K Number
K130199

Validate with FDA (Live)

Date Cleared
2013-11-22

(298 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

755mm: The CLARITY LPC Laser System is indicated for temporary hair reduction. Stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I-VI) including tanned skin. Treatment of benign pigmented lesions. Treatment of wrinkles. The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).

1064nm: The CLARITY LPC Laser System is indicated for the removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques. The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles.

Device Description

The CLARITY LPC Laser System contains two separate laser resonators, an Alexandrite resonator and an Nd: YAG (Neodymium-doped Yttrium Aluminum Garnet) resonator which generates pulsed laser energy at the nominal wavelength 755 nm and 1064 nm, respectively. The outputs of each laser generator and the aiming beam are optically combined on the laser rail so that their beam paths are identical as they exit the laser system. This allows the use of a single delivery system for either the 755 nm or 1064 nm wavelength. The control panel is equipped with an LCD touch screen so that users can easily adjust parameters for optimal settings. The software included provides all the function which is necessary to use the device. The CLARITY LPC laser system is configured with an ICD (Intelligent Cooling Device) skin cooling device. The CLARITY LPC Laser System delivers laser energy with various pulse durations from 0.35 milliseconds to 300 milliseconds. The output of this laser is delivered to the area of treatment by means of a lens coupled user replaceable optical fiber with a treatment handpiece attached to its distal end. A trigger switch (finger switch or footswitch) is used to control the delivery of laser pulses. The user may choose to deliver a single pulse each time the trigger switch is depressed, or pulses may be delivered repetitively as long as the switch is depressed, at repetition rates up to 10 pulses per second (depending on the chosen pulse duration). The laser output energy is delivered via an optical fiber to a handpiece which produces circular beams with diameters from 2 to 20 millimeters on the skin. Operators of the laser select parameters such as desired energy density (fluence) level and repetition rate and monitor operation via a touch screen and display panel. The touch screen panel can also be used to enable or disable the triggering of the laser, to initiate the calibration feature and to obtain feedback from the system.

AI/ML Overview

The provided document is a 510(k) summary for the Lutronic Corporation CLARITY LPC Laser System. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, not typically for proving clinical efficacy through extensive studies with specific acceptance criteria in the same way a PMA (Pre-Market Approval) or clinical trial would.

Therefore, the document explicitly states: "7. Performance Data None presented."

This clearly indicates that the submission does not contain a detailed study proving the device meets specific acceptance criteria based on performance metrics such as accuracy, sensitivity, or specificity. Instead, the submission relies on demonstrating substantial equivalence to a previously cleared device (Candela Corporation GentleMAX Family of Laser Systems, K112715) by comparing technological characteristics (wavelengths, laser source, indications for use, treatment parameters, etc.).

Given this explicit statement, I cannot provide the requested information about acceptance criteria, a study proving it, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or training set details because this information is not present in the provided text.

The document's purpose is to show that the new device is as safe and effective as the predicate device without needing new clinical performance data, as the technological characteristics and intended uses are highly similar.

{0}------------------------------------------------

KI30199

NOV 2 2 2013

Page I of II

510(k) Summary for the Lutronic Corporation CLARITY LPC Laser System

This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.

1. General Information

Submitter: Lutronic Corporation 219 Sowon-ro Haengsin-dong, Deogyang-gu Goyang-si, Gyeonggi-do 410-722 Republic of Korea Tel: (82) 31-908-3440 Fax: (82) 31-907-3440 Jhung Won Vojir, Ph.D Preparer and Contact Person: Global Regulatory Officer Lutronic, Inc. Six Neshaminy Interplex, Suite 100 Trevose, PA 19053 Telephone: 215-205-2219 Fax: 609-488-6958 October 21, 2013 Summary Preparation Date: 2. Names CLARITY LPC Laser System Trade Name: Common Name: Dermatology Laser Laser Instrument, Surgical, Powered Classification Name: Product Code: GEX Panel: General & Plastic Surgery Regulatory Class: II Regulation Number: 21 CFR 878.4810

3. Predicate Devices

The CLARITY LPC Laser System is substantially equivalent to the Candela Corporation GentleMAX Family of Laser Systems (K112715).

