truth for the Training Set Was Established

This section is N/A for this submission, as there is no mention of an algorithm training set in the AI/ML sense.

{0}------------------------------------------------

AgaMatrix.

AgaMatrix, Inc. · 7C Raymond Avenue · Salem, NH 03079 USA

510(k) Summary

K 130535

This summary of 510k) substantial equivalence is being submitted in accordance with the requirements of 21 CFR 807.92.

Prepared:March 28, 2013
Submitter:AgaMatrix, Inc.
Address:7C Raymond Ave.Salem, NH 03079Phone: (603) 328-6000
Contact:William H. McGrailExecutive Director of Regulatory & Clinical Affairswmcgrail@agamatrix.comPhone: 603-328-6051Fax: 603-836-4025
Device Name:Trade/Proprietary Name: iBGStar Blood Glucose Monitoring SystemCommon Name: Glucose Test SystemProduct Name: iBGStar Diabetes Manager ApplicationCommon Name: Diabetes Management Software

Device Classification:

Product CodeClassificationRegulation SectionPanel
NBW - system, test, bloodglucose, over the counterClass II21 CFR 862.134575, Clinical Chemistry
JQP - Calculator/data processingmodule for clinical use.Class I21 CFR 862.210075, Clinical Chemistry

Predicate Device:

iBGStar Blood Glucose Monitoring System, iBGStar Diabetes Manager Application, 510(k) K103544

AgaMatrix, Inc. 7C Raymond Avenue TEL FAX WEB Salem, New Hampshire 03079 USA +1 (603) 328 6000 +1 (617) 588 0430 www.agamatrix.com

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "AgaMatrix" in a stylized font. The word is written in black ink, and there is a curved line underneath the "Aga" portion of the word. Above the "Matrix" portion of the word, there is a series of dots that form an arc.

· 7C Raymond Avenue · Salem, NH 03079 USA

Device Description:

The iBGStar Blood Glucose Monitoring System (BGMS) consists of:

  • iBGStar Blood Glucose Meter .
  • . BGStar Test Strips
  • BGStar Control Solution .

The iBGStar. Diabetes Manager Application (DMA) is an optional software accessory for the iBGStar Blood Glucose Monitoring System. It is a digital logbook and diabetes tool designed to operate using an iPhone or iPod touch. An individual can manually enter blood glucose readings or can download readings directly to the DMA installed on an iPhone or iPod touch from the iBGStar meter.

Labeling and Intended Use:

The iBGStar BGMS Owner's guide has been edited to include instructions on how to connect the meter to a device with the Lightning adapter. The iBGStar BGMS carton has been updated to include a statement that the meter will work with the iPhone 5 and iPod touch 5th generation when using the Lightning adapter.

Intended Use

The iBGStar™ Blood Glucose Monitoring System is intended for the quantitative measurement of blood glucose levels in fresh capillary whole blood samples drawn from the fingertip, palms (at the base of the thumb), or forearms. It is intended to be used by a single patient and should not be shared. The iBGStar™ Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The iBGstar Blood Glucose Monitoring System is not for the diagnosis of, or screening for diabetes, and is not intended for use with neonates.

The iBGStar Diabetes Manager Application is intended for use in the capability of sending glucose readings through email to an individual's healthcare professional in the review, analysis and evaluation of glucose test results to support an effective diabetes management program. It is an optional data management software accessory for use with the iBGStar Blood Glucose Monitoring System.

Technological Characteristics:

There were no changes to the fundamental scientific technology.

AgaMatrix, Inc. 7C Raymond Avenue
Salem, New Hampshire 03079 USA

+1 (603) 328 6000

+1 (617) 588 0430

ww.agamatrix.com

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "AgaMatrix." in a stylized font. The word is in bold black letters. There is a curved line under the "Aga" portion of the word, and a series of dots above the "trix" portion of the word.

AgaMatrix, Inc. · 7C Raymond Avenue · Salem. NH 03079 USA

Comparison to Predicate:

Blood Glucose Monitoring System (meter, strips, controls)

There have been no modifications to the iBGStar BGMS to allow it to operate with the iPhone 5 and iPod touch 5th generation. The Meter, Test Strips and Control Solution are identical to the previously cleared system. The iBGStar BGMS has the all the same technological and performance characteristics as the predicate. There were minor modifications to the labeling for the iBGStar BGMS to clarify the use of the meter with the iPhone 5 and iPod touch 5th generation. The table below illustrates the same intended use and characteristics between the predicate device and the candidate device (no differences between predicate and candidate).

