K Number
K123812

Validate with FDA (Live)

Date Cleared
2013-03-01

(80 days)

Product Code
Regulation Number
892.5050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the SIEMENS branded ARTISTE™, ONCOR™ and PRIMUS™ family of linear accelerator systems is to deliver X-Ray photon and electron radialion for the therapeulic treatment of cancer.

COHERENCE™ RT Therapist is a component of the linear accelerator system and is based on the syngo@ architecture. It enables patient management, patient selection/selup, palient positioning verification, treatment delivery/verification, and treatment recording.

COHERENCE™ RT Therapist Workspace Connect v2.3 can be Interfaced with 3rd party devices conforming to the DICOM standard.

Device Description

COHERENCE™ RT Therapist Connect Workspace, v2.3:

The COHERENCE™ RT Therapist Connect Workspace v2.3 release is intended to update customers with the currently released COHERENCE™ RT Therapist Connect Workspace with versions v2.1a (ONCOR / PRIMUS systems). The technological characteristics of the COHERENCE™ RT Therapist Connect Workspace v2.3 remain unchanged from the currently cleared products.

The syngo® Software Architecture:

The COHERENCE™ RT Therapist Connect software utilizes the proprietary syngo® software architecture design provides a method of delivering customized software applications based on the modality as clinically supporting packages. From these applications SIEMENS utilizes, as part of the Oncology clinical focus package, multiple applications for patient set-up and position verification, treatment localization, treatment verification, portal imaging as well as data processing, image reformatting, display and printing. The currently cleared COHERENCE™ and syngo@ products also include an array of image-oriented software tools, support for DICOM connectivity, Siemens Remote Service, and virus protection features.

AI/ML Overview

The provided text describes the COHERENCE™ RT Therapist Connect Workspace, v2.3, and its substantial equivalence to predicate devices, but it does not contain the specific information requested about acceptance criteria and a study proving the device meets those criteria, as typically found in a clinical performance study summary.

This document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than providing detailed clinical performance study results with specific acceptance criteria, sample sizes, and ground truth methodologies. The device described is a software update for a radiation therapy system, and the testing performed primarily consists of bench testing (unit, integration, system integration, and regression testing) to ensure functionality and safety.

Therefore, many of the requested fields cannot be filled from the provided text.

Here's an attempt to answer based on the information available:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria CategoryReported Device Performance
Functional RequirementsSuccessfully verified and traced in accordance with the Siemens product development (lifecycle) process (PDP). All testable requirements in the Product Requirements Specifications (PRS) for the Sys_VA32a project were met. Specific requirements for the implementation of the third-party OIS were also verified.
Safety RequirementsRisk management ensured via risk analysis to identify potential hazards and mitigations. These hazards are controlled by software means, user instructions, verification of requirements, and validation of clinical workflow. Adheres to recognized and established industry practice and relevant international standards to minimize electrical, mechanical, and radiation hazards.
Verification & ValidationSoftware verification and regression testing performed successfully to meet previously determined acceptance criteria. System-level validation and regression testing performed successfully, demonstrating that the software meets acceptance criteria as noted in the system test plans.
Consensus StandardsTested to meet requirements for conformity (where applicable) to IEC 60601-1-4:1996+ A1: 1999, IEC 62304:2006, IEC 61217 (2007), and IEC 62274:2005.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not specified. The document refers to "testable requirements" and "system test level on production prototype devices" but does not quantify the number of cases or datasets used.
  • Data Provenance: Not specified. The testing described appears to be internal validation of software rather than a study using external clinical data. It is implied that the testing was performed by Siemens personnel.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Number of Experts: Not specified.
  • Qualifications of Experts: The testing was performed by "appropriately trained and knowledgeable test personnel." No further details on specific expert qualifications (e.g., medical professionals) are provided, as this was functional and safety testing of software, not a clinical diagnostic study.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Adjudication Method: Not specified. The verification and validation processes are mentioned but not the specific methods for resolving discrepancies if they arose during testing.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, an MRMC comparative effectiveness study was not done. This document describes a software update for a radiation therapy system, not a diagnostic AI tool requiring assessment of human reader performance.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: The testing described (unit, integration, system integration, and regression testing) is essentially an assessment of the algorithm's functionality and adherence to specifications, which could be considered a form of standalone testing in the context of software validation. However, the document doesn't present "performance" in terms of clinical metrics but rather in terms of meeting functional and safety requirements.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Type of Ground Truth: The "ground truth" for this software validation was the Product Requirements Specifications (PRS) and "previously determined acceptance criteria" specified in system test plans. The software's performance was measured against these predefined functional and safety requirements, rather than against external clinical ground truth like pathology or patient outcomes.

