(140 days)
The MectaLIF Anterior is an anterior interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The interior of the spacer component of the MectaLIF Anterior can be packed with autograft or autologous bone graft.
DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
The MectaLIF Anterior is a system intended to be used with the integrated bone screws provided and requires additional supplementary fixation such as pedicle screws and rods.
These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s).
These patients should be skeletally mature and have had six months of non-operative treatment.
The MectaLIF Anterior is an anterior interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. MectaLIF Anterior is intended to be used with bone screws provided and requires additional supplementary fixation. MectaLIF Anterior consists of a disc spacer made of PEEK-OPTIMA LT1: Implant Grade Polyetheretherketone (ASTM F 2026) which contains three Tantalum Markers (ISO 13782 / ASTM F 560), bone screws made of Titanium: Ti6Al4V ELI (ISO 5832-3/ASTM F 136), and a plate made of Titanium: Ti6Al4V ELI (ISO 5832-3/ASTM F 136). The interior of the disc spacer can be packed with autograft or autologous bone graft. The plate comes in three different designs (Flush, Long, and L5-S1) and is secured to the disc spacer via an interlocking mechanism. The disc spacer and attached plate are secured to the vertebral body with the bone screws. The Flush and Long plates are used with four bone screws while the L5-S1 plate is used with three bone screws.
The presented document describes the performance testing for the MectaLIF Anterior, an anterior interbody fusion device, in the context of a 510(k) premarket notification. The goal of the testing was to demonstrate substantial equivalence to predicate devices.
1. Table of acceptance criteria and reported device performance:
| Test Method | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| ASTM F2077 (Static and Dynamic Axial Compression, Static and Dynamic Compression-Shear, Static and Dynamic Torsion, Subsidence, Expulsion) | The MectaLIF Anterior, tested using the worst-case device, should not perform worse than predicate device data. This implies that the MectaLIF Anterior should meet or exceed the worst-case mechanical performance of legally marketed predicate devices for similar intervertebral fusion applications. | "The performance testing demonstrated that the MectaLIF Anterior is not worst case compared to predicate data." This statement indicates that for all the listed tests, the MectaLIF Anterior performed at least as well as, if not better than, the mechanical performance of the worst-case predicate devices. |
| ASTM F2267 (Likely also includes similar mechanical tests based on the context of interbody fusion devices) | The MectaLIF Anterior, tested using the worst-case device, should not perform worse than predicate device data. | "The performance testing demonstrated that the MectaLIF Anterior is not worst case compared to predicate data." |
2. Sample size used for the test set and the data provenance:
- Sample size: The document does not specify the exact sample size. It states that testing was performed "using the worst-case device for each of the following tests." This suggests that potentially one "worst-case" device was selected and subjected to each test. However, standards like ASTM F2077 and F2267 typically require multiple samples for statistical validity. Without specific numbers, it's impossible to confirm the sample size.
- Data provenance: The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. Given that this is a 510(k) submission for a device manufactured by Medacta International SA (Switzerland) and represented by Medacta USA, the testing would likely have been conducted in a laboratory setting, possibly in Switzerland or the US, to meet regulatory requirements. The nature of the performance tests (mechanical testing) indicates this is prospective laboratory test data, not patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not applicable to the supplied document. The "tests" described are mechanical performance tests of a medical device, not clinical studies involving expert interpretation of patient data or images. The "ground truth" for these tests is the quantitative measurement obtained through standardized mechanical testing protocols (ASTM F2077 and F2267).
4. Adjudication method for the test set:
- This information is not applicable. Adjudication methods (like 2+1, 3+1) are used in clinical studies or image review where there are multiple human readers whose interpretations need to be reconciled. The performance testing outlined here is mechanical.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This information is not applicable. The document describes mechanical performance testing of an interbody fusion device, not an AI-powered diagnostic or assistive technology for human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- This information is not applicable. The document describes mechanical performance testing of an interbody fusion device, not an algorithm or AI system.
7. The type of ground truth used:
- For the mechanical performance tests, the "ground truth" is defined by the measured mechanical properties (e.g., load-bearing capacity, resistance to shear, subsidence, expulsion force) as determined by the standardized test methods ASTM F2077 and ASTM F2267. These standards define the experimental setup, loading conditions, and measurement techniques. The acceptance criteria are then set relative to predicate device performance.
8. The sample size for the training set:
- This information is not applicable. The document describes mechanical performance testing of a physical medical device, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established:
- This information is not applicable, as there is no training set mentioned or implied in this document for the device's performance evaluation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Medacta International. The logo consists of two black triangles, the word "Medacta" in black, a black line underneath the word, the word "International" in black, and a black plus sign. The logo is simple and modern.
