K Number
K123840

Validate with FDA (Live)

Date Cleared
2013-02-14

(63 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Connect™ for da Vinci® Surgical System(s) is an accessory intended for use by trained surgical proctors to (1) communicate and provide surgical advice to the operating surgeon when using da Vinci Surgical System(s) and (2) view surgical procedures related to the use of da Vinci Surgical System(s) using the Remote Proctor Interface. The wireless connectivity option provides a suitable alternative for the wired Ethernet connection between the da Vinci Surgical System(s) and the hospital's Internet Protocol (IP) infrastructure.

Device Description

Connect is a software accessory, with a Remote Proctor Interface, intended for use by trained surgical proctors to communicate and provide surgical advice to operating surgeons when using the da Vinci Surgical System. By using the Remote Proctor Interface, proctors can also view da Vinci Surgical Procedures. Communication can be established through either a wired or wireless Ethernet connection between the da Vinci Surgical System(s) and the hospital's Internet Protocol (IP) infrastructure.

The modified Connect software developed by ISI has the following capabilities:
• Bi-directional audio
• Video from the OR to the RPI
• Enable/Disable connection on the da Vinci Touchscreen
• Image selection from RPI
• Remote Telestration
• Remote Pointer

AI/ML Overview

The provided document, K123840, is a 510(k) summary for the "Connect for da Vinci Surgical System(s)," which is a software accessory. This submission is a Special 510(k), indicating modifications to an already cleared device. Special 510(k)s often rely on verification that changes do not adversely affect functionality, rather than extensive clinical studies.

Here's an analysis of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Functional Performance (General)All test cases met requirements, demonstrating the subject device performed as expected.
Maintenance of Intended UseThe modified Connect accessory has the same Intended Use as the predicate devices.
Technological CharacteristicsThe fundamental technological characteristics are identical to the predicate device(s). Both encode audio/video for internet transport and utilize embedded da Vinci system software.
Safety and Effectiveness (Impact of Changes)Risk analysis identified no new issues related to safety and effectiveness. Verification and validation results demonstrate the modified software performs as expected and in accordance with functional specifications.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state a "sample size" in terms of patient data or clinical cases for the testing. Instead, it refers to "Software Verification Testing and Validation Testing." This type of testing typically involves:

  • Test Cases: The document states "All test cases met requirements." The number of specific test cases is not provided.
  • Data Provenance: Not applicable in the context of software verification and validation testing, which focuses on device functionality rather than clinical data from patients. The testing would have been performed by Intuitive Surgical, Inc.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Not applicable. Software verification and validation testing generally relies on predefined functional specifications, engineering standards, and test protocols to determine if the software performs as designed. It does not typically involve human experts establishing a "ground truth" for clinical outcomes in the same way a diagnostic AI might. Approval is based on functional equivalence to predicates.

4. Adjudication Method for the Test Set

Not applicable. As described above, the testing involved software verification and validation against specified requirements, not the adjudication of clinical outcomes by experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The device is a software accessory for communication and viewing surgical procedures, not a diagnostic or interventional AI that requires human reader comparisons for effectiveness. The submission focuses on functional equivalence and safety of software modifications.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, in a sense, a standalone evaluation of the software's performance was done through "Software Verification Testing and Validation Testing." This type of testing evaluates the algorithm and its interaction with the system's software independently of a human user's performance, ensuring the software features (bi-directional audio, video, remote telestration, etc.) function correctly. However, this is distinct from a standalone clinical performance study for a diagnostic AI.

7. The Type of Ground Truth Used

The "ground truth" in this context is the functional specification of the software. The testing verified that the modified software performed according to these engineering and functional specifications and that changes had no adverse impact on existing functionalities.

8. The Sample Size for the Training Set

Not applicable. The "Connect" device is a software accessory for communication and viewing surgical procedures. It is not an artificial intelligence/machine learning (AI/ML) device that would involve a "training set" of data to learn patterns or make predictions. The software operates based on deterministic programming and design specifications.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

K123840'

Intuitive Surgical, Inc.

Special 510(k) Connect for da Vincio Surgical System(s)

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.

