K Number
K130506

Validate with FDA (Live)

Device Name
IMSE P-TLIF
Date Cleared
2013-05-13

(75 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IMSE P-TLIF is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). IMSEP-TLIF is to be used with autologous bone graft and implanted via an open transforaminal or posterior approach. IMSEP-TLIF implants are to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

Device Description

Institute of Musculoskeletal Science and Education, P-TLIF is a device for interbody fusion of the anterior column of the spine. These cages are hollow so that bone can grow through the device, fusing the adjacent bony surfaces. IMSE, P-TLIF is a hollow device with texture on two opposing convex sides, and is offered in various lengths, widths, heights and shapes. Institute of Musculoskeletal Science and Education designed the IMSE, P-TLIF to be placed through a transforaminal or posterior approach and to address vertebrae in the lumbosacral region of the spine.

AI/ML Overview

The provided documentation describes the IMSE P-TLIF device, an intervertebral fusion device. The focus of the submission is to demonstrate substantial equivalence to predicate devices, primarily through nonclinical testing.

Here's an analysis of the acceptance criteria and the study results based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Test Performed)Reported Device Performance
Static and Dynamic Compression Test (per ASTM F2077)Met the acceptance criteria.
Static and Dynamic Compression Shear (per ASTM F2077)Met the acceptance criteria.
Subsidence Test (per ASTM F2267)Met the acceptance criteria.
Wear Debris (per ASTM F2077 and ASTM F1877)Met the acceptance criteria.
Static Expulsion TestMet the acceptance criteria.
Sterility Assurance Level (SAL) of 10^-6 (per ISO 17665) for Autoclaveable ImplantsValidation testing conducted (using the half-cycle method) to achieve this SAL.
Sterility Assurance Level (SAL) of 10^-6 (per ISO 17665) for Autoclaveable Instruments and CaseValidation testing conducted (using the half-cycle method) to achieve this SAL.

2. Sample Size Used for the Test Set and Data Provenance

The provided summary does not specify the sample size for the nonclinical tests (e.g., number of devices tested for compression, shear, etc.). It only states that these tests were performed to demonstrate substantial equivalence.

Similarly, the data provenance (e.g., country of origin, retrospective or prospective) is not mentioned for these nonclinical tests. As these are laboratory tests, provenance in the clinical sense is not applicable.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not applicable as the studies performed were nonclinical (mechanical and material testing), not clinical studies requiring expert ground truth for interpretation of patient data.

4. Adjudication Method for the Test Set

This information is not applicable as the studies performed were nonclinical, not clinical studies requiring adjudication of outcomes.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a MRMC comparative effectiveness study was not performed. The document explicitly states: "No clinical tests were performed."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable as the device (IMSE P-TLIF) is a physical intervertebral fusion device, not an algorithm or AI system.

7. The Type of Ground Truth Used

The ground truth for the nonclinical tests was based on standardized mechanical and material testing protocols (e.g., ASTM F2077, ASTM F2267, ASTM F1877, ISO 17665). The "ground truth" in this context refers to the specified performance parameters and thresholds defined within these standards, which the device had to meet.

8. The Sample Size for the Training Set

This question is not applicable as the device is a physical medical implant, not an AI or machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable for the same reason as point 8.

{0}------------------------------------------------

Page 1 of 4

MAY 1 3 2013

Image /page/0/Picture/2 description: The image is a logo that is divided into four squares. The top left square contains the letter 'i', the top right square contains the letter 'm', the bottom left square contains the letter 's', and the bottom right square contains the letter 'e'. Below the four squares is the text 'The Institute of Musculoskeletal Science & Education'.

510(k) Summary

as required by section 807.92(c).

P-TLIF

K130506

Institute of Musculoskeletal Science & Education

90 S. Newtown Street Rd., Suite #10

Newtown Square, PA 19073

John Moran

CEO

Phone: 484-420-4219

Fax: 484 318-8031

Email: mberwyn@aol.com

Submitted 4/4/13

{1}------------------------------------------------

MAY 1 3 2013

Page 2 of 4

K130506

.

