K Number
K123753

Validate with FDA (Live)

Date Cleared
2013-07-17

(223 days)

Product Code
Regulation Number
866.1640
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IMDx VanR for Abbott m2000 assay is an in vitro diagnostic assay that uses polymerase chain reaction (PCR) amplification for the qualitative detection of nucleic acids encoding the vancomycin resistance genes vanA and/or vanB. The assay is performed directly on human perirectal swabs, rectal swabs, or stool specimens from patients at risk for Vancomycin Resistant Enterococcus (VRE) colonization. The IMDx VanR for Abbott m2000 assay detects the presence of vanA and vanB genes that can be associated with vancomycin-resistant enterococci. The IMDx VanR for Abbott m2000 assay can be used as an aid to identify. prevent and control vancomycin-resistant colonization in healthcare settings. The IMDx VanR for Abbott m2000 assay is not intended to diagnose VRE infection nor to guide or monitor treatment of infection. Culture methods are necessary to recover organisms for epidemiology typing and confirmation testing.

Device Description

The IMDx VanR for Abbott m2000 assay is a PCR-based assay that targets regions unique to the vanA and vanB vancomycin resistance genes that may be associated with vancomycin resistant Enterococcus (VRE). Detection of the vanA and vanB genes is measured by the presence of fluorescently-labeled oligonucleotide probes that generate a fluorescent signal when specifically bound to amplified vanA and/or vanB PCR products. Differentiation of vanA from vanB is attained by labeling the oligonucleotide probes with different colored fluorescent dyes. The amplification cycle at which fluorescent signal is detected by the Abbott m2000rt is inversely proportional to the vanA and vanB DNA target level present in the sample.

The IMDx VanR for Abbott m2000 assay includes reagents for the detection of the assay process control, which contains inactivated bacteria, unrelated to enterococci, and is introduced into each specimen during sample preparation. The process control (also acting as an internal control (IC)) is co-extracted with the specimen and co-amplified in the same PCR reaction as the vanA and vanB targets. The IC demonstrates that the entire assay process has proceeded within specification.

AI/ML Overview

Here's an analysis of the provided information, focusing on the acceptance criteria and the study proving the device meets them:

Acceptance Criteria and Device Performance for IMDx VanR for Abbott m2000

The document provided, K123753, describes the IMDx VanR for Abbott m2000 assay, an in vitro diagnostic device for detecting vancomycin resistance genes. The acceptance criteria are implicitly derived from the performance goals demonstrated in the clinical studies, as explicit pre-defined acceptance criteria are not directly stated in the summary. The performance is reported as sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) based on comparison to a "gold standard" method.

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Implied)Reported Device Performance (95% CI)
Peri-rectal Swab Specimens
SensitivityHigh sensitivity for detecting vanA and/or vanB genes.94.0% (83.8% - 97.9%)
SpecificityHigh specificity for differentiating between detected vanA/vanB and negative samples.93.9% (91.5% - 95.6%)
Positive Predictive ValueAdequate PPV for indicating presence of vanA/vanB.58.8% (47.8% - 68.9%)
Negative Predictive ValueHigh NPV for ruling out vanA/vanB presence.99.4% (98.3% - 99.8%)
Rectal Swab Specimens (Prospective)
SensitivityHigh sensitivity for detecting vanA and/or vanB genes.96.8% (89.1% - 99.1%)
SpecificityHigh specificity for differentiating between detected vanA/vanB and negative samples.91.7% (88.3% - 94.2%)
Positive Predictive ValueAdequate PPV for indicating presence of vanA/vanB.68.5% (58.3% - 77.2%)
Negative Predictive ValueHigh NPV for ruling out vanA/vanB presence.99.4% (97.7% - 99.8%)
Rectal Swab Specimens (Retrospective)
Positive Percent AgreementHigh agreement with the reference method for positive samples.100.0% (91.6% - 100.0%)
Negative Percent AgreementHigh agreement with the reference method for negative samples.0.0% (0.0% - 65.8%) Note: This low NPA is due to the specific composition of the retrospective cohort, where all "negative" samples by the IMDx assay were also negative by the reference standard. The total NEG for the IMDx assay was 0, meaning no calculation was possible. This requires careful interpretation.
Stool Specimens
SensitivityHigh sensitivity for detecting vanA and/or vanB genes.87.0% (77.0% - 93.0%)
SpecificityHigh specificity for differentiating between detected vanA/vanB and negative samples.85.8% (82.0% - 88.8%)
Positive Predictive ValueAdequate PPV for indicating presence of vanA/vanB.51.3% (42.3% - 60.2%)
Negative Predictive ValueHigh NPV for ruling out vanA/vanB presence.97.4% (95.2% - 98.6%)

