(92 days)
The Gridlock Plating System is intended for use in trauma and reconstructive procedures of the small bones in the hand/foot, ankle, and other bones appropriate for the size of the device.
The plates (implant), screws (implant), olive wires (instrument), and guide wires (instrument) are intended for single use only.
Gridlock Plating System consists of various shape and size plates for the management of orthopedic osteotomies, reconstruction, and trauma. Features include a low profile, limited contact, capability of dynamic/manual compression, and angulated-locking threaded screw holes. The system also consists of multiple locking/standard screw lengths and diameters and the necessary instruments to facilitate the placement of these implants.
The provided text describes a physical medical device (Gridlock Plating System), not an AI/software device. Therefore, the questions related to AI/software performance criteria, ground truth, expert consensus, and multi-reader studies are not applicable.
This 510(k) submission is for a Special 510(k), meaning it's for a modification to a previously cleared device (Gridlock Plating System, K121452). The change from the predicate is the addition of two specific plates for the first metatarsal phalangeal joint. The submission aims to demonstrate substantial equivalence to the predicate device.
Here's the relevant information based on the provided text, modified to address the nature of the device:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria were based on substantial equivalence to predicate devices, particularly regarding mechanical performance.
| Acceptance Criteria Category | Specific Test / Standard | Performance Goal | Reported Device Performance |
|---|---|---|---|
| Mechanical Performance | Static compression testing | Equivalent to predicate devices | The results of this testing indicate that the current Gridlock Plating System is equivalent to predicate devices. |
| Mechanical Performance | Dynamic compression testing per ASTM F382-99, Annexes 1 and 2 | Equivalent to predicate devices | The results of this testing indicate that the current Gridlock Plating System is equivalent to predicate devices. |
| Materials | CP Titanium per ASTM F67 | Same as predicate devices | Manufactured from the same material. |
| Materials | Titanium alloy per ASTM F136 | Same as predicate devices | Manufactured from the same material. |
| General Design | Features (low profile, limited contact, dynamic/manual compression, angulated-locking threaded screw holes) | Same as predicate devices | Systems are similar in principles of operation, technology, and indications for use. |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: Not explicitly stated as a "sample size" in the context of clinical data. For mechanical testing, the number of samples tested per condition is not provided.
- Data Provenance: The mechanical testing was performed in a laboratory setting. There is no mention of data from human subjects or retrospective/prospective studies.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not Applicable. For a physical medical device proving mechanical equivalence, "ground truth" related to expert assessment of data is not relevant. The relevant "experts" would be engineering/materials scientists conducting and interpreting the mechanical tests according to standards.
4. Adjudication Method for the Test Set
Not Applicable. Adjudication methods (like 2+1, 3+1) are used for resolving discrepancies in expert assessment of clinical data, which is not relevant here.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done
No. No clinical studies were performed. This submission relies on mechanical testing and substantial equivalence to a predicate device.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
Not Applicable. This is a physical medical device, not an algorithm.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance demonstration is based on established engineering standards (ASTM F382-99) and comparison against the known performance of legally marketed predicate devices. In essence, the "truth" is that the new plates perform mechanically equivalently to the previously cleared plates.
8. The Sample Size for the Training Set
Not Applicable. No training set for an algorithm is involved.
9. How the Ground Truth for the Training Set Was Established
Not Applicable. No training set for an algorithm is involved.
{0}------------------------------------------------
1/2
510(k) Summary for the Gridlock Plating System
FEB 1 5 2013
In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the Gridlock Plating System
1. GENERAL INFORMATION
Date Prepared: November 12, 2012
Trade Name: Gridlock Plating System
Common Name: bone plate & screws
Single/multiple component metallic bone fixation appliances and accessories Classification Name: Smooth or threaded metallic bone fixation fastener
Class: II
Product Code: HRS / HWC
CFR section: 21 CFR section 888.3030 / 888.3040
Device panel: OrthopedicOrthopedic
Legally Marketed Gridlock Plating System (K121452) Predicate Device: OsteoMed Foot Plating System (K091614)
Submitter: Trilliant Surgical LTD 6721 Portwest Dr. Suite 160 Houston, TX 77024 1-800-495-2919 Tele
Contact: J.D. Webb 1001 Oakwood Blvd Round Rock, TX 78681 512-388-0199 Tele
512-692-3699 Fax
e-mail: jdwebb@orthomedix.net
DEVICE DESCRIPTION 2.
Gridlock Plating System consists of various shape and size plates for the management of orthopedic osteotomies, reconstruction, and trauma. Features include a low profile, limited contact, capability of dynamic/manual compression, and angulated-locking threaded screw holes. The system also consists of multiple locking/standard screw lengths and diameters and the necessary instruments to facilitate the placement of these implants.
Change from Predicate:
This Special 510(k) is submitted in order to gain clearance for the Gridlock MPJ plates.
Materials:
CP Titanium per ASTM F67 Titanium alloy per ASTM F136
3. INTENDED USE
The Gridlock Plating System is intended for use in trauma and reconstructive procedures of the small bones in the hand/foot, ankle, and other bones appropriate for the size of the device.
{1}------------------------------------------------
The plates (implant), screws (implant), olive wires (instrument), and guide wires (instrument) are intended for single use only.
SUMMARY OF TECHNOLOGICAL CHARACTERISTICS / SUBSTANTIAL EQUIVALENCE 4.
The Gridlock Plating System is substantially equivalent to the predicate devices in terms of intended use, design, materials used, mechanical safety and performances. Two bone plates have been added to the Gridlock Plating System. These two plates are for use in treating the first metatarsal phalangeal joint and are manufactured from the same material, use the same screws and instruments, and are packaged and sterilized the using the same methods as the predicate Gridlock plates.
NON-CLINICAL TEST SUMMARY 5.
The following tests were performed:
- Static and dynamic compression testing per ASTM F382-99, Standard Test Method for . Determining the Bending Fatigue Properties of Metallic Bone Plates, Annexes 1 and 2
The results of this testing indicate that the current Gridlock Plating System is equivalent to predicate devices.
6. CLINICAL TEST SUMMARY
No clinical studies were performed
7. CONCLUSIONS NONCLINICAL AND CLINICAL
Trilliant Surgical LTD considers the current Gridlock Plating System to be equivalent to the predicate device listed above. This conclusion is based upon the devices' similarities in principles of operation, technology, materials and indications for use.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, with three stylized lines representing the wings and body.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 15, 2013
Trilliant Surgical Ltd % The Orthomedix Group, Incorporated Mr. J.D. Webb 1001 Oakwood Boulevard Round Rock, Texas 78681
Re: K123525
Trade/Device Name: Gridlock Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: December 14, 2012 Received: January 30, 2013
Dear J.D. Webb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
Page 2 - Mr. J.D. Webb
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/3/Picture/7 description: The image shows the name "Erin-I.D.Keith" in a stylized font. The letters "Erin" and "eith" are in a bold, sans-serif font. The letters "I.D.K" are stylized with a pattern of lines and curves.
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
INDICATIONS FOR USE
510(k) Number (if known): __K123525
Device Name: _Gridlock Plating System
Indications for Use:
The Gridlock Plating System is intended for use in trauma and reconstructive procedures of the small bones in the hand/foot, ankle, and other bones appropriate for the size of the device.
The plates (implant), screws (implant), olive wires (instrument), and guide wires (instrument) are intended for single use only.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/4/Picture/11 description: The image shows the name "Michael C. Owens" with a stylized logo in the background. The logo appears to be a geometric design, possibly representing the initials "MCO". The text is in a clear, bold font, making it easily readable against the more intricate background design.
Division of Orthopedic Devices
N/A