Submitter:Institute of Musculoskeletal Science & Education
90 S. Newtown Street Rd., Suite #10
Newtown Square, PA 19073
Contact PersonJohn Moran
CEO
Phone: (484) 420-4219
Email: mberwyn@aol.com
Trade NameIMSE P-TLIF
Common NameP-TLIF
Device ClassClass II
Classification Nameand NumberIntervertebral Fusion Device With Bone Graft, Lumbar21 CFR 888.3080
Classification Panel:Orthopedic
Product CodeMAX
Reason for 510kNew Device
Predicate DevicesCamber Cage (K121254), Corelink Foundation Cage (K 073440),Stryker AVS PL (K093704) & K2M Aleutian (K113138)
Device DescriptionInstitute of Musculoskeletal Science and Education, P-TLIF is a devicefor interbody fusion of the anterior column of the spine. These cagesare hollow so that bone can grow through the device, fusing theadjacent bony surfaces.IMSE, P-TLIF is a hollow device with texture on two opposing convexsides, and is offered in various lengths, widths, heights and shapes.Institute of Musculoskeletal Science and Education designed theIMSE, P-TLIF to be placed through a transforaminal or posteriorapproach and to address vertebrae in the lumbosacral region of thespine.

.

2

{2}------------------------------------------------

K130506 Page 3 of 4

The IMSE P-TLIF is indicated for intervertebral body fusion procedures in
Intended Useskeletally mature patients with degenerative disc disease (DDD) of the
lumbar spine at one or two contiguous levels from L2-S1. DDD is defined as
discogenic pain with degeneration of the disc confirmed by history and
radiographic studies. These DDD patients may also have up to Grade I
spondylolisthesis or retrolisthesis at the involved level(s). IMSEP-TLIF is to
be used with autologous bone graft and implanted via an open
transforaminal or posterior approach.
IMSEP-TLIF implants are to be used with supplemental fixation. Patients
should have at least six (6) months of non-operative treatment prior to
treatment with an intervertebral cage.
Materials:The implant is manufactured from ASTM2026 Solvay ZenivaZA-500 implant grade Polyetheretherketone (PEEK)
-------------------------------------------------------------------------------------------------------------------------
Statement ofTechnologicalComparisonIMSE, P-TLIF and its predicate devices have the same indications for use,similar design, and test results. Both devices are manufactured usingmaterials with a long history of use in orthopaedic implants.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

.

{3}------------------------------------------------

. K130506 Page 4 of 4

Nonclinical TestThe following tests were performed to demonstrate that the IMSEP-TLIF is substantially
Summaryequivalent to other predicate devices.
Static and Dynamic Compression Test per ASTM F2077
Static and Dynamic Compression Shear ASTM F2077
Subsidence Test per ASTM F2267
Wear Debris ASTM F2077 and ASTM F1877
Static Expulsion Test
The results of these studies showed that the IMSEP-TLIF met the acceptance criteria.
Clinical Test
SummaryNo clinical tests were performed.
Sterilization Information
ImplantsThe Implant will be shipped non-sterile and will be autoclaveable, validation testing of
the process was conducted (using the half-cycle method) to a Sterility Assurance Level
(SAL) of 10-6 per ISO 17665.
Instruments andThe instrument and case will be shipped non-sterile and will be autoclaveable,
Casevalidation testing of the process was conducted (using the half-cycle method) to a
Sterility Assurance Level (SAL) of 10-6 per ISO 17665.
ConclusionThe IMSE, P-TLIF is substantially equivalent to its predicate devices. This conclusion isbased upon the fact the Lumbar Cage and its predicate devices have the sameindications for use, have a similar design and technical characteristics, similar testresults, and any differences do not raise question of safety and effectiveness.
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

4

:

·

:

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, with three wavy lines below the eagle, possibly representing water or movement.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Institute of Musculoskeletal Science & Education % Mr. John Moran CEO 90 South Newtown Street Road, Suite #10 Newtown Square, Pennsylvania 19073

Re: K130506

Trade/Device Name: IMSE P-TLIF Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Codes: MAX Dated: April 5, 2013 Received: April 5, 2013

Dear Mr. Moran:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

Letter dated: May 13, 2013

{5}------------------------------------------------

Page 2 - Mr. John Moran

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark NMelkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number: IMSE P-TLIF Device Name:

K130506

Indications:

The Institute of Musculoskeletal Science & Education P-TUF is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). institute of Musculoskeletal Science & Education P-TLIF is to be used with autologous bone graft and implanted via an open transforaminal or posterior approach.

Institute of Musculoskeletal Science & Education P-TLIF implants are to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.

AND/OR Over-the-counter Prescription Use X (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Anton E. Dmitriev, PhD Division of Orthopedic Devices

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.