K Number
K130268

Validate with FDA (Live)

Date Cleared
2013-08-23

(200 days)

Product Code
Regulation Number
866.3860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD ProbeTec™ Trichomonas vaginalis (TV) Q* Amplified DNA Assay, when tested with the BD Viper™ System in Extracted Mode, uses Strand Displacement Amplification technology for the direct, qualitative detection of Trichomonas vaginalis DNA in clinician-collected female endocervical swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and female urine specimens. The assay is indicated for use with asymptomatic and symptomatic females to aid in the diagnosis of trichomoniasis.

Device Description

The BD ProbeTec Trichomonas vaginalis (TV) Q* Amplified DNA Assay (TVQ Assay) is based on the simultaneous amplification and detection of target DNA using amplification primers and fluorescently-labeled detector probes. The reagents for SDA are dried in two separate disposable microwells: the Priming Microwell contains the amplification primers, fluorescently-labeled detector probe, nucleotides and other reagents necessary for amplification, while the Amplification Microwell contains the two enzymes (a DNA polymerase and a restriction endonuclease) that are required for SDA. The BD Viper™ System pipettes a portion of the purified DNA solution from each Extraction Tube into a Priming Microwell to rehydrate the contents. After a brief incubation, the reaction mixture is transferred to a corresponding, prewarmed Amplification Microwell which is sealed to prevent contamination and then incubated in one of the two thermally-controlled fluorescent readers. The presence or absence of Trichomonas vaginalis DNA is determined by calculating the peak fluorescence (Maximum Relative Fluorescent Units (MaxRFU) over the course of the amplification process and by comparing this measurement to a predetermined threshold value.

In addition to the fluorescent probe used to detect amplified TV target DNA, a second labeled oligonucleotide is incorporated in each reaction Control (EC) oligonucleotide is labeled with a different dye than that used for detection of the TV specific target and is used to confirm the validity of the extraction process. The EC is dried in the Extraction Tubes and is rehydrated upon addition of the specimen and extraction reagents. At the end of the extraction process, the EC fluorescence is monitored by the BD Viper System and an automated algorithm is applied to both the EC and target-specific signals to report results as positive, negative, or EC failure.

AI/ML Overview

BD ProbeTec™ Trichomonas Vaginalis (TV) Q* Amplified DNA Assay - Acceptance Criteria and Study Details

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state pre-defined acceptance criteria for clinical performance (sensitivity and specificity). However, it reports the clinical performance observed in the study. Based on the "Conclusion" section stating that the "analytical and clinical study results... support the determination of substantial equivalence in accordance with the intended use," it can be inferred that the reported performance met the implicit or explicit requirements for regulatory approval.

MetricSpecimen Type (Status)Reported Performance (95% CI)Implied Acceptance Criteria (Based on Approval)
SensitivityNeat Urine (Asymptomatic)93.1% (78.0%, 98.1%)High sensitivity, likely >90%
Neat Urine (Symptomatic)96.4% (89.9%, 98.8%)High sensitivity, likely >90%
Neat Urine (Total)95.5% (90.0%, 98.1%)High sensitivity, likely >90%
Vaginal (Asymptomatic)93.5% (79.3%, 98.2%)High sensitivity, likely >90%
Vaginal (Symptomatic)100.0% (95.7%, 100.0%)Very high sensitivity, likely >95%
Vaginal (Total)98.3% (93.9%, 99.5%)Very high sensitivity, likely >95%
Endocervical (Asymptomatic)92.2% (81.5%, 96.9%)High sensitivity, likely >90%
Endocervical (Symptomatic)98.8% (93.5%, 99.8%)Very high sensitivity, likely >95%
Endocervical (Total)96.3% (91.6%, 98.4%)Very high sensitivity, likely >95%
All Specimen Types (Asymptomatic)92.8% (86.4%, 96.3%)High sensitivity, likely >90%
All Specimen Types (Symptomatic)98.4% (96.0%, 99.4%)Very high sensitivity, likely >95%
Overall96.7% (94.3%, 98.1%)Very high sensitivity, likely >95%
SpecificityNeat Urine (Asymptomatic)99.6% (97.9%, 99.9%)Very high specificity, likely >98%
Neat Urine (Symptomatic)98.1% (96.1%, 99.1%)Very high specificity, likely >98%
Neat Urine (Total)98.7% (97.5%, 99.3%)Very high specificity, likely >98%
Vaginal (Asymptomatic)99.0% (97.2%, 99.7%)Very high specificity, likely >98%
Vaginal (Symptomatic)99.0% (97.5%, 99.6%)Very high specificity, likely >98%
Vaginal (Total)99.0% (98.0%, 99.5%)Very high specificity, likely >98%
Endocervical (Asymptomatic)99.1% (97.8%, 99.7%)Very high specificity, likely >98%
Endocervical (Symptomatic)99.8% (98.6%, 100.0%)Very high specificity, likely >98%
Endocervical (Total)99.4% (98.6%, 99.8%)Very high specificity, likely >98%
All Specimen Types (Asymptomatic)99.2% (98.5%, 99.6%)Very high specificity, likely >98%
All Specimen Types (Symptomatic)99.0% (98.2%, 99.4%)Very high specificity, likely >98%
Overall99.1% (98.6%, 99.4%)Very high specificity, likely >98%