{1}------------------------------------------------

4. Device Description

The CLARITY LPC Laser System contains two separate laser resonators, an Alexandrite resonator and an Nd: YAG (Neodymium-doped Yttrium Aluminum Garnet) resonator which generates pulsed laser energy at the nominal wavelength 755 nm and 1064 nm, respectively. The outputs of each laser generator and the aiming beam are optically combined on the laser rail so that their beam paths are identical as they exit the laser system. This allows the use of a single delivery system for either the 755 nm or 1064 nm wavelength.

The control panel is equipped with an LCD touch screen so that users can easily adjust parameters for optimal settings. The software included provides all the function which is necessary to use the device.

The CLARITY LPC laser system is configured with an ICD (Intelligent Cooling Device) skin cooling device.

The CLARITY LPC Laser System delivers laser energy with various pulse durations from 0.35 milliseconds to 300 milliseconds. The output of this laser is delivered to the area of treatment by means of a lens coupled user replaceable optical fiber with a treatment handpiece attached to its distal end. A trigger switch (finger switch or footswitch) is used to control the delivery of laser pulses. The user may choose to deliver a single pulse each time the trigger switch is depressed, or pulses may be delivered repetitively as long as the switch is depressed, at repetition rates up to 10 pulses per second (depending on the chosen pulse duration).

The laser output energy is delivered via an optical fiber to a handpiece which produces circular beams with diameters from 2 to 20 millimeters on the skin.

Operators of the laser select parameters such as desired energy density (fluence) level and repetition rate and monitor operation via a touch screen and display panel. The touch screen panel can also be used to enable or disable the triggering of the laser, to initiate the calibration feature and to obtain feedback from the system.

5. Indications for Use

The CLARITY LPC Laser System is indicated for the following:

755mm

The CLARITY LPC Laser System is indicated for temporary hair reduction. Stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I-VI) including tanned skin. Treatment of benign pigmented lesions.

Treatment of wrinkles. The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).

{2}------------------------------------------------

l 064nm

The CLARITY LPC Laser System is indicated for the removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.

The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.

The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.

Treatment of wrinkles.

6. Substantial Equivalence

The CLARITY LPC Laser System is substantially equivalent to the GentleMAX Family of Laser Systems cleared in K112715. The CLARITY LPC Laser System has the same technological characteristics as the predicate device. Both devices have two separate laser resonators, an Alexandrite resonator and an Nd: YAG (Neodymium-doped Yttrium Aluminum Garnet) resonator which generates pulsed laser energy at the nominal wavelength of 755 nm and of 1064 nm, respectively.

Additionally, they have the same intended uses, similar treatment parameters, similar operating principles, including similar spot sizes and the same maximum delivered fluence as the predicate device. On the basis of similarities in methods of assembly, method of operation, and intended uses, CLARITY LPC Laser System is substantially equivalent to the predicate device.

{3}------------------------------------------------

Page 4 of 11

:

1

DeviceCLARITY-LPC Laser SystemGentleMAX Family of Laser Systems
ManufacturerLutronic CorporationCandela Corporation
510(k) #K130199K112715
Wavelength755 nm755 nm
Laser SourceAlexandriteAlexandrite
IndicationThe CLARITY LPC Laser System is indicated fortemporary hair reduction. Stable long-term orpermanent reduction through selective targeting ofmelanin in hair follicles. Permanent hair reduction isdefined as the long-term, stable reduction in the numberof hairs regrowing when measured at 6, 9, and 12months after the completion of a treatment regime. Onall skin types (Fitzpatrick I-VI) including tanned skin.Treatment of benign pigmented lesions.Treatment of wrinkles.The photocoagulation of dermatological vascularlesions (such as port-wine stains, hemangiomas,telangiectasias)The GentleMAX Family of Laser Systems is indicated fortemporary hair reduction. Stable long-term or permanentreduction through selective targeting of melanin in hairfollicles. Permanent hair reduction is defined as long-termstable reduction in the number of hairs regrowing after atreatment regime. On all skin types (Fitzpatrick 1-VI)including tanned skin.Treatment of benign pigmented lesions.Treatment of wrinkles.The photocoagulation of dermatological vascular lesions(such as port-wine stains, hemangiomas, telangiectasias)

{4}------------------------------------------------

Page5of11
-----------------

.

