K Number
K123364

Validate with FDA (Live)

Date Cleared
2013-07-15

(256 days)

Product Code
Regulation Number
862.1825
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DIAsource 25 OH Vitamin D Total ELISA is intended for the quantitative measurement of 25hydroxy vitamin D2 and D3 (25 OH-D2 and 25 OH-D3) in human serum. The results are to be used in conjunction with other clinical and laboratory findings to assess the Vitamin D status of a patient.

Device Description

The DIAsource 25 OH Vitamin D Total ELISA Test is an enzyme linked immunosorbent assay to detect total 250H Vitamin D (D2 and D3) present in human serum. During the first incubation at room temperature, 250H Vitamin D is dissociated from binding serum proteins to fix on binding sites of a specific monoclonal antibody. After washing, a fixed amount of 250H Vitamin D-labeled with biotin in presence of horseradish peroxidase (HRP) compete with unlabeled 250H Vitamin D2 and 250H vitamin D3 present on the binding sites of the specific monoclonal antibody. After another incubation at room temperature, the microtiterplate is washed to stop the competition reaction. A chromogenic solution (TMB) is added and then stopped with a Stop Solution after the last incubation period. The amount of substrate turnover is determined colorimetrically by measuring the absorbance which is inversely proportional to the total 250H Vitamin D concentration.

AI/ML Overview

The DIAsource 25OH Vitamin D Total ELISA Test is an in vitro diagnostic device for the quantitative measurement of 25-hydroxy vitamin D2 and D3 in human serum. The device's performance was evaluated through various non-clinical tests and a clinical comparison study to demonstrate substantial equivalence to a legally marketed predicate device (ImmunoDiagnostic Systems 25-Hydroxy Vitamin D EIA, K021163).

