K Number
K130284

Validate with FDA (Live)

Date Cleared
2013-12-20

(318 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The On Call Sharp Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood from the fingertips, forearm and palm by people with diabetes at home as an aid in monitoring the effectiveness of diabetes control programs. Alternative site testing should be done only during steady-state times (when blood glucose level is not changing rapidly). The On Call Sharp Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared. It is for in vitro diagnostic use only.

The On Call Sharp Blood Glucose Monitoring System in not intended for the diagnosis of or screening for diabetes, nor intended for use on neonates.

The On Call Sharp Blood Glucose Test Strips are used with the On Call Sharp Blood Glucose Meter in the quantitative measurement of glucose in capillary blood from the fingertips, forearm and palm."

The On Call Sharp Blood Glucose Control Solutions are for use with the On Call Sharp Blood Glucose Meter and On Call Sharp Blood Glucose Test Strips to check that the meter and test strips are working together properly and the test is performing correctly.

For In Vitro Diagnostic Use

Device Description

The On Call Sharp Blood Glucose Monitoring System is a quantitative assay for the detection of glucose in capillary whole blood sampled from the fingertip, palm and forearm. The glucose measurement is achieved by using the amperometric detection method.

The test strip has a reagent system including flavin adenine dinucleotide-glucose dehydrogenase (GDH-FAD) enzyme chemistry and a mediator that reacts with glucose in the whole blood sample to produce an electrical current. This current is measured by the meter, and after calculation by the meter, the blood glucose concentration reading is displayed on the meter display, calibrated to a plasma reference.

AI/ML Overview

The provided text describes the "On Call Sharp Blood Glucose Monitoring System" and its performance evaluation. Here's a summary of the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

The submission mentions compliance with EN ISO 15197:2003 "In vitro diagnostic test systems - Requirements for in vitro whole blood glucose monitoring systems intended for use by patients for self testing in management of diabetes mellitus." This standard typically sets the accuracy requirements for blood glucose monitoring systems. While the exact numerical criteria from ISO 15197 are not explicitly stated in the provided text, the conclusion affirms that the device "meets the accuracy requirements per EN ISO 15197."

Based on the information provided, a table detailing specific acceptance criteria and reported device performance from the clinical study, particularly regarding accuracy, would typically present comparisons between device readings and a reference method (like YSI Model 2300 STAT PLUS). However, the document only broadly states that the device "meets the accuracy requirements per EN ISO 15197."

General Features and Specifications (as acceptance criteria for technical characteristics):

FeatureAcceptance Criteria (Specification)Reported Device Performance (as per specification)
Measurement Range20 to 600 mg/dL (1.1-33.3 mmol/L)20 to 600 mg/dL (1.1-33.3 mmol/L)
Result CalibrationPlasma-equivalentPlasma-equivalent
SampleFresh capillary whole bloodFresh capillary whole blood
Minimum Sample Size0.8 µL0.8 µL
Test Time5 seconds5 seconds
Battery LifeMinimum of 1,000 measurementsMinimum of 1,000 measurements
Hematocrit Range25-70%25-70%
Operating Temperature10-45°C (50-113°F)10-45°C (50-113°F)
Operating Relative Humidity10-90% (non-condensing)10-90% (non-condensing)
CodingAuto Coding by meter automatic recognitionAuto Coding by meter automatic recognition
Meter MemoryUp to 500 records with time and dateUp to 500 records with time and date
User SatisfactionUsers satisfied with ease of operation and overall performanceAcknowledged via lay person questionnaire

The specific accuracy data (e.g., mean bias, percentage of readings within certain error margins) required by ISO 15197 is not detailed in this summary. The summary only concludes that the device meets these requirements.

2. Sample size used for the test set and the data provenance

  • Sample Size: The document does not specify the exact numerical sample size (number of participants or measurements) used in the clinical studies for the test set. It mentions studies were conducted with "lay persons and trained laboratory technicians."
  • Data Provenance: The document does not explicitly state the country of origin. It indicates that the studies were performed using the ACON Clinical Study Protocol for the Blood Glucose Monitoring System, implying an internal or sponsored study. The clinical studies were performed to evaluate system accuracy compared to a reference method. It's a prospective study as it involves conducting clinical studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: The document mentions "trained laboratory technicians" performed comparative readings. It does not specify the number of technicians involved.
  • Qualifications of Experts: The experts are described as "trained laboratory technicians." Specific qualifications (e.g., years of experience, certifications) are not provided.

4. Adjudication method for the test set

The document does not describe a formal adjudication method (e.g., 2+1, 3+1). The "trained laboratory technicians" used the YSI Model 2300 STAT PLUS as a reference method. The comparison was between the device readings by lay persons and the results obtained by these technicians using the reference device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no indication that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was performed, nor is there any mention of AI assistance. This device is a standalone blood glucose monitoring system, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device itself is a standalone system for measuring blood glucose. The performance evaluation includes both its internal accuracy (laboratory testing) and its use by "lay persons" (human-in-the-loop). The "algorithm only" performance would be equivalent to the device's accuracy when operated correctly, which is implied by the laboratory testing and comparison to the YSI reference. The clinical study also involved assessing accuracy when used by lay persons, which is a key aspect of real-world "human-in-the-loop" performance for a self-monitoring device.

