K Number
K123156

Validate with FDA (Live)

Date Cleared
2012-12-13

(65 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GORE® Flow Reversal System is intended to provide embolic protection during carotid artery angioplasty and stenting for the patients diagnosed with carotid artery stenosis and who have the appropriate anatomy described below:

  • Adequate iliac/femoral access
  • Common carotid artery diameters between 6 and 12 mm
  • External carotid artery diameters less than 6 mm
Device Description

The GORE® Flow Reversal System consists of three primary components:

  • GORE® Balloon Sheath
  • GORE® Balloon Wire
  • GORE® External Filter
    When assembled together, the GORE® Flow Reversal System reverses the flow of blood at the treatment site of the internal carotid artery (ICA), directing embolic particles away from the neurovascular circulation and removing them through an external filter.
AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the GORE® Flow Reversal System (K123156), based only on the information given:

1. Table of Acceptance Criteria and Reported Device Performance

Unfortunately, the provided document does not explicitly state specific acceptance criteria in terms of numerical thresholds or performance targets. Instead, it focuses on demonstrating substantial equivalence to a predicate device.

The reported device performance primarily revolves around qualitative assessments:

Acceptance Criteria (Implied)Reported Device Performance
BiocompatibilityTested in accordance with ISO 10993-1, results demonstrate substantial equivalence.
SterilizationTested, results demonstrate substantial equivalence.
Shelf LifeTested, results demonstrate substantial equivalence.
Functional PerformanceTested, results demonstrate substantial equivalence.
Mechanical PerformanceTested, results demonstrate substantial equivalence.
Physical PerformanceTested, results demonstrate substantial equivalence.
Design CharacteristicsSubstantially equivalent to the predicate device.
Material CharacteristicsSubstantially equivalent to the predicate device.
Intended UseSubstantially equivalent to the predicate device.

2. Sample Size Used for the Test Set and Data Provenance

The document does not provide information on a "test set" or any clinical study involving patient data (retrospective or prospective). The evaluations mentioned are described as "non-clinical evaluations."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

This information is not mentioned in the document, as no clinical test set requiring expert ground truth establishment is described.

4. Adjudication Method for the Test Set

This information is not mentioned in the document, as no clinical test set requiring adjudication is described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

No MRMC comparative effectiveness study is mentioned in the provided text. The evaluation is focused on demonstrating substantial equivalence through non-clinical testing, not on comparative effectiveness with human readers.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

This question is not applicable as the device is a physical medical device (catheter system) and not an algorithm or AI system.

7. The Type of Ground Truth Used

The concept of "ground truth" as typically applied to performance studies (e.g., expert consensus, pathology, outcomes data) is not directly applicable to the "non-clinical evaluations" described. The "truth" in this context is based on engineering specifications, regulatory standards (e.g., ISO 10993-1), and comparison to the predicate device's established characteristics through physical and functional testing.

8. The Sample Size for the Training Set

This information is not applicable as the device is a physical medical device and not an AI system that undergoes training.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as the device is a physical medical device and not an AI system.

{0}------------------------------------------------

K12 3156

W. L. Gore & Associates, Inc. Special 510(k) GORE® Flow Reversal System

510(k) SUMMARY [21 CFR 807.92(a)] 5

DEC 1 3 2012 1

    1. 510(k) Owner's Contact Information: name, address, phone and fax numbers, name of contact person, and date the summary was prepared [807.92(a){1)]
      Applicant: W. L. Gore & Associates, Inc. 32360 N. North Valley Parkway Phoenix, AZ 85085

Contact:

.

Sharon Alexander Requlatory Affairs Direct: (623) 234-5440 Fax: (623) 234-5105 Email: salexand@wlgore.com

Date Prepared:

October 3, 2012

    1. Name of the Device: including the trade or proprietary name, if applicable, the common or usual name, and the classification name, if known [807.92(a)(2)] 1
    • GORE® Flow Reversal System . Trade name:
    • Common name: .

Occlusion catheter

  • Classification name: Temporary carotid catheter for embolic capture
  • Classification: 21 CFR 870.1250, NTE, Class II C
    1. Device Predicate [807.92(a)(3)]

K083300, GORE® Flow Reversal System

  1. Description of the Device [807.92(a)(4)1

The GORE® Flow Reversal System consists of three primary components:

  • GORE® Balloon Sheath .
  • GORE® Balloon Wire .
  • GORE® External Filter .

When assembled together, the GORE® Flow Reversal System reverses the flow of blood at the treatment site of the internal carotid artery (ICA), directing embolic particles away from the neurovascular circulation and removing them through an external filter.

    1. Intended Use [807.92(a)(4)]
      The GORE® Flow Reversal System is intended to provide embolic protection during carotid artery angioplasty and stenting for the patients

GORE and Design are Trademarks of W. L. Gore & Associates, Inc.

{1}------------------------------------------------

W. L. Gore & Associates, Inc. Special 510(k) GORE® Flow Reversal System

5 510(k) SUMMARY [21 CFR 807.92(a)]

diagnosed with carotid artery stenosis and who have the appropriate anatomy described below:

  • . Adequate iliac/femoral access
  • Common carotid artery diameters between 6 and 12 mm .
  • External carotid artery diameters less than 6 mm .
    1. Summary of Similarities and Difference in Technological Characteristics, Performance and Intended Use:

The modified GORE® Flow Reversal System is substantially equivalent to the currently marketed predicate device in its intended use, technological characteristics and performance.

    1. Predicate Device Comparison [807,92(a)(6)]
      Non-clinical evaluations of the modified GORE® Flow Reversal System included biocompatibility testing in accordance with ISO 10993-1, sterilization, shelf life, and performance testing, including functional, mechanical and physical evaluations. The results of these tests demonstrate that the technological characteristics such as product performance, design, materials and the intended use are substantially equivalent to the currently marketed predicate device.

8. Conclusion

The modified GORE® Flow Reversal System is substantially equivalent to the predicate GORE® Flow Reversal System (K083300) in terms of design, material composition, intended use and performance attributes.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three abstract human figures, representing the department's focus on health and human well-being. The text is in bold, sans-serif font.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -- WO66-G609 Silver Spring, MD 20993-0002

DEC 1 3 2012

W.L. Gore & Associates, Inc. c/o Sharon Alexander, Regulatory Affairs Associate Medical Products Division 32360 N North Valley Parkway Phoenix, AZ 85085 United States

Re: K123156

Trade/Device: Gore Flow Reversal System Model Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NTE Dated: October 3, 2012 Received: November 13, 2012

Dear Ms. Alexander:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Ms. Sharon Alexander

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

W. L. Gore & Associates, Inc. Special 510(k) GORE® Flow Reversal System

INDICATIONS FOR USE STATEMENT 4

Indications for Use

510(k) Number (if known):

K123156

Device Name:

GORE® Flow Reversal System

Indications for Use:

The GORE® Flow Reversal System is intended to provide embolic protection during carotid artery angioplasty and stenting for the patients diagnosed with carotid artery stenosis and who have the appropriate anatomy described below:

  • Adequate iliac/femoral access .
  • Common carotid artery diameters between 6 and 12 mm ●
  • External carotid artery diameters less than 6 mm .

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER : PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sinn-Ori) 510 h; Number

GORE and Design are Trademarks of W. L. Gore & Associates, Inc. Confidential Information

16

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).