(134 days)
Retrograde coronary sinus cardioplegia catheters with and without Duraflo coating are intended for delivery of blood or cardioplegia solution intraoperatively to avoid cardiac damage and aid in myocardial protection.
Retrograde coronary sinus cardioplegia catheters may be used in pediatric or adult populations based on the requirements of individual patient anatomy.
This device is for short-term use only (≤ 6 hours).
Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.
Edwards Retrograde Coronary Sinus Cardioplegia Catheters (or Retrograde Cardioplegia Cannulae) are sterile, non-pyrogenic, single-use catheters made of flexible and non-flexible polymeric materials. They are intended to provide retrograde cardioplegia to the coronary sinus during cardiopulmonary bypass procedures.
Retrograde Cardioplegia Catheters have two or three lumens. One lumen delivers cardioplegia. A second lumen monitors pressure within the coronary sinus. If a third lumen is present, it either facilitates a manually inflated balloon or a stiffening wire.
Catheters include manually inflating or self-inflating smooth or textured occlusion balloons. A curved semi-rigid or malleable insertion stylet is provided with each catheter.
Edwards Retrograde Cardioplegia Catheters are also available with Duraflo™ (heparin) coating.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Device Performance for Edwards Lifesciences® Retrograde Cardioplegia Catheters
1. Table of Acceptance Criteria and Reported Device Performance
| Testing | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Sterility | Per ISO11135-1, Sterilization of health care products - Ethylene oxide - Part 1 | Met acceptance criteria |
| Ethylene oxide sterilization residuals | ISO 10993-7, Biological evaluation of medical devices - Part 7 | Met acceptance criteria |
| Biocompatibility | Per ISO 10993-1 for External communicating device, direct circulating blood path, duration ≤ 24 hours. | Met acceptance criteria |
| Conical Fittings | Fittings must be compatible with standard connections. | Met acceptance criteria |
| Wire Encapsulation | Cannula body wire reinforcement shall be fully encapsulated. | Met acceptance criteria |
| Assembly Leak | Pressure drop must meet minimum requirement. | Met acceptance criteria |
| Kink | The cannulae shall not kink at a pre-determined diameter. | Met acceptance criteria |
| Balloon Burst | Minimum pressures, diameter, and volume at burst must be maintained. | Met acceptance criteria |
| Balloon Leak | Balloon must not leak when inflated with pre-determined volumes. | Met acceptance criteria |
| Pressure Monitoring | Flow testing will confirm patency of the pressure monitoring lumen. | Met acceptance criteria |
| Corrosion | The metallic components shall show no signs of corrosion. | Met acceptance criteria |
| Tensile | Confirmation of the bond strength of the catheter assembly must meet pre-determined loads. | Met acceptance criteria |
General Study Information
The provided document describes a 510(k) submission (K123187) for Edwards Lifesciences® Retrograde Cardioplegia Catheters. The study conducted to demonstrate the device meets acceptance criteria is primarily Non-Clinical Testing (bench and laboratory testing).
Detailed Study Information:
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: The document does not specify the exact sample size for each non-clinical test. It generally states that "Bench and laboratory testing was performed."
- Data Provenance: The data is generated from non-clinical bench and laboratory testing. The document does not indicate a specific country of origin for this data, but it is presumably conducted by Edwards Lifesciences LLC or their contracted testing facilities. This is prospective testing, designed to evaluate the new device against established criteria.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
- Not Applicable. For non-clinical bench and laboratory testing of medical devices like catheters, the "ground truth" is established by adherence to recognized international standards (e.g., ISO standards) and pre-determined engineering specifications. These standards define the objective criteria for performance, and the tests are designed to measure against these objective criteria rather than relying on expert consensus for ground truth.
4. Adjudication Method for the Test Set:
- Not Applicable. As the ground truth is based on objective measurements against pre-defined standards and specifications, there is no need for an adjudication method typically used for subjective assessments (like image interpretation by multiple experts).
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
- No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic or interpretive devices where human readers are involved. This submission is for a physical medical device (catheter).
- Effect Size of Human Readers with AI vs. Without AI Assistance: Not applicable, as no AI component or human reader interaction with an AI system is mentioned.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
- Not Applicable. This device is a physical catheter, not an algorithm. Therefore, standalone algorithm performance is not relevant.
7. The Type of Ground Truth Used:
- The ground truth for the non-clinical testing is based on objective engineering specifications and recognized international standards. These standards (e.g., ISO 11135-1 for sterility, ISO 10993-7 for ETO residuals, ISO 10993-1 for biocompatibility) define the acceptable performance parameters. The "ground truth" is that the device must meet or exceed these physical and biological performance thresholds.
- For the pediatric indication, a "clinical assessment based on published studies" was used, which implies literature review/outcomes data from existing clinical evidence.
8. The Sample Size for the Training Set:
- Not Applicable. This submission describes a physical medical device, not a machine learning algorithm. Therefore, there is no "training set" in the context of AI.
9. How the Ground Truth for the Training Set Was Established:
- Not Applicable. As there is no training set, this question is not relevant.