Note on Acceptance Criteria: The document does not explicitly state numerical acceptance criteria (e.g., "Sensitivity must be >X%"). Instead, the clinical performance study aims to demonstrate that the device performs comparably and effectively for its intended use, implying that the reported percentages of sensitivity, specificity, PPV, and NPV are deemed acceptable for supporting substantial equivalence to the predicate device.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set (Clinical Performance Study):
    • Peri-rectal swabs: 587 samples
    • Rectal swabs: 444 samples (further broken down into prospective and retrospective collections)
      • Prospective: 400 samples
      • Retrospective: 44 samples
    • Stool specimens: 469 samples
    • Total Clinical Samples: 587 + 444 + 469 = 1500 samples
  • Data Provenance: Samples were collected from "five geographically diverse test sites within the United States." The study included both prospective and retrospective collection for rectal swabs.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

The document does not specify the number or qualifications of experts used to establish the ground truth. It states that the ground truth was "enriched culture combined with confirmation of the molecular basis of vancomycin resistance of isolates using an alternate PCR method." This implies laboratory personnel with expertise in microbiology and molecular diagnostics, but no specific number or detailed qualifications are provided.

4. Adjudication Method for the Test Set

The document does not describe a formal adjudication method (e.g., 2+1, 3+1, none) for the test set. The ground truth was established by a combination of enriched culture and an alternate PCR method, implying a direct comparison against these established laboratory techniques rather than a human consensus process.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This study focuses on the in vitro diagnostic performance of the assay itself (device vs. reference method) rather than evaluating human reader performance with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the clinical performance study evaluates the IMDx VanR for Abbott m2000 assay in a standalone manner. The assay is an automated PCR-based diagnostic test, and its results are compared directly to the ground truth (enriched culture + alternative PCR). There is no human interpretation of images or other data being assisted by the algorithm; the assay itself generates the result.

7. The Type of Ground Truth Used

The ground truth used was a combination of enriched culture combined with confirmation of the molecular basis of vancomycin resistance of isolates using an alternate PCR method. This is a highly robust ground truth for detecting the presence of vanA and/or vanB genes and is considered a reference standard in microbiology.

8. The Sample Size for the Training Set

The document does not explicitly state the sample size for a "training set." This type of molecular diagnostic assay typically undergoes extensive analytical validation (e.g., LOD, reactivity, cross-reactivity) using characterized strains and spiked samples, and then its clinical performance is evaluated on a distinct clinical test set. It's possible that data from these extensive analytical studies could be considered analogous to training data for method optimization, but a defined "training set" in the context of machine learning is not applicable here as it's a rule-based PCR assay, not an AI/ML algorithm.

9. How the Ground Truth for the Training Set Was Established

As noted above, a traditional "training set" as understood in AI/ML is not described. However, for the analytical studies (e.g., Analytical Reactivity, Challenge Study), the ground truth was established using "well-characterized vancomycin-resistant Enterococcus strains and/or clinical isolates" and other organisms. These strains would have their vanA/vanB status, and other genetic characteristics, confirmed through standard microbiological and molecular methods (e.g., sequencing, biochemical tests, established PCR assays).

