K Number
K123803

Validate with FDA (Live)

Date Cleared
2013-07-02

(204 days)

Product Code
Regulation Number
870.3300
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Reverse Medical™ Micro Vascular Plug (MVP™) System is intended for use to obstruct or reduce the rate of blood flow in the peripheral vasculature.

Device Description

The Reverse Medical Micro Vascular Plug (MVP) is a micro vascular occlusion device comprised of a detachable embolic plug attached to a composite delivery wire and designed for delivery via a microcatheter (0.021" ID). The MVP is a self-expandable, ovoid-shaped device made from Nitinol with an ePTFE partial cover. The device is secured at both ends with platinum marker bands. The Reverse Medical MVP is intended to reduce or occlude vascular blood flow of vessels having a diameter of 1.5 – 3.0mm. The proximal marker band attaches to a delivery wire that pushes through a commercially available catheter to the intended treatment site. The Reverse Medical Detachment Box regulates detachment of the implant device from the delivery wire by electrolytic means during deployment, and monitors, detects, signals and measures the time of detachment. The Reverse Medical Cable Set – 275 cm length (Model RMCS – 2.75US) is provided sterile. The cable set connects to the Detachment Box through a bayonet type dual pin connector that ensures correct polarity. The Reverse Medical Cable Set and Detachment Box will be sold separately. One 9-volt battery and a sterile needle (20 G or 22 G) will also be needed for use with the Reverse Medical Micro Vascular Plug (MVP).

AI/ML Overview

This document is a 510(k) Summary for the Reverse Medical MVP™ Micro Vascular Plug System, which focuses on demonstrating substantial equivalence to existing predicate devices. It does not contain a study that establishes acceptance criteria for specific device performance metrics in the way a clinical trial or a detailed performance study with defined endpoints would. Instead, it relies on non-clinical data (biocompatibility, sterilization, and bench-top testing) to show that the device performs as intended and is comparable to predicate devices.

Therefore, many of the requested sections (sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable to this type of regulatory submission, as it's not a clinical effectiveness study of an AI/ML device.

Here's a breakdown of the available information:

1. A table of acceptance criteria and the reported device performance

The document provides a table for biocompatibility tests, which acts as acceptance criteria for the biological safety of the materials. It also lists various design verification tests (bench-top testing) that were performed, indicating that "All tests performed passed successfully." However, specific numerical acceptance criteria and reported performance values for each of these mechanical/performance tests are not provided in this summary.

Test TypeAcceptance Criteria (Stated)Reported Device Performance (Summary)
Biocompatibility
CytotoxicityNon-CytotoxicNon-Cytotoxic
Kligman Maximization Sensitization TestNon-SensitizingNon-Sensitizing
Intracutaneous InjectionNon-IrritantNon-Irritant
Systemic InjectionNon-ToxicNon-Toxic
Material-Mediated PyrogenicityNon-PyrogenicNon-Pyrogenic
Genotoxicity/MutagenicityNon-MutagenicNon-Mutagenic
In Vitro Mouse Lymphoma AssayNon-MutagenicNon-Mutagenic
In Vivo Mouse LymphomaNon-MutagenicNon-Mutagenic
HemolysisNon-HemolyticNon-Hemolytic
Complement Activation C3a and SC5ba-9No greater biological response than corresponding controlNo greater biological response than corresponding control
Inactivated Partial Thromboplastin TimeMinimal, passed acceptance criteriaNon-activator, passed acceptance criteria
Platelet and Leukocyte CountsNo range or acceptable level established (Test Articles)(No specific numeric result; assumed within acceptable range for testing)
Muscle ImplantationPassed acceptance criteria (4 and 13 week)Intramuscular Implantation- 4 and 13 week, passed acceptance criteria
Design Verification (Bench-Top Testing)
Visual Inspection(Implied to meet specifications)Passed successfully
Dimensional Inspection(Implied to meet specifications)Passed successfully
Tensile Strength(Implied to meet specifications)Passed successfully
USP Particulate(Implied to meet specifications)Passed successfully
Radial Force(Implied to meet specifications)Passed successfully
Microcatheter Compatibility(Implied to meet specifications)Passed successfully
Detachment Time(Implied to meet specifications)Passed successfully
Torque Strength(Implied to meet specifications)Passed successfully
Plug Foreshortening(Implied to meet specifications)Passed successfully
Nickel Release(Implied to meet specifications)Passed successfully
Corrosion Resistance(Implied to meet specifications)Passed successfully
Flow Occlusion/Reduction(Implied to meet specifications)Passed successfully
Magnetic Resonance Compatibility(Implied to meet specifications)Passed successfully
Labeling(Implied to meet specifications)Passed successfully
Packaging(Implied to meet specifications)Passed successfully
Shelf Life(Implied to meet specifications)Passed successfully
SterilitySterility Assurance Level (SAL) of 10⁻⁶Validated to provide SAL of 10⁻⁶
Biocompatibility(As detailed above)Passed successfully
Detachment Box and Cable Set(Implied to meet specifications)Passed successfully