Similarities
CharacteristiciBGStar BGMS (predicate)iBGStar BGMS (candidate)
Indications for UseBlood Glucose MonitoringSame
Intended UseHome UseSame
CalibrationNo coding requiredSame
Test Principle/EnzymeGlucose OxidaseSame
Control LevelsTwo levelsSame
System CharacteristicsOperating Temp, Test Time, Test Range, Sample Size, Test strips, Control SolutionSame
BacklightNoSame
Number of results stored300Same
Power SourcePolymer lithium-ion rechargeable batteriesSame
Optional Software accessoryiBGStar Diabetes Manager ApplicationSame
Differences
None
AgaMatrix, Inc. 7C Raymond AvenueSalem, New Hampshire 03079 USATEL+1 (603) 328 6000FAX+1 (617) 588 0430WEBwww.agamatrix.com
---------------------------------------------------------------------------------------------------------------------------------------------------

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "AgaMatrix." in a stylized font. There is a curved line under the "Aga" portion of the word. There are also a series of dots above the "Matrix" portion of the word.

AgaMatrix, Inc. · 7C Raymond Avenue · Salem, NH 03079 USA

Diabetes Manager Application

The iBGStar DMA has the same intended use as the previously cleared system. The iBGStar DMA has the all the same technological and performance characteristics as the predicate. There is no modification to the labeling for the iBGStar DMA. To transfer glucose readings from the iBGStar meter to the iBGStar DMA which has been downloaded on Apple's new iPhone 5 and/or iPod touch 500 generation, an Apple Lightning to 30-pin Adapter or the Apple Lightning to 30-pin Adapter (0.2m) is required The table below lists the similarities and differences between candidate and predicate device.

Similarities
CharacteristiciBGStar Diabetes Manager App(predicate)toiPhone 3G,3GS,4, 4SiPod touch 2,3,4iBGStar Diabetes ManagerApp (candidate)toiPhone 5 and iPod touch 5
Indications forUseDownload glucose readings to a datamanagement system to aid in theeffective management of diabetes.Same
Intended UseHome UseSame
ManagementToolsLogbook and Trend ChartsSame
Upload ToDevice compatible with the iPhoneOperating System platformSame
Differences
CharacteristiciBGStar Diabetes Manager App(predicate)toiPhone 3G,3GS,4, 4SiPod touch 2,3,4iBGStar Diabetes ManagerApp (candidate)toiPhone 5 and iPod touch 5
Transfer ofGlucose ReadingsNo connector needed(see Figure 1)Apple Lightning to 30-pinAdapter(see Figure 2)orApple Lightning to 30-pinAdapter (0.2m) cable(see Figure 3)
AgaMatrix. Inc. 7C Raymond AvenueTELFAX------------------------------------------------------------------------------------------------------------------------------------------------------------------------------WEB
Salem, New Hampshire 03079 USA------------------------------------------------------------------------------------------------------------------------------------------------------------------------------+1 (603) 328 6000+1 (617) 588 0430www.agamatrix.com

{4}------------------------------------------------

Figure 1: iPhone to iBGStar-no connector

Image /page/4/Picture/2 description: The image shows a close-up of a device, possibly an iPod or similar gadget, displaying a screen with health-related information. The screen indicates a reading of '147 mg/dL', which is likely a blood sugar level. The date displayed on the screen is 'Monday, January 10, 2011, 9:30 AM', and the context is labeled as 'Post Breakfast'.

Figure 2: Apple Lightning to 30-pin Adapter

Image /page/4/Picture/4 description: The image shows a smartphone displaying a blood glucose level of 147 mg/dL. The screen indicates that the reading was taken post-lunch on Thursday, January 17, 2013, at 2:11 PM EDT. The phone is connected to a device, possibly a blood glucose meter, via a port at the bottom. The device is attached to the phone.

