K Number
K123148

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-11-19

(410 days)

Product Code
Regulation Number
870.1875
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EKOSCOPE device is intended for medical diagnostic purposes only. It has an electronic stethoscope function with three modes of operation. It also has an electrocardiograph (ECG) function with three modes of operation. The Bell mode processes sounds having a frequency range of 20 Hz -500 Hz. Secondly, the Diaphragm mode processes sound within a range of 20 Hz to 700 Hz. The third mode is the Diaphragm Plus mode. The Diaphragm Plus Mode processes mechanical heart sounds between 1KHz to 20 KHz. The stethoscope function is used for the detection and amplification of sounds from the heart, lungs, arteries, veins and other internal organs during physical assessment. It has a built-in penlight accessory with adjustable brightness used to visualize pupil, nasal and oral cavity.

The ECG mode utilizes one of three modes to detect cardiac electrical activities, specifically from the limb leads, to determine the state of the heart beyond the abilities of a stethoscope. The ECG are detectable by a series of electrodes marked as #1 and #2 as referred in the user manual to user manual to register, record and display the electrical activity of the cardiac cycle in real time. This assessment will allow for the medical professional to then analyze, interpret and diagnose heart rate and rhythm based on recorded limb lead electrocardiogram information. This ECG is not a standard 12-lead ECG and is not suitable for detection of ischemia or infarction. The EKOSCOPE is suitable for adults and pediatrics but not intended for use on infants and in utero. The device is intended for use in hospitals or health care facilities by doctors and trained health care professionals.

Device Description

The Ekoscope iS100 Device is an electronic stethoscope with 6 leads EKG capability. The electronic microphone can detect sounds between 1.0 Hz to 2KHz from the heart, vessels, lungs and abdomen. The detected sounds are then processed and filtered to reduce ambient noises and transmitted to a removable headset that contains a built in speaker. The sound may also be amplified or attenuated upon demand. The visceral sounds are processed via three stethoscope modes. The first is the Bell Mode. The Bell mode processes sounds having a frequency range of 20 Hz. Secondly, the Diaphragm mode processes sound within a range of 20 Hz to 700 Hz. The third mode is the Diaphragm Plus mode. The Diaphragm Plus Mode processes mechanical heart sounds between 1KHz to 20 KHz. The acquired visceral sounds, along with EKG data, are stored in 30 seconds increments into the device's 2 Gigabyte SD card for playback, delete or upload via USB to a PC using Microsoft Window® version Operating System. The Ekoscope iS100 is powered by a rechargeable Lithium battery. The battery is rechargeable via a micro USB. It has a built-in penlight accessory with adjustable brightness used to visualize pupil, nasal and oral cavity.

The Ekoscope iS 100 also features 6 electrodes. The electrodes are positioned on the head of the device and body of the device. The strategically placed electrodes are used for the detection of 6 channel electrocardiograph (ECG) signals. The ECG signals are from the chest, hands and left leg, via three designated modes: A, B and C. The acquired data is first stored on the Ekoscope iS 100, and can be uploaded to a compatible PC for printing or sending to another location via the internet.

The sample frequency of ECG signal is 200Hz and the bit resolution is 16bit: Adopts 50/60Hz interference suppression. The sample frequency of heart sounds signal is 48kHz and bit resolution is 16bit. Different digital filters will be used in the Bell Mode, Diaphragm Mode and Diaphragm Plus Mode. The device embedded software, proprietary filter and Digital Signal Processing (DSP) technologies, attenuate ambient noises of sounds within a frequency range of 1.0 Hz to 2.0KHz. The sounds are attenuated while preserving native auscultated sounds of interest.

The display speed of the ECG wave and heart sound wave is different on the device's TFT color screen. The ECG wave has two display speed 12.5mm/s and 25mm/s. The display speed of heart sound is 153mm/s. The display range of the ECG wave is selectable and they are 'x0.5', 'x1' and 'x2'. The display range of heart sound wave is not alterable. ECG display is marked with 'ECG' while the Sound Wave display is marked with 'Sound'.