2. Sample Size for the Test Set and Data Provenance

  • Test Set Sample Size:
    • Urine specimens: 735 compliant subjects
    • Vaginal specimens: 832 compliant subjects
    • Endocervical specimens: 995 compliant subjects
    • All specimen types combined: 2568 compliant subjects (This refers to total tests, not unique subjects)
  • Data Provenance: Prospective collection from subjects attending family planning, OB/GYN, and sexually transmitted disease clinics at seven geographically diverse clinical sites in North America.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document specifies the use of a "composite reference method" for establishing ground truth, consisting of Wet Mount and InPouch TV Culture. These are laboratory-based methods, not expert human readers for image interpretation. Therefore, the concept of "number of experts" and "qualifications of those experts" as it might apply to an AI image recognition device is not directly applicable here. The ground truth relies on the performance of these established microbiological diagnostic techniques.

4. Adjudication Method for the Test Set

The document describes a "composite reference method" for establishing ground truth. The first two clinician-collected vaginal swabs were randomized: one for Wet Mount and the other for InPouch TV Culture. While the specific adjudication rules are not detailed (e.g., if one test was positive and the other negative, how a "composite" result was derived), the setup implies that the results from these two reference methods were combined to form the final ground truth. It is not explicitly stated if a third, independent expert or method was used for discordant results between Wet Mount and InPouch TV Culture.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted as this is a diagnostic assay for the direct detection of DNA, not an AI-assisted interpretation of images or other data by human readers. The study focuses on the standalone performance of the assay compared to established laboratory reference methods.

6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

Yes, a standalone performance study was conducted. The main clinical performance evaluation (sensitivity and specificity) presented in Table 4 ("TVQ ASSAY VS COMPOSITE REFERENCE RESULT CULTURE OF INPOUCH TV TABLE 4: CULTURE AND WET MOUNT (BY SPECIMEN TYPE AND SYMPTOMATIC STATUS)") represents the performance of the BD ProbeTec TVQ Assay (algorithm only) against the composite reference method. There is no human-in-the-loop component mentioned for the interpretation of the assay's results. The BD Viper System provides an automated result (positive, negative, or EC failure).

7. Type of Ground Truth Used

The ground truth used was a composite reference method consisting of:

  • Wet Mount
  • InPouch TV Culture

These are considered established microbiological diagnostic methods for Trichomonas vaginalis.

8. Sample Size for the Training Set

The document does not provide information regarding a specific training set size. This indicates that the device's development likely used a different methodology, such as being designed and optimized based on known biological principles and analytical performance rather than a large machine learning training dataset. For an in vitro diagnostic (IVD) assay like this, analytical studies (LOD, specificity, interference) and subsequent clinical validation are typical, rather than a "training set" in the machine learning sense.

9. How the Ground Truth for the Training Set Was Established

As no specific training set is mentioned in the context of machine learning, there is no information provided on how ground truth for such a set was established. The development of the assay would have involved standard molecular biology techniques and validation against known positive and negative controls (analytical truth) rather than a clinical training set with "ground truth" derived from patient outcomes or expert consensus on clinical samples.