1. Table of Acceptance Criteria and Reported Device Performance:

Test CategoryAcceptance CriteriaReported Device Performance
Stability (Accelerated)1. OD of Calibrator 0 and Calibrator 6: +/- 50% of values obtained with component/kit kept at 4°C. 2. (OD for Calibrator 1 / OD for calibrator 0) X 100: +/- 15% of the value obtained with the component kept at 4°C. 3. Values of Positive and Negative controls: within the announced range with the component/kit kept at 4°C.Accelerated stability studies indicated that the kit and components met the predefined criteria, supporting a 24-month stability for the kit and 48 months for Incubation Buffer and Conjugate Buffer. An expiration of 18 months at 4°C is given to the kits.
Open Vial StabilityReconstituted Calibrators and Controls: Acceptable compared to a new kit after 7 days at 4°C, or after one, two, and three months at -20°C.Results indicated that reconstituted Calibrators and Controls were acceptable after 7 days at 4°C and after one, two, and three months at -20°C. Therefore, Calibrators and Controls can be used up to 7 days when stored at 4°C or up to three months when stored at -20°C.
PrecisionCV% values typically below 15-20% for inter-assay and intra-assay precision, with values decreasing at higher concentrations (general laboratory practice).Intra-Assay (singlicate, 3 different lots, 20 days): Sample A (5.6 ng/ml): CV (%) 7.8 Sample B (27.4 ng/ml): CV (%) 5.5 Sample C (43.0 ng/ml): CV (%) 2.7 Sample D (81.2 ng/ml): CV (%) 2.5 Inter-Assay (singlicate, 3 different lots, 20 days): Sample A (17.7 ng/ml): CV (%) 7.4 Sample B (26.3 ng/ml): CV (%) 4.7 Sample C (42.0 ng/ml): CV (%) 4.5 Sample D (85.4 ng/ml): CV (%) 9.4 Reproducibility (60 serum samples, 3 concentrations, duplicate, 5 days, twice a day, 3 sites, 2 technicians/site): Sample 1 (25.5 ng/ml): Total CV 10.2% Sample 2 (52.9 ng/ml): Total CV 9.8% Sample 3 (124.8 ng/ml): Total CV 5.0%
Limit of Detection (LoD)LoD should be sufficiently low to detect clinically relevant low concentrations of 25OH Vitamin D.LoB: 1.69 ng/ml LoD: 2.81 ng/ml LoQ: 4.32 ng/ml
RecoveryTypically within 80-120% for spiked samples.Added 25OH-Vit D3: 25 ng/ml: 95% 50 ng/ml: 92% Added 25OH-Vit D2: 25 ng/ml: 105% 50 ng/ml: 95%
LinearityR² value close to 1.0, and recovery within an acceptable range (e.g., 90-110%) across the measurable range.Sample 1: Slope: 1.015, Y-Intercept: -0.298, R²: 0.99 Recoveries: 91.7% - 103% Sample 2: Slope: 1.005, Y-Intercept: 0.435, R²: 0.99 Recoveries: 97.4% - 105% Linear range: 7.7 ng/ml to 122.9 ng/ml.
Time DelayAssay results remain accurate, indicating minimal impact of time delay on dispensing.Results showed minimal change in concentration for samples tested at 0, 10, and 20 minutes delay, indicating accuracy is maintained.
Analytical Specificity(Cross-Reactivity)% Cross Reaction should be low for structurally similar but distinct compounds.1,25(OH)2-Vitamin D3: 20.3% 1,25(OH)2-Vitamin D2: 1.9% Vitamin D3: 2.9% Vitamin D2: 1.3% 24,25(OH)2-Vitamin D3: >100% 25,26(OH)2-Vitamin D3: >100% 3-epi-25 hydroxy Vitamin D3: 0.07% 25 OH Vitamin D3 (reference): 100% 25 OH Vitamin D2: 83%
Analytical Specificity(Interference)Interference should be less than 10%.Hemoglobin: -0.5% (at 250 mg/dL and 500 mg/dL) Bilirubin Conjugated: -3.5% (at 50 mg/dL and 100 mg/dL) Bilirubin Unconjugated: 2.5% (at 50 mg/dL and 100 mg/dL) Triglyceride: -4.3% (at 7.5, 125, 250 mg/dL) Vitamin C: 4.5% (at 7.5, 125, 500 mg/dL) Biotin: 4.6% (at 1, 250, 500 mg/dL) Zemplar: -4.3% (at 10, 100, 1000 ng/mL) All tested substances resulted in interference less than 10%.
Method ComparisonCorrelation coefficient (R-squared) should be high (e.g., >0.90) between the new device and the predicate device. Slope should be close to 1.0, and y-intercept close to 0.Correlation coefficient (R) = 0.917 (95% CI: 87.6% - 93.6%) Slope = 0.954 Y-intercept = 3.05 Range of samples: 8.0 ng/ml to 123.0 ng/ml.
Reference RangeEstablish a representative reference range for the intended population.Based on 150 healthy individuals from different US regions, collected in winter, not taking supplements, and without relevant medical conditions: Highest Conc. (ng/mL): 88.6 Lowest Conc. (ng/ml): 4.9 Median Conc. (ng/ml): 17.2 Used Central 95% (2.5% - 97.5%) of the results.

2. Sample size used for the test set and the data provenance:

  • Precision: 4 serum samples for intra-assay and inter-assay (n varies, but typically 10-42 replicates); 60 serum samples (each at 3 concentrations) for reproducibility.
  • Limit of Detection: Not explicitly stated as a separate test set, but calculations involved measuring blanks several times and testing 5 low-value samples 10 times.
  • Recovery: Not explicitly stated as a dedicated test set, but involved adding different levels of 25OH Vitamin D to samples.
  • Linearity: 2 samples (with concentrations known to be distributed throughout the measurable range).
  • Time Delay: Not explicitly stated if a specific test set of samples was used; likely involved testing a few representative samples.
  • Analytical Specificity (Cross-Reactivity): Sera with spiked and unspiked cross-reactants. Number of samples not specified.
  • Analytical Specificity (Interference): Serum samples with different 25OH Vitamin D concentrations spiked with various interfering substances. Number of samples not specified for each interferent, but likely a small set per substance.
  • Method Comparison: 356 samples.
  • Reference Range: 150 apparently healthy individuals.

Data Provenance:

  • Method Comparison: Implies human serum samples (not specified if retrospective or prospective or country of origin, but compared against a commercially available kit).
  • Reference Range: Human serum samples obtained from a certified commercial source (Dx Biosamples, San Diego, CA.), collected from an FDA Licensed Donor Center with informed consent. Samples were from Northern US (Pennsylvania), Central US (Tennessee), and Southern US (Florida). Collected in the Winter season (January, February, March). This is prospective in nature as specific criteria for collection were followed.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

For an ELISA test system like this, expert consensus or qualified experts are generally not used to establish the "ground truth" for quantitative measurements in the same way they would be for image interpretation. Instead, the ground truth for calibrators and controls is established through reference methods and rigorous validation.