{0}------------------------------------------------

K123753

27 IMDx VanR for Abbott m2000 Supplement IMDx Response to FDA Request for Additional Information

510(k) SUMMARY

Date of Summary:July 10, 2013JUL 17 2013
Product NameIMDx VanR for Abbott m 2000
SponsorIntelligent Medical Devices, Inc.19 Blackstone StreetCambridge, MA 02139
CorrespondentMDC Associates, LLCFran White, Regulatory Consultant180 Cabot StreetBeverly, MA 01915
Device Identification
Trade or Proprietary Name:Common or Usual Name:IMDx VanR for Abbott m 2000Vancomycin-resistant enterococci detection assay

Product Code: NIJ, OOI Regulation Section: 21 CFR 866.1640 Antimicrobial susceptibility test powder Device Class: Class II Panel: 83; Microbiology

Device Description

The IMDx VanR for Abbott m2000 assay is a PCR-based assay that targets regions unique to the vanA and vanB vancomycin resistance genes that may be associated with vancomycin resistant Enterococcus (VRE). Detection of the vanA and vanB genes is measured by the presence of fluorescently-labeled oligonucleotide probes that generate a fluorescent signal when specifically bound to amplified vanA and/or vanB PCR products. Differentiation of vanA from vanB is attained by labeling the oligonucleotide probes with different colored fluorescent dyes. The amplification cycle at which fluorescent signal is detected by the Abbott m2000rt is inversely proportional to the vanA and vanB DNA target level present in the sample.

The IMDx VanR for Abbott m2000 assay includes reagents for the detection of the assay process control, which contains inactivated bacteria, unrelated to enterococci, and is introduced into each specimen during sample preparation. The process control (also acting as an internal control (IC)) is co-extracted with the specimen and co-amplified in the same PCR reaction as the vanA and vanB targets. The IC demonstrates that the entire assay process has proceeded within specification.

Intended Use

The IMDx VanR for Abbott m2000 assay is an in vitro diagnostic assay that uses polymerase chain reaction (PCR) amplification for the qualitative detection of nucleic acids encoding the

{1}------------------------------------------------

IMDx Response to FDA Request for Additional Information

vancomycin resistance genes vanA and/or vanB. The assay is performed directly on human perirectal swabs, rectal swabs, or stool specimens from patients at risk for Vancomycin Resistant Enterococcus (VRE) colonization. The IMDx VanR for Abbott m2000 assay detects the presence of vanA and vanB genes that can be associated with vancomycin-resistant enterococci. The IMDx VanR for Abbott m2000 assay can be used as an aid to identify. prevent and control vancomycin-resistant colonization in healthcare settings. The IMDx VanR for Abbott m2000 assay is not intended to diagnose VRE infection nor to guide or monitor treatment of infection. Culture methods are necessary to recover organisms for epidemiology typing and confirmation testing.

Substantial Equivalency

The IMDx VanR for Abbott m2000 assay is substantially equivalent to the BD GeneOhm™ VanR Assay (K102416). Table 1 compares the characteristics of the IMDx VanR for m2000 assay (New Device) and the GeneOhm assay (Predicate Device).

