K Number
K130306

Validate with FDA (Live)

Date Cleared
2013-04-11

(63 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Incite Anchored Cervical Interbody Device is indicated for use in skeletally mature patients with degenerative disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The Incite Anchored Cervical Interbody Device is to be used with autogenous bone graft and implanted via an open, anterior approach. Supplemental fixation, i.e. an anterior cervical plate, is required to properly utilize this system.

Device Description

The Incite Anchored Cervical Interbody Fusion Device acts as a spacer to maintain proper Intervertebral and vertebral body spacing and angulation. The Incite Anchored Cervical Interbody Fusion Device is manufactured from PEEK and Ti6A14V titanium alloy with tantalum radiopaque markers

AI/ML Overview

The provided text describes a Special 510(k) Summary for a medical device called the "Incite Anchored Cervical Interbody Fusion (ACI) Device." This type of submission focuses on demonstrating substantial equivalence to a predicate device, often for minor modifications rather than entirely new devices requiring extensive clinical trials for efficacy.

Based on the provided document, here's a breakdown of the acceptance criteria and the study used:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Mechanical Performance: Static CompressionDeemed applicable from predicate device testing (K122008) because the new size (16x14mm) "does not introduce a new worst case condition."
Mechanical Performance: Static TorsionDeemed applicable from predicate device testing (K122008) because the new size (16x14mm) "does not introduce a new worst case condition."
Mechanical Performance: SubsidenceDeemed applicable from predicate device testing (K122008) because the new size (16x14mm) "does not introduce a new worst case condition."
Mechanical Performance: ExpulsionDeemed applicable from predicate device testing (K122008) because the new size (16x14mm) "does not introduce a new worst case condition."
Material CompositionSame as the currently cleared predicate device: PEEK and Ti6A14V titanium alloy with tantalum radiopaque markers. (Demonstrated by simple comparison)
Intended UseSame as the currently cleared predicate device. (Demonstrated by comparison in the document).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document explicitly states that "additional mechanical testing on the new size is not warranted." This means there was no new physical test set specifically for the 16x14mm device being cleared in this 510(k). The "test set" for the performance criteria refers to the testing data from the predicate device (K122008). The document does not specify the sample size (e.g., number of devices tested for static compression, torsion, etc.) used for the predicate device's original testing.
  • Data Provenance: The data provenance is retrospective, as it relies entirely on the performance testing previously conducted and accepted for the predicate device (K122008). The country of origin of the data is not specified, but it would have been generated as part of the predicate device's original regulatory submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • This information is not applicable and not provided in the document. The study is a mechanical performance comparison based on engineering principles and previous device testing, not a clinical or diagnostic study involving human expert interpretation for ground truth.

4. Adjudication Method for the Test Set

  • This information is not applicable and not provided. As above, this is a mechanical performance study, not a study requiring adjudication of expert opinions.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic or AI-assisted interpretation devices to assess the impact on human reader performance. This submission is for an implantable medical device, and the demonstration of substantial equivalence relies on mechanical performance data and comparison of physical attributes.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • No, a standalone performance study was not done. This device is a physical implant, not an algorithm or AI system.

7. The Type of Ground Truth Used

  • The "ground truth" for the mechanical performance aspects is based on engineering and biomechanical testing standards and the results obtained from the predicate device's testing against those standards. It is not expert consensus, pathology, or outcomes data. The ground for "substantial equivalence" is the direct comparison of design, materials, and intended use to the predicate device, along with the reasoning that the new size does not introduce a "new worst case condition."

8. The Sample Size for the Training Set

  • This information is not applicable and not provided. There is no "training set" in the context of this device and its regulatory submission, as it is not an AI/ML algorithm.

9. How the Ground Truth for the Training Set Was Established

  • This information is not applicable and not provided. As there is no training set, there is no ground truth to establish for it.