  • Calibrators Value Assignment: "The master calibrator stock is an ethanolic solution prepared in house by weighing 25OH Vitamin D3. The value assignment of the stock is made by UV absorbance at 254 nm using a molar extinction coefficient of 18000 L mol" cm". The stock solution of 25OH Vitamin D is then used to make calibrators by diluting into horse serum. Calibrator values have been determined using native serum samples that have been assayed by LC/MS-MS. The LC/MS-MS method has been validated against a reference method." The LC/MS-MS method acts as the "ground truth" reference here, but no specific number or qualifications of "experts" are mentioned for this process.
  • Controls Value Assignment: "The mean values of 30 replicates obtained from 10 independent runs were assigned as the control target values." This is a statistical assignment, not expert consensus.

4. Adjudication method for the test set:

Not applicable in the context of a quantitative ELISA test. Adjudication, like 2+1 or 3+1, is typically used for qualitative or interpretation-based assessments (e.g., in medical image analysis) where inter-reader variability needs to be resolved to establish ground truth. For this device, values are measured quantitatively and compared against established methods or statistical criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is a standalone diagnostic test (ELISA kit) for measuring a biomarker, not a device designed to assist human readers (e.g., radiologists) in interpretation. Therefore, no MRMC study or assessment of human reader improvement with/without AI assistance was performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Yes, the entire submission describes the standalone performance of the DIAsource 25OH Vitamin D Total ELISA Test. This device is an analytical instrument and reagent system that provides quantitative results directly, without human interpretive input or human-in-the-loop performance in the clinical decision-making process for its output. The system itself generates the result.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The ground truth for the analytical performance of the device (calibration, linearity, accuracy/recovery, etc.) is primarily established through:

  • Reference Methods: Specifically, LC/MS-MS for calibrator value assignment, which was itself "validated against a reference method."
  • Statistical Derivation: For control value assignment.
  • Spiked Samples: For recovery and linearity.
  • Pharmacological Standards: For cross-reactivity.
  • Comparison to a Predicate Device: For clinical correlation, where the predicate device's established performance serves as a comparative "truth."

8. The sample size for the training set:

For an ELISA assay like this, there isn't a traditional "training set" in the context of machine learning. The assay's performance characteristics (calibration curve, reagent formulations, incubation times, etc.) are developed and optimized through iterative experimental work much like an engineering process rather than a discrete "training phase" on a dataset. The stability, precision, linearity, and other studies are part of the validation and verification of the developed assay.

  • However, if we consider any data used to develop the assay's parameters:
    • Calibrators and Controls: Developed using in-house weighed standards and dilutions in horse serum, with values traced to UV absorbance and validated against LC/MS-MS results from native human serum. This development involves experimental data, but not a "training set" in the machine learning sense. Statistical data (e.g., 30 replicates from 10 independent runs for controls) are used to assign values, but these are part of the validation of the developed system, not a training set to build an algorithm.

9. How the ground truth for the training set was established:

As noted above, there isn't a distinct "training set" for an ELISA in the machine learning paradigm. The "ground truth" for the developmental stages of the assay (e.g., optimizing reagent concentrations, incubation times, measuring ranges) would be established through:

  • Chemical/Biological Principles: Adherence to known antigen-antibody reaction kinetics and colorimetric detection.
  • Gravimetric/Spectrophotometric Standards: For preparing primary calibrator stocks (e.g., weighing 25OH Vitamin D3 and measuring UV absorbance).
  • Higher Order Reference Methods: The use of LC/MS-MS as a reference method to validate the calibrator values.
  • Empirical Optimization: Iterative laboratory experiments to determine optimal conditions for sensitivity, specificity, and performance across the measuring range.

{0}------------------------------------------------

K123364

DIA Source

JUL 1 5 2013

510(k) Summary

This 510(k) summary information is being submitted in accordance with the requirement of SMDA 1990 and 21 CFR 807.92.