Similarities
CharacteristicPredicate DeviceBD GeneOhm VanR Assay(K102416)New DeviceIMDx VanR forAbbott m2000
Intended UseThe BD GeneOhm VanR Assay is aqualitative in vitro test for the rapid detectionof vancomycin-resistant ( vanA and vanB )genes directly from perianal or rectal swabs.The BD GeneOhm VanR Assay detects thepresence of the vanA and vanB genes that canbe associated with vancomycin-resistantenterococci (VRE). The assay is performed onan automated real-time PCR instrument withperianal or rectal swabs from individuals atrisk for VRE colonization. The BD GeneOhmVanR Assay can be used as an aid to identify,prevent and control vancomycin-resistantcolonization in healthcare settings. The BDGeneOhm VanR Assay is not intended todiagnose VRE infections nor to guide ormonitor treatment for VRE infections.Concomitant cultures are necessary to recoverorganisms for epidemiological, susceptibilitytesting and for further confirmatoryidentification.The IMDx VanR for Abbott m2000 assayis an in vitro diagnostic assay that usespolymerase chain reaction (PCR)amplification for the qualitative detectionof nucleic acids encoding the vancomycinresistance genes vanA and/or vanB . Theassay is performed directly on human peri-rectal swabs, rectal swabs, or stoolspecimens from patients at risk forVancomycin Resistant Enterococcus(VRE) colonization. The IMDx VanR forAbbott m2000 assay detects the presenceof vanA and vanB genes that can beassociated with vancomycin-resistantenterococci. The IMDx VanR for Abbottm2000 assay can be used as an aid toidentify, prevent and control vancomycin-resistant colonization in healthcare settings.The IMDx VanR for Abbott m2000 assayis not intended to diagnose VRE infectionnor to guide or monitor treatment ofinfection. Culture methods are necessaryto recover organisms for epidemiologytyping and confirmation testing.
Sample TypePerianal or rectal swab specimensRectal and peri-rectal swabsor stool specimens
Test PrincipleReal-time PCR DNA amplificationReal-time PCR DNA amplification

Table 1: Substantial Equivalence.

{2}------------------------------------------------

IMDx Response to FDA Request for Additional Information

Similarities
CharacteristicPredicate DeviceBD GeneOhm VanR Assay(K102416)New DeviceIMDx VanR forAbbott m2000
TargetsDetectedvanA and vanBvanA and vanB
ControlsPositive ControlNegative ControlProcess ControlPositive ControlNegative ControlProcess Control
Differences
CharacteristicPredicate DeviceBD GeneOhm VanR Assay(K102416)New DeviceIMDx VanR for Abbott m2000
InstrumentCepheid SmartCycler SystemAbbott m2000 System
SamplePreparationManualAutomated

The differences between the IMDx VanR for Abbott m2000 assay and the BD GeneOhm VanR Assay do not impact substantial equivalence. Both assays detect vanA and vanB nucleic acids from similar specimen types and have comparable intended uses. The differences noted above do not change the intended use of the IMDx VanR for Abbott m2000 and do not raise questions regarding the safety and effectiveness of the device.

Performance Characteristics Analytical Performance

Precision (Repeatability/Reproducibility)

Assay precision was measured in four independent studies: within laboratory repeatability, userto-user reproducibility, lot-to-lot reproducibility and instrument reproducibility using a seven-member panel consisting of one vanA and one vanB VRE strain at varving concentrations: Positive (2-3X LoD), Low Positive (LoD) and High Negative (<1X LoD). The final panel member was a true negative sample (negative specimen matrix alone).

Table 2 lists %CV values for precision studies. Table 3 lists % agreement values for precision studies.

PrecisionPanel MemberReproducibilityRepeatabilityLot to LotInstrument to Instrument
vanAPositive3.3%3.6%0.7%1.0%0.8%
vanALow Positive3.9%4.7%2.8%1.5%1.7%

Table 2. Precision (%CV) for vanA/vanB based on CN values.

{3}------------------------------------------------

IMDx Response to FDA Request for Additional Information

PrecisionReproducibilityRepeatabilityLot to LotInstrument to Instrument
vanAHigh Negative3.4%3.3%3.6%3.5%2.8%
vanBPositive2.0%2.3%0.6%1.1%0.8%
vanBLow Positive2.6%2.1%1.5%1.3%1.5%
vanBHigh Negative3.2%2.1%1.7%2.4%2.6%
NegativeN/AN/AN/AN/AN/A

Table 3. Precision (% Agreement).

