K Number
K130459

Validate with FDA (Live)

Device Name
TRIA FAN SYSTEM
Manufacturer
Date Cleared
2013-10-25

(245 days)

Product Code
Regulation Number
878.4810
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Tria FAN is indicated for the treatment of periorbital wrinkles, which may result in smoother appearing skin in the treated area.

Device Description

The Tria FAN System is a semiconductor diode laser system that delivers infrared light at a wavelength of 1450 nm ± 50 nm.

AI/ML Overview

The provided text describes the Tria FAN System, a diode laser for treating periorbital wrinkles. However, it does not contain detailed acceptance criteria for the device's performance, nor does it provide a specific study with statistical results that prove the device meets such criteria in terms of quantitative metrics (e.g., accuracy, sensitivity, specificity, wrinkle reduction percentage).

The text focuses on demonstrating substantial equivalence to predicate devices based on safety and effectiveness.

Here's an analysis of the provided information against your requested points:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Safety: (Implicitly: no new safety concerns compared to predicates)"The Tria FAN System was found to be as safe...as the predicate non-ablative fractional devices."
Effectiveness: (Implicitly: as effective as predicates for periorbital wrinkles)"The Tria FAN System was found to be as...effective as the predicate non-ablative fractional devices." "indicated for the treatment of periorbital wrinkles, which may result in smoother appearing skin in the treated area."
Substantial Equivalence: (Implicitly: similar intended uses, indications, tech characteristics, and operations)"The Tria FAN System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device." "The minor technological differences...raise no new issues of safety or effectiveness."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: 60 subjects
  • Data Provenance: Prospective, open-label study. Country of origin not specified, but the applicant is based in Dublin, CA, USA.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not provided. The text only states that the study evaluated safety and effectiveness, but it doesn't describe how "effectiveness" (e.g., wrinkle reduction) was objectively measured or by whom.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a laser system, not an AI or diagnostic imaging device involving "readers" or "AI assistance." The clinical study focused on the device's direct effect on patients.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a medical device (laser system), not an algorithm. The clinical study evaluated the device being used by humans.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The "ground truth" for effectiveness in the clinical study would likely be based on subjective or objective assessments of wrinkle severity before and after treatment. However, the specific method (e.g., physician global aesthetic improvement scale, profilometry, patient satisfaction) is not detailed in the provided summary. "Smoother appearing skin" is the reported outcome, implying a visual assessment.

8. The sample size for the training set

Not applicable. This is a medical device, not an AI model that requires a "training set."

9. How the ground truth for the training set was established

Not applicable.

Summary of missing information:

The provided 510(k) summary is high-level and lacks detailed information typically found in a clinical study report, especially regarding:

  • Specific, quantifiable acceptance criteria for effectiveness (e.g., a specific percentage reduction in wrinkle count or depth, or a percentage of subjects achieving a certain improvement score).
  • The methodology for assessing wrinkle improvement (how "smoother appearing skin" was measured and by whom).
  • Any statistical results from the 60-subject clinical study (e.g., p-values, confidence intervals, specific improvement scores). It only states the device was "as safe and effective as the predicate."

{0}------------------------------------------------

K130459

510(k) SUMMARY

Tria Beauty's Tria FAN System

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Tria Beauty, Inc. 4160 Dublin Blvd, Ste 200 Dublin, CA 94568 Phone: 925-452-2500 Facsimile: 925-452-2598 Contact Person: Tobin Island, Ph.D. Date Prepared: September 12, 2013

Name of Device and Name/Address of Sponsor

Tria FAN System Tria Beauty, Inc. 4160 Dublin Blvd, Ste 200 Dublin, CA 94568

Common or Usual Name

OCT 2 5 2013

Diode Laser

Classification Name

Laser Instrument, Surgical, Powered Regulation Number: 21 CFR§878.4810 Product Code: ONG

Predicate Devices

Palomar LOI System (K090525) Cynosure Iluminage Diode Laser (K111454) Cynosure Affirm Laser Family (K080006; K101601) Solta Fraxel Re:Fine Laser (K091420; K060310; K063808; K062303; K070284; K050841; K042319; K040617; K031795)

Intended Use / Indications for Use

Tria FAN is indicated for the treatment of periorbital wrinkles, which may result in smoother appearing skin in the treated area.

Technological Characteristics

The Tria FAN System is a semiconductor diode laser system that delivers infrared light at a wavelength of 1450 nm ± 50 nm.

{1}------------------------------------------------

Performance Data

Performance data was submitted with this 510(k) notification to support the determination of substantial equivalence for the Tria FAN System relative to predicate devices.

Performance Testing (Non-Clinical):

Performance testing was conducted to demonstrate that the Tria FAN System performs according to specifications and functions as intended.

Animal:

A GLP compliant histology study was conducted. Hairless guinea pigs were treated with the Tria FAN System and tissue sampling occurred immediately, 5 days, and 14 days post-treatment. The tissue response was found to be equivalent to the predicate non-ablative fractional devices.

Clinical:

A 60-subject, prospective, open-label safety and effectiveness study was conducted to evaluate the Tria FAN System for the treatment of periorbital wrinkles. The Tria FAN System was found to be as safe and effective as the predicate non-ablative fractional devices.

Substantial Equivalence

The Tria FAN System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the Tria FAN System and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the Tria FAN System is as safe and effective as the predicate devices. Thus, the Tria FAN System is substantially equivalent.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Tria Beauty, Incorporated % Mr. Tobin Island, Ph.D. 4160 Dublin Boulevard. Suite 200 Dublin, California 94568

Re: K130459

Trade/Device Name: Tria Fan Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONG Dated: September 23, 2013 Received: September 23, 2013

Dear Mr. Island:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

Image /page/2/Picture/10 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three wavy lines instead of snakes.

October 25, 2013

{3}------------------------------------------------

Page 2 - Mr. Tobin Island, Ph.D.

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mark Ni Melkerson -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K130459

Device Name: Tria Fan

Indications For Use:

Tria FAN is indicated for the treatment of periorbital wrinkles, which may result in smoother appearing skin in the treated area.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

× Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Neil R Oggen
2013.10.25 113:17:42 -04'00' (Division Sign-Off) for MXM Division of Surgical Devices 510(k) Number K130459

Page 1 of __ 1 __ 1

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.