Submitter's Name:DIAsource ImmunoAssays
Address:Rue du Bosquet, 2 B-1348 Louvain-la Neuve, Belgium
Phone Number:+32(0)1084991
Contact Person:Napoleon Monce (Tel: 530-759-8000)
Date:May 29, 2013
2) Name of Device:25OH Vitamin D Total ELISA Test
Trade Name:25OH Vitamin D Total ELISA Test
Common Name:Vitamin D Assay
Device Classification Name:Vitamin D Test System
Product Code:MRG - Vitamin D Test System
Panel:Chemistry (75)
Regulation Number:21 CFR 862.1825 - Vitamin D Test System - Class I
    1. Legally marketed device to which the submitter claims equivalence (Predicate Device): Immuno diagnostic systems 25-Hydroxy Vitamin D EIA for the quantitative determination of 25hydroxy vitamin D and other hydroxylated metabolites in serum or plasma. K021163.

4) Description of the device:

The assay requires a total of 165 minutes incubation time. The DIAsource 25 OH Vitamin D Total ELISA Test is an enzyme linked immunosorbent assay to detect total 250H Vitamin D (D2 and D3) present in human serum. During the first incubation at room temperature, 250H Vitamin D is dissociated from binding serum proteins to fix on binding sites of a specific monoclonal antibody. After washing, a fixed amount of 250H Vitamin D-labeled with biotin in presence of horseradish peroxidase (HRP) compete with unlabeled 250H Vitamin D2 and 250H vitamin D3 present on the binding sites of the specific monoclonal antibody. After another incubation at room temperature, the microtiterplate is washed to stop the competition reaction. A chromogenic solution (TMB) is added and then stopped with a Stop Solution after the last incubation period. The amount of substrate turnover is determined colorimetrically by measuring the absorbance which is inversely proportional to the total 250H Vitamin D concentration.

{1}------------------------------------------------

5) Intended use of the device:

The DIAsoure 25 OH Vitamin D Total ELISA is intended for the quantitative measurement of 25hydroxy vitamin D2 and D3 (25 OH-D2 and 25 OH-D3) in human serum. The results are to be used in conjunction with other clinical and laboratory findings to assess the Vitamin D status of a patient.

6) Comparison with the predicate device:

The DIAsource 25OH Vitamin D Total ELISA Test Kit was compared to a commercially available 25OH Vitamin D ELISA kit manufactured by ImmunoDiagnostic Systems (IDS) (K021163). Below is a table comparing the two kits.

The DIAsource 25OH Vitamin D Total ELISAIDS 25OH Vitamin D ELISA
Intended Use - For the quantitative determination of25-hydroxyvitamin D2 and D3 (25-OH D2 and 25-OH D3) in human serum. The results are to be usedin conjunction with other clinical and laboratoryfindings to assess the Vitamin D status of a patient.Intended Use - For the quantitativedetermination of 25-hydroxyvitamin D (25-OHD) in human serum. The results are to be usedin conjunction with other clinical and laboratorydata to assist the clinician in the assessment ofVitamin D sufficiency in adult populations.
Platform - ELISA plate read on a plate readerSame
Microtiter Plate - 96 wells coated with monoclonalanti 25OH D₂ and D₃Microtiter Plate - 96 wells coated with sheeppolyclonal anti 25OH D
Wash Solution Concentrate - 200xWash Solution Concentrate - 20x
Incubation Buffer - Proprietary buffer fordissociating vitamin DSame
Conjugate Concentrate - 100xNone
HRP Concentrate - 200xHRP Conjugate - Ready to Use
Conjugate Buffer - Use to dilute Conjugate andHRP ConcentrateNone
Substrate -- Tetramethylbenzidine (TMB)Same
Stop Solution - Acid mixtureSame
Calibrators 0-5 - LyophilizedCalibrators 0-6 - Ready to Use
Controls 1-2 - LyophilizedSame
Interpretation of Results - Standard CurveSame
Expected Values: Deficient <10 ng/ml; Insufficient10-29 ng/ml; Sufficient 20-100 ng/ml; PotentialToxicity >100ng/mlSame

Table 1: Kit Comparison

Nonclinical tests: 6(b1)

Calibrators and Controls:

Traceboility/Calibrators Value Assignment - The master calibrator stock is an ethanolic solution prepared in house by weighing 250H Vitamin D3. The value assignment of the stock is made by UV absorbance at 254 nm using a molar extinction coefficient of 18000 L mol" cm". The stock solution of 250H Vitamin D is then used to make calibrators by diluting into horse serum. Calibrator values

{2}------------------------------------------------

have been determined using native serum samples that have been assayed by LC/MS-MS. The LC/MS-MS method has been validated against a reference method.