K124034 Page 1 of 3
MAY 1 7 2013
510(k) Summary
Applicant/Sponsor: Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Mr. Adam Gross Contact Person: Director of Regulatory, Quality and Compliance Medacta USA 4725 Calle Quetzal, Unit B Camarillo, CA, 93012 Phone: (805)437-7085 Fax: (805)437-7553 Email: AGross@medacta.us.com
December 14, 2012 Date Prepared:
DEVICE INFORMATION
Trade/Proprietary Name: MectaLIF Anterior Common Name: Anterior Intervertebral Body Fusion Device Classification Name: intervertebral fusion device with integrated fixation, lumbar
21 CFR 888.3080 Class II Device Product Codes: OVD
Predicate Devices:
K101310 Vu aPOD, Integra Spine K073109 Stalif TT, Centinel Spine K101301 Stalif Midline, Centinel Spine K110927 MectaLIF, Medacta International K072253 SynFix-LR, Synthes K022791 ATB System, Synthes K071726 Zuma, SeaSpine K082252 Independence, Globus Medical K013665 Pyramid plate, Medtronic
MectaLIF Anterior 510(k)
Section 5 - Page 2 of 4
{1}------------------------------------------------
Product Description
The MectaLIF Anterior is an anterior interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. MectaLIF Anterior is intended to be used with bone screws provided and requires additional supplementary fixation. MectaLIF Anterior consists of a disc spacer made of PEEK-OPTIMA LT1: Implant Grade Polyetheretherketone (ASTM F 2026) which contains three Tantalum Markers (ISO 13782 / ASTM F 560), bone screws made of Titanium: Ti6Al4V ELI (ISO 5832-3/ASTM F 136), and a plate made of Titanium: Ti6Al4V ELI (ISO 5832-3/ASTM F 136). The interior of the disc spacer can be packed with autograft or autologous bone graft. The plate comes in three different designs (Flush, Long, and L5-S1) and is secured to the disc spacer via an interlocking mechanism. The disc spacer and attached plate are secured to the vertebral body with the bone screws. The Flush and Long plates are used with four bone screws while the L5-S1 plate is used with three bone screws.
Indications for Use
The MectaLIF Anterior is an anterior interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondvlolisthesis or retrolisthesis at the involved level(s). The interior of the spacer component of the MectaLIF Anterior can be packed with autograft or autologous bone graft.
DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
The MectaLIF Anterior is a system intended to be used with the integrated bone screws provided and requires additional supplementary fixation such as pedicle screws and rods.
These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s).
These patients should be skeletally mature and have had six months of non-operative treatment.
Comparison to Predicate Devices
The indications for use, design features and materials of the MectaLIF Anterior are substantially equivalent to those of the predicate devices. The fundamental scientific technology of the modified devices has not changed relative to the predicate devices. The safety and effectiveness of the MectaLIF Anterior are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.
{2}------------------------------------------------
K124034 Page 3 of 3
Performance Testing
The MectaLIF Anterior was tested per ASTM F2077 and ASTM F2267 using the worstcase device for each of the following tests:
-
- Static Axial Compression
-
- Dynamic Axial Compression
-
- Static Compression-Shear
-
- Dynamic Compression-Shear
-
- Static Torsion
-
- Dynamic Torsion
-
- Subsidence
-
- Expulsion
The performance testing demonstrated that the MectaLIF Anterior is not worst case compared to predicate data.
Conclusion:
Based on the above information, the MectaLIF Anterior can be considered as substantially equivalent to its predicate devices.
MectaLIF Anterior 510(k)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Letter dated: May 17, 2013
Medacta International SA % Medacta USA Mr. Adam Gross Director of Regulatory, Ouality and Compliance 4725 Calle Quetzal, Unit B Camarillo, California 93012
Re: K124034
Trade/Device Name: MectaLIF Anterior Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD Dated: April 11, 2013 Received: April 15, 2013
Dear Mr. Gross:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Mr. Adam Gross
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mark Night erson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K124034 Page 1 of 1
Indications for Use
510(k) Number: K124034
Device Name: MectaLIF Anterior
Indications for Use:
The MectaLIF Anterior is an anterior interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The interior of the spacer component of the MectaLIF Anterior can be packed with autograft or autologous bone graft.
DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
The MectaLIF Anterior is a system intended to be used with the integrated bone screws provided and requires additional supplementary fixation such as pedicle screws and rods.
These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s).
These patients should be skeletally mature and have had six months of non-operative treatment.
Prescription Use × (21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Ronald P. Jean -S
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K124034
MectaLIF Anterior 510(k) April 11, 2013
N/A