{0}------------------------------------------------
510(k) Summary
| Submitter: | Edwards Lifesciences LLC |
|---|---|
| Contact Person: | Karen Jones, Senior Manager, Regulatory Affairs12050 Lone Peak PkwyDraper, UT 84020(801) 565-6231 |
| Date Prepared: | October 10, 2012 |
| Trade Name: | Edwards Lifesciences® Retrograde Cardioplegia Catheterswith and without Duraflo Coating |
| Classification Name: | Catheter, Cannula and Tubing, Vascular, CardiopulmonaryBypass, 21 CFR Part 870.4210, Product Code DWF, Class II |
| Predicate Device: | K880103 - Retrograde Coronary Sinus Cardioplegia CannulaeK991170 - Retrograde Cardioplegia Cannulae with DurafloTreatmentK001565 - Steerable Retrograde Cardioplegia Cannula with andwithout Duraflo Treatment |
Device Description:
Edwards Retrograde Coronary Sinus Cardioplegia Catheters (or Retrograde Cardioplegia Cannulae) are sterile, non-pyrogenic, single-use catheters made of flexible and non-flexible polymeric materials. They are intended to provide retrograde cardioplegia to the coronary sinus during cardiopulmonary bypass procedures.
Retrograde Cardioplegia Catheters have two or three lumens. One lumen delivers cardioplegia. A second lumen monitors pressure within the coronary sinus. If a third lumen is present, it either facilitates a manually inflated balloon or a stiffening wire.
{1}------------------------------------------------
Catheters include manually inflating or self-inflating smooth or textured occlusion balloons. A curved semi-rigid or malleable insertion stylet is provided with each catheter.
Edwards Retrograde Cardioplegia Catheters are also available with Duraflo™ (heparin) coating.
Indications for Use:
Retrograde coronary sinus cardioplegia catheters with and without Duraflo coating are intended for delivery of blood or cardioplegia solution intraoperatively to avoid cardiac damage and aid in myocardial protection.
Retrograde coronary sinus cardioplegia catheters may be used in pediatric or adult populations based on individual patient anatomy.
This device is for short-term use only (≤ 6 hours)
Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.
Comparative Analysis:
The Retrograde Cardioplegia Catheters have the same fundamental scientific technology and principles of operation as the predicate devices. Minor differences in features relate to customer reference rather than the clinical performance of the device. The product range of sizes now includes 9 to 14 French, 8 to 12.5 inches in length, and balloons from 9mm to 3cm. Size variations relate directly to patient anatomy. Balloons have a smooth or textured surface, and can either be self inflating or manually inflating. Additionally, the polymeric material formulations in the devices have evolved, but material types remain the same as those originally cleared in predicate submissions.
Non Clinical Testing:
Bench and laboratory testing was performed and assures that the product meets its specifications per the table below. The performance testing met the acceptance criteria.
{2}------------------------------------------------
| Testing | Criteria |
|---|---|
| Sterility | Per ISO11135-1, Sterilization of health care products - Ethylene oxide - Part 1: |
| Ethylene oxidesterilization residuals | ISO 10993-7, Biological evaluation of medical devices - Part 7: |
| Biocompatibility | Per ISO 10993-1 for External communicating device, directcirculating blood path, duration ≤ 24 hours. |
| Conical Fittings | Fittings must be compatible with standard connections. |
| Wire Encapsulation | Cannula body wire reinforcement shall be fully encapsulated. |
| Assembly Leak | Pressure drop must meet minimum requirement. |
| Kink | The cannulae shall not kink at a pre determined diameter. |
| Balloon Burst | Minimum pressures, diameter, and volume at burst must bemaintained. |
| Balloon Leak | Balloon must not leak when inflated with pre determined volumes. |
| Pressure Monitoring | Flow testing will confirm patency of the pressure monitoring lumen. |
| Corrosion | The metallic components shall show no signs of corrosion. |
| Tensile | Confirmation of the bond strength of the catheter assembly mustmeet pre determined loads. |
The Retrograde Cardioplegia Catheters conform to the following standards:
- · ISO 10993-1:2009, Biological evaluation of medical devices -- Part 1: Evaluation and testing.
- · 1SO 10993-7, Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals
- · ISO11135-1, Sterilization of health care products Ethylene oxide Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- · ISO14971, Medical devices Application of risk management to medical devices
Clinical Assessment:
A clinical assessment based on published studies was provided to support the pediatric indication.
Conclusion:
The Retrograde Cardioplegia Catheters are substantially equivalent to the cited predicate devices. The nonclinical tests and clinical assessment demonstrate that the devices are as safe and as effective as the legally marketed devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 22, 2013
Edwards Lifesciences LLC Karen Jones, Senior Principal Project Manager 12050 Lone Peak Pkwy. Draper, UT 84020
Re: K123187
Trade/Device Name: Retrograde Cardioplegia Catheter Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary bypass vascular catheter, cannula or tubing Regulatory Class: Class II Product Code: DWF Dated: January 25, 2013 · Received: January 28, 2013
Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Ms. Karen Jones
CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Matthew G. Hillebrenner
for Bram D. Zuckerman, M.D.
Director
Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Statement of Indications for Use
510(k) Number (if known): K123187
Device Name: Edwards Lifesciences® Retrograde Cardioplegia Catheter
Indications for Use:
Retrograde coronary sinus cardioplegia catheters with and without Duraflo coating are intended for delivery of blood or cardioplegia solution intraoperatively to avoid cardiac damage and aid in myocardial protection.
Retrograde coronary sinus cardioplegia catheters may be used in pediatric or adult populations based on the requirements of individual patient anatomy.
This device is for short-term use only (≤ 6 hours).
Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.
Prescription Use × (Per 21 CFR 801.109)
OR Over-The-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office Of Device Evaluation (ODE)
Matthew GM lebrenner
§ 870.4210 Cardiopulmonary bypass vascular catheter, cannula, or tubing.
(a)
Identification. A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.(b)
Classification. Class II (performance standards).