K Number
K130489

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-04-19

(53 days)

Product Code
Regulation Number
876.4300
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PRECISION Endoscopic Infrared Coagulator with single-use disposable MAXiguide flexible light guide is intended for treatment of hemorrhoids and other lesions in the colon and rectum by the transmission of infrared energy when used as an accessory to a colonoscope, flexible sigmoidoscope, or other flexible gastrointestinal endoscope.

Device Description

This device is designed to coagulate blood and tissue, specifically for treatment of hemorrhoids and small lesions in the colon and rectum, when used as an accessory to a flexible endoscope.

AI/ML Overview

This 510(k) submission (K130489) is for a design and labeling change to an existing device, the PRECISION Endoscopic Infrared Coagulator™ with SINGLE-USE DISPOSABLE MAXI-GUIDE FLEXIBLE LIGHTGUIDE. The submission explicitly states that there are no technology differences between the submission device and its predicate devices, and no changes to the intended use or indications for use.

Therefore, the study performed is focused on demonstrating that the design changes do not affect the safety and effectiveness of the device, rather than proving performance against new acceptance criteria for a novel technology.

Here's a breakdown of the requested information based on the provided text:


1. Table of Acceptance Criteria and Reported Device Performance

Given that this is a 510(k) for a design change with no technological differences or changes to intended use, the acceptance criteria are implicitly that the device continues to meet the safety and effectiveness profile of the predicate devices. The reported performance is verification that these standards are maintained.

Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred from 510(k) context for design changes)Reported Device Performance
SafetyDevice design changes do not introduce new safety risks."all tests passed"
EffectivenessDevice design changes do not negatively impact the ability to coagulate blood and tissue for its stated indications."all tests passed"
Substantial EquivalenceDevice remains substantially equivalent to predicate devices, maintaining previous performance characteristics."The submission device is substantially equivalent and as safe and effective as the predicate devices."

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a quantitative sample size for the "bench testing" performed. It generally states that "Verification testing results for the design changes were completed and all tests passed."

  • Sample Size: Not explicitly stated (likely refers to a number of test articles/units for bench testing).
  • Data Provenance: The testing was "Bench testing," implying it was conducted in a laboratory or engineering setting by the manufacturer (Optim LLC in Sturbridge, MA). It is retrospective in the sense that it's verifying changes to an existing design, not gathering new clinical data on patients.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This type of information is not applicable to this 510(k) submission.

  • This submission is not for an AI/ML device, diagnostic device, or a device that requires human interpretation of outputs to establish ground truth.
  • The "verification testing" described is bench testing, which assesses engineering and functional performance against defined specifications, not against expert-established ground truth related to clinical outcomes or interpretations.

4. Adjudication Method for the Test Set

Not applicable for the reasons stated above. There's no interpretive test set requiring adjudication.


5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is an infrared coagulator, not an AI-assisted diagnostic or therapeutic tool that would involve human readers or image interpretation.


6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device does not involve an algorithm with standalone performance.


7. The Type of Ground Truth Used

The "ground truth" for this type of submission is typically defined by engineering specifications, established performance parameters of the predicate device, and compliance with relevant consensus standards. For example, ground truth for power output might be a specific wattage as measured by calibrated equipment, or ground truth for light guide flexibility might be its ability to withstand a certain bend radius without damage. The document indicates verification was against the safety and effectiveness of the predicate devices.


8. The Sample Size for the Training Set

Not applicable. This is not an AI/ML device that requires a training set.


9. How the Ground Truth for the Training Set was Established

Not applicable. This is not an AI/ML device that requires a training set.

{0}------------------------------------------------

K 130489 102

PR 1 9 2013

Image /page/0/Picture/1 description: The image shows the logo for Optim LLC, which is a company. The logo is black and white and features the company name in a bold, sans-serif font. Below the company name is the text "A Juno Company" in a smaller font. The logo is simple and professional.

7.0 510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirement of 21 CFR 807.92.