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not specified for individual bench tests; generally involves representative samples of manufactured devices.
  • Data Provenance: Not explicitly stated, but all testing described is non-clinical (bench-top, in vitro/ex vivo lab testing, and animal testing for biocompatibility). Given the company location (Irvine, CA), it's highly likely this testing was conducted in the US or by US-based contractors, but this is not explicitly stated as "country of origin for data." All data is prospective insofar as it was generated specifically for this 510(k) submission, not gathered from past clinical procedures.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. This is a medical device 510(k) for a physical implant, not an AI/ML diagnostic device requiring expert interpretation for ground truth. Ground truth for these tests is based on objective measurements, chemical analysis, and standardized biological observations.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. Ground truth is established through laboratory test protocols, not physician adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a physical implant, not an AI/ML diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is a physical implant, not an AI/ML algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For biocompatibility: Ground truth is established by standardized biological assays and observation of biological responses, compared against control samples where relevant.
  • For design verification: Ground truth is established by engineering specifications, objective physical measurements (e.g., tensile strength, dimensions, detachment time), and functional tests (e.g., flow occlusion in a mock vessel).

8. The sample size for the training set

  • Not applicable. This device does not involve a "training set" in the context of machine learning.

9. How the ground truth for the training set was established

  • Not applicable. As above, no machine learning training set is involved.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Reverse Medical. The logo is in black and white and features the words "Reverse Medical" in a bold, sans-serif font. To the upper right of the word "Medical" is the trademark symbol. Underneath the words is a curved line that starts thin on the left and gradually thickens as it curves upward to the right, ending in an arrowhead.

K123803

510(k) Summary

Pursuant to Section 12, Part (a){i)(3A) of the Safe Medical Devices Act of 1990, Reverse Medical Corporation is providing the summary of Substantial Equivalence for the Reverse Medical MVP™ Micro Vascular Plug System.

Device Trade or Proprietary Name

Reverse Medical MVP™Micro Vascular Plug System

Sponsor /Applicant Name and Address

Reverse Medical Corporation 13700 Alton Parkway Suite 167 Irvine, CA 92618

Sponsor Contact Information

Linda D'Abate Reverse Medical Vice President, RA/CA/QA

Date of Preparation of 510(k) Summary

June 24, 2013

Device Common/Usual or Classification Name

Device Embolization, Vascular (21 CFR 870.3300, Product Code: KRD)

Identification of the Legally Marketed Devices to which Equivalence is Being Claimed:

Name of PredicateDevicesManufacturer510(k) Number
AMPLATZER® Vascular Plug 4AGA CorporationPlymouth, MNK113658
Azur Peripheral HydroCoilEndovascular EmbolizationSystem - Detachable 35MicroVention, Inc.Tustin, CAK093002
Axium Detachable Coil Systemev3 Endovascular, Inc.Irvine, CAK081465
InZone Detachment Systemwith the IZDS Connecting CableBoston Scientific, IncFreemont, CAK103008
Guglielmi Detachable CoilPower SupplyBoston Scientific, IncFreemont, CAK001083

