K Number
K130463

Validate with FDA (Live)

Date Cleared
2013-05-15

(82 days)

Product Code
Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The UCP Home Drug Screening Test Cards, UCP Home Drug Screening Test Cups are rapid, qualitative, competitive binding immunoassays for the detection of the following drugs and their metabolites in human urine: Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Marijuana, Methadone, Methamphetamine, MDMA, Morphine, Opiates 2000, Oxycodone, Phencyclidine, Propoxyphene, Tricyclic Antidepressant. The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The tests are intended for over-the-counter (OTC) users as the first step in a two step process to provide consumers, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing - the second step in the process, along with the materials for shipping the urine specimen to the laboratory, is provided. The test is also intended for prescription use. The tests will yield preliminary positive results when the prescription drugs Barbiturates, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressants are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressant in urine. The tests provide only preliminary data, which should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test results, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring drug levels. For Over-The-Counter (OTC) use and prescription use For In Vitro Diagnostics only.

Device Description

UCP Home Drug Screening Tests are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Amphetamine, Barbiturates, Benzodiapines, Buprenorphine, Cocaines, Marijuana, Methamphetamine, MDMA, Methadone, Opiates, Opiates 300, Oxycodone, Phencyclidine, Propoxyphene, Tricyclic Antidepressants and their metabolites at the cut-off levels as indicated. The tests can be performed without the use of an instrument.

AI/ML Overview

The UCP Home™ Drug Screening Test Cards and Cups are rapid, qualitative, competitive binding immunoassays for the detection of drugs and their metabolites in human urine. The acceptance criteria for this device are based on its ability to correctly identify the presence or absence of various drugs at specific cut-off levels, with a high agreement rate with GC/MS results.

Here's an breakdown of the information requested:

1. Table of Acceptance Criteria and Reported Device Performance

The device is intended to provide preliminary positive results when drug concentrations are at or above the specified cut-off levels. The performance is measured by the agreement rate with confirmatory laboratory methods.

Drug TestCalibratorCut-off (ng/mL)Acceptance Criteria (Agreement with GC/MS)Reported Device Performance (Agreement with GC/MS)
AmphetamineD-Amphetamine1000≥ 97%≥ 97%
BarbituratesSecobarbital300≥ 97%≥ 97%
BenzodiazepinesOxazepam300≥ 97%≥ 97%
BuprenorphineBuprenorphine10≥ 97%≥ 97%
CocaineBenzoylecgonine300≥ 97%≥ 97%
MarijuanaDelta-9-THC-COOH50≥ 97%≥ 97%
MethadoneMethadone300≥ 97%≥ 97%
MethamphetamineD-Methamphetamine1000≥ 97%≥ 97%
MDMAMDMA500≥ 97%≥ 97%
MorphineMorphine300≥ 97%≥ 97%
Opiates 2000Morphine2000≥ 97%≥ 97%
OxycodoneOxycodone100≥ 97%≥ 97%
PhencyclidinePhencyclidine25≥ 97%≥ 97%
PropoxyphenePropoxyphene300≥ 97%≥ 97%
Tricyclic AntidepressantNortriptyline1000≥ 97%≥ 97%

2. Sample Size Used for the Test Set and Data Provenance

The consumer study, which evaluates the device's performance when used by lay persons, involved 115 lay persons.
The urine samples for this study were prepared with various drug concentrations (strong negative, very weak negative, weak negative, very weak positive, weak positive, high positive) by spiking pure drugs or metabolites into drug-free human urine. This suggests the data provenance is prospective as samples were specifically prepared for the study. The country of origin of the data is not explicitly stated, but given the US FDA submission, it is likely the study was conducted in the United States.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The ground truth for the test set (urine samples with various drug concentrations) was established by Gas Chromatography/Mass Spectrometry (GC/MS) for all drugs except Tricyclic Antidepressants (TCA), for which High-Performance Liquid Chromatography (HPLC) was used. This indicates laboratory methods, not human visual experts, established the ground truth for drug concentrations.