{0}------------------------------------------------

K130268

510(k) Summary BD ProbeTec"M Trichomonas Vaginalis (TV) Q* Amplified DNA Assay

ApplicantBD Diagnostic Systems7 Loveton CircleSparks, MD 21152
Establishment Registration No.1119779
Contact PersonEmily Howardtel. 410-316-4428fax. 410-316-4041Emily_Howard@bd.com
Summary DateAugust 15, 2013
Proprietary NameBD ProbeTecTM Trichomonas vaginalis (TV) Qx AmplifiedDNA Assay
Generic NameDNA probe, nucleic acid amplification, Trichomonas vaginalis
ClassificationClass II
Classification NameTrichomonas vaginalis nucleic acid amplification test system
Regulation Number21 CFR 866.3860
Product CodeOUY
Predicate DeviceAPTIMA Trichomonas vaginalis Assay (K102911)

Device Description

The BD ProbeTec Trichomonas vaginalis (TV) Q* Amplified DNA Assay (TVQ Assay) is based on the simultaneous amplification and detection of target DNA using amplification primers and fluorescently-labeled detector probes. The reagents for SDA are dried in two separate disposable microwells: the Priming Microwell contains the amplification primers, fluorescently-labeled detector probe, nucleotides and other reagents necessary for amplification, while the Amplification Microwell contains the two enzymes (a DNA polymerase and a restriction endonuclease) that are required for SDA. The BD Viper™ System pipettes a portion of the purified DNA solution from each Extraction Tube into a Priming Microwell to rehydrate the contents. After a brief incubation, the reaction mixture is transferred to a corresponding, prewarmed Amplification Microwell which is sealed to prevent contamination and then incubated in one of the two thermally-controlled fluorescent readers. The presence or absence of Trichomonas vaginalis DNA is determined by calculating the peak fluorescence (Maximum Relative Fluorescent Units (MaxRFU) over the course of the amplification process and by comparing this measurement to a predetermined threshold value.

{1}------------------------------------------------

In addition to the fluorescent probe used to detect amplified TV target DNA, a second labeled oligonucleotide is incorporated in each reaction Control (EC) oligonucleotide is labeled with a different dye than that used for detection of the TV specific target and is used to confirm the validity of the extraction process. The EC is dried in the Extraction Tubes and is rehydrated upon addition of the specimen and extraction reagents. At the end of the extraction process, the EC fluorescence is monitored by the BD Viper System and an automated algorithm is applied to both the EC and target-specific signals to report results as positive, negative, or EC failure.

Intended Use

The BD ProbeTec™ Trichomonas vaginalis (TV) Q* Amplified DNA Assay, when tested with the BD Viper™ System in Extracted Mode, uses Strand Displacement Amplification technology for the direct, qualitative detection of Trichomonas vaginalis DNA in clinician-collected female endocervical swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and female urine specimens. The assay is indicated for use with asymptomatic and symptomatic females to aid in the diagnosis of trichomoniasis.

Summary and Principles of Operation

When used with the BD Viper System, the BD ProbeTec TV Q* Amplified DNA Assay involves automated extraction of DNA from clinical specimens through the chemical lysis of cells, followed by binding of DNA to para-magnetic particles, washing of the bound nucleic acid, and elution in an amplification-compatible buffer. When present, TV DNA is then detected by Strand Displacement Amplification (SDA) of the specific target sequence in the presence of a fluorescently-labeled detector probe.

Analytical Performance Characteristics

Limit of Detection (Analytical Sensitivity)

The Limit of Detection (LOD) for the TV Q* Assay on the BD Viper System was determined for two strains of T. vaginalis (one Metronidazole-sensitive and one Metronidazole-resistant) by diluting in specimen matrices (vaginal swab, endocervical swab, and urine) and in clean BD O* Swab Diluent to create an LOD panel consisting of six target levels of organism. For each strain and target level, twenty replicates of each panel member were tested on different BD Viper Systems for a total of 60 replicates. LOD was confirmed using 20 replicates at the estimated LOD for each matrix type. The LOD for the TV Q* Assay in clean BD Q* Swab Diluent was determined to be 54.5 and 55.5 trichomonads/mL for ATCC Strains 30001 and 50143. respectively. The LOD for urine and swab matrices are presented in Table 1. The TV Q* Assay on the BD Viper System in extracted mode could detect with ≥95% proportion positive four additional ATCC strains (30237, 50144, 30184, and 30185) in urine matrix and three ATCC strains (30185, 30237, and 50144) in vaginal swab specimen matrix at a concentration of 122,1 TV/mL. ·

{2}------------------------------------------------

Specimen TypeATCC StrainLOD(TV/mL)
Neat Urine30001109.7
50143108.2
Vaginal3000174.4
5014388.4
30184*152.8
Endocervical3000164.8
5014376.2

Table 1: LoD Estimates for TV Q* Assay

*LOD for T. vaginalis ATCC strain 30184 was determined in vaginal swab specimen matrix only.