Skin TypePulse DurationSpot SizeFluenceSkin TypePulse DurationSpot SizeFluence
I – II3 ms8 mm30-40 J/cm²I – II3 ms8 mm30-40 J/cm²
3 ms10 mm26-34 J/cm²3 ms10 mm25-35 J/cm²
3 ms12 mm26-30 J/cm²3 ms12 mm26-36 J/cm²
3 ms15 mm20-40 J/cm²
3 ms18 mm14-20 J/cm²
III3 ms8 mm26-40 J/cm²3 ms8 mm25-40 J/cm²
3 ms10 mm15-30 J/cm²3 ms10 mm15-30 J/cm²
3 ms12 mm18-26 J/cm²III3 ms12 mm18-30 J/cm²
TreatmentParametersfor hairremoval3 ms15 mm16-19 J/cm²3 ms15 mm16-25 J/cm²
3 ms18 mm10-13 J/cm²3 ms18 mm10-16 J/cm²
IV3 ms8 mm20-30 J/cm²3 ms8 mm20-30 J/cm²
3 ms10 mm15-24 J/cm²3 ms10 mm15-25 J/cm²
3 ms12 mm16-20 J/cm²IV3 ms12 mm16-20 J/cm²
3 ms15 mm10-15 J/cm²3 ms15 mm10-20 J/cm²
3 ms18 mm8-11 J/cm²3 ms18 mm8-16 J/cm²
V3 ms8 mm20-24 J/cm²3 ms8 mm20-25 J/cm²
3 ms10 mm16-20 J/cm²3 ms10 mm16-20 J/cm²
3 ms12 mm14-20 J/cm²V3 ms12 mm14-20 J/cm²
3 ms15 mm10-15 J/cm²3 ms15 mm10-20 J/cm²
3 ms18 mm6-11 J/cm²3 ms18 mm6-16 J/cm²
VI3 ms8 mm20-24 J/cm²3 ms8 mm20-25 J/cm²
3 ms10 mm16-20 J/cm²3 ms10 mm16-20 J/cm²
3 ms12 mm10-20 J/cm²VI3 ms12 mm10-20 J/cm²
3 ms15 mm6-13 J/cm²3 ms15 mm6-20 J/cm²
3 ms18 mm6-10 J/cm²3 ms18 mm6-16 J/cm²
Skin TypeI – IV (solarlentigines,benignmelanocyticnevi)3 ms8 mm30-40 J/cm²Skin TypeI – IV (solarlentigines,benignmelanocyticnevi)3 ms8 mm30-40 J/cm²TreatmentParametersfor pigmentedlesions
Skin TypeI – IV (solarlentigines,benignmelanocyticnevi)3 ms8 mm30-40 J/cm²Skin TypeI – IV (solarlentigines,benignmelanocyticnevi)3 ms8 mm30-40 J/cm²
TreatmentParametersfor pigmentedlesions

and the comments of the comments of the country of

{5}------------------------------------------------

Page 6 of 1 I

I – IV(Seborrheickeratoses)3 ms10 mm30-40 J/cm²I – IV(seborrheickeratoses)3 ms10 mm80-100 J/cm²
3 ms12 mm80-100 J/cm²3 ms12 mm
Skin TypePulse DurationSpot SizeFluenceSkin TypePulse DurationSpot SizeFluence
TreatmentParametersfor wrinklesI - IV3 ms12 mm26-30 J/cm²I – IV3 ms12 mm25-40 J/cm²
FluenceMax 600 J/cm²520 J/cm²
Pulse Width0.35 ms to 300 ms0.25 ms to 300 ms
Average Power(Watts)55.95 W50.84 W
MaximumEnergyDelivered (J)Max 55.95 J (Pulse width: 300 ms @ 18 mm)Max 50.87 J (Pulse width: 300 ms @ 18 mm)
Pulse RateUp to 10 HzUp to 10 Hz
Spot Size2, 3, 5, 8, 10, 12, 15, 18, 20 mm6, 8, 10, 12, 15, and 18 mm
DeliverySystemOptical fiber with handpieceOptical fiber with handpiece
CoolingWater to AirWater to Air