4. Adjudication Method for the Test Set

The document does not describe an explicit adjudication method for the comparison between the device results and the GC/MS/HPLC ground truth. The comparison appears to be a direct assessment of agreement between the device's qualitative results and the quantitative ground truth.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The UCP Home™ Drug Screening Test Cards and Cups are in-vitro diagnostic devices designed for direct user interpretation (either by lay users or healthcare professionals). There is no "AI assistance" component to improve human reader performance in this type of qualitative chemical assay. The consumer study evaluated the human user's ability to interpret the test and instructions, not their diagnostic performance with or without AI.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

This is not explicitly applicable in the typical sense of a "standalone algorithm" for image analysis or similar AI applications. However, the device itself, when used by a lay person, functions as a standalone qualitative test. The "performance studies" mentioned (method comparison, specificity, cut-off study) assess the inherent chemical performance of the device itself, independent of user interpretation, through comparison with GC/MS/HPLC. The results for these studies demonstrated the device "performs satisfactorily when used according to the package inserts."

7. The Type of Ground Truth Used

The ground truth used was laboratory gold standard methods:

  • Gas Chromatography/Mass Spectrometry (GC/MS) for Amphetamine, Barbiturates, Benzodiazepines, Buprenorphine, Cocaine, Marijuana, Methadone, Methamphetamine, MDMA, Morphine, Opiates 2000, Oxycodone, Phencyclidine, Propoxyphene.
  • High-Performance Liquid Chromatography (HPLC) for Tricyclic Antidepressants (TCA).

8. The Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of machine learning or AI. This device is an immunoassay, not a software algorithm that requires a training set. The performance studies used prepared urine samples to evaluate the device's analytical performance across different concentrations.

9. How the Ground Truth for the Training Set Was Established

As there is no distinct "training set" in the AI sense, this question is not applicable. The chemical properties and performance of the immunoassay are inherent to its design and manufacturing. The cut-off levels are predefined. The "ground truth" for evaluating its performance (as described in point 7) was established through established laboratory analytical techniques (GC/MS, HPLC).

{0}------------------------------------------------

10. 510(K) SUMMARY

This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR 807.92.

The Assigned 510(k) number is K130463

UCP Biosciences, Inc Submitter: 1445 Koll Circle, Ste 111 San Jose, CA 95014 Tel: 408-392-0064 Fax: 408-392-0163

Date: May 13, 2013

Contact Person: Dr. Nancy Chen

Trade Name: UCP Home™ Drug Screening Test Cards, UCP Home™ Drug Screening Test Cups

Common Name:

Amphetamine Test System Methamphetamine Test System Cocaine Test System Barbiturate Test System Benzodiazepine Test System Buprenorphine Test System Methamphetamine Test System (MDMA) Opiates Test System Methadone Test System Opiates Test System (Oxycodone) Amphetamine Test System (Enzyme Immunoassay Phencyclidine) Cannabinoid Test System Propoxyphene Test System Tricyclic Antidepressant Test System

Product Code: DKZ, DIO, DIS, JXM, DJC, DJR, DJG, LCM, LDJ, LFG, JXN

Regulation Section:

CFR 21 § 862.3100 CFR 21 § 862.3150 CFR 21 & 862.3170 CFR 21 § 862.3250 CFR 21 § 862.3610 CFR 21 § 862.3620 CFR 21 § 862.3650

{1}------------------------------------------------

CFR 21 § 862.3870 CFR 21 § 862.3910 CFR 21 § 862.3700 Unclassified, Enzyme immunoassay, Phencyclidine

Panel: Toxicology (91)

Device Classification: II =

Substantially Equivalent Devices:

UCP Home™ Drug Screening Test Cards/UCP Home™ Drug Screening Test Cups (K091588), UCP Drug Screening Test Cups (K110515)

Product Description:

UCP Home Drug Screening Tests are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Amphetamine, Barbiturates, Benzodiapines, Buprenorphine, Cocaines, Marijuana, Methamphetamine, MDMA, Methadone, Opiates, Opiates 300, Oxycodone, Phencyclidine, Propoxyphene, Tricyclic Antidepressants and their metabolites at the cut-off levels as indicated. The tests can be performed without the use of an instrument.