7. The type of ground truth used

The ground truth for the accuracy evaluation was established using a reference method, specifically the YSI Model 2300 STAT PLUS (K913806), which is a laboratory-grade analyzer for glucose measurement.

8. The sample size for the training set

The document does not provide information about a separate "training set" or its sample size. For medical devices like this, performance is typically evaluated against established standards and reference methods in validation studies, rather than a machine learning training/test set paradigm.

9. How the ground truth for the training set was established

As no specific "training set" for a machine learning algorithm is mentioned, information on how its ground truth was established is not applicable or provided. The device relies on established electrochemical principles and calibration against reference methods.

{0}------------------------------------------------

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The Assigned 510(k) number is K130284

Submitter's Identification:

ACON Laboratories, Inc.

10125 Mesa Rim Road

San Diego. California 92121

Tel .: 858-875-8019 Fax: 858-875-8099 Date Prepared: December 12, 2013

Contact Person:

Qiyi Xie Senior Staff, Clinical & Regulatory Affairs

Proprietary Name of the Device:

On Call Sharp Blood Glucose Monitoring System

Common Name:

Glucose Test System

ಡಿ

Classification Name:

Class II §862.1345 Glucose Test System

Predicate Device:

On Call® Vivid Blood Glucose Monitoring System ACON Laboratories. Inc., located at 10125 Mesa Rim Road. CA 92121. USA. 510(k) Number: K112653

Device Name: On Call Sharp Blood Glucose Monitoring System

Proprietary NameClassificationProductCodeDescriptionCommon Name
On Call SharpBlood GlucoseMonitoring System862.1345Class II75 NBWSystem, Test,Blood Glucose,PrescriptionGlucose TestSystem

{1}------------------------------------------------

On Call® SharpBlood GlucoseMeter and On Call®Sharp BloodGlucose Test Strips862.1345Class II75 LFRGlucose MonitorGlucose Meter &Test Strips
On Call® SharpGlucose ControlSolution862.1660Class I,reserved75 JJXSingle AnalyteControlControl Solution

Description:

The On Call Sharp Blood Glucose Monitoring System is a quantitative assay for the detection of glucose in capillary whole blood sampled from the fingertip, palm and forearm. The glucose measurement is achieved by using the amperometric detection method.

The test strip has a reagent system including flavin adenine dinucleotide-glucose dehydrogenase (GDH-FAD) enzyme chemistry and a mediator that reacts with glucose in the whole blood sample to produce an electrical current. This current is measured by the meter, and after calculation by the meter, the blood glucose concentration reading is displayed on the meter display, calibrated to a plasma reference.

Intended Use:

The On Call Sharp Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood from the fingertips. forearm and palm by people with diabetes at home as an aid in monitoring the effectiveness of diabetes control programs. Alternative site testing should be done only during steady-state times (when blood glucose level is not changing rapidly). The On Call Sharp Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared. It is for in vitro diagnostic use only.

The On Call Sharp Blood Glucose Monitoring System in not intended for the diagnosis of or screening for diabetes, nor intended for use on neonates.

The On Call Sharp Blood Glucose Test Strips are used with the On Call Sharp Blood Glucose Meter in the quantitative measurement of glucose in capillary blood from the fingertips, forearm and palm."

The On Call Sharp Blood Glucose Control Solutions are for use with the On Call Sharp Blood Glucose Meter and On Call Sharp Blood Glucose Test Strips to check that the meter and test strips are working together properly and the test is performing correctly.

For In Vitro Diagnostic Use

{2}------------------------------------------------

Technological Characteristics:

.

FeatureSpecification
Measurement Range20 to 600 mg/dL (1.1-33.3 mmol/L)
Result CalibrationPlasma-equivalent
SampleFresh capillary whole blood
Minimum SampleSize0.8 µL
Test Time5 seconds
Power SourceTwo (2) CR 2032 3.0V coin cell batteries
Battery LifeMinimum of 1,000 measurements (without considering datatransfer and test reminder alarms)
Glucose Units ofMeasureThe meter is pre-set at time of manufacturing to either millimolesper liter (mmol/L) or milligrams per deciliter (mg/dL) dependingon the standard of your country. The meter will be set to mg/dLby default when sold in the United States.
MemoryUp to 500 records with time and date
Meter Size3.58" x 2.28" x 0.83"
Display Size1.58" x 1.42"
WeightApproximately 60 g (without battery installed)
OperatingTemperature10-45°C (50-113°F)
Operating RelativeHumidity10-90% (non-condensing)
Hematocrit Range25-70%
Data Port9600 baud, 8 data bits, 1 stop bit, no parity

Specification of Blood Glucose Meter:

{3}------------------------------------------------

Comparison to Predicate Devices:

The On Call Sharp Blood Glucose Monitoring System is substantially equivalent to
The On Call Vivid Blood Glucose Monitoring System. K I 12653.