Device Description

The Reverse Medical Micro Vascular Plug (MVP) is a micro vascular occlusion device comprised of a detachable embolic plug attached to a composite delivery wire and designed for delivery via a microcatheter (0.021" ID). The MVP is a self-expandable, ovoid-shaped device made from Nitinol with an

2 2013 JUL

{1}------------------------------------------------

ePTFE partial cover. The device is secured at both ends with platinum marker bands. The Reverse Medical MVP is intended to reduce or occlude vascular blood flow of vessels having a diameter of 1.5 – 3.0mm.

The proximal marker band attaches to a delivery wire that pushes through a commercially available catheter to the intended treatment site. The Reverse Medical Detachment Box regulates detachment of the implant device from the delivery wire by electrolytic means during deployment, and monitors, detects, signals and measures the time of detachment. The Reverse Medical Cable Set – 275 cm length (Model RMCS – 2.75US) is provided sterile. The cable set connects to the Detachment Box through a bayonet type dual pin connector that ensures correct polarity. The Reverse Medical Cable Set and Detachment Box will be sold separately. One 9-volt battery and a sterile needle (20 G or 22 G) will also be needed for use with the Reverse Medical Micro Vascular Plug (MVP).

Intended Use

The Reverse Medical MVP is intended for use to obstruct or reduce the rate of blood flow in the peripheral vasculature.

New DevicePredicate Devices
Reverse MedicalMVP SystemAMPLATZER®Vascular Plug 4Azur PeripheralHydroCoilEndovascularEmbolizationSystem –Detachable 35Axium DetachableCoilsInZone DetachmentSystem with the IZDSConnecting CableGuglielmi DetachableCoil Power Supply
510(k) No.K123803 Class IIKRD 870:3300K113658 Class IIKRD 870:3300K093002 Class IIKRD 870:3300K081465 Class IIKRD 870:3300K103008 Class IIKRD 870:3300 andHCGK001083 Class IIHCG 882.5950
Indication foruseIndicated for useto obstruct orreduce the rate ofblood flow in theperipheralvasculature.Indicated for arterialand venousembolizations in theperipheralvasculature.Intended to reduceor block the rate ofblood flow invessels of theperipheralvasculature.The AXIUMDetachable Coils arealso indicated forarterial and venousembolizations in theperipheralvasculatureIntended for use withall versions of BSCDetachable Coils in theembolization ofintracranial aneurysmsand other vascularmalformations of theneuro and peripheralvasculature.Intended for use withall versions of BostonScientific/Target'sGuglielmi DetachableCoils in theembolization ofintracranialaneurysms and othervascularmalformations of theneuro and peripheralvasculature
Method ofPlacementDelivery wirethrough a 0.021" IDmicrocatheter.Delivery wireDelivery wireDelivery wire(pusher)Delivery wireDelivery wire
RadiopaquemarkersPlatinum markerbands at each endRadiopaque markerbands at each endN/ARadiopaque positionmarkerRadiopaque positionmarkerRadiopaque positionmarker
Proximal EndConfigurationProximal markerband andattachment forpusher wireRadiopaque markerband and microscrew attachmentN/AN/AN/AN/A
DetachmentSystemYes - ElectrolyticYes - MechanicalYes - ThermalYes - MechanicalYes - ElectrolyticYes - Electrolytic
BatteryOperatedYesNoYesNoYesYes

Comparison to Predicate Devices

{2}------------------------------------------------

Summary of Non-Clinical Data

Biocompatibility and Sterilization

The device was characterized as an implant, internal communicating device, which contacts circulating blood for exposure ≥ 30 days.