Intended Use:

UCP Home™ Drug Screening Test Cards, UCP Home™ Drug Screening Test Cups:

The UCP Home Drug Screening Test Cups, UCP Home Drug Screening Test Cups are rapid, qualitative, competitive binding immunoassays for the detection of the following drugs and their metabolites in human urine:

TestCalibratorCut-off
AmphetamineD-Amphetamine1000 ng/mL
BarbituratesSecobarbital300 ng/mL
BenzodiazepinesOxazepam300 ng/mL
BuprenorphineBuprenorphine10 ng/mL
CocaineBenzoylecgonine300 ng/mL
MarijuanaDelta-9-THC-COOH50 ng/mL
MethadoneMethadone300 ng/mL
MethamphetamineD-Methamphetamine1000 ng/mL
MDMAMDMA500 ng/mL
MorphineMorphine300 ng/mL
Opiates 2000Morphine2000 ng/mL
OxycodoneOxycodone100 ng/mL
PhencyclidinePhencyclidine25 ng/mL

{2}------------------------------------------------

PropoxyphenePropoxyphene300 ng/mL
Tricyclic AntidepressantNortriptyline1000 ng/mL

The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The tests are intended for over-the-counter (OTC) use as the first step in a two step process to provide the consumers with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing - the second step in the process, along with the materials for shipping the urine specimen to the laboratory, is provided. The test is also intended for prescription use. The test also is intended for prescription use.

The tests will vield preliminary positive results when the prescription drugs Barbiturates, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressants are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene Tricyclic Antidepressant in urine. The tests provide only preliminary data, which should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS).

Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring the drug levels.

For Over-The-Counter (OTC) use and prescription use For In Vitro Diagnostics only

Comparison to Predicate Devices:

When compared to the predicate (K091588), UCP Home Drug Screening Tests can qualitatively detect Amphetamine, Barbiturates, Benzodiazepine, Cocaine, Marijuana, Methadone, Methamphetamine, MDMA, Morphine, Opiates 2000, Oxycodone, Phencyclidine, Tricyclic Antidepressant and their metabolites in human urine. Both devices utilize the same cutoff concentrations. Both devices are immunochromatographic, lateral flow assays for the qualitative detection of drugs with visual, qualitative end results. Both tests are intended to provide preliminary analytical test results. Both devices are intended for health care professionals use and for OTC consumers use. The UCP Home Drug Screening Tests can detect up to 14 drugs (by also detecting buprenorphine and propoxyphene), whereas the predicates can detect up to 12 drugs.

Safety and Effectiveness Data:

Performance Studies:

{3}------------------------------------------------

The method comparison, precificity (interference and cross-reactivity), cut-off study of each drug test in UCP Home Drug Screening Tests can be found in K091612, K091588, and K061457.

Consumer Studies

The study design and protocol in the consumer study of UCP Home Drug Screening Tests is the same as that described in K091588, K110515, was conducted among 115 lay persons in three geographic regions. Fifty seven females and fifty eight males from ages between 18 and 77 years have participated the study. Fifty seven participants had high school education or less, fifty eight participants had finished college courses. None of the participants had experiences using drug testing products before. The urine samples were prepared to contain strong negative (0% of cutoff), a very weak negative (50% of cutoff), a weak negative (75% of cutoff), a very weak positive (125% of cutoff), a weak positive (150% of cutoff) and high positive (300% of Cutoff). The urine samples with various drug concentrations were prepared by spiking pure drugs or drug metabolites into drug free human urine, the final drug concentrations in each urine sample were confirmed by GC/MS but TCA, TCA concentrations in the urine samples was confirmed by HPLC. The test results performed by the lay users showed 97% or above agreement rate with GC/MS results and indicate the lay users can perform UCP Home Drug Screening Tests satisfactorily by following the test instruction. The post-study survey was conducted to determine if the lay users can understand the test instruction, the meaning of the test " results and how to interpret the test results. Consumers were asked 9 questions including whether the test was easy to run, the results was easy to read, how to interpret the test results, importance of confirmatory test and some medicines and foods may affect the test results. Participant responses support that the lay users can understand how to run the test, interpret the test results, the importance of confirmatory test, and some issues concerning certain prescription medicines and foods may affect the test results.