AI/ML Overview

Acceptance Criteria and Device Performance Study for PARACE MD® Ekoscope™ Model iS100

This document outlines the acceptance criteria and the study approach used to demonstrate the substantial equivalence of the PARACE MD® Ekoscope™ Model iS100, an electronic stethoscope with electrocardiograph (ECG) function.

1. Table of Acceptance Criteria and Reported Device Performance

The PARACE MD® Ekoscope™ Model iS100 did not establish specific acceptance criteria for a new clinical study. Instead, its performance was compared to existing predicate devices to demonstrate substantial equivalence. The table below summarizes the key performance features against predicate devices.

Performance FeaturesPredicate Device Stethoscope (Littmann Model 4100 / 3200)PARACE MD® Ekoscope™ iS100 (New Device)
Stethoscope Frequency ResponseBell mode (20-200 Hz), Diaphragm mode (100-500 Hz)Bell mode: 20 Hz - 550Hz, Diaphragm mode: 20Hz - 700Hz, Diaphragm plus mode: 1KHz - 20KHz (Mechanical heart valves typically 1.3KHz to 20KHz)
Stethoscope AmplificationUp to 25 dB acoustic gain, equivalent to 18 times amplificationUp to 30 dB acoustic gain, equivalent to 21.6 times amplification
Maximum Sound Level140 dB SPL Max87 dB SPL Max (Note that 100dB is the tolerable maximum human threshold for sound)
Power SourceTwo (2) AAA alkaline batteriesLithium-ion Battery
Displays Heart RateYesYes
Record and Playback SoundsYesYes
Permits Data Transfer to PCYesYes
Volume Control8 step volume control8 step volume control
On/Off SwitchYesYes
Automatic Shut-off by ElectronicsPresent but not specifiedPresent but not specified
Performance FeaturesPredicate Device ECG (Edan SE-3)PARACE MD® Ekoscope™ iS100 (New Device)
Three Channel ECGYesYes
No. of Leads6 to 12 Leads6 Leads
Power SourceRechargeable Lithium-ion, AC/DC power supplyRechargeable Lithium-ion, AC/DC power supply
Intended UseAcquire ECG signals from adult and pediatric patients through body surface ECG electrodesAcquire ECG signals from adult and pediatric patients through body surface ECG electrodes
Display of ECG Signal WaveformsLCDTFT LCD
Results Recording/StorageThermal printer or USB printerUSB printer and internal 2GB storage
Data TransmissionVia PC Ethernet portVia PC email system
Patient Data TransferUSB interfaceUSB interface
LCD ScreenFoldable LCD screenTFT LCD with touch key interface
ECG Recording Modes1 ECG Recording Mode3 ECG Recording Modes: Mode A: 6 Lead (Chest), Mode B: 6 Leads (Hands + Legs), Mode C: 1 Lead (Hands)

2. Sample Size Used for the Test Set and Data Provenance

The submission explicitly states that no clinical studies were required to support substantial equivalence. The determination was based on comparisons to predicate devices and adherence to voluntary standards. Therefore, there was no "test set" in the traditional sense of a clinical trial involving patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Since no new clinical studies were conducted, there was no separate test set requiring ground truth establishment by experts in the context of a new clinical evaluation. The assessment relied on the established performance characteristics and safety profiles of the predicate devices.

4. Adjudication Method for the Test Set

As no clinical test set was used, there was no adjudication method employed for this purpose.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. The submission indicates that clinical studies were not required for the clearance of this device.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device's functionality is described as a medical device for use by healthcare professionals (human-in-the-loop). While the device incorporates proprietary filters and Digital Signal Processing (DSP) technologies to attenuate ambient noises, this is part of the device's inherent design and not a separate "standalone algorithm only" performance evaluation as might be seen for AI-driven diagnostic software. The focus of the submission was on the overall device's safety and effectiveness in the hands of a user, compared to predicate devices.