Controls Value Assignment -- The mean values of 30 replicates obtained from 10 independent runs were assigned as the control target values. The controls target ranges were assigned as mean +/- 20% if the observed CV's are less than 7%; mean +/- 25% if observed CV's are 7-10%; and mean +/- 30% if the observed CV's are 10-15%.

Stability Study Protocol and Acceptance Criteria – Accelerated stability studies were performed on the kit and components. Kits and components were placed in an incubator at 37℃ for 30 days and for 2 components 42 days and 56 days then tested. Following DIAsource experience that 7 days at 37℃ is equivalent to six months at 4℃, 30 days ,42 days and 56 days at 37℃ is equivalent to 24, 36 and 48 months, respectively. The results are accepted if they fit into 3 predefined criteria when compared to an unopened kit stored at 4ºC.

These criteria are :

  1. The OD of the Calibrator 0 and the Calibrator 6 have to be at +/- 50% of values obtained with the component (or kit) kept at 4°C

  2. The (OD for Calibrator 1 / OD for calibrator 0 ) X 100 has been at +/- 15% of the value obtained with the component kept at 4°C

  3. The values of the Positive and Negative controls have to be in the announced range with the component (or kit) kept at 4ºC

In conclusion the stability of the kit and components was found to be 24 months with the Incubation Buffer and Conjugate Buffer at 48 months.

An expiration of 18 months at 4℃ is given to the kits.

Ongoing real time stability is being conducted.

Open Vial Stability - Calibrators and Controls come in lyophilized form. Once reconstituted with DI water, they were stored at 4℃ for 7 days, and frozen at -20℃ for one months and three months with one thawing. The results after 7 days at 4℃ and after one, two, and three months at -20℃ were acceptable, when compared to a new kit.

In conclusion the Calibrators and Controls can be used up to 7 days when stored at 4℃ or up to three months when stored at -20℃

Precision:

Precision studies were evaluated according to the Clinical and Laboratory Standards Institute (CLSI) EP5-A Guideline.

Precision was calculated by running four serum samples at the given n value for 20 days on 3 different lots in singlicate. The results are summarized in the table below:

{3}------------------------------------------------

Intra-AssayInter-Assay
Samplen ± SD (ng/ml)CV (%)Samplen ± SD (ng/ml)CV (%)
A245.6 ± 0.47.8A4217.7 ± 1.37.4
B3527.4 ± 1.55.5B1026.3 ± 1.34.7
C3543.0 ± 1.22.7C1042.0 ± 1.94.5
D2481.2 ± 2.02.5D2185.4 ± 7.89.4

The reproducibility of the assay was done by testing 60 serum samples each at 3 concentrations duplicate for five days, twice a day, at three sites with two technicians per site. The mean results are summarized in the table below:

Samplenng/mlWithin-RunBetween-RunBetween-DayBetween-TechBetween-SiteTotal
16025.5SD0.2170.6110.9751.5372.2062.593
CV0.3%0.9%3.8%6.0%8.7%10.2%
26052.9SD0.6381.5711.1082.2854.3105.192
CV0.9%2.3%2.1%4.3%8.2%9.8%
360124.8SD1.001.7351.8343.3914.9066.190
CV1.4%2.5%1.5%2.7%3.9%5.0%

Limit of Detection:

The Limit of Blank (LoB), Limit of Detection (LoD), and the Limit of Quantitation (LoQ), were determined in accordance with the CLSI guideline EP17-A.

The LoB was calculated by measuring the blank several times and calculating the 95th percentile of the distribution of the test values. The LoB was calculated to be 1.69ng/ml.

The LoD was calculated by using the formula LoD = LoB + 1.65 x SDs (from the CLSI EP17-A) where SDs is the standard deviation of a low value serum. The LoD was calculated to be 2.8 I ng/ml.

The LoQ was calculated by testing 5 samples of low value 10 times in different test. The mean value was put in the X axis and the CV values on the Y axis. The LoQ was calculated to be 4.32ng/ml.