Other Information about Performance Characteristics:

The performance characteristics of UCP Home Drug Screening Tests including the precision study, sensitivity study, specificity and cross reactivity study, interference study and stability study have been also established. The results have demonstrated that UCP Home Drug Screening Tests performs satisfactorily when used according to the package inserts.

Conclusion:

The performance data in this submission supports UCP Home™ Drug Screening Test Cards and UCP Home"M Drug Screening Test Cups are substantially equivalent to the predicate device UCP Home™ Drug Screening Test Cards/ UCP Home™ Drug Screening Test Cups (K091588).

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail. The eagle is positioned to the right of a circular seal that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 15, 2013

UCP Biosciences, Inc. C/O Ms. Nancy Chen 1445 Koll Circle, Ste. 111 SAN JOSE CA 95112

Re: K130463

Trade/Device Name: UCP Home Drug Screening Test Cards UCP Home Drug Screening Test Cups Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: II Product Code: DKZ, DIO, DIS, JXM, DJC, DJR, DJG, LCM, LDJ, LFG, JXN Dated: March 13, 2013 Received: March 15, 2013

Dear Ms. Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Carol Benson -S
for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K130463

Device Name: UCP Home Drug Screening Test Cards UCP Home Drug Screening Test Cups

Indications for Use:

The UCP Home Drug Screening Test Cards, UCP Home Drug Screening Test Cups are rapid, qualitative, competitive binding immunoassays for the detection of the following drugs and their metabolites in human urine:

TestCalibratorCut-off
AmphetamineD-Amphetamine1000 ng/mL
BarbituratesSecobarbital300 ng/mL
BenzodiazepinesOxazepam300 ng/mL
BuprenorphineBuprenorphine10 ng/mL
CocaineBenzoylecgonine300 ng/mL
MarijuanaDelta-9-THC-COOH50 ng/mL
MethadoneMethadone300 ng/mL
MethamphetamineD-Methamphetamine1000 ng/mL
MDMAMDMA500 ng/mL
MorphineMorphine300 ng/mL
Opiates 2000Morphine2000 ng/mL
OxycodoneOxycodone100 ng/mL
PhencyclidinePhencyclidine25 ng/mL
PropoxyphenePropoxyphene300 ng/mL
Tricyclic AntidepressantNortiptyline1000 ng/mL

The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The tests are intended for over-the-counter (OTC) users as the first step in a two step process to provide consumers, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample.

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Denise Johnson-lyless=55333 Division Sign-Off

Office of In Vitro Diagnostics and Radiological Health

510(k) K130463

{7}------------------------------------------------

Information regarding confirmatory testing - the second step in the process, along with the materials for shipping the urine specimen to the laboratory, is provided. The test is also intended for prescription use.

The tests will yield preliminary positive results when the prescription drugs Barbiturates, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressants are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepines, Buprenorphine, Oxycodone, Propoxyphene, Tricyclic Antidepressant in urine. The tests provide only preliminary data, which should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS).

Clinical considerations and professional judgment should be applied to any drug of abuse test results, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring drug levels.

Prescription Use x(21 CFR Part 801 Subpart D)And/OrOver the Counter Use x(21 CFR Part 801 Subpart C)
------------------------------------------------------------------------------------------------------------------

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Denise Johnson=lyless=55
2013.05.14 12:32:35-04-00'

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

K130463 510(k)

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).