FeaturesOn Call® Sharp Blood GlucoseMonitoring SystemOn Call® Vivid Blood GlucoseMonitoring System (K112653)
Similarities
Result CalibrationPlasma-equivalentSame
Test Time5 secondsSame
Sample TypeFresh capillary whole bloodSame
Glucose Units ofMeasuremg/dLSame
Operating RelativeHumidity10-90%Same
Data PortOne Serial data portSame
MeasurementRange20 to 600 mg/dL (1.1-33.3mmol/L)Same
Automatic ShutoffTwo minutes after last user actionSame
Battery LifeMinimum of 1,000 measurements(without considering data transferand test reminder alarms)Same
Minimum SampleSize0.8 µLSame
CodingAuto Coding by meter automaticrecognition of the intended codingafter strip insertionSame
Meter MemoryUp to 500 records with time anddateSame
Power SourceTwo (2) CR 2032 3.0 V coin cellbatteriesSame
Meter Size3.53" x 2.28" x 0.85"(89.6mm x 58mm x 21.7mm)Same
Meter WeightApprox. 60 g (with batteryinstalled)Same
Differences
Hematocrit Range25-70%20-70%
OperatingTemperature10-45°C (50-113°F)5-45°C (41-113°F)
Assay MethodFAD-dependent GlucoseDehydrogenase biosensorGlucose Oxidase biosensor
Meter DisplayBacklightNoYes
Meter Strip PortLightNoYes

{4}------------------------------------------------

Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

Guidance documents included the *FDA Guidance for Industry In Vitro Diagnostic Glucose Test System" and "FDA Guidance for Total Product Life Cycle for Portable Invasive Blood Glucose Monitoring Systems" as well as "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

Compliance to applicable voluntary standards includes EN ISO 15197:2003 "In vitro diagnostic test systems - Requirements for in vitro whole blood glucose monitoring systems intended for use by patients for self testing in management of diabetes mellitus."

Laboratory Testing:

The performance characteristics of the On Call Sharp Blood Glucose Monitoring System were evaluated by performing the following studies: repeatability precision, intermediate precision, linearity, interfering agents, hematocrit effect, temperature effect evaluation - blood & control solution. low battery effect, altitude effect, sample volume, humidity effect, simulated shipping study - test strip & control solution, control value assignment, meter testing, software validation testing, electromagnetic compatibility and electrical safety testing as part of meter and strip validation testing.

Discussion of Clinical Tests Performed:

Clinical studies were conducted with lay persons and trained laboratory technicians using the On

Call Sharp Blood Glucose Monitoring System. The study data were presented evaluating the system accuracy of the On Call Sharp Blood Glucose Monitoring System compared to the YSI Model 2300 STAT PLUS (K913806) per the ACON Clinical Study Protocol for the Blood Glucose Monitoring System. Study results indicate that nonprofessional, inexperienced lay persons were able to obtain comparable blood glucose readings when using the On Call Sharp Blood Glucose Monitoring System as compared to the results obtained by the trained technicians. In addition, the participating lay persons were questioned and responded as satisfied with the ease of operation by following the Instructions for Use in the User's Manual and the overall performance of the On Call Sharp Blood Glucose Monitoring System.

Conclusion:

The laboratory testing and clinical study results demonstrate that the On Call Sharp Blood Glucose Monitoring System is safe, effective and easy-to-use. It also demonstrates that the On Call Sharp Blood Glucose Monitoring System meets the accuracy requirements per EN ISO 15197 and as such is substantially equivalent to the On Call® Vivid Blood Glucose Monitoring System, currently sold on the U.S. market (K112653).

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 20, 2013

ACON LABORATORIES, INC. QIYI VIE, M.D., MPH 10125 MESA RIM ROAD SAN DIEGO CA 92121

Re: K130284

Trade/Device Name: On Call Sharp Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, LFR, JJX Dated: December 18, 2013 Received: December 18, 2013

Dear Dr. Vie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

Page 2-Dr. Vie

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDeyices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Carol C. Benson -S for

Courtney Lias, Ph. D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known) k 130284

Davice Name

. •

On Call Sharp Blood Glucose Monitoring System

Indications for Use (Describe)

The On Call Sharp Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood from the fingertips, forearm and palm by people with disbetes at home as an aid in monitoring the effectivencss of diabetes control programs. Altemative site testing should be done only daring steady-state times (when blood glucose level is not changing rapidly). The On Call Sharp Blood Gluccee Monitoring System is intended to be used by a single patient and should not be shared. It is for in vitro diagnostic use only.

The On Call Sharp Blood Glucose Monitoring System in not intended for the diagnosis of or samended for use on neonates.

The On Call Sharp Blood Glucose Test Strips are used with the On Call Sharp Blood Glucose Meter in the quantitative measurement of glucose in capillary blood from the fingertips, forearm and palm."

The On Call Sharp Blood Glucose Control Solutions are for use with the On Call Sharp Blood Glucose Meter and On Call Sharp Blood Glucose Test Strips to check that the meter and test strips are working together properly and the lest is performing correctly,

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A BEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CORH) (Signature)

Stavce Beck

FORM FDA 3881 (9/13)

Page 1 of J

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.