The Reverse Medical MVP materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993-1 guidelines "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." The Reverse Medical MVP successfully passed all of the following biocompatibility tests, demonstrating that the materials are biocompatible:

TestResults/Acceptance
CytotoxicityNon-Cytotoxic
Kligman MaximizationSensitization TestNon-Sensitizing
Intracutaneous InjectionNon-Irritant
Systemic InjectionNon-Toxic
Material-Mediated PyrogenicityNon-Pyrogenic
Genotoxicity/MutagenicityNon-Mutagenic
In Vitro Mouse Lymphoma AssayNon-Mutagenic
In Vivo Mouse LymphomaNon-Mutagenic
HemolysisNon-Hemolytic
Complement Activation C3a andSC5ba-9No greater biological response than corresponding control
Inactivated PartialMinimal, passed acceptance criteria
Thromboplastin TimeNon-activator, passed acceptance criteria
Platelet and Leukocycte CountsTest articles: No range or acceptable level established.
Muscle ImplantationIntramuscular Implantation- 4 and 13 week, passed acceptance criteria

Test Result Summary

Sterilization conditions have been validated according to ANSI / AAMI / ISO 11135, Sterilization of Health Care Products-Part 1: Requirement, validation and routine control of a sterilization process for medical devices to provide a Sterility Assurance Level (SAL) of 10 °.

Design Verification (Bench-Top Testing)

The physical, mechanical, and performance testing of the Reverse Medical MVP System demonstrate that the product is substantially equivalent to the currently marketed predicate devices. Design verification testing was conducted to evaluate the physical and mechanical properties of the Reverse Medical MVP. All studies were conducted in accordance with Reverse Medical Design Control procedures. All testing was performed on units that were sterilized and met all inspection criteria. Tests on the Reverse Medical MVP System included:

{3}------------------------------------------------

Verification and Test Summary

  • Visual Inspection .
  • . Dimensional Inspection
  • Tensile Strength .
  • USP Particulate
  • . Radial Force
  • . Microcatheter Compatibility
  • . Detachment Time
  • . Torque Strength
  • Plug Foreshortening ●
  • . Nickel Release
  • Corrosion Resistance (potentiodynamic and galvanic).
  • Flow Occlusion/Reduction �
  • Magnetic Resonance Compatibility .
  • Labeling ●
  • Packaging .
  • Shelf Life
  • Sterility
  • Biocompatibility .
  • Detachment Box and Cable Set

All tests performed passed successfully. The physical, mechanical, and performance testing of the Reverse Medical MVP System demonstrate that the product is safe and effective for its labeled indications and is Substantially Equivalent to the currently marketed predicate devices

Substantial Equivalence

The performance of the Reverse Medical MVP System demonstrates that the product is substantially equivalent to the performance of the predicate devices. The equivalence was shown through comparison of component materials and specifications, performance, biocompatibility testing, animal testing, and sterilization validation.

The Reverse Medical MVP System is substantially equivalent in intended use, design, technology/principles of operation, materials, and performance to the predicate devices. Differences between the devices do not raise any significant issues of safety or effectiveness.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is depicted with a simple, flowing design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 2, 2013

Reverse Medical Corporation c/o Mr. Jeffrey Valko 13700 Alton Parkway, Suite 167 Irvine, CA 92618

Re: K123803

Trade/Device Name: Reverse Medical Micro Vascular Plug System Regulation Number: 21 CFR 870.3300 Regulation Name: Device. Vascular, For Promoting Embolization Regulatory Class: Class II Product Code: KRD Dated: June 10, 2013 Received: June 11, 2013

Dear Mr.. Valko:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate. commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2 - Mr. Jeffrey Valko

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please . contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDeyices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman -S

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

  1. Indications for Use

510(k) Number (if known):_

Device Name: Reverse Medical™ Micro Vascular Plug (MVP™)

Indication for Use:

: ・

The Reverse Medical™ Micro Vascular Plug (MVP™) System is intended for use to obstruct or reduce the rate of blood flow in the peripheral vasculature.

Prescription Use__X (Part 21 CFR 801 Subpart D) AND/OR

Over the Counter Use_ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bram D. Zuckerman -S
2013.07.02 17:24:16
-04'00'

§ 870.3300 Vascular embolization device.

(a)
Identification. A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).