510(k) Number:

DateFebruary 12, 2013
SubmitterIntuitive Surgical, Inc.
1266 Kifer Road
Sunnyvale, CA 94086
ContactCherece L. Jones
Sr. Regulatory Affairs Specialist
Telephone: (408) 523 - 6925
Fax: (408) 523 - 8907
e-mail: cherece.jones@intusurg.com
SubjectDeviceTrade Name: ConnectTM for da Vinci® Surgical System(s)
Common Name: System, surgical, computer controlledinstrument
Classification Name: Endoscope and Accessories, 21 CFR876.1500
Predicate DevicesIntuitive Surgical® da Vinci® STM Surgical System, Model IS2000with da Vinci OnSiteTM and da Vinci ConnectTM (legally marketedunder K081207).
Intuitive Surgical® da Vinci® STM Surgical System, Model IS2000with ConnectTM OnSiteTM and Wireless Connectivity Option (legallymarketed under K101581).
DeviceDescriptionConnect is a software accessory, with a Remote Proctor Interface,intended for use by trained surgical proctors to communicate andprovide surgical advice to operating surgeons when using the daVinci Surgical System. By using the Remote Proctor Interface,proctors can also view da Vinci Surgical Procedures.Communication can be established through either a wired or
wireless Ethernet connection between the da Vinci Surgical
System(s) and the hospital's Internet Protocol (IP) infrastructure.
The modified Connect software developed by ISI has thefollowing capabilities:
• Bi-directional audio
• Video from the OR to the RPI
• Enable/Disable connection on the da VinciTouchscreen
• Image selection from RPI
• Remote Telestration
• Remote Pointer
Indications forUseConnect™ for da Vinci® Surgical System(s) is an accessoryintended for use by trained surgical proctors to (1) communicateand provide surgical advice to the operating surgeon when usingda Vinci Surgical System(s) and (2) view surgical proceduresrelated to the use of da Vinci Surgical System(s) using the RemoteProctor Interface. The wireless connectivity option provides asuitable alternative for the wired Ethernet connection between theda Vinci Surgical System(s) and the hospital's Internet Protocol(IP) infrastructure.
TechnologicalCharacteristicsThe fundamental technological characteristics of the subjectdevice are identical to the predicate device(s). Both the subjectand predicate device(s):• encode audio and video for transport over the internet, and• utilize the embedded da Vinci system software to facilitateConnect's interaction with the system.
PerformanceDataSoftware Verification Testing and Validation Testing wasperformed to confirm that the modified Connect accessoryperformed as intended and that changes made to the software hadno adverse impact on the functionality of the system. All testcases met requirements demonstrating that the subject deviceperformed as expected.
ConclusionThe modified Connect accessory has the same Intended Use andtechnological characteristics and similar Indications for Use andfunctional characteristics as the predicate devices (K081207 and

Page | 1

{1}------------------------------------------------

. .

.

·

'

{2}------------------------------------------------

K101581).

The Risk analysis did not identify any new issues related to the safety and effectiveness of the device. In addition, verification and validation results demonstrate that the modified software performs as expected and in accordance to the functional specification, thereby also supporting that the Connect accessory is substantially equivalent to the predicate devices.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of a human figure, represented by three curved lines, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the upper portion of the circle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Intuitive Surgical, Incorporated % Ms. Cherece L. Jones Senior Regulatory Affairs Specialist 1266 Kifer Road Sunnyvale, California 94086

February 14, 2013

Re: K123840

Trade/Device Name: Connect ™ for da Vinci@ Surgical System(s) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: NAY Dated: January 31, 2013 Received: February 01, 2013

Dear Ms. Jones:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 -- Ms. Cherece L. Jones

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Peter D. Rumm -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Intuitive Surgical, Inc.

INDICATIONS FOR USE

510(k) Number if known: K123840

Device Name: Connect" for da Vincio Surgical System(s)

Connect™ for da Vincio Surgical System(s) is an accessory intended for use by trained surgical proctors to (1) communicate and provide surgical advice to the operating surgeon when using da Vinci Surgical System(s) and (2) view surgical procedures related to the use of da Vinci Surgical System(s) using the Remote Proctor Interface. The wireless connectivity option provides a suitable alternative for the wired Ethernet connection between the da Vinci Surgical System(s) and the hospital's Internet Protocol (IP) infrastructure.

Prescription Use X (Per 21 CFR 801 Subpart D) Subpart C)

AND/OR ...

Over-the-Counter Use (Per 21 CFR 807

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Image /page/5/Picture/11 description: The image shows the name "Long Chen" in large, bold font on the left side. To the right of the name is a graphic design element that includes the letter "S". The right side of the image contains a digital signature block with text such as "Digitally signed by Long H. Chen -5" and a date: "2013.02.20 08:25:27-05'00'".

(Division Sign-Off)

Division of Surgical Devices

510(k) Number

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.