PrecisionPanelExpectedPositivityReproducibilityRepeatabilityLot toLotInstrument toInstrument
Memberm2000spm2000rt
vanAPositive100%100.0%97.2%100.0%100.0%100.0%
vanALow Positive95%96.7%95.8%100.0%100.0%100.0%
vanAHighNegative20 - 80%66.7%55.6%44.4%55.6%55.6%
vanBPositive100%100.0%100.0%100.0%100.0%100.0%
vanBLow Positive95%100.0%100.0%100.0%100.0%100.0%
vanBHighNegative20 - 80%35.6%27.8%55.6%16.7%27.8%
Negative0%0.0%0.0%0.0%0.0%0.0%

Analytical Sensitivity (Limit of Detection)

Limit of Detection (LoD) studies were conducted to determine the lowest concentration of each target analyte that could be detected ≥ 95% of the time. Six strains of vancomycin-resistant Enterococcus, four VanA-type and two VanB-type, were tested. Dilutions of quantified bacterial stocks were tested in replicates of sixty (60). The LoD of the assay was estimated using probit analysis for each target (vanA and vanB). The observed LoD for VRE strains tested is provided in Table 4.

{4}------------------------------------------------

IMDx Response to FDA Request for Additional Information

StrainGenotypeLimit of Detection (95% CI)
E. faecium ATCC 51559vanA1010.7 CFU/swab (975.1 - 1047.5)
E. faecium ATCC 700221vanA4300.6 CFU/swab (3862.6 - 4788.2)
E. faecium ATCC BAA-2318vanA889.0 CFU/swab (777.1 - 1017.0)
E. faecium ATCC BAA-2320vanA2435.4 CFU/swab (2043.9–2901.7)
E. faecalis ATCC 51575vanB810.1 CFU/swab (571.7-1147.8)
E. faecalis ATCC 700802vanB1610.0 CFU/swab (1569.1 – 1652.0)

Table 4. Limit of Detection

Analytical Reactivity

Eighty-eight (88) well-characterized vancomycin-resistant Enterococcus strains and/or clinical isolates (43 VanA-type and 45 VanB-type) were evaluated with the IMDx VanR for Abbott m2000 assay. All strains were detected by the assay.

Challenge Study

A challenge study was conducted using a panel of 72 well-characterized strains of Enterococcus: 23 vanA strains, 25 vanB strains, 2 strains with both vanA and vanB genes, 1 strain with both vanA and vanC genes, 5 vanC strains, 3 vanD strains, 1 vanG strains, 1 vanG strain, and 10 vancomycin-susceptible strains. All Enterococcus strains harboring vanA or vanB resistance genes were detected. All Enterococcus strains harboring vanC, vanE, or vanG resistance genes and all vancomycin-sensitive Enterococcus strains were not detected by the IMDx VanR for Abbott m2000 assay.

Cross-Reactivity/Microbial Interference

The IMDx VanR for Abbott m2000 assay was evaluated for potential cross-reactivity and/or interference using a panel of 96 organisms that may be present in rectal. peri-rectal and stool samples. Included in the panel were 13 vancomycin-sensitive Enterococcus strains and 10 vancomycin-resistant (non-vanAlvanB) Enterococcus strains. Bacteria were tested at a concentration of ≥ 1 x 106 CFU/mL, and viruses at a concentration of ≥ 1 x 10 TCIDs6/mL. None of the organisms tested were found to cross-react or interfere with the IMDx VanR for Abbott m2000 assay.

Vancomycin-Resistant Staphylococcus aureus (VRSA) strains

Twelve (12) VRSA (VanA-type) isolates from the Network on Antimicrobial Resistance in Staphylococcus aureus (NARSA) were tested with the IMDx VanR for Abbott m2000 assay. The vanA gene was detected in all strains. The vanB gene was not detected in any of the strains. Since S. aureus may harbor vanA and vanB vancomycin resistance genes, the IMDx VanR for Abbott m2000 assay may produce detected results if S. aureus organisms harboring these genes are present in the clinical specimen.