·

{3}------------------------------------------------

Analytical Specificity

The DNA from 54 organisms was extracted on the BD Viper System and tested with the BD ProbeTec TV Q Amplified DNA Assay. All potential cross-reactive species were tested at approximately ≥ 1x106 CFU/mL (bacteria and yeast), ≥ 1x106 vp/mL (viral particles), or organisms/mL (viruses and pathogens), except where noted. Results are summarized in Table 2.

Organism
Acinetobacter baumanniiHPV-18
Actinomyces israeliiHuman Immunodeficiency Virus 1 (HIV-1)
Atopobium vaginaeKlebsiella oxytoca
Bacteroides fragilisLactobacillus acidophilus
Bifidobacterium bifidumLactobacillus jensenii
Campylobacter jejuniLactobacillus vaginalis
Candida albicansListeria monocytogenes
Candida glabrataMobiluncus curtisii
Candida parapsilosisMoraxella catarrhalis (Branhamella sp.)
Candida tropicalisMycobacterium smegmatis
Chlamydia trachomatisMycoplasma genitalium
Clostridium difficileMycoplasma hominis
Clostridium perfringensNeisseria gonorrhoeae
Corynebacterium genitalium biovar 1Pentatrichomonas hominis1
Cryptococcus neoformansPeptostreptococcus anaerobius
Enterobacter aerogenesPrevotella bivia
Enterobacter cloaceaeProteus mirabilis
Enterococcus faecalisPseudomonas aeruginosa
Escherichia coliStaphylococcus aureus, non-protein A
Fusobacterium nucleatumStaphylococcus aureus, protein-A producing
Gardnerella vaginalisStaphylococcus epidermidis
Haemophilus ducreyiStaphylococcus saprophyticus
Herpes Simplex Virus Type 1Streptococcus pyogenes (Group A)
Herpes Simplex Virus Type 2Streptococcus agalactiae (Group B)
HPV-6Trichomonas tenax
HPV-11Ureaplasma urealyticum
HPV-16Veillonella parvula

Table 2: Potential Cross-Reactants

I Tested at 3.39x10' organisms/mL

It was determined that Trichomonas tenax is a cross-reactant at levels above 1.0 x 10° organisms/mL of the TVQ Assay when tested on the BD Viper System. There were no other organisms found to cross-react with the TVQ Assay when tested on the BD Viper System.

{4}------------------------------------------------

Interfering Substances

Potential interfering substances which may be encountered in endocervical, vaginal, and urine specimens were tested in the absence and presence of TV target (223.2 trichomonads/mL from ATCC Strain 30001 in Vaginal Specimens and 165 trichomonads/mL from ATCC Strain 50143 in female urine specimens) and tested with the BD ProbeTec TV Q* Amplified DNA Assay on the BD Viper System. Results are summarized in Table 3.

InterpretationSwabUrine matrix
No interferenceWhole Blood (≤ 60%)Phenazopyridine Hydrochloride
observed at levelsSeminal FluidWhole Blood (≤ 1% v/v)
listedMucusAcidic urine (pH 5.0)
Over the counter vaginalAlkaline urine (pH 9.0)
products and contraceptivesHoromone pool
Hemorrhoidal creamAnalgesic pool
Prescription vaginal treatmentsAntibiotics
Leukocytes (1x106 cells/mL)Bilirubin
Intravaginal hormonesMucus
Albumin (≤ 1mg/mL)
Glucose
Semen (5% v/v)
Over the counter deodorant spray
and powder
Leukocytes (2.5x106 cells/mL)
May cause ExtractionBlood (> 60%)Not applicable
Control (EC) failures
May cause falseNot applicableNot applicable
negative results

POTENTIAL INTERFERING SUBSTANCES TABLE 3:

{5}------------------------------------------------

Clinical Performance Characteristics

First void urine specimens, clinician-collected endocervical swab specimens, and self-collected (in a clinical setting) vaginal swab specimens were collected from 1,222 compliant female subjects attending family planning, OB/GYN, and sexually transmitted disease clinics at seven geographically diverse clinical sites in North America. Specimens were collected from subjects presenting with symptoms of trichomoniasis (symptomatic) or from subjects during routine visits to their healthcare provider (asymptomatic). Subjects were excluded from the data analysis due to collection site workflow, opting to withdraw from the study after initially consenting, transport/handling/storage errors, informed consent issues, subjects failing to meet the inclusion/exclusion criteria, collection errors, shipping errors, instrument errors, operating condition excursions or labeling errors. Therefore, the final data analysis included 735 compliant subjects for urine specimens. 832 compliant subjects for vaginal specimens, and 995 compliant subjects for endocervical specimens.

For each compliant subject, clinical specimens were collected in the following order: (1) a first void urine, (2) a patient-collected vaginal swab, (3) three clinician-collected vaginal swabs, and (4) an endocervical swab. The three clinician collected vaginal swabs were used for reference and discrepant testing. The first two of these were randomized, one swab was tested for the Wet Mount (reference method) and the other swab was used for the InPouch TV Culture (reference method). The third swab was used for testing using a commercially available NAAT. The aliquot of neat urine, the self-collected vaginal swab, and the endocervical swab were tested on the BD Viper System in extracted mode with the TVO assay.

All sensitivity and specificity calculations were based on the total number of BD ProbeTec TVQ Assay results for endocervical, vaginal, and urine specimens, as compared to a composite reference method. Sensitivity and specificity by specimen type (comparing the TVO Assay with the composite reference method) are presented in Table 4. The initial instrument error rate during the clinical study was 0.1%, or 3 indeterminate results out of 2568 tests. The final error rate after repeat testing was performed on indeterminate results was 0.04%. or 1 indeterminate result out of 2568 tests.

{6}------------------------------------------------

TVQ ASSAY VS COMPOSITE REFERENCE RESULT CULTURE OF INPOUCH TV TABLE 4: CULTURE AND WET MOUNT (BY SPECIMEN TYPE AND SYMPTOMATIC STATUS)

SpecimenTypeStatusnSensitivity95% C.I.Specificity95% C.I.
Neat UrineA28993.1% (27/29)(78.0%, 98.1%)99.6% (259/260)(97.9%, 99.9%)
S44696.4% (80/83)(89.9%, 98.8%)98.1% (356/363)(96.1%, 99.1%)
Total73595.5% (107/112)A(90.0%, 98.1%)98.7% (615/623)B(97.5%, 99.3%)
VaginalA34393.5% (29/31)(79.3%, 98.2%)99.0% (309/312)(97.2%, 99.7%)
S495100.0% (85/85)(95.7%, 100.0%)99.0% (406/410)(97.5%, 99.6%)
Total83898.3% (114/116)(93.9%, 99.5%)99.0% (715/722)C(98.0%, 99.5%)
EndocervicalA50592.2% (47/51)(81.5%, 96.9%)99.1% (450/454)(97.8%, 99.7%)
S49098.8% (82/83)(93.5%, 99.8%)99.8% (406/407)(98.6%, 100.0%)
Total99596.3% (129/134)(91.6%, 98.4%)99.4% (856/861)D(98.6%, 99.8%)
All SpecimenTypesCombinedA113792.8% (103/111)(86.4%, 96.3%)99.2% (1018/1026)(98.5%, 99.6%)
S143198.4% (247/251)(96.0%, 99.4%)99.0% (1168/1180)(98.2%, 99.4%)
Overall256896.7% (350/362)(94.3%, 98.1%)99.1% (2186/2206)(98.6%, 99.4%)

.

A = asymptomatic, S = symptomatic

{7}------------------------------------------------

A frequency distribution of the initial MaxRFU values for the TVQ Assay with an assay cutoff of 125 MaxRFU is shown in Figure A.

Image /page/7/Figure/1 description: The image is a bar chart showing the frequency of MaxRFU values. The x-axis represents the MaxRFU ranges, and the y-axis represents the frequency. The highest frequency is observed in the 0-49 range, with a value of 2182. There is also a notable frequency in the >=800 range, with a value of 344.

FREQUENCY DISTRIBUTION OF MAXRFU FOR THE TVQ ASSAY (ALL SPECIMEN FIGURE A: TYPES)

.