7. The Type of Ground Truth Used

The ground truth for demonstrating the device's conformance and substantial equivalence was based on:

  • Performance specifications of legally marketed predicate devices. This implies that the 'ground truth' for the expected performance was derived from the established and cleared characteristics of the 3MTM Littmann® Electronic Stethoscope (Models 4100 and 3200) and the Edan SE-3 Electrocardiograph.
  • Compliance with international accepted standards on safety and effectiveness. These standards (e.g., IEC 60601-1, IEC 60601-1-2, ANSI/AMMI EC 38, ISO 10993-5, ISO 10993-10) serve as a de facto "ground truth" for general device safety, electromagnetic compatibility, electrical safety, and biocompatibility.

8. The Sample Size for the Training Set

Since the device did not undergo a clinical trial or machine learning development through a 'training set' in the traditional sense, this information is not applicable. The device's design and engineering would have been informed by general medical device design principles and relevant scientific literature, rather than a specific dataset for algorithm training.

9. How the Ground Truth for the Training Set Was Established

As there was no specific training set for a machine learning algorithm, there was no ground truth establishment for such a set. The device's functionality is based on established electronic signal processing and hardware capabilities, with performance benchmarked against equivalent, already-approved medical devices.

{0}------------------------------------------------

1

K123148 P 1/7

K123148 510k Summary Statement (In accordance with 21 CFR 807.92) Date Prepared: 15 July 2013

Sponsor Information:

PARACE LLC PO Box 171 Yorktown Heights NY 10598 USA

NOV 1 9 2013

Contact Person:VanDette Carter, PA-CPresident and Chief Executive Officer
Mailing Address:2310 Eastchester RoadBronx, NY, 10469
Telephone Number:914-356-4848
Facsimile Number:1-718-519-7898
E-mail:ekoscope@parace.com
Alternate Contact Person:Fernand A. De Los Reyes, RN, MAChief Operations Officer
Telephone Number:845-270-9289
Facsimile Number:1-718-519-7898
E-mail:dean.dir@parace.com
Alternate Contact Person:Alma L. Carter, RNChief Compliance Officer
Telephone Number:877-727-2231
Facsimile Number:1-718-519-7898
E-mail:paracecorp@aol.com

Device Name and Classification:

Common or Usual Name:Electronic Stethoscope with Electrocardiograph Function
Proprietary Name:PARACE MD® Ekoscope™ Model iS100
Classification Name:Electronic Stethoscope Class II (Two), 21 CFR 870.1875Electrocardiograph Class II (Two), 21 CFR

870.2340 Performance Standards: None

Predicate Devices:

3MTM Littmann ® Electronic Stethoscope, Model 4100, K051790 3MTM Littmann® Electronic Stethoscope, Model 3200, K083903 Edan SE-3 Series Electrocardiograph, K091513

Device Description:

The Ekoscope iS100 Device is an electronic stethoscope with 6 leads EKG capability. The electronic microphone can detect sounds between 1.0 Hz to 2KHz from the heart, vessels, lungs and abdomen. The detected sounds are then processed and filtered to reduce ambient

{1}------------------------------------------------

K123148 P 2/7

noises and transmitted to a removable headset that contains a built in speaker. The sound may also be amplified or attenuated upon demand. The visceral sounds are processed via three stethoscope modes. The first is the Bell Mode. The Bell mode processes sounds having a frequency range of 20 Hz. Secondly, the Diaphragm mode processes sound within a range of 20 Hz to 700 Hz. The third mode is the Diaphragm Plus mode. The Diaphragm Plus Mode processes mechanical heart sounds between 1KHz to 20 KHz. The acquired visceral sounds, along with EKG data, are stored in 30 seconds increments into the device's 2 Gigabyte SD card for playback, delete or upload via USB to a PC using Microsoft Window® version Operating System. The Ekoscope iS100 is powered by a rechargeable Lithium battery. The battery is rechargeable via a micro USB. It has a built-in penlight accessory with adjustable brightness used to visualize pupil, nasal and oral cavity.