Recovery:

Recovery was assessed by adding different levels of 250H Vitamin D to samples. The results are summarized in the table below:

Recovery Test
Added 25OH-Vit D3(ng/ml)Recovery(%)

{4}------------------------------------------------

0100
2595
5092
Added 25OH-Vit D₂(ng/ml)Recovery(%)
0100
25105
5095

Linearity:

Two samples with concentrations known to be distributed throughout the measurable range were tested at equidistant dilutions to determine the linear range of the assay. A linear regression analysis was performed. The results are summarized in the following table:

Sample 1

SampleDilutionTheoreticalConcentration(ng/ml)MeasuredConcentration(ng/ml)SlopeY-InterceptR2Recovery(%)
1/196.796.7100
1/248.547.698.1
1/424.224.51.015-0.2980.99101.2
1/812.111.191.7
1/166.06.2103

Sample 2

SampleDilutionTheoreticalConcentration(ng/ml)MeasuredConcentration(ng/ml)SlopeY-InterceptR2Recovery(%)
1/1122.9122.91.0050.4350.99100
1/261.564.5105
1/430.731.5103
1/815.415.097.4
1/167.77.698.7

The linear range of the assay was found to be 7.7 ng/ml to122.9 ng/ml.

Time Delay:

Time delay test between the last Calibrator and sample dispensing results is shown in the following table.

Time Delay
0 min(ng/ml)10 min(ng/ml)20 min(ng/ml)
Sample 127.930.530.2
Sample 249.547.549.0

{5}------------------------------------------------

Assay results remain accurate even when incubation buffer is dispensed 10 and 20 minutes after the Calibrator has been added in the coated wells.

Analytical Specificity:

Cross reactivity of the 250H Vitamin D Total ELISA assay was determined by testing sera with spiked and unspiked cross reactants. The results are summarized in the following table:

Spiked Vitamin DUnspiked Vitamin% Cross
Compound and Concentration(ng/ml)D (ng/ml)Reaction
1,25(OH)2-Vitamin D3 at 200 ng/ml57.316.720.3
1,25(OH)2-Vitamin D2 at 690 ng/ml29.916.71.9
Vitamin D3 at 200 ng/ml22.516.72.9
Vitamin D2 at 200 ng/ml19.316.71.3
24,25(OH)2-Vitamin D3 at 20 ng/ml87.916.7>100
25,26(OH)2-Vitamin D3 at 4 ng/ml31.116.7>100
3-epi-25 hydroxy Vitamin D 3 at20 µg/ml31.5816.70.07
25 OH Vitamin D3 at 10 ng/ml26.716.7100
25 OH Vitamin D2 at 10 ng/ml25.016.783

The effect of potential interfering substances on samples using the DIAsoure 25 OH Vitamin D Total ELISA test was evaluated. Different levels of Hemoglobin, Bilirubin, Triglyceride, Vitamin C, Bilirubin Conjugate and Unconjugated and Zemplar in serum samples were tested on samples with different 250H Vitamin D Concentration. Our acceptance criteria was to have interference of less than 10%. The tested substances did not affect the performance of the DIAsoure 25 OH Vitamin D Total ELISA test.

Substance25OH Vitamin DConcentration (ng/ml)Concentration ofInterferent(mg/dL)Mean PercentVariation
Hemoglobin7.6250-0.5%
500
29.3250
500
42.5250
500
BilirubinConjugated6.050-3.5%
100
21.550
100
38.650
100
BilirubinUnconjugated7.6502.5%
29.3100
42.550
Trygliceride7.67.5-4.3%
29.3125
42.5250
Vitamin C6.05004.5%
21.57.5
38.6125
Biotin8.72504.6%
19.8500
36.11
Zemplar17.610-4.3%
33.5100

{6}------------------------------------------------

.

.

{7}------------------------------------------------

6(b2): Clinical Comparison:

Method Comparison:

The performance of the DIAsource 250H Vitamin D Total ELISA test was determined by conducting a correlation study tested at three different sites using a total of 356 samples. The samples were tested on both the DIAsource 250H Vitamin D Total ELISA test and a commercially available 250H Vitamin D ELISA test manufactured by ImmunoDiagnostics System (IDS). The results ranged from 8.0ng/ml to 123.0 ng/ml, the correlation coefficient between the two methods was 0.917, with the 95% confidence interval of 87.6% to 93.6%, a slope of 0.954 and the y-intercept of 3.05. The following plots summarizes the results:

Image /page/7/Figure/3 description: This image is a scatter plot. The title of the plot is "Scatter Plot". The equation of the trend line is y = 0.954x + 3.05, and the R-squared value is 0.917.

