K Number
K123892

Validate with FDA (Live)

Date Cleared
2013-04-05

(108 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Mepilex® Transfer Ag dressing is indicated for the management of a wide range of exuding wounds such as leg and foot ulcers, pressure ulcers, partial thickness burns, traumatic and surgical wounds. Mepilex® Transfer Ag can also be used under compression bandaging.

Device Description

Mepilex® Transfer Ag is a soft silicone wound contact layer that absorbs and transfers exudate, maintains a moist wound environment and has antimicrobial properties. A moist wound environment is shown to be beneficial for wound healing. Mepilex® Transfer Ag contains silver sulphate which acts as a preservative to reduce or minimize growth of microorganisms within the dressing. Mepilex® Transfer Ag has been shown to inactivate microorganisms for up to 14 days in vitro. Mepilex® Transfer Ag consists of: a Safetac® adhesive layer which is a unique and a patented adhesive technology; a compressed polyurethane foam containing silver sulphate and activated carbon.

AI/ML Overview

This document describes the Mepilex® Transfer Ag wound and burn dressing. Here's an analysis of the provided text regarding acceptance criteria and supporting studies:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Performance Testing)Reported Device Performance
Antimicrobial effect against 16 strains"All areas performed as expected to provide a level of efficacy deemed necessary for the intended use of this device."
Antimicrobial effect inside the dressing against P.a, S.a and C.a"All areas performed as expected to provide a level of efficacy deemed necessary for the intended use of this device."
Preservation to reduce or minimize growth of microorganisms within the dressing"Mepilex® Transfer Ag has been shown to inactivate microorganisms for up to 14 days in vitro."

2. Sample Size Used for the Test Set and Data Provenance

The document mentions "Antimicrobial effect against 16 strains" and "Antimicrobial effect inside the dressing against P.a, S.a and C.a." However, it does not specify the sample sizes (e.g., number of dressings, replicates, or experiments) used for these in vitro tests.

The data provenance is in vitro (laboratory-based) testing. The country of origin of the data is not specified. It is inherently retrospective as the tests were completed before the 510(k) submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The "ground truth" for the in vitro antimicrobial tests would be established by the testing methodology and controls used in the laboratory experiments, rather than expert consensus on data interpretation.

4. Adjudication Method for the Test Set

This information is not provided and is not applicable to the type of in vitro testing performed. Adjudication methods are typically used in clinical studies or image-based diagnostic evaluations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No, an MRMC comparative effectiveness study was not done. This device is a wound dressing, not an AI-powered diagnostic tool, so such a study would not be relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

No, a standalone algorithm performance study was not done. This device is a medical product, not a software algorithm.

7. The Type of Ground Truth Used

The ground truth used for the performance evaluation was based on in vitro antimicrobial testing standards and methodologies. This involves observing and quantifying the reduction or inactivation of specific microorganisms under controlled laboratory conditions.

8. The Sample Size for the Training Set

This is not applicable. The Mepilex® Transfer Ag is a physical medical device (wound dressing), not an AI model, so there is no "training set."

9. How the Ground Truth for the Training Set was Established

This is not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

K123892

APR 0 5 2013

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(c).

Date Prepared:March 20, 2013
Submitter:Mölnlycke Health Care US, LLC
5550 Peachtree Parkway, Suite 500
Norcross, GA 30092
Official Correspondent:Angela L. Bunn, RAC
Director, Regulatory Affairs of the Americas
Tel: 678-250-7930
Fax: 678-250-7981
e-mail: angela.bunn@molnlycke.com
Trade/Proprietary Name:Mepilex® Transfer Ag
Common Name:Wound and Burn Dressing

und and Burn Dressing Classification Name: Dressing, Wound, Drug Device Class: Unclassified Product Code: FRO Mepilex® Border Ag Predicate Device Name(s):

Description of Device:

Mepilex® Transfer Ag is a soft silicone wound contact layer that absorbs and transfers exudate, maintains a moist wound environment and has antimicrobial properties. A moist wound environment is shown to be beneficial for wound healing.

Mepilex® Transfer Ag contains silver sulphate which acts as a preservative to reduce or minimize growth of microorganisms within the dressing.

Mepilex® Transfer Ag has been shown to inactivate microorganisms for up to 14 days in vitro.

Mepilex® Transfer Ag consists of:

  • a Safetac® adhesive layer . o which is a unique and a patented adhesive technology
  • a compressed polyurethane foam containing silver sulphate and activated carbon .

{1}------------------------------------------------

Intended Use/Indication for Use:

Mepilex® Transfer Ag dressing is indicated for the management of a wide range of exuding wounds such as leg and foot ulcers, pressure ulcers, partial thickness burns, traumatic and surgical wounds. Mepilex® Transfer Ag can be used under compression bandaging.

In vitro Data:

The following performance testing was completed on the proposed device:

  • Antimicrobial effect against 16 stains .
  • Antimicrobial effect inside the dressing against P.a, S.a and C.a .

All areas performed as expected to provide a level of efficacy deemed necessary for the intended use of this device.

Clinical Testing:

No clinical data was required.

Conclusion:

Based on the information presented in this submission, it can be concluded that the Mepilex Transfer Ag is equivalent to the Mepilex® Border Ag (K100029) predicate with respect to intended use, materials, design, and technological characteristics. The only difference in the two materials is the removal of the backing film from the predicate device to the proposed device.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

MöInlycke Health Care US, LLC % Angela L. Bunn. RAC Director, Regulatory Affairs of the Americas 5550 Peachtree Parkway, Suite 500 Norcross, Georgia 30092

Re: K123892

Trade/Device Name: Mepilex® Transfer Ag Regulatory Class: Unclassified Product Code: FRO Dated: February 20, 2013 Received: February 25, 2013

Dear Ms. Bunn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

Image /page/2/Picture/10 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the image.

Letter dated: April 5, 2013

{3}------------------------------------------------

Page 2 - Angela L. Bunn, RAC

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.

Sincerely yours,

Mark Nighterson -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known): _ _ _ _ _ _ K 123892

Device Name: Mepilex® Transfer Ag

Indications For Use:

Mepilex® Transfer Ag dressing is indicated for the management of a wide range of exuding wounds such as leg and foot ulcers, pressure ulcers, partial thickness burns, traumatic and surgical wounds.

Mepilex® Transfer Ag can also be used under compression bandaging.

Prescription Use X X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Jiyoung Dang -S

(Division Sign-Off) Division of Surgical Devices 510(k) Number: K123892

N/A