The Ekoscope iS 100 also features 6 electrodes. The electrodes are positioned on the head of the device and body of the device. The strategically placed electrodes are used for the detection of 6 channel electrocardiograph (ECG) signals. The ECG signals are from the chest, hands and left leg, via three designated modes: A, B and C. The acquired data is first stored on the Ekoscope iS 100, and can be uploaded to a compatible PC for printing or sending to another location via the internet.

The sample frequency of ECG signal is 200Hz and the bit resolution is 16bit: Adopts 50/60Hz interference suppression. The sample frequency of heart sounds signal is 48kHz and bit resolution is 16bit. Different digital filters will be used in the Bell Mode, Diaphragm Mode and Diaphragm Plus Mode. The device embedded software, proprietary filter and Digital Signal Processing (DSP) technologies, attenuate ambient noises of sounds within a frequency range of 1.0 Hz to 2.0KHz. The sounds are attenuated while preserving native auscultated sounds of interest.

The display speed of the ECG wave and heart sound wave is different on the device's TFT color screen. The ECG wave has two display speed 12.5mm/s and 25mm/s. The display speed of heart sound is 153mm/s. The display range of the ECG wave is selectable and they are 'x0.5', 'x1' and 'x2'. The display range of heart sound wave is not alterable. ECG display is marked with 'ECG' while the Sound Wave display is marked with 'Sound'.

Indications for Use:

The EKOSCOPE device is intended for medical diagnostic purposes only. It has an electronic stethoscope function with three modes of operation. It also has an electrocardiograph (ECG) function with three modes of operation. The Bell mode processes sounds having a frequency range of 20 Hz -500 Hz. Secondly, the Diaphragm mode processes sound within a range of 20 Hz to 700 Hz. The third mode is the Diaphragm Plus mode. The Diaphragm Plus Mode processes mechanical heart sounds between 1KHz to 20 KHz. The stethoscope function is used for the detection and amplification of sounds from the heart, lungs, arteries, veins and other internal organs during physical assessment. It has a built-in penlight accessory with adjustable brightness used to visualize pupil, nasal and oral cavity.

{2}------------------------------------------------

K123148 P 3/7

3

The ECG mode utilizes one of three modes to detect cardiac electrical activities, specifically from the limb leads, to determine the state of the heart beyond the abilities of a stethoscope. The ECG are detectable by a series of electrodes marked as #1 and #2 as referred in the user manual to user manual to register, record and display the electrical activity of the cardiac cycle in real time. This assessment will allow for the medical professional to then analyze, interpret and diagnose heart rate and rhythm based on recorded limb lead electrocardiogram information. This ECG is not a standard 12-lead ECG and is not suitable for detection of ischemia or infarction. The EKOSCOPE is suitable for adults and pediatrics but not intended for use on infants and in utero. The device is intended for use in hospitals or health care facilities by doctors and trained health care professionals.

Caution: Federal Law restricts this device to sale by or on the order of a licensed health care practitioner.

Effectiveness and Safety Considerations

Combined in functionality with an electronic stethoscope, the Ekoscope™ iS100 is a safe, accurate, and a reliable device for recording ECG data. The Ekoscope™ iS100 device has generally the same technological characteristics and intended use as its predicate device 3M Littmann Model 3200 and 4100 and Edan SE-3 electrocardiograph. The Ekoscope TM iS100 as compared to its predicate devices is more efficient and clinically practical in terms of ease of use and reliability. There are no new questions of safety or effectiveness.

