K Number
K123478

Validate with FDA (Live)

Date Cleared
2013-05-10

(178 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Designed to grasp, manipulate and coagulate selected tissue for use in general surgical procedures. They are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. Bipolar forceps must only be used with bipolar coagulation current. Coagulation is achieved using electrosurgical energy generated by the electrosurgical generator unit and activated by a footswitch. The Stingray Forceps have not been shown to be effective for tubal stenlization or tubal coagulation for sterilization procedures and should not be used for these procedures.

Device Description

These devices are bipolar forceps design for use in general surgical procedures. They are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. The forceps are designed to grasp and manipulate selected tissues. Coagulation is achieved using electrosurgical energy generated by the electrosurgical generator unit and activated by a footswitch. They are constructed of stainless steel, a nylon coating, and a non-stick tip. The devices are provided non-sterile and must be autoclaved prior to use.

AI/ML Overview

Here's an analysis of the provided text regarding acceptance criteria and the supporting study:

Important Note: The provided text only describes a 510(k) summary for a medical device (Stingray Electrosurgical Forceps). A 510(k) submission primarily demonstrates substantial equivalence to a predicate device, not necessarily detailed performance criteria or clinical efficacy studies as would be required for a novel device or a PMA. The information requested in the prompt, such as acceptance criteria table with reported performance, sample sizes for test/training sets, ground truth establishment, expert qualifications, adjudication methods, and MRMC studies, are not present in the provided document.

Therefore, the following response will indicate where the requested information is absent.


Acceptance Criteria and Study Details for Stingray Electrosurgical Forceps (K123478)

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance CriteriaReported Device Performance
Mechanical IntegrityNot SpecifiedValidated through twenty (20) sterilization cycles. Additional limit testing to forty (40) sterilization cycles showed no degradation observed.
Coagulation EfficacyNot Specified"Coagulation is achieved using electrosurgical energy generated by the electrosurgical generator unit and activated by a footswitch." (This describes the mechanism, not specific performance metrics or acceptance criteria).
BiocompatibilityNot SpecifiedNot explicitly mentioned but generally implied by use of "stainless steel, a nylon coating, and a non-stick tip."
Sterilization EfficacyNot Specified"Substantial changes to the Instructions for Use are proposed within this submission including a change of steam sterilization method and an update to the cleaning/reprocessing instructions." (This indicates a change in method, not acceptance criteria for sterilization efficacy itself).
Useful LifetimeNot SpecifiedEstimated to be two (2) years. Validated through twenty (20) sterilization cycles (exceeds estimated amount of cycles for a single device in two years).

Explanation: The 510(k) summary focuses on demonstrating substantial equivalence to a predicate device (K083162) and changes to instructions for use, particularly sterilization and cleaning. It does not provide explicit, quantitative acceptance criteria for device performance or a detailed study demonstrating that these criteria have been met in the format typically seen for novel product performance. The "reported device performance" in the table above is derived from statements about validation and testing mentioned in the document.


2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not specified. The document mentions validation through "twenty (20) sterilization cycles" and "forty (40) sterilization cycles" for an unspecified number of devices, which relates to the device's durability rather than a clinical test set.
  • Data Provenance: Not specified. The description of testing (sterilization cycles) suggests laboratory-based validation, not clinical data provenance (country of origin, retrospective/prospective).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not applicable/not specified. The provided text describes engineering/technical validation (sterilization cycles), not a study requiring expert-established ground truth.
  • Qualifications of Experts: Not applicable/not specified.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable/not specified. The validation described (sterilization cycles) does not involve adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study Done: No. The device is an electrosurgical forceps, a surgical instrument. MRMC studies are typically performed for diagnostic imaging or AI-assisted diagnostic devices to assess how AI affects human interpretation. This type of study is not relevant or described for this device.
  • Effect Size of Human Reader Improvement: Not applicable.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Standalone Study Done: No. This device is a manual surgical instrument. The concept of an "algorithm only" or "standalone" performance without human-in-the-loop is not applicable to an electrosurgical forceps.

7. Type of Ground Truth Used

  • Type of Ground Truth: Not applicable for clinical ground truth. The "ground truth" implied by the testing described relates to the physical integrity and continued functionality of the device after repeated sterilization, determined by engineering assessments, not expert consensus, pathology, or outcomes data in a clinical sense.

8. Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable/not specified. This device does not involve a "training set" in the context of machine learning or AI models.

9. How Ground Truth for the Training Set Was Established

  • How Ground Truth Was Established: Not applicable/not specified. This concept is not relevant to the described device or the 510(k) submission.

{0}------------------------------------------------

K123478

510(k) Summary

The device trade name and common/classifications name are:

Device Trade Name

Stingray Electrosurgical Forceps

Common Name

Bipolar Forceps

Classification Name

Electrosurgical cutting and coagulation device and accessories

Address, Registration #, and Contact Information

Device Name

The address and registration number of the manufacturer site for the electrosurgical forceps is:

Manufacturer

Stingray Surgical Products LLC 156 NW 16 Street Boca Raton, FL 33432

FDA Registration #: 3006059917

Contact Information

Name: Brian McBrinn Position: Quality Manager Phone: (561) 210-7582 Fax: (561) 210-5608 brian@stingraysurgical.com

Device Class and ClassificationStingray Electrosurgical Forceps have been classified as:
--------------------------------------------------------------------------------------------
Classification: Class IIClassification Panel: General and Plastic Surgical DevicesProduct Code: GEIC.F.R Section: 878.4400
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------

No performance standards have been established under Section 513/514 of the Food, Drug and Cosmetic Act for electrosurgical forceps.

{1}------------------------------------------------

PredicateDeviceInformationThe predicate device is the Stingray Electrosurgical Forceps [K083162,concurrence date 01/12/2009].Useful Life: The lifetime of the device has been best estimated based on the average age of devices returned for repair. When a device requires repair, it has exceeded its useful lifetime. Therefore the lifetime of the devices produced by Stingray Surgical Products LLC has been determined to be two (2) years. The performance characteristics of the device have been validated through twenty (20) sterilization cycles. Additional limit testing was conducted by subjecting the device to forty (40) sterilization cycles with no degradation observed. It is estimated that twenty sterilization cycles exceeds the estimated amount of cycles for a single device in use for two years.
Labeling andIntended UseDraft “Instructions for Use” can be found in Attachment B.No changes have been made to the device labeling and no modifications to the device labeling are included in this submission. Therefore no labeling has been provided within this submission.Substantial changes to the Instructions for Use are proposed within this submission including a change of steam sterilization method and an update to the cleaning/reprocessing instructions. Additional instructions are proposed for cautions, precautions, use, handling, storage, warranty, and repair.The intended use of the modified device, as described in its labeling, has not changed as a result of the modifications.Intended Use: Designed to grasp, manipulate and coagulate selected tissue for use in general surgical procedures. They are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. Bipolar forceps must only be used with bipolar coagulation current. Coagulation is achieved using electrosurgical energy generated by the electrosurgical generator unit and activated by a footswitch. The Stingray Forceps have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures and should not be used for these procedures.This is the same intended use as previously cleared for the Stingray Electrosurgical Forceps, K083162.The statement of Indications for Use can be found in section 6.
DeviceDescription andComparisonDevice Description:These devices are bipolar forceps design for use in general surgical procedures. They are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. The forceps are designed to grasp and manipulate selected tissues. Coagulation is achieved using electrosurgical energy generated by the electrosurgical generator unit and activated by a footswitch. They are constructed of stainless steel, a nylon coating, and a non-stick tip. The devices are provided non-sterile and must be autoclaved prior to use.Comparison: The Change Summary can be found in section 10 which provides the comparison to the predicate device

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is an abstract symbol that resembles a stylized caduceus, a symbol often associated with medicine and healthcare.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 10, 2013

Stingray Surgical Products LLC % Mr. Brian McBrinn Quality Manager 156 NW 16th Street Boca Raton, Florida 33432

Re: K123478

Trade/Device Name: Stingray Electrosurgical Forceps Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: April 09, 2013 Received: April 16, 2013

Dear Mr. McBrinn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 - Mr. Brian McBrinn

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.

Sincerely yours, FOR

Peter Dig Rumm -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K123478

Device Name: Stingray Electrosurgical Forceps

Indications for Use:

Designed to grasp, manipulate and coagulate selected tissue for use in general surgical procedures. They are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. Bipolar forceps must only be used with bipolar coagulation current. Coagulation is achieved using electrosurgical energy generated by the electrosurgical generator unit and activated by a footswitch. The Stingray Forceps have not been shown to be effective for tubal stenlization or tubal coagulation for sterilization procedures and should not be used for these procedures.

Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Joshua C. Nipper -S

(Division Sign-Off)

For

Division of Surgical Devices

510(k) Number K123478

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.