K Number
K123092

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-12-21

(80 days)

Product Code
Regulation Number
872.5470
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CERAMIC BRACKETS Z are indicated for orthodontic movement on natural teeth surfaces.

Device Description

The CERAMIC BRACKETS Z are designed to move teeth to improve their alignment.

The CERAMIC BRACKETS Z are bonded to natural teeth by dental professionals to connect with orthodontic wires to cause tooth movement to a more preferred position.

No accessories are marketed with the CERAMIC BRACKETS Z . The dental clinician is free to choose the bonding cement, supplemental ligatures and orthodontic wires for use with the brackets.

The CERAMIC BRACKETS Z are composed of a polycrystalline zirconia ceramic bracket, which includes an archwire slot and tiewings. The self-ligating CERAMIC BRACKETS Z have a metal clip so that no other ligation of the archwire is needed.

AI/ML Overview

Here's a summary of the acceptance criteria and study information for the CERAMIC BRACKETS Z, based on the provided 510(k) summary:

This device is an orthodontic bracket. The acceptance criteria and testing focus on the physical performance of the brackets, as is typical for this type of medical device. There is no information provided about studies involving human subjects, reader performance, or clinical outcomes that would typically be seen in studies for diagnostic devices or AI-powered devices.


Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Test PerformedReported Device Performance
Mechanical StrengthShear Bond TestsPerformed and validated
Mechanical StrengthTorque (Fracture) TestsPerformed and validated

Note: The 510(k) summary states that "Bench testing was performed to ensure that the CERAMIC BRACKETS Z performance was achieved and validated, including shear bond tests and torque (fracture) tests." However, specific numerical acceptance criteria (e.g., minimum bond strength in MPa, minimum torque resistance in Ncm) and the corresponding measured device performance values are not provided in this summary. It only indicates that the tests were performed and the performance was achieved and validated.


Study Details (Based on available information)

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not specified.
    • Data Provenance: The tests are described as "Bench testing," implying laboratory-based testing on the physical devices rather than patient data. The manufacturer is based in Japan (TOMY, Inc., Tokyo, Japan).
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This device does not involve expert interpretation or ground truth derived from human assessments in the context of diagnostic or AI performance. The "ground truth" for mechanical testing would be the physical properties measured by testing equipment.
  3. Adjudication method for the test set:

    • Not applicable. As noted above, this does not involve human interpretation or adjudication.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a physical medical device (orthodontic bracket), not a diagnostic device or an AI-powered system designed to assist human readers. Therefore, an MRMC study is not relevant or performed for this product.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No. This is a physical medical device, not an algorithm or AI system.
  6. The type of ground truth used:

    • The "ground truth" in this context refers to the measured mechanical properties (shear bond strength, fracture torque) obtained through standard laboratory testing methods for orthodontic brackets. These are objective physical measurements.
  7. The sample size for the training set:

    • Not applicable. This device does not involve machine learning or an AI algorithm, so there is no "training set."
  8. How the ground truth for the training set was established:

    • Not applicable as there is no training set.

{0}------------------------------------------------

510(k) Summary

per 21CFR807.92

CONTACT:

Mr. Jinichi Watanabe Manager, Legal Sec. TOMY, Inc. Tenko Bldg. 3-16-7 Midoricho, Fuchu City, Tokyo 183-0006 Japan

DATE PREPARED: November 2, 2012

TRADE OR PROPRIETARY NAME: CERAMIC BRACKETS Z

CLASSIFICATION NAME: Bracket, Ceramic, Orthodontic, 872.5470

PREDICATE DEVICES:

Clarity Advance Ceramic Brackets (K102803) In-Ovation C ceramic brackets (K060837)

DEVICE DESCRIPTION: The CERAMIC BRACKETS Z are designed to move teeth to improve their alignment.

The CERAMIC BRACKETS Z are bonded to natural teeth by dental professionals to connect with orthodontic wires to cause tooth movement to a more preferred position.

No accessories are marketed with the CERAMIC BRACKETS Z . The dental clinician is free to choose the bonding cement, supplemental ligatures and orthodontic wires for use with the brackets.

INTENDED USES: The CERAMIC BRACKETS Z are indicated for orthodontic movement on natural teeth surfaces.

TECHNOLOGICAL CHARACTERISTICS: The CERAMIC BRACKETS Z are composed of a polycrystalline zirconia ceramic bracket, which includes an archwire slot and tiewings. The self-ligating CERAMIC BRACKETS Z have a metal clip so that no other ligation of the archwire is needed.

Bench testing was performed to ensure that the CERAMIC BRACKETS Z performance was achieved and validated, including shear bond tests and torque (fracture) tests.

DEC 2 1 2012

{1}------------------------------------------------

The CERAMIC BRACKETS Z were not evaluated for biocompatibility because zirconia has a history of safe use in medical devices. All of the components have been used in legally marketed predicate CERAMIC BRACKETS Z.

The CERAMIC BRACKETS Z device is substantially equivalent in safety and effectiveness to the predicate device.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing a globe.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

December 21, 2012

TOMY, Inc. C/O Carolyn M. Primus, PhD Consultant Primus Consulting 7046 Owl's Nest Terrace Bradenton, FL 34203

Re: K123092

Trade/Device Name: Ceramic Brackets Z Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: II Product Code: NJM Dated: September 28, 2012 Received: October 25, 2012

Dear Dr. Primus:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 – Dr. Primus

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Kwame Ó. Ulmer

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices

Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K12 3 64 2

Device Name: CERAMIC BRACKETS Z

Indications For Use: CERAMIC BRACKETS Z are indicated for orthodontic movement on natural teeth surfaces.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1

Page 16 of 54
-----------------
Susan Runner DDS, MA
2012.12.19
Premarket To:06:27:005'00'
Tomy, Inc.
CERAMIC BRACKETS Z
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
510(k) Number:K123092

§ 872.5470 Orthodontic plastic bracket.

(a)
Identification. An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.(b)
Classification. Class II.