Further, the Ekoscope iS100 is compliant and conforms to international accepted standards on safety and effectiveness. The following are the standards which the Ekoscope iS100 conforms and complies to:

  • IEC 60601-1 Medical equipment/medical electrical equipment Part 1: General . requirements for basic safety and essential performance
  • . IEC 60601-1-2: 2007 Medical electrical equipment-Part 1-2: General requirements for safety-Collateral standard: Electromagnetic compatibility - Requirements and test
  • ANSI/AMMI EC 38:2007 Medical electrical equipment Part 2-47: Particular . requirements for the safety, including essential performance of ambulatory electrocardiographic systems
  • ISO 10993-5 Biological evaluation of medical devices -- Part 5: Tests for in vitro . cytotoxicity

{3}------------------------------------------------

  • ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for . irritation and skin sensitization

Comparative Discussion for Determining Substantial Equivalence of New Device to Predicate Device:

Information provided in this 510(k) submission shows that Ekoscope™ Model iS100 is substantially equivalent to the predicate device 3M™ Littmann® Electronic Stethoscope, Model 3200, cleared under K083903 and Littmann® Electronic Stethoscope, Model 4100, cleared under K051790 in terms of intended use, indications for use, composition, physical properties and technological characteristics. In addition, Ekoscope iS100 is substantially equivalent to the predicate device Edan SE-3, cleared under K091513 for its electrocardiograph functions. The device is substantially equivalent to predicate devices in technological characteristics and performance requirements. Comparison of performance features of the Ekoscope iS100 are presented in the two succeeding tables.

4

{4}------------------------------------------------

5

K123148
K123148 P 5/7
Comparison of Performance Features Ekoscope iS100Stethoscope versus Littmann® Electronic Stethoscope Model 4100 and Littmann® Electronic Stethoscope Model 3200
PerformanceFeaturesLittmann Model 4100*(Predicate Device K051790)Ekoscope iS100( New Device K123148)
FrequencyResponseBell mode (20-200 HZ)and Diaphragm mode(100-500 Hz)Bell mode: 20 Hz - 550HzDiaphragm mode: 20Hz - 700HzDiaphragm plus mode: 1KHz - 20KHz- Mechanical heart valves are typicallybetween 1.3KHz to 20KHz
AmplificationUp to 25 dB acoustic gain,equivalent to 18 timesamplificationUp to 30 dB acoustic gain,equivalent to 21.6 timesAmplification
Maximum soundlevel140 dB SPL Max87 dB SPL Max(Note that 100dB is the tolerablemaximum human threshold for sound)
Power sourceTwo (2) AAA alkalinebatteriesLithium-ion Battery
Displays heart rateYesYes
Record and playbacksoundsYesYes
Permits data transferof stored digitalsignals to and froma Personal ComputerYesYes
Volume Control8 step volume control8 step volume control
On/Off SwitchYesYes
Automatic shut-offby Electronics
*Approved and cleared under predicate device Littman Model 3200 under K083903

.

:

·

·

.

{5}------------------------------------------------

K123148 P 6/7

Comparison of Performance Features Ekoscope iS100
versus Edan SE-3 Electrocardiograph
Performance FeaturesEdan SE-3(Predicate Device K091513)Ekoscope iS100(New Device K123148)
Three channel ECGYesYes
No of Leads6 to 12 Leads6 Leads
Power sourceRechargeable Lithium-ionAC/DC power supplyRechargeable Lithium-ionAC/DC power supply
Intended useAcquire ECG signalsfrom adult and pediatricpatients through bodysurface ECG electrodesAcquire ECG signalsfrom adult and pediatricpatients through bodysurface ECG electrodes
Display of ECG signal waveformsLCDTFT LCD
Results recording/storageThermal printer orUSB printerUSB printer andinternal 2GB storage
Data transmissionVia PC Ethernet portVia PC email system
Patient data transferUSB interfaceUSB interface
LCD screenFoldable LCD screenTFT LCD withtouch key interface
ECG Recording Modes1 ECG Recording Mode3 ECG Recording Modes:Mode A: 6 Lead (Chest)Mode B: 6 Leads (Hands + Legs)Mode C: 1 Lead (Hands)

