K Number
K130317

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-12-11

(306 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Genesys Spine Apache™ IBFD Star Cervical System is indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from the C2-C3 disc to the C7-T1 disc. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.

Device Description

The Genesys Spine Apache™ IBFD Star Cervical System will be offered in various device configurations based on surgical approach and patient anatomy, and consist of a Genesys Spine cervical interbody fusion device, which may be implanted as a single device via an anterior approach. The Genesys Spine Interbody Fusion System implant components are made of polyether ether ketone (Invibio PEEK Optima LTI) that conforms to ASTM F2026. Additionally, the devices contain tantalum markers (ASTM F560) to assist the surgeon with proper placement of the device.

AI/ML Overview

The Genesys Spine Apache™ IBFD Star Cervical System is a medical device designed for intervertebral body fusion in the cervical spine. The provided document describes the acceptance criteria and the study conducted to demonstrate the device's substantial equivalence to legally marketed predicate devices.

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Genesys Spine Apache™ IBFD Star Cervical System were based on demonstrating substantial equivalence to predicate devices through non-clinical testing. The performance was evaluated against established ASTM standards for intervertebral body fusion devices.

Acceptance Criteria (based on ASTM Standards)Reported Device Performance
Static Testing (ASTM F2077-11)Device was tested in static axial compression and torsion.
Dynamic Testing (ASTM F2077-11)Device was tested in dynamic axial compression and torsion.
Subsidence Testing (ASTM F2267-04)Device was tested for static subsidence.
Expulsion Testing (ASTM draft standard F-04.25.02.02)Device was tested for expulsion.
Material ConformanceImplants made of PEEK Optima LTI conform to ASTM F2026.
Tantalum MarkersTantalum markers conform to ASTM F560.

Conclusion: The reported device performance indicates that the Genesys Spine Apache™ IBFD Star Cervical System met the acceptance criteria by successfully undergoing all specified non-clinical tests according to the relevant ASTM standards. The overall technological characteristics and mechanical performance data led to the conclusion of substantial equivalence to the predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

This submission pertains to a non-clinical study for a medical implant, not a clinical study involving patient data. Therefore, the concepts of "sample size for the test set," "data provenance (e.g., country of origin of the data, retrospective or prospective)," "number of experts used to establish the ground truth," and "adjudication method" are not applicable in the context of this device's performance evaluation. The "test set" here refers to the physical devices and materials subjected to mechanical and material tests.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable, as this was a non-clinical evaluation of mechanical and material properties against standardized tests, not a clinical study requiring expert interpretation of patient data. The "ground truth" is defined by the ASTM standards themselves and the measurable outcomes of the physical tests.

4. Adjudication Method for the Test Set

Not applicable, as this was a non-clinical evaluation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

Not applicable. This is a medical implant, not an AI-powered diagnostic or imaging device. Therefore, no MRMC study, human readers, or AI assistance is relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This is a medical implant, not an algorithm or software. Its performance is evaluated through physical and material testing.

7. The Type of Ground Truth Used

The "ground truth" for the performance evaluation of this device was established based on pre-defined engineering standards and material specifications:

  • ASTM F2077-11: Standard Test Methods for Intervertebral Body Fusion Devices. This standard defines the acceptable mechanical performance (static and dynamic axial compression and torsion).
  • ASTM F2267-04: Standard Test Method for Measuring Load Induced Subsidence of an Intervertebral Body Fusion Device Under Static Axial Compression. This standard defines the acceptable subsidence performance.
  • ASTM draft standard F-04.25.02.02: Related to expulsion testing (specific details of the draft standard not provided, but it serves as the ground truth for expulsion resistance).
  • ASTM F2026: Standard Specification for Polyetheretherketone (PEEK) Polymers for Surgical Implant Applications. This standard defines the material properties for the PEEK components.
  • ASTM F560: Standard Specification for Unalloyed Tantalum for Surgical Implant Applications. This standard defines the material properties for the tantalum markers.

The device's performance against these quantified, objective standards represents the "ground truth."