{5}------------------------------------------------

Interfering Substances

The IMDx VanR for Abbott m2000 assay was challenged with twenty-three (23) substances that may be present in rectal, peri-rectal or stool specimens. The substances included: anti-fungal/ anti-itch vaginal cream, suppositories, anti-hemorrhoid creams/ointments, antacids, enemas. condoms with spermicidal lubricant, anti-diarrheal medication, laxatives, antibiotics (oral and topical), non-steroidal anti-inflammatory medications, moist towelettes, fecal components (e.g. blood, mucus, fecal lipid), and MRI contrast agent. No assay interference was observed for any of the substances.

Clinical Performance Characteristics

The performance of the IMDx VanR for Abbott m2000 assay was assessed by comparison to enriched culture combined with confirmation of the molecular basis of vancomycin resistance of isolates using an alternate PCR method. Samples were collected from five geographically diverse test sites within the United States. The patient distribution, by age, is shown in Figure 1. Samples enrolled for this study included a total of 587 peri-rectal swabs. 444 rectal swabs, and 469 stool specimens. Assay performance, by specimen type, is listed in Tables 5-10. Tables 11-12 presents Assay performance by genotype detected.

Image /page/5/Figure/5 description: The image is a bar graph showing the number of subjects versus patient age in years. The x-axis shows the patient age in years, broken down into groups such as less than or equal to 10, 11-20, 21-30, 31-40, 41-50, 51-60, 61-70, 71-80, 81-90, and greater than or equal to 91. The y-axis shows the number of subjects, ranging from 0 to 300. The number of subjects is highest in the 61-70 age range, with approximately 275 subjects.

Image /page/5/Figure/6 description: The image shows the title of a figure and a description of the patient population. The title of the figure is "Figure 1. Patient Population." The description states that the subjects ranged in age from less than 1 to 98 years old.

{6}------------------------------------------------

Enriched Culture + Alternative PCR
VanA-typeEnterococcusVanB-typeEnterococcusVanA-type andVanB-typeEnterococcusNegativeTotal
IMDx VanRfor Abbottm2000vanA38001452
vanB0001515
vanA and vanB900413
Not Detected300504507
Total5000537587

Table 5. Peri-rectal Swab Specimens: IMDx vs. Enriched Culture and Alternative PCR.

Resulting Truth Table for Peri-rectal Swab Specimens: IMDx vs. Enriched Table 6. Culture and Alternative PCR.

Enriched Culture + Alternative PCR
POSNEGTotal
IMDx VanRfor Abbottm2000POS473380
NEG3504507
Total50537587
95% CI
Sensitivity94.0% (83.8% - 97.9%)
Specificity93.9% (91.5% - 95.6%)
Positive Predictive Value58.8% (47.8% - 68.9%)
Negative Predictive Value99.4% (98.3% - 99.8%)
Prevalence8.5%

Table 7. Rectal Swab Specimens (Prospective Collection): IMDx vs. Enriched Culture and Alternative PCR.

VanA-typeEnterococcusVanB-typeEnterococcusVanA-type andVanB-typeEnterococcusNegativeTotal
IMDx VanRfor Abbottm2000vanA51001162
vanB0001616
vanA and vanB910111
Not Detected200309311
Total6210337400

Enriched Culture + Alternative PCR

{7}------------------------------------------------

IMDx Response to FDA Request for Additional Information

Table 8. Resulting Truth Table from Rectal Swab Specimens (Prospective Collection): IMDx vs. Enriched Culture and Alternative PCR.

Enriched Culture + Alternative PCR
POSNEGTotal
IMDx VanRfor Abbottm2000POS612889
NEG2309311
Total63337400
95% CI
Sensitivity96.8% (89.1% - 99.1%)
Specificity91.7% (88.3% - 94.2%)
Positive Predictive Value68.5% (58.3% - 77.2%)
Negative Predictive Value99.4% (97.7% - 99.8%)
Prevalence15.8%

Table 9. Rectal Swab Specimens (Retrospective Collection): IMDx vs. Enriched Culture and Alternative PCR.