8. The sample size for the training set

  • Sample Size for Training Set: Not applicable. This document is about a software update for an existing system, not a machine learning or AI algorithm that would typically require a training set.

9. How the ground truth for the training set was established

  • Ground Truth for Training Set Establishment: Not applicable, as there is no mention of a training set or machine learning in the provided document.

{0}------------------------------------------------

Siemens Medical Solutions USA, Inc. Radiation Oncology

K123812

Section 5

MAR 1 2013

Page Sect. 5/1

510(k) Summary

Date Prepared:Date UpdatedDecember 05, 2012February 12, 2013
Submitter:Siemens Medical Solutions USA, Inc.Radiation Oncology757A Arnold DriveMartinez, CA 94553
Contact:Christine DunbarSenior Manager, Regulatory Affairs
Phone:Fax:Email:(925)293-5442(925)602-8019christine.dunbar@siemens.com
Proprietary Name:COHERENCE™ RT Therapist Connect Workspace, v2.3And the optional 3rd party OIS Connectivity System
Common Name:Accessory To; Medical Charged-Particle Radiation Therapy System
Classification:892.5050
Product Code:IYE

Substantial Equivalence Claimed To:

Product510(k)Clearance /DateClaim of Equivalence for:
COHERENCE™ RTTherapist v2.1a withControl Console 9.2.K090683 /May 08,2009COHERENCE™ RT Therapist Connect Workspace, forRT Therapist with Software update Sys_VA32a (RTTCv2.3) and Control Console 9.2.
ONCOR™ with 160MLCOption with RTT v2.2 & withControl Console 11.0.K092145 /March 13,2009COHERENCE™ RT Therapist Connect Workspace forRT Therapist with Software update Sys_VA32a (RTTCv2.3) and Control Console 11.0.
ONCOR™ Expression withRTT v2.0+ & with ControlConsole 9.0+K060226 /Mar. 15,2006COHERENCE™ RT Therapist Connect Workspace forRT Therapist with Software update Sys_VA32a (RTTCv2.3) and Control Console 9.2.

The update to the COHERENCE™ RT Therapist Connect Workspace, v2.3 as described in this premarket notification has the same intended use and fundamental scientific technical characteristics as the predicate devices listed above.

Description Summary

COHERENCE™ RT Therapist Connect Workspace, v2.3:

510(k) for SIEMENS COHERENCE™ RT Therapist Connect Update +30 Party OIS

{1}------------------------------------------------

Technological Characteristics:

The COHERENCE™ RT Therapist Connect Workspace v2.3 release is intended to update

customers with the currently released COHERENCE™ RT Therapist Connect Workspace with versions v2.1a (ONCOR / PRIMUS systems). The technological characteristics of the COHERENCE™ RT Therapist Connect Workspace v2.3 remain unchanged from the currently cleared products.

The syngo® Software Architecture:

The COHERENCE™ RT Therapist Connect software utilizes the proprietary syngo® software architecture design provides a method of delivering customized software applications based on the modality as clinically supporting packages. From these applications SIEMENS utilizes, as part of the Oncology clinical focus package, multiple applications for patient set-up and position verification, treatment localization, treatment verification, portal imaging as well as data processing, image reformatting, display and printing. The currently cleared COHERENCE™ and syngo@ products also include an array of image-oriented software tools, support for DICOM connectivity, Siemens Remote Service, and virus protection features.

General Safety and Effectiveness:

The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.

Risk management is ensured via a risk analysis, which is used to identify potential hazards and mitigations. These potential hazards are controlled by software means, user instructions, verification of requirements and validation of the clinical workflow to ensure that the product meets its intended uses. To minimize electrical, mechanical and radiation hazards, SIEMENS adheres to recognized and established industry practice and relevant international standards.

Intended Use:

The intended use of the SIEMENS branded ARTISTE™, ONCOR™ and PRIMUS™ family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer:

COHERENCE™ RT Therapist is a component of the linear accelerator system and is based on the syngo® architecture. It enables patient management, patient selection/setup, patient positioning verification, treatment delivery/verification, and treatment recording.

COHERENCE™ RT Therapist Workspace Connect v2.3 can be interfaced with 3rd party devices conforming to the DICOM standard.

Substantial Equivalence:

The Substantial Equivalence comparison chart demonstrates the comparison of the technological characteristics of the COHERENCE™ RT Therapist Connect Workspace v2.3 update to the currently cleared predicate devices.