{6}------------------------------------------------

Page 7 of 11

i,

Predicate Comparison Table for 1064nm

DeviceCLARITY LPC Laser-SystemGentleMAX Family of Laser Systems
ManufacturerLutronic CorporationCandela Corporation
510(k) #K130199K112715.
Wavelength1064 nm1064 nm
Laser SourceNd: YAGNd: YAG
IndicationsThe CLARITY LPC Laser System is indicated for theremoval of unwanted hair, for stable long term orpermanent hair reduction and for treatment of PFB.Permanent hair reduction is defined as the long-term.stable reduction in the number of hairs regrowing whenmeasured at 6, 9, and 12 months after the completion ofa treatment regime. The lasers are indicated on all skintypes Fitzpatrick I-VI including tanned skin.Photocoagulation and hemostasis of pigmented andvascular lesions such as but not limited to port winestains, hemangioma, warts, telangiectasia, rosacea,venus lake, leg veins and spider veins. Coagulation andhemostasis of soft tissue. Benign pigmented lesions suchas, but not limited to, lentigos (age spots), solar lentigos(sun spots), cafe au lait macules, seborrheic keratoses,nevi, chloasma, verrucae, skin tags, keratoses, tattoos(significant reduction in the intensity of black and/orblue-black tattoos) and plaques.The laser is indicated for pigmented lesions to reducelesion size, for patients with lesions that wouldThe GentleMAX Family of Laser Systems is indicated for theremoval of unwanted hair, for stable long term or permanenthair reduction and for treatment of PFB. The lasers areindicated on all skin types Fitzpatrick I-VI including tannedskin. Photocoagulation and hemostasis of pigmented andvascular lesions such as but not limited to port wine stains.hemangioma, warts, telangiectasia, rosacea, venus lake, legveins and spider veins. Coagulation and hemostasis of softtissue. Benign pigmented lesions such as, but not limited to.lentigos (age spots), solar lentigos (sun spots), cafe au laitmacules, seborrheic keratoses, nevi, chloasma, verrucae, skintags, keratoses, tattoos (significant reduction in the intensityof black and/or blue-black tattoos) and plaques.The laser is indicated for pigmented lesions to reduce lesionsize, for patients with lesions that would potentially benefitfrom aggressive treatment, and for patients with lesions thathave not responded to other laser treatments.The laser is also indicated for the reduction of redpigmentation in hypertrophic and keloid scars wherevascularity is an integral part of the scar.

{7}------------------------------------------------

potentially benefit from aggressive treatment, and forTreatment of wrinkles.
patients with lesions that have not responded to other laser treatments. The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles.
Skin TypePulse DurationSpot SizeFluenceSkin TypePulse DurationSpot SizeFluence
I - II20 ms8 mm90-100 J/cm2I - II3, 10, 20 ms8 mm90-100 J/cm2
10 mm70-80 J/cm210 mm70-80 J/cm2
12 mm60-70 J/cm212 mm60-70 J/cm2
15 mm40-44 J/cm215 mm40-44 J/cm2
18 mm26-30 J/cm218 mm26-30 J/cm2
III20 ms8 mm80-90 J/cm2III3, 10, 20 ms8 mm80-90 J/cm2
10 mm50-70 J/cm210 mm50-70 J/cm2
12 mm40-60 J/cm212 mm40-60 J/cm2
15 mm30-40 J/cm215 mm30-40 J/cm2
18 mm24-30 J/cm218 mm24-30 J/cm2
IV20 ms8 mm60-70 J/cm2IV3, 10, 20 ms8 mm60-70 J/cm2
10 mm40-50 J/cm210 mm40-50 J/cm2
12 mm20-45 J/cm212 mm20-46 J/cm2
15 mm20-30 J/cm215 mm20-30 J/cm2
18 mm14-20 J/cm218 mm14-20 J/cm2
V20 ms8 mm50-60 J/cm2V3, 10, 20 ms8 mm50-60 J/cm2
10 mm30-50 J/cm210 mm30-50 J/cm2
12 mm18-30 J/cm212 mm18-30 J/cm2
15 mm10-30 J/cm215 mm10-30 J/cm2
18 mm6-20 J/cm218 mm6-20 J/cm2
VI20 ms8 mm40-50 J/cm2VI3, 10, 20 ms8 mm40-50 J/cm2
10 mm30-40 J/cm210 mm30-40 J/cm2
12 mm18-30 J/cm212 mm18-30 J/cm2
15 mm6-30 J/cm215 mm6-30 J/cm2
18 mm6-20 J/cm218 mm6-18 J/cm2