8. The Sample Size for the Training Set

Not applicable. This is a non-clinical evaluation of a physical medical implant, not a machine learning model. Therefore, no training set for an algorithm is involved.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set involved in the performance evaluation of this physical device.

{0}------------------------------------------------

DEC 1 1 2013

5.510(K) SUMMARY

Submitter's Name:Genesys Spine
Submitter's Address:1250 Capital of Texas Highway SouthBuilding Three, Suite 600Austin, Texas 78746
Submitter's Telephone:512-381-7080
Submitter's Fax:800-817-4938
Contact Name:William W. Sowers
Date Summary wasPrepared:4-Feb-13
Trade or Proprietary Name:Genesys Spine Apache™ IBFD Star Cervical System
Common or Usual Name:Intervertebral Body Fusion Device, Cervical
Classification:Class II per 21 CFR §888.3080
Product Codes:ODP
Classification Panel:Orthopedic and Rehabilitation Devices Panel
Legally Marketed(unmodified) device:Genesys Spine Apache™ Cervical Interbody System(K103034)BAK/Cervical Interbody Fusion System (P980048)Eminent Spine Copperhead IBFD (K090064)

Description of the Device Subject to Premarket Notification:

The Genesys Spine Apache™ IBFD Star Cervical System will be offered in various device configurations based on surgical approach and patient anatomy, and consist of a Genesys Spine cervical interbody fusion device, which may be implanted as a single device via an anterior approach.

INDICATIONS FOR USE

The Genesys Spine Apache™ IBFD Star Cervical System is indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from the C2-C3 disc to the C7-T1 disc. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.

{1}------------------------------------------------

TECHNICAL CHARACTERISTICS

The Genesys Spine Apache™ IBFD Star Cervical System is comprised of various device configurations designed to accommodate patient anatomy and provide the surgeon with different surgical approach options.

The Genesys Spine Interbody Fusion System implant components are made of polyether ether ketone (Invibio PEEK Optima LTI) that conforms to ASTM F2026. Additionally, the devices contain tantalum markers (ASTM F560) to assist the surgeon with proper placement of the device. The additional implant offering being proposed has similar technological characteristics and identical indications as the currently cleared product line.

Results of non-clinical testing relative to static and dynamic testing per ASTM F2077-03, subsidence testing per ASTM F2267-04, and expulsion testing per ASTM draft standard F-04.25.02.02 were used to determine substantial equivilence.

PERFORMANCE DATA

The Genesys Spine Apache™ IBFD Star Cervical System was tested in static and dynamic axial compression and torsion per ASTM F2077-11, static subsidence per ASTM F2267-04, and expulsion testing per ASTM draft standard F-04.25.02.02.

CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that Genesys Spine Apache™ IBFD Star Cervical System is substantially equivalent to the Genesys Spine Apache™ Cervical Interbody System, the Centerpulse BAK/Cervical Interbody Fusion System, and the Eminent Spine Copperhead IBFD

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle emblem, with its wings forming a distinctive, flowing design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 11, 2013

Genesys Spine Mr. William W. Sowers Vice President of Quality and Regulatory 1250 Capital of Texas Highway South, Building Three, Suite 600 Austin, Texas 78746

Re: K130317

Trade/Device Name: Genesys Spine Apache™ IBFD Star Cervical System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP Dated: October 25, 2013 Received: November 1. 2013

Dear Mr. Sowers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. William W. Sowers

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYowIndustry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part . 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Ronaldingean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

4. INDICATIONS FOR USE STATEMENT

Device Name: Genesys Spine Apache™ IBFD Star Cervical System

The Genesys Spine Apache™ Interbody Fusion Device Star Cervical System is indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from the C2-C3 disc to the C7-T1 disc. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Patients should have at least six (6) weeks of nonoperative treatment prior to treatment with an intervertebral cage.

Prescription Use ਮ (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Anton E. Dmitriev, PhD Division of Orthopedic Devices

Genesys Spine Apache™ IBFD Star Cervical System

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.