{8}------------------------------------------------

Reproducibility

Reproducibility of the BD Viper System using the BD ProbeTec TVQ Assay was evaluated at three test sites on one BD Viper System per site. Each site was provided with four identical reproducibility panels, each of which consisted of samples prepared using either Vaginal Matrix in O Swab Diluent or urine matrix, Each panel member was either left unspiked (0x), or was spiked with known amounts of TV organism (ATCC strain 30001) at 0.3x, 1x, and 3x LOD for each respective specimen matrix. Three replicates of each panel member were tested every day for five days on each BD Viper System. A summary of the reproducibility data for the TVQ Assay is summarized in Table 5.

Within RunBetween RunWithin DayBetween DayWithin SiteBetween Site
SpecimentypePanelAgreement95% CIMeanSD%CVSD%CVSD%CVSD%CV
VaginalZero100.0%(89/89)(95.9%,100.0%)3.894.02103.532.0151.640.9223.574.69120.60
High Negative(0.3X LOD)43.3%(39/90)(33.6%,53.6%)795.66814.68102.39136.9717.21287.3436.11190.1523.90
Low Positive(IX LOD)96.7%(87/89)(90.7%,98.9%)1632.07457.8328.050.000.000.000.00110.136.75
High Positive(3X LOD)100.0%(90/90)(95.9%,100.0%)1756.68297.7816.950.000.00104.515.95260.4514.83
UrineMatrixZero100.0%(90/90)(95.9%,100.0%)8.7011.33130.180.000.000.000.008.2995.33
High Negative(0.3X LOD)39.3%(35/89)*(29.8%,49.7%)976.96913.8793.540.000.000.000.000.000.00
Low Positive(IX LOD)92.1%(82/89)*(84.6%,96.1%)1574.67681.1443.260.000.000.000.000.000.00
High Positive(3X LOD)100.0%(90/90)(95.9%,100.0%)1822.52364.4620.000.000.000.000.00172.889.49

Table 5: Summary of Reproducibility Data on the BD Viper System for the TVQ Assay

  • Four non-reportable results were due to extraction control failures which caused a reduction in the full number of replicates at one test site.

Conclusion:

The analytical and clinical study results for the BD ProbeTec Trichomonas vaginalis (TV) Q* Amplified DNA Assay support the determination of substantial equivalence in accordance with the intended use as stated in the product labeling.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 23, 2013

EMILY HOWARD REGULATORY AFFAIRS SPECIALIST BECTON, DICKINSON AND COMPANY 7 LOVETON CIRCLE SPARKS MD 21152

Re: K130268

Trade/Device Name: BD ProbeTec"™ Trichomonas vaginalis (TV) Q* Amplified DNA Assav Regulation Number: 21 CFR 866.3860 Regulation Name: Trichomonas vaginalis nucleic acid amplification test system Regulatory Class: II Product Code: OUY Dated: August 12, 2013 Received: August 13, 2013

Dear Ms. Howard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{10}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Sally 緑◎jvat -S

Sally Hojvat, M.Sc., Ph.D. Director, Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure .

{11}------------------------------------------------

Indications for Use

510(k) Number (if known): K130268

Device Name: BD ProbeTec™ Trichomonas vaginalis (TV) Q* Amplified DNA Assay

Indications for Use:

The BD ProbeTec™ Trichomonas vaginalis (TV) Q* Amplified DNA Assay, when tested with the BD Viper™ System in Extracted Mode, uses Strand Displacement Amplification technology for the direct, qualitative detection of Trichomonas vaginalis DNA in clinician-collected female endocervical swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and female urine specimens. The assay is indicated for use with asymptomatic females to aid in the diagnosis of trichomoniasis.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Tamara V. Feldblyum -S 2013.08.21 1 1:06:01 -04'00'

Page 1 of 1

§ 866.3860

Trichomonas vaginalis nucleic acid assay.(a)
Identification. ATrichomonas vaginalis nucleic acid assay is a device that consists of primers, probes, enzymes, and controls for the amplification and detection of trichomonas nucleic acids in endocervical swabs, vaginal swabs, and female urine specimens, from women symptomatic for vaginitis, cervicitis, or urethritis and/or to aid in the diagnosis of trichomoniasis in asymptomatic women. The detection of trichomonas nucleic acids, in conjunction with other laboratory tests, aids in the clinical laboratory diagnosis of trichomoniasis caused byTrichomonas vaginalis .(b)
Classification. Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Nucleic Acid Amplification Assays for the Detection ofTrichomonas vaginalis; Guideline for Industry and Food and Drug Administration Staff.” See § 866.1(e) for information on obtaining this document.