In summary:

The study presented here is a Special 510(k) submission based on substantial equivalence to a predicate device (K122008). The "study" for acceptance criteria is primarily a justification based on engineering principles and previously accepted mechanical test data from the predicate device. The key argument is that the dimensional differences of the new device (a larger footprint 16x14mm) do not create a "new worst-case condition" compared to the predicate device, thus making the predicate's performance data applicable to the new size. No new testing was performed for this specific submission for the new size.

{0}------------------------------------------------

SPECIAL 510(K) SUMMARY

APR 1 1 2013

Submitter Information:

Submitter's Name:Address:Incite Innovation LLCP.O. Box 153881500 Main Street, Ste 2410Springfield, MA 01115-5707
Telephone:413-382-0212
Contact Person:John Kirwan
Date Prepared:February 6, 2013
Device Trade Name:Incite Anchored Cervical Interbody Fusion (ACI) Device
Common/Usual Name:Spinal Intervertebral body fixation orthosis
Classification:21 CFR §888.3080
Class:II
Product Code:OVE
Predicate Device:Incite Anchored Cervical Interbody Device K122008

Intended Use:

The Incite Anchored Cervical Interbody Device is indicated for use in skeletally mature patients with degenerative disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The Incite Anchored Cervical Interbody Device is to be used with autogenous bone graft and implanted via an open, anterior approach. Supplemental fixation, i.e. an anterior cervical plate, is required to properly utilize this system.

Device Description:

The Incite Anchored Cervical Interbody Fusion Device acts as a spacer to maintain proper Intervertebral and vertebral body spacing and angulation. The Incite Anchored Cervical Interbody Fusion Device is manufactured from PEEK and Ti6A14V titanium alloy with tantalum radiopaque markers

{1}------------------------------------------------

Comparison to Predicate Device(s):

The indication for use and m aterial composition of the 16mm wide x14mm deep Incite Anchored Cervical Interbody Fusion (ACI) Device are the sam e as the currently cleared predicate device, Incite Anchored Cervi cal Interbody F usion (ACI) Device. The only difference is the addition of a larger footprint implant, 16x14mm:

Incite Anchored Cervical Interbody Fusion(ACI) Device(K122008)(W x D)Incite Anchored Cervical Interbody Fusion(ACI) Device(W x D)
EXISTING SIZES: 14x14mm, 14x12mmNEW SIZE: 16 x 14mm

The dimensional differences between the subject device and the predicate are not considered significant because the new size does not introduce a n ew worst case conditio n (e.g. introduction of a ne w region with decreased cro ss sectional area compared to the predicate) and therefore does not raise new questions regarding safety and effectiveness of the device.

Performance Standards:

In consideration of design control activities including risk analysis the non-clinical performance testing performed on the predicate device. Incite Anchored Cervical Interbody Fusion (ACI) Device (K122008), including static compression, static torsion, subsidence, and expulsion are applicable in the characterization of the new size because the new size does not introduce a new worst case condition (e.g. introduction of a new region with decreased cross sectional area compared to the predicate), therefore additional mechanical testing on the new size is not warranted.

Performance and SE Determination:

Based on the supporting documentation within this premarket notification, the subject device demonstrates substantial equivalence to the listed predicate device.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure with three flowing lines extending from the figure's head.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 11, 2013

Incite Innovations, LLC % Mr. John Kirwan President P.O. Box 15388 1500 Main Street, Suite 2410 Springfield, Massachusetts 01115

· Re: K130306

Trade/Device Name: Incite Anchored Cervical Interbody Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVE Dated: March 14, 2013 Received: March 15, 2013

Dear Mr. Kirwan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Mr. John Kirwan

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/3/Picture/7 description: The image shows the name "Erin J. Keith" in a stylized, bold font. The letters have a textured or outlined effect, giving them a distinct appearance. The name is presented in a clear and legible manner.

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K130306

The Incite Anchored Cervical Interbody Device is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The Incite Anchored Cervical Interbody Device is to be used with autogenous bone graft and implanted via an open, anterior approach. Supplemental fixation, i.e. an anterior cervical plate, is required to properly utilize this system.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Anton E. Dmitriev, PhD Division of Orthopedic Devices

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.