Jeff Barrett, CEO Optim LLC 64 Technology Park Rd. Sturbridge, MA 01566

Date Prepared: 17 April 2013

Device Trade Name: Precision Endoscopic Infrared Coagulator; Model 5100 Single-Use disposable Maxi-guide Flexible Lightguide; Model 5100-2

Device Classification Name: Electrode, Flexible Suction Coagulator Registration Number: 1218141 Regulation Number: 876.4300 Product Code: FEH Classification: Class II Review Panel: Gastroenterology / Urology Predicate Device Information

  • . Precision Endoscopic Infrared Coagulator; Model 5100 Single-Use disposable Maxi-guide Flexible Lightguide; Model 5100-2 (K122593); Precision Endoscopic Technologies, Inc.
  • Precision Endoscopic Infrared Coagulator; Model 5100 Single-Use disposable Maxi-guide . Flexible Lightguide; Model 5100-2 (K083275); MAX Endoscopy, Inc.

Submission Device Description:

This device is designed to coagulate blood and tissue, specifically for treatment of hemorrhoids and small lesions in the colon and rectum, when used as an accessory to a flexible endoscope.

Statement of Intended Use

The PRECISION Endoscopic Infrared Coagulator with single-use disposable MAXiguide flexible light guide is intended for treatment of hemorrhoids and other lesions in the colon and rectum by the transmission of infrared energy when used as an accessory to a colonoscope, flexible sigmoidoscope, or other flexible gastrointestinal endoscope.

Substantial Equivalence Summary

There are no technology differences between the predicate devices and the submission device. This change is a design and labeling change to address customer needs. There are no changes to the intended use or indications for use. The purpose of the devices is to provide infrared energy to coagulate blood and tissue. The predicate devices and the submission device deliver infrared energy from a lamp using a light guide for treatment of hemorrhoids and small lesions in the colon and rectum.

{1}------------------------------------------------

K130489 20/2

Image /page/1/Picture/1 description: The image shows the word "Optim" in bold, black font. To the right and slightly above the "m" is the superscript "LLC". Below the word "Optim" is the phrase "A Juno Company" in a smaller, less bold font. The text is all black against a white background.

Non Clinical Verification Testing

Verification testing results for the design changes were completed and all tests passed. Bench testing was performed to verify that the submission device is substantially equivalent to the predicate devices and does not affect the safety and effectiveness of the device.

Conclusion: The submission device is substantially equivalent and as safe and effective as the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name arranged in a circular pattern around a stylized emblem. The emblem consists of a stylized caduceus, a symbol often associated with medicine and healthcare, with a double-stranded snake winding around a staff.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 19, 2013

Optim, LLC % Mr. Jeff Barrett CEO 64 Technology Park Road STURBRIDGE MA 01566

Re: K130489

Trade/Device Name: PRECISION Endoscopic Infrared Coagulator™ with SINGLE-USE DISPOSABLE MAXI-GUIDE FLEXIBLE LIGHTGUIDE Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: FEH Dated: April 10, 2013 Received: April 16, 2013

Dear Mr. Barrett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{3}------------------------------------------------

Page 2 - Mr. Jeff Barrett

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Optim LLC. The logo is in black and white and features the word "Optim" in a bold, sans-serif font. The letters are closely spaced together, and the "m" is slightly taller than the other letters. Below the word "Optim" is the phrase "A Juno Company" in a smaller font. To the right of the "m" is the letters "LLC" in a smaller font and stacked on top of each other.

Indications for Use

510(k) number (if known): K130489

Device name: PRECISION Endoscopic Infrared Coagulator™ with SINGLE- USE DISPOSABLE MAXI-GUIDE FLEXIBLE LIGHTGUIDE

Statement of Indication of Use

The PRECISION Endoscopic Infrared Coagulator with single-use disposable MAXiguide flexible light guide is intended for treatment of hemorrhoids and other lesions in the colon and rectum by the transmission of infrared energy when used as an accessory to a colonoscope, flexible sigmoidoscope, or other flexible gastrointestinal endoscope.

Prescription Use_ ਮ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use_ (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Benjamin 2013.04.19

K130489

FDA 510(k) Page 17

§ 876.4300 Endoscopic electrosurgical unit and accessories.

(a)
Identification. An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.(b)
Classification. Class II (performance standards).