Page 8 of 11

{8}------------------------------------------------

Page 9 of 11

LesionPulse DurationSpot SizeFluenceLesionPulse DurationSpot SizeFluence
TreatmentParametersfor pigmentedlesionsTelangiectasia< 1 mm40-50 ms3 mmStart with220 J/cm²Telangiectasia< 1 mm40-60 ms3 mmStart with220 J/cm²
Telangiectasia(Facial)Vessels< 1 mm30 ms3 mmStart with130 J/cm²Telangiectasia(Facial)Vessels< 1 mm30 ms3 mmStart with130 J/cm²
Telangiectasia(Leg)Vessels< 1.5 mm40-50 ms3 mmStart with200 J/cm²Telangiectasia(Leg)Vessels< 1.5 mm40-60 ms3 mmStart with200 J/cm²
Telangiectasia(Leg)Vessels1.5-3.0 mm50 ms3 mmStart with180 J/cm²Telangiectasia(Leg)Vessels1.5-3.0 mm60 ms3 mmStart with180 J/cm²
AreaPulse DurationSpot SizeFluenceAreaPulse DurationSpot SizeFluence
TreatmentParametersfor wrinklesFace & Neck50 ms10 mm50 J/cm²Face & Neck50 ms10 mm50 J/cm²
Forehead50 ms10 mm40-50 J/cm²Forehead50 ms10 mm40-50 J/cm²
FluenceMax. 600 J/cm²Max. 600 J/cm²
Pulse Width0.35 ms to 300 ms0.25 ms to 300 ms
Average Power(Watts)96.08 W76.30 W
MaximumEnergyDelivered (J)Max 96.08 J (Pulse width: 300 ms @ 18 mm)Max 76.30 J (Pulse width: 300 ms @ 18 mm)
Pulse RateUp to 10 HzUp to 10 Hz

·

·

ﺘﺮ

{9}------------------------------------------------

Page 10 of 11

.

ﻟﻤﺴﺎﺑﻘﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤ

Spot Size2, 3, 5, 8, 10, 12, 15, 18, 20 mm1.5, 3, 6, 8, 10, 12, 15, and 18mm
DeliverySystemOptical fiber with handpieceOptical fiber with handpiece
CoolingWater to AirWater to Air

.

.

.

.

·

.

.

{10}------------------------------------------------

.

  1. Performance Data None presented.

ﺀ ﺴﺴ

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three wavy lines instead of snakes.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Lutronic Corporation Jhung Won Voiir, M.S. Ph.D. Global Regulatory Officer 6 Neshaminy Interplex, Suite 100 Trevose, Pennsylvania 19053

November 22, 2013

Re: K130199

Trade/Device Name: CLARITY LPC Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: October 18, 2013 Received: October 18, 2013

Dear Dr. Vojir:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{12}------------------------------------------------

Page 2 - Jhung Won Vojir

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Joshual@Nipper -S

For

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{13}------------------------------------------------

Indications for Use

510(k) Number (if known): K130199

Device Name: CLARITY LPC Laser System

Indications for Use:

755mm

The CLARITY LPC Laser System is indicated for temporary hair reduction. Stable longterm or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I-VI) including tanned skin. Treatment of benign pigmented lesions. Treatment of wrinkles.

The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).

1064nm

The CLARITY LPC Laser System is indicated for the removal of unwanted hair, for

stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.

The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.

The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.

Treatment of wrinkles.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over The Counter Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Page I of 1

Neil R Oaden 2013.11 22 09:1 00.500

(Division Sign-Off) for BSA Division of Surgical Devices

510(k) Number K130199

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.