K Number
K130076

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-08-22

(220 days)

Product Code
Regulation Number
872.3760
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Merz Dental GmbH Denture Base Resins with Prosthetic Color System (Promolux, Promolux Hi, Weropress and PremEco PCS) are indicated for:

  • Relining a denture surface that contacts tissue ●
  • Repairing a fractured denture, or
  • Forming a new denture base
Device Description

Merz Dental Denture Base Resins with Prosthetic Color System are composed of the following products:
• Promolux: A heat processed PMMA denture base resin
• Promolux Hi: A heat-processed PMMA denture base resin formulated for greater strength
· Weropress: A two part cold curing PMMA denture base resin
: PremEco PCS: A two part cold curing PMMA denture base resin with individual tints for aesthetic coloration of gingival portion of a denture base
These products are intended for the preparation of denture bases. They are composed of PMMA and cured by standard heat or cold setting methods. The Prosthetic Color System is a three part system composed of the following: (1) PMMA polymer with various tints in a powder form, (2) MMA liquid for mixing with the powder, and (3) a cold-casting resin (PMMA and MMA) for final preparation of the denture base. The hot and cold curing denture materials contain standard dental colorants and catalyst systems. One of the hot cured dental base materials, Promolux Hi, is formulated to increase impact resistance.

AI/ML Overview

This document is a 510(k) Summary for Merz Dental GmbH Denture Base Resins with Prosthetic Color System. It focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific acceptance criteria through a study with performance metrics in the way a medical imaging AI would. Therefore, many of the requested categories (e.g., sample size for test/training, number of experts, adjudication method, MRMC studies, standalone performance) are not applicable to this type of submission.

Here's a breakdown of the available information:

1. A table of acceptance criteria and the reported device performance

The acceptance criteria are that the device materials meet the requirements of two international ISO standards for denture base polymers. The reported device performance is that the materials met these standards.

Acceptance Criteria (ISO Standard)Reported Device Performance
ISO 1567:2000 "Dentistry -Base polymers - Part 1: Denture base polymers" (for cold curing -- Type II: Class 1)The materials "met the physical and residual methyl methacrylate monomer requirements" of this standard.
ISO 20795-1 "Dentistry - Base polymers - Part 1: Denture base polymers" (for hot curing-Type II, Class 2)The materials "met the physical and residual methyl methacrylate monomer requirements" of this standard.
ISO 10993-5 (In Vitro Cytotoxicity)The materials "were tested and found to comply with in vitro cytotoxicity."
ISO 10993-10.2 (Guinea Pig Sensitization)The materials "were tested and found to comply with guinea pig sensitization."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. Performance testing for material properties typically involves laboratory-controlled testing of material samples, not patient data. The provenance for such tests would usually be the testing laboratory, which is not specified beyond being performed by the manufacturer.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable/provided. Ground truth, in the context of this 510(k), refers to material property measurements against established international standards (ISO). This typically involves laboratory technicians and machinery, not clinical experts establishing a ground truth from patient data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable/provided. Adjudication methods are relevant for clinical assessments, especially in cases where human readers are establishing ground truth from medical images. For material property testing, the measurements are objective.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-assisted image interpretation, which is not the subject of this 510(k) submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. This submission details physical material properties of denture resins, not the performance of an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the material performance testing was based on references to international standards (ISO 1567:2000, ISO 20795-1, ISO 10993-5, ISO 10993-10.2). The materials were tested to ensure they met the specified requirements in these standards for physical and biological properties.

8. The sample size for the training set

This information is not applicable/provided. There is no "training set" in the context of material property testing for denture resins. The submission focuses on demonstrating that the final product meets specified standards.

9. How the ground truth for the training set was established

This information is not applicable/provided. As there is no training set, there is no ground truth established for one.

{0}------------------------------------------------

مسمسلمسلمسلسل

510(k) Summary

Merz Dental GmbH Denture Base Resins with Prosthetic Color System

Submitter

Company Name:Merz Dental GmbH
Address:Eetzweg 20Lutjenburg, Germany D-24321
Contact Name:Dr. Med. Claudia Bobrowski
Telephone No.(011) 43 81/4 03-4 11
Fax No.(011) 43 81/4 09-1 07
e-mail:Claudia.bobrowski@merz-dental.de
Date of Summary:July 17, 2013

AUG 2 2 2013## Device Name

Proprietary name:Promolux, Promolux Hi, Weropress, PremEco PCS
Common name:Denture Base Resin
Classification name:Resin, Denture, Relining, Repairing, Rebasing(21CFR 872.3760 Product Code EBI)
Predicate Device:Vertex hot-curing denture base material (K102654)Vertex cold-curing denture base material (K102640)

Device Description:

Merz Dental Denture Base Resins with Prosthetic Color System are composed of the following products:

• PromoluxA heat processed PMMA denture base resin
• Promolux HiA heat-processed PMMA denture base resin formulated forgreater strength
· WeropressA two part cold curing PMMA denture base resin
: PremEco PCSA two part cold curing PMMA denture base resin withindividual tints for aesthetic coloration of gingival portion ofa denture base

These products are intended for the preparation of denture bases. They are composed of PMMA and cured by standard heat or cold setting methods. The Prosthetic Color System

{1}------------------------------------------------

is a three part system composed of the following: (1) PMMA polymer with various tints in a powder form, (2) MMA liquid for mixing with the powder, and (3) a cold-casting resin (PMMA and MMA) for final preparation of the denture base. The hot and cold curing denture materials contain standard dental colorants and catalyst systems. One of the hot cured dental base materials, Promolux Hi, is formulated to increase impact resistance.

Intended Use

The Merz Dental GmbH Denture Base Resins and Prosthetic Color System are for use in:

  • · Relining a denture surface that contacts tissue
  • · Repairing a fractured denture, or
  • · Forming a new denture base

Technological characteristics and substantial equivalence

The technological characteristics of the Merz Dental GmbH Denture Base Resins with Prosthetic Color System, and those of the predicate devices, are shown in Tables 1 and 2.

The Merz Dental GmbH Denture Base Resins and the predicate devices are composed of the same acrylic material (PMMA), utilize the same curing methods (heat or cold cured), and have similar physical properties. They also have the same indications for use (fabrication of denture bases). They differ with regard to impact resistance (Promolux Hi has a higher impact resistance than the heat-cured predicate). The Merz Dental GmbH Prostheic Color System also adds flexibility in the technician's ability to vary the coloring of the gingival portion of the denture base.

{2}------------------------------------------------

Table 1: Comparison of Promolux and Promolux Hi to the Predicate Vertex Rapid Simplified

DeviceVertex RapidPromoluxPromolux Hi
Heat-CuredSimplified
TechnologyPMMA Acrylic PlasticPMMA Acrylic PlasticPMMA Acrylic Plastic
FeaturesHeat curableHeat curableHeat curable
MaterialsTwo part mix ofTwo part mix ofTwo part mix of
polymethylmethacrylatepolymethylmethacrylatepolymethylmethacrylate
powder andpowder andpowder and
methacrylate monomermethacrylate monomermethacrylate monomer
liquidliquidliquid
OperationsMixed into dough andMixed into dough andMixed into dough and
pressed into flaskpressed into flaskpressed into flask
Indications· Relining a denture• Relining a denture· Relining a denture
surface that contactssurface that contactssurface that contacts
tissuetissuetissue
· Repairing a fractured· Repairing a fractured· Repairing a fractured
denture, ordenture, ordenture, or
· Forming a new· Forming a new· Forming a new
denture basedenture basedenture base
PhysicalImpact-resistanceImpact-resistanceImpact-resistance
PropertiesFlexural strengthFlexural strengthFlexural strength
Flexural modulusFlexural modulusFlexural modulus
Water sorptionWater sorptionWater sorption
SolubilitySolubilitySolubility
Bio-Composed ofComposed ofComposed of
compatibilitybiocompatible PMMAbiocompatible PMMAbiocompatible PMMA
with colorants.with colorants.with colorants.

.

{3}------------------------------------------------

Device Cold-CuredVertex CastavariaWeropressPremEco PCS
TechnologyPMMA Acrylic plasticPMMA Acrylic plasticPMMA Acrylic plastic
FeaturesCold CuringPMMA Acrylic plasticPMMA Acrylic plastic
MaterialsTwo part mix ofpolymethylmethacrylatepowder andmethacrylate monomerliquidTwo part mix ofpolymethylmethacrylatepowder andmethacrylate monomerliquidTwo part mix ofpolymethylmethacrylatepowder andmethacrylate monomerliquid
OperationsMixed dough andpressed into flaskMixed dough andpressed into flaskMix powder and liquidand brush on
Indications• Relining a denturesurface that contactstissue• Repairing a fractureddenture,• Forming a newdenture base• Relining a denturesurface that contactstissue• Repairing a fractureddenture,• Forming a newdenture base• Relining a denturesurface that contactstissue• Repairing a fractureddenture,• Forming a newdenture base
PhysicalPropertiesImpact-resistanceFlexural strengthFlexural modulusWaterSolubilityImpact-resistanceFlexural strengthFlexural modulusWaterSolubilityImpact-resistanceFlexural strengthFlexural modulusWaterSolubility
Bio-compatibilityComposed ofbiocompatible PMMAwith colorants.Composed ofbiocompatible PMMAwith colorants.Composed ofbiocompatible PMMAwith colorants.

Table 2: Comparison of Weropress and PremEco PCS to Vertex Castavaria

Performance Testing

Tests of the cured polymer base materials was performed according to ISO 1567:2000 "Dentistry -Base polymers - Part 1: Denture base polymers" (for cold curing -- Type II: Class 1), and ISO 20795- (for hot curing-Type II, Class 2) The materials met the physical and residual methyl methacrylate monomer requirements.

These well known denture base materials were tested and found to comply with in vitro cytotoxicity (ISO 10993-5), guinea pig sensitization (ISO 10993-10.2), physical properties (ISO 1567:2000 for cold-curing polymers; and ISO 20795-1 for hot-curing polymers), and residual monomer requirements. These materials have a long history of safe use and have been extensively tested for physical and biological properties.

{4}------------------------------------------------

Conclusions:

Merz Dental GmbH denture base resins (Promolux Hi, Weropress and PremEco PCS) have similar technological and performance characteristics as the currently marketed predicate denture base resins (Vertex Rapid Simplified and Vertex Castavaria). The products are composed of the same biocompatible acrylic plastic (PMMA), utilize the same preparation and casting techniques, and have identical indications for use. Their physical properties are similar for each type of curing system (hot or cold) with the exception that the hot-curing Merz Dental Promolux Hi has a higher impact resistance than the hot-curing Vertex Rapid Simplified. In addition, the Merz Dental PremEco PCS product allows greater flexibility in the coloring of the gingival portion of the denture than the predicate device. Based on the above similarities and differences, it is concluded that the Merz Dental GmbH denture base resin products (Promolux, Promolux Hi, Weropress and PremEco PCS) are substantially equivalent to the predicate denture base resin products (Vertex Rapid Simplified and Vertex Castavaria).

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight, composed of three curved lines.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 22, 2013

Merz Dental GmbH C/O Richard G. Hunter, M.S., RAC Principal Washington Regulatory Consultants 5616 Mariola Place NE ALBUQUERQUE, New Mexico 87111

Re: K130076

Trade/Device Name: Merz Denture Base Resins with Prosthetic Color System Regulation Number: 21 CFR 872.3760 Regulation Name: Resin Denture. Relining, Repairing, Rebasing Regulatory Class: II Product Code: EBI Dated: July 17, 2013 Received: July 18, 2013

Dear Mr. Hunter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Oosmetie for (110) the device. subject to the general controls provisions of the Act. The r ou may, therefore, many of the Act include requirements for annual registration, listing of general controls proficioning practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it n your device is elaboriise (boo assns). Existing major regulations affecting your device can be may be subject to additions. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Please be advised that FDA *s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address hup://www.fda.gov/MedicalDevices/Resources/orYou/Industry/default.htm. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincercly yours.

Mary S. Runner -S

Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

4. Indications for Use

510(k) Number (if known): ¥13007 ص

Device Name: Merz Denture Base Resins with Prosthetic Color System

Indications for Use:

The Merz Dental GmbH Denture Base Resins with Prosthetic Color System (Promolux, Promolux Hi, Weropress and PremEco PCS) are indicated for:

  • Relining a denture surface that contacts tissue ●
  • Repairing a fractured denture, or
  • Forming a new denture base

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR .

Over-The-Counter Use

(21 CFR Part 801 Subpart D)

(21 CFR Part 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Andrew I. Steen -S. پياز -2013.08.21 11:08:54 -04'00'

(Division Sign-Off) Division of Anesthesiology, General Hospital infection Control, Dental Devices

Page 1 of __ 1_

510(k) Number:_K | 30076

§ 872.3760 Denture relining, repairing, or rebasing resin.

(a)
Identification. A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.(b)
Classification. Class II.