Image /page/7/Figure/4 description: The image is a difference plot showing the difference between two devices, DiaSource and Predicate Device, on the y-axis and the mean of concentration on the x-axis. The y-axis ranges from -20 to 20, while the x-axis ranges from 0 to 140. The plot shows a scatter of data points, with most points clustered between 0 and 60 on the x-axis. The title of the plot is "Difference Plot".

2.8

{8}------------------------------------------------

Reference Range:

To determine the reference range, the total 250H Vitamin D of 150 apparently healthy individuals were measured. The individual patient serum samples used were obtained from a certified commercial source (Dx Biosamples, San Diego, CA.) and were collected from an FDA Licensed Donor Center with informed consent. 50 samples were from Northern US (Pennsylvania), 50 samples were from Central US (Tennessee), and 50 samples were from Southern US (Florida).

All samples met the following criteria:

  • Age between 21-90 years old. .
  • Subjects from three different geographical locations. .
  • Samples collected in the Winter season (January, February, and March) .
  • Subjects were not taking any vitamin D supplements .
  • Subjects were of different skin tones .
  • Subjects had no family history of parathyroid, or calcium regulatory disease. .
  • Subjects had no history or Kidney, Liver, Parathyroid, Calcium related disease or bariatric . surgery.
  • Subjects were not taking any medications known to affect absorption or catabolism of Vitamin D. . The following results were obtained:
  • Ages between 20-62 years old. .
  • Population consisted of 75 light skin (50%) and 75 dark skin (50%). .
  • No subjects were taking vitamin D supplements. .
  • No subjects had a family history of parathyroid, or calcium regulatory disease. .
  • No subjects had a history of Kidney, Liver, Parathyroid, Calcium related disease or bariatric . surgery.
  • No subjects were taking any medications known to affect absorption or catabolism of Vitamin D. .
  • The following table is the summary or results: .
FloridaTennesseePennsylvaniaOverall
Highest Conc. (ng/mL)88.671.754.688.6
Lowest Conc. (ng/ml)6.14.95.94.9
Median Conc. (ng/ml)20.815.914.317.2

Only Central 95% (2.5% - 97.5%) of the results observed were used.

6(b3) Conclusion:

From the data and comparison above, it is our contention that the DIAsource 250H Vitamin D Total ELISA Test Kit is substantially equivalent to the commercially available 250H Vitamin D ELISA kit manufactured by ImmunoDiagnostic Systems (IDS) (K021163).

{9}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/9/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized symbol that resembles a human figure. The figure is composed of three curved lines that form the head, body, and legs.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 15, 2013

DIAsource Immunoassays, S.A. c/o Napoleon Monce c/o Gold Standard Diagnostics 2851 Spafford Street DAVIS CA 95618

Re: K123364

Trade/Device Name: 250H Vitamin D Total ELISA Test Regulation Number: 21 CFR 862.1825 Regulation Name: Vitamin D test system Regulatory Class: II Product Code: MRG Dated: June 12, 2013 Received: June 13, 2013

Dear Mr. Monce:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{10}------------------------------------------------

Page 2-Mr. Monce

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 You desite specific advisour PDA/CentersOffices/CDRH/CDRHOffices/ucm | 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Carol C. Benson -S for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

Indications for Use

510(k) Number (if known): K123364

Device Name: 250H Vitamin D Total ELISA Test

Indications for Use:

The DIAsource 250H Vitamin D Total ELISA Test is intended for the quantitative determination of 25-hydroxyvitamin D2 and D3 (25-OH D2 and 25-OH D3) in human serum. The results are to be used in conjunction with other clinical and laboratory findings to assess the Vitamin D status of a patient.

Prescription Use _ X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Ruth A. Chesler -S

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

510(k) K123364

Page 1 of_1_

§ 862.1825 Vitamin D test system.

(a)
Identification. A vitamin D test system is a device intended for use in clinical laboratories for the quantitative determination of 25-hydroxyvitamin D (25-OH-D) and other hydroxylated metabolites of vitamin D in serum or plasma to be used in the assessment of vitamin D sufficiency.(b)
Classification. Class II (special controls). Vitamin D test systems must comply with the following special controls:(1) Labeling in conformance with 21 CFR 809.10 and
(2) Compliance with existing standards of the National Committee on Clinical Laboratory Standards.