Enriched Culture + Alternative PCR
VanA-typeEnterococcusVanB-typeEnterococcusVanA-type andVanB-typeEnterococcusNegativeTotal
vanA3000232
IMDx VanRfor Abbottm2000vanB00000
vanA and vanB1200012
Not Detected00000
Total4200244

E-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Table 10. Resulting Truth Table from Rectal Swab Specimens (Retrospective Collection): IMDx vs. Enriched Culture and Alternative PCR.

Enriched Culture + Alternative PCR
POSNEGTotal
IMDx VanRfor Abbottm2000POS42244
NEG000
Total42244

{8}------------------------------------------------

IMDx Response to FDA Request for Additional Information

95% CI
Positive Percent Agreement100.0%(91.6% - 100.0%)
Negative Percent Agreement0.0%(0.0% - 65.8%)

Table 11. Stool Specimens: IMDx vs. Enriched Culture and Alternative PCR.

VanA -typeEnterococcusVanB -typeEnterococcusVanA -type andVanB -typeEnterococcusNegativeTotal
IMDx VanRfor Abbottm2000vanA5000757
vanB2*004547
vanA and vanB.901515
Not Detected700343350
Total6801400469

Enriched Culture + Alternative PCR

  • Considered as a false negative result for the 2x2 table below

Table 12. Resulting Truth Table from Stool Specimens: IMDx vs. Enriched Culture and Alternative PCR.

Enriched Culture + Alternative PCR
IMDx VanRfor Abbottm2000POSNEGTotal
POS6057117
NEG9343352
Total69400469
95% CI
Sensitivity87.0% (77.0% - 93.0%)
Specificity85.8% (82.0% - 88.8%)
Positive Predictive Value51.3% (42.3% - 60.2%)
Negative Predictive Value97.4% (95.2% - 98.6%)
Prevalence14.7%

Conclusions

.

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is positioned to the right of a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

INTELLIGENT MEDICAL DEVICES, INC. C/O FRAN WHITE REGULATORY CONSULTANT MDC ASSOCIATES 180 CABOT STREET BEVERLY MA 01915

July 17, 2013

Re: K123753

Trade/Device Name: IMDx VanR for Abbott m2000 Assay Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial susceptibility test powder Regulatory Class: II Product Code: NIJ, OOI Dated: June 13, 2013 Received: June 14, 2013

Dear Ms. White:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{10}------------------------------------------------

Page 2-Ms. White

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 80) and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometries/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Uwe Scherf -S (or

Sally A. Hojvat. M.Sc., Ph.D. Director, Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

Indications for Use

K123753 510(k) Number:

Device Name:

IMDx VanR for Abbott m2000

Indications for Use:

The IMDx VanR for Abbott m2000 assay is an in vitro diagnostic assay that uses polymerase chain reaction (PCR) amplification for the qualitative detection of nucleic acids encoding the vancomycin resistance genes vanA and/or vanB. The assay is performed directly on human perirectal swabs, rectal swabs, or stool specimens from patients at risk for Vancomycin-Resistant Enterococcus (VRE) colonization. The IMDx VanR for Abbott m2000 assay detects the presence of vanA and vanB genes that can be associated with vancomycin-resistant enterococci. The IMDx VanR for Abbott m2000 assay can be used as an aid to identify, prevent and control vancomycin-resistant colonization in healthcare settings. The IMDx VanR for Abbott m2000 assay is not intended to diagnose VRE infection nor to guide or monitor treatment of infection. Culture methods are necessary to recover organisms for epidemiology typing and confirmation testing.

Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

AND/OR

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Image /page/11/Picture/10 description: The image shows the text "John Hobson-S" at the top, followed by "2013.07.16" and "13:11:52-04'00'". The text is overlaid on a patterned background that includes the letters "FDA" in a triangular shape. The text appears to be a timestamp or record of some kind, possibly related to the FDA.

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).