510(k) for SIEMENS COHERENCE™ RT Therapist Connect Update +310 Party OIS

{2}------------------------------------------------

The COHERENCE™ RT Therapist Connect Workspace v2.3, does not change the intended use of the original COHERENCE™ RT Therapist Connect Workspace v2.1a or the Siemens branded Linear Accelerator Systems.

Bench Testing:

Bench testing in the form of Unit, Integration and System Integration testing was performed to evaluate the performance and functionality of the software update v2.3 for the RT Therapist Connect and regression testing the Control Console, versions 9.2 and 11.0. All testable requirements in the Product Requirements Specifications (PRS) for the Sys_VA32a project, and additionally the specific requirements for the implementation of the third party OIS marketed by have been successfully verified and traced in accordance with the Siemens product development (lifecycle) process (PDP).

The software verification and regression testing has been performed successfully to meet their previously determined acceptance criteria.

Non-Clinical Test Results:

Validation of the COHERENCE™ RT Therapist Connect Workspace v2.3, implementation of the optional third party OIS with the low voltage electronic switch and regression testing with Control Consoles 9.2.18 and 11.0.212 has been performed at the System test level on production prototype devices by appropriately trained and knowledgeable test personnel. System level validation and regression testing has been performed successfully, demonstrating that the software meets the acceptance criteria as noted in the system test plans.

Testing to Consensus Standards:

The COHERENCE™ RT Therapist Connect Workspace, v2.3, Control Console and the new electronic switch to support the third party OIS implementation have been tested to meet the requirements for conformity (where applicable) to the following standards:

  • IEC 60601-1-4:1996+ A1: 1999 Medical Electrical Equipment: Part 1-4: General . requirements for Collateral Standard: Programmable Electrical Medical Systems
  • IEC 62304:2006 Medical Device Software Software Life Cycle Processes ●
  • IEC 61217 (2007), Radiotherapy equipment Coordinates, movements, and scales .
  • IEC 62274:2005 Medical Electrical Equipment, Safety of Radiotherapy Record and . Verify Systems

Substantial Equivalence to Predicates:

The verification testing to the Product requirements for the AT Therapist Connect workspace, validation of the intended use, and the regression testing to the existing RT Therapist Connect software and Control Console functional requirements, is intended to support the claim of substantial equivalence to the following predicates:

  • The COHERENCE™ RT Therapist Connect v2.1a. & CC 9.2 (K090683), .
  • The ONCOR with 160MLC option with the RTT v2.2 & CC 11.0 (K092145) and ◆

510(k) for SIEMENS COHERENCE™ RT Therapist Connect Update +31d Party OIS Page Sect. 5/3

{3}------------------------------------------------

. The ONCOR and PRIMUS with the RTT v2.0 and 2.1 & CC 9.0+(K060226) Summary:

In summary, it is SIEMENS' opinion that the COHERENCE™ RT Therapist Connect Workspace v2.3 update including the optional 30 party OIS Connectivity systems requiring the low voltage electronic switch where configured, does not introduce any new potential safety risks and is substantially equivalent to, and performs as well as, the predicate devices.

Image /page/3/Picture/4 description: The image shows the text "Page Sect. 5/4" on the top line and "0026" on the second line. The top line appears to indicate a page and section number, while the second line likely represents a document or record identifier. The text is written in a clear, legible font, making it easy to read. The image is likely a snippet from a larger document or record.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. The eagle is black, and the text is also black.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 1, 2013

Christine Dunbar Senior Manager, Regulatory Affairs Siemens Medical Solutions USA, Inc. Radiation Oncology 757A Arnold Drive MARTINEZ CA 94553

Re: K123812

Trade/Device Name: COHERENCE™ RT Therapist Connect Workspace Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: December 7. 2012 Received: January 15, 2013

Dear Ms. Dunbar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2-Ms. Dunbar

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michaela Hara

for Janine M. Morris, M.S. Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K123812

Device Name: COHERENCETM RT Therapist Connect Workspace, v2.3

Indications for Use:

The intended use of the SIEMENS branded ARTISTE™, ONCOR™ and PRIMUS™ family of linear accelerator systems is to deliver X-Ray photon and electron radialion for the therapeulic treatment of cancer.

COHERENCE™ RT Therapist is a component of the linear accelerator system and is based on the syngo@ architecture. It enables patient management, patient selection/selup, palient positioning verification, treatment delivery/verification, and treatment recording.

COHERENCE™ RT Therapist Workspace Connect v2.3 can be Interfaced with 3rd party devices conforming to the DICOM standard.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR