K Number
K123373

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-03-06

(125 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Internal fixation implants are load-sharing devices intended to stabilize and maintain alignment until normal healing occurs. Implants are not intended to replace normal body structures or bear the weight of the body in the presence of incomplete bone healing.

The NAUTILUS Spinal System, when properly used, is intended for posterior pedicle screw fixation of the non-cervical posterior spine in skeletally mature patients. It provides stabilization and immobilization of spinal segments as an adjunct to fusion.

When used as a posterior spine thoracic/lumbar system, the NAUTILUS Spinal System is indicated for one or more of the following: (1) degenerative disc disease (is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), (2) trauma (i.e. fracture or dislocation), (3) curvatures (scoliosis, kyphosis, and/or lordosis), (4) spinal tumor. (5) failed previous fusion (6) pseudoarthrosis, (7) spinal stenosis, (8) spondylolisthesis.

Device Description

The NAUTILUS Thoracolumbar Spinal System consists of an assortment of rods, screws, cross connectors, and axial and offset connectors. The bone screw, head, and taper lock are assembled together during manufacturing to create the NAUTILUS Thoracolumbar Spinal System screw assembly component. The cross, axial, and offset connectors are also assembled during manufacturing. The NAUTILUS Thoracolumbar Spinal System implant components are made from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136. Do not use any of the NAUTILUS Thoracolumbar Spinal System components with the components from any other system or manufacturer.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the NAUTILUS Spinal System, asserting its substantial equivalence to predicate devices rather than presenting a performance study with acceptance criteria in the typical sense of a diagnostic or AI device.

Therefore, many of the requested categories (such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and training set details) are not applicable to this type of submission.

The "acceptance criteria" here are based on demonstrating comparable mechanical performance to existing, legally marketed devices.

Here's a breakdown of the relevant information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (from Predicate Devices)Reported Device Performance (NAUTILUS Spinal System)
Static Torsion (equivalent performance)Static Torsion (presented to demonstrate substantial equivalence)
Static Compression Bending (equivalent performance)Static Compression Bending (presented to demonstrate substantial equivalence)
Dynamic Compression Bending (equivalent performance)Dynamic Compression Bending (presented to demonstrate substantial equivalence)
Material: Titanium Alloy (Ti-6Al-4V ELI) as per ASTM F136Material: 6AL-4V-ELI titanium manufactured according to ASTM F136
Design (substantially equivalent to predicates)Design (substantially equivalent to predicates)
Indications for Use (substantially equivalent to predicates)Indications for Use (substantially equivalent to predicates)
Sizing (substantially equivalent to predicates)Sizing (substantially equivalent to predicates)

2. Sample size used for the test set and the data provenance:

  • Not Applicable. This is a mechanical performance study based on standardized tests (ASTM F1717) comparing the device to predicate devices, not a clinical study with patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. Ground truth in this context refers to the established mechanical properties and performance standards outlined in ASTM F1717, which are not set by human experts in the same way clinical ground truth is.

4. Adjudication method for the test set:

  • Not Applicable. The evaluation is based on meeting the specifications of ASTM F1717 and demonstrating equivalence to predicate devices, not on human adjudication of test results.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This is not an AI or diagnostic device and therefore no MRMC study was performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This refers to a medical device's mechanical performance, not an algorithm's standalone performance.

7. The type of ground truth used:

  • Engineering Standards and Predicate Device Performance. The "ground truth" for this submission is based on established engineering standards (ASTM F1717) and the known performance and characteristics of the predicate devices (K111953, K080767, K061600). The device seeks to demonstrate "substantial equivalence" to these benchmarks.

8. The sample size for the training set:

  • Not Applicable. There is no "training set" as this is not an AI/machine learning device. The device's design and manufacturing processes are likely informed by prior engineering knowledge and predicate device designs.

9. How the ground truth for the training set was established:

  • Not Applicable. As there is no training set for an AI model, this question is not relevant.

The Study that Proves the Device Meets the Acceptance Criteria:

The study proving the device meets the "acceptance criteria" (i.e., demonstrates substantial equivalence) is the "Performance Data" section described in the 510(k) summary. It states:

  • "Static torsion, static compression bending, and dynamic compression bending testing per ASTM F1717 was presented to demonstrate the substantial equivalency of the NAUTILUS Spinal System."

The conclusion further reinforces this:

  • "The information presented demonstrates the substantial equivalency of the Nautilus Spinal System."

This indicates that a series of mechanical tests, following the ASTM F1717 standard, were conducted on the NAUTILUS Spinal System. The results of these tests were then compared to the known performance of the predicate devices, and the FDA determined that the NAUTILUS Spinal System's performance was "substantially equivalent" for the stated indications for use.

{0}------------------------------------------------

MAR 0 6 2013

510(k) Summary NAUTILUS Spinal System

Submitted By:

Life Spine, Inc. 2401 W. Hassell Road, Suite 1535 Hoffman Estates, IL 60169 Telephone: 847-884-6117 Fax: 847-884-6118

510(k) Contact:

Randy Lewis Life Spine 2401 W. Hassell Road, Suite 1535 Hoffman Estates, IL 60169 Telephone: 847-884-6117 Fax: 847-884-6118

Date Prepared:

Trade Name:

Common Name:

Classification:

Predicate Device:

NKB, CFR 888.3070, Class III MNH, CFR 888.3070, Class II MNI, CFR 888.3070, Class II

NAUTILUS Spinal System (K111953) CONQUEST Spinal System (K080767) ARX Spinal System (K061600)

October 25th, 2012

NAUTILUS Spinal System

Pedicle Screw Spinal System

Device Description:

The NAUTILUS Thoracolumbar Spinal System consists of an assortment of rods, screws, cross connectors, and axial and offset connectors. The bone screw, head, and taper lock are assembled together during manufacturing to create the NAUTILUS Thoracolumbar Spinal System screw assembly component. The cross, axial, and offset connectors are also assembled during manufacturing. The NAUTILUS Thoracolumbar Spinal System implant components are made from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136. Do not use any of the NAUTILUS Thoracolumbar Spinal System components with the components from any other system or manufacturer.

{1}------------------------------------------------

Intended Use of the Device:

Internal fixation implants are load-sharing devices intended to stabilize and maintain alignment until normal healing occurs. Implants are not intended to replace normal body structures or bear the weight of the body in the presence of incomplete bone healing.

The NAUTILUS Spinal System, when properly used, is intended for posterior pedicle screw fixation of the non-cervical posterior spine in skeletally mature patients. It provides stabilization and immobilization of spinal segments as an adjunct to fusion.

When used as a posterior spine thoracic/lumbar system, the NAUTILUS Spinal System is indicated for one or more of the following: (1) degenerative disc disease (is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), (2) trauma (i.e. fracture or dislocation), (3) curvatures (scoliosis, and/or lordosis), (4) spinal tumor, (5) failed previous fusion (6) pseudoarthrosis, (7) spinal stenosis, (8) spondylolisthesis.

Technological Characteristics:

The NAUTILUS Spinal System is substantially equivalent to the predicate systems in terms of design, materials, indications for use and sizing.

Material:

The NAUTILUS Spinal System is 6AL-4V-ELI titanium manufactured according to ASTM F136. The device is comprised of a variety of non-sterile titanium, single use components.

Performance Data:

Static torsion, static compression bending, and dynamic compression bending testing per ASTM F1717 was presented to demonstrate the substantial equivalency of the NAUTILUS Spinal System.

Conclusion:

The information presented demonstrates the substantial equivalency of the Nautilus Spinal System

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" arranged in a circle around the symbol. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 6, 2013

Life Spine, Incorporated % Mr. Randy Lewis Director, Regulatory Affairs/Quality Assurance 2401 West Hassell Road, Suite 1535 Hoffman Estates, Illinois 60169

Re: K123373

Trade/Device Name: NAUTILUS Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH Dated: February 14, 2013 Received: February 15, 2013

Dear Mr. Lewis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Randy Lewis

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, Misbranding by reference to premarket notification (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htmfor the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Ering Keith

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) number (if known): K123373 -

Device Name: NAUTILUS Spinal System

Internal fixation implants are load-sharing devices intended to stabilize and maintain alignment until normal healing occurs. Implants are not intended to replace normal body structures or bear the weight of the body in the presence of incomplete bone healing.

The NAUTILUS Spinal System, when properly used, is intended for posterior pedicle screw fixation of the non-cervical posterior spine in skeletally mature patients. It provides stabilization and immobilization of spinal segments as an adjunct to fusion.

When used as a posterior spine thoracic/lumbar system, the NAUTILUS Spinal System is indicated for one or more of the following: (1) degenerative disc disease (is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), (2) trauma (i.e. fracture or dislocation), (3) curvatures (scoliosis, kyphosis, and/or lordosis), (4) spinal tumor. (5) failed previous fusion (6) pseudoarthrosis, (7) spinal stenosis, (8) spondylolisthesis.

Prescription Use x (Part 21 CFR 801 Subpart D) And/Or

Over-the-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ronald P: Jean -S

(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K123373