Determination of Substantial Equivalence

The Ekoscope™ iS100 complies with voluntary standards as detailed in this pre-market submission. The following quality assurance measures and related measures cited in the attached reports were undertaken and applied to the Ekoscope™ iS100:

  • . Risk Analysis
  • . Reliability Test
  • . Safety Test
  • Electromagnetic and Biocompatibility Test .
  • Performance Test voluntarily against AAMI EC38 Performance .
  • . Product Specification Verification
  • Software Verification and Validation .

17

{6}------------------------------------------------

Summary of Clinical Tests and Conclusion:

The Ekoscope™ iS100 which is the subject of this premarket notification submission and bench test as presented in the attachments for this submission, did not require clinical studies to support substantial equivalence.

PARACE LLC's Medical Division considers the electronic stethoscope Ekoscope™ iS100 with electrocardiograph function (ECG) as a product to be as safe, as effective, more compact and efficient and substantially equivalent in performance to its predicate devices. Above all, there are no significant differences that affect the safety or effectiveness of the Ekoscope iS100™ compared to the predicate devices as earlier mentioned.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a symbol that resembles a stylized caduceus, which is a symbol often associated with medicine and healthcare.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 19, 2013

Parace LLC c/o Mr. Vandette Carter President and Chief Executive Officer P.O. Box 171 Yorktown Heights, NY 10598 US

Re: K123148

Trade/Device Name: PARACE MD Ekoscope iS100 Regulation Number: 21 CFR 870.1875 Regulation Name: Electronic Stethoscope Regulatory Class: II (two) Product Code: DQD, DPS Dated: September 30, 2013 Received: October 8, 2013

Dear Mr. Carter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{8}------------------------------------------------

Page 2 - Mr. Vandette Carter

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Owen P. Faris -S

for Bram D. Zuckerman, M.D. I. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Indications for Use

510(k) Number (if known): K123148

Device Name: PARACE MD Ekoscope iS100

Indications For Use:

The EKOSCOPE device is intended for medical diagnostic purposes only. It has an electronic stethoscope function with three modes of operation. It also has an electrocardiograph (ECG) function with three modes of operation. The Belle mode processes sounds having a frequency range of 20Hz to 500Hz. Secondly the Diaphragm mode processes sound within a range of 20 Hz to 700 Hz. The third mode is the Diaphragm plus mode. The Diaphragm plus mode processes mechanical heart sounds between 1KHz to 20KHz. The stethoscope function is used for the detection and amplification of sounds from the heart, lungs, arteries, veins and other internal organs during physical assessment. It has a built-in penlight accessory with adjustable brightness used to visualize pupil, nasal and oral cavity.

The ECG mode utilizes one of three modes to detect cardiac electrical activities, specifically from the limb leads, to determine the state of the heart beyond the abilities of a stethoscope. The ECG are detectable by a series of electrodes marked as #1 and #2 as referred in the user manual to user manual to register, record and display the electrical activity of the cardiac cycle in real time. This assessment will allow for the medical professional to then analyze, interpret and diagnose heart rate and thythm based on recorded limb lead electrocardiogram information. This ECG is not a standard 12-lead ECG and is not suitable for detection of ischemia or infarction. The EKOSCOPE is suitable for adults and pediatrics but not intended for use on infants and in utero. The device is intended for use in hospitals or healthcare facilities by doctors and trained health care professionals.

Caution: Federal Law restricts this device to sale by or on the order of a licensed health care practitioner.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Digitally signed by
Owen P. Faris -5
Date: 2013.11.19
14:38:57 -05'00'

Page 1 of

§ 870.1875 Stethoscope.

(a)
Manual stethoscope —(1)Identification. A manual stethoscope is a mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.(b)
Electronic stethoscope —(1)Identification. An electronic stethoscope is an electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.