Figure 3: Apple Lightning to 30-pin Adapter (0.2m) Cable

Image /page/4/Picture/6 description: The image shows a white iPhone connected to a white charger. The phone is displaying the number 147 on the screen. The phone and charger are against a gray wall. The charger is plugged into a black device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "AgaMatrix" in a stylized font. The letters are bold and slightly italicized, with a curved line underneath the "Aga" portion of the word. Above the "Matrix" portion of the word, there is a series of dots that appear to be arranged in an arc. A period is at the end of the word.

AgaMatrix, Inc. · 7C Raymond Avenue · Salem. NH 03079 USA

Assessment of Performance:

Disinfection Study:

Per pre-submission Q120016 (See Attachment 2 of original submission), FDA recommended that disinfection and robustness testing be performed on the new components.

Disinfection testing on the surface materials of the iPhone 5, iPod touch 5, Apple Lightning to 30-pin Adapter, and Apple Lightning to 30-pin Adapter (0.2m) cable was performed by an outside lab. The results with PDI® Super Sani-Cloth® Germicidal Disposable Wipes with EPA number 9480-4 demonstrated complete inactivation of live virus (see Attachments 3 and 4 of original submission).

Robustness testing was performed with the iPhone 5, iPod touch 5th generation, Lightning to 30-pin Adapter, and Lightning to 30-pin Adapter (0.2m) cable using the PDI® Super Sani-Cloth® Germicidal Disposable Wipes with EPA number 9480-4. The results demonstrate there was no change in performance or material of the tested devices at the conclusion of the disinfection and robustness testing which consisted of 260 cleaning/disinfection cycles designed to simulate 5 years of use (see Attachments 5,6 and 7 of original submission).

Verification and Validation:

Verification and validation protocols were successfully executed to validate the use of the Apple Lightning to 30-pin Adapter and the Apple Lightning to 30-pin Adapter (0.2m) cable with the iBGStar DMA. Results demonstrate substantial equivalence to the predicate system.

Human Factors Study:

A human factors study (see Attachments 8 and 9 of original submission) was conducted to evaluate the ability of users to connect and transfer readings from the iBGStar DMA via the new Apple Lightning to 30-pin Adapter and the Apple Lightning to 30-pin Adapter (0.2m) cable. Results demonstrate substantial equivalence to the predicate system.

Conclusion:

The results of the disinfection and robustness testing and the performance assessments demonstrated that the candidate iBGStar Blood Glucose Monitoring System and the iBGStar Diabetes Manager Application using the Lightning to 30-pin Adapter or 30-pin Adapter (0.2m) cable performs in a substantially equivalent manner to that of the predicate. We conclude that the iBGStar Blood Glucose Monitoring system and the iBGStar Diabetes Manager Application are substantially equivalent the predicate devices.

+1 (603) 328 6000

1 (617) 588 0430

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The text is in all capital letters and is evenly spaced around the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 17, 2013

AgaMatrix, Inc. C/O William H. McGrail 7C Raymond Ave SALEM NH 03079

Re: K130535

Trade/Device Name: iBGStar Blood Glucose Monitoring System iBGStar Diabetes Manager Application Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, JQP Dated: March 28, 2013 Received: April 02, 2013

Dear Mr. McGrail:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. "Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carol (Beingson -S för

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K130535

Device Name: jBGStar Blood Glucose Monitoring System, iBGStar Diabetes Manager Application

Indications for Use:

-The iBGStar™ Blood Glucose Monitoring System

The iBGStar™ Blood Glucose Monitoring System is intended for the quantitative measurement of blood glucose levels from fresh capillary whole blood samples drawn from the fingertip, palms (at the base of the thumb), or forearms. It is intended to be used by a single patient and should not be shared. The iBGStar™ Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The iBGStar™ Blood Glucose Monitoring System is not for the diagnosis of, or screening for diabetes, and is not intended for use with neonates.

The iBGStar Diabetes Manager Application - Home Use

The iBGStar Diabetes Manager Application is intended for use in the home with the capability of sending glucose readings through email to an individual's healthcare professional in the review, analysis and evaluation of glucose test results to support an effective diabetes management program. It is an optional data management software accessory for use with the iBGStar Blood Glucose Monitoring System.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use × (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Devices and Radiologic Health (OIR)

Katherine Serrano

Division Sign-Off Office of In Vitro Diagnostic Devices and Radiologic Health

510(k) K130535

Page 1 of 1

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.