K Number
K123026

Validate with FDA (Live)

Date Cleared
2012-11-16

(49 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Powder free, examination gloves made of latex and colored blue. The gloves are provided in sizes small, medium, large and extra large.

AI/ML Overview

The provided document describes the acceptance criteria and study results for the "Textured, Blue, Latex, Powder Free Examination Gloves with Protein Labeling Claim (50µg/dm2 Or Less of Water Soluble Protein)" in the context of an Abbreviated 510(k) Premarket Notification.

Here's the breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

CharacteristicAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM D3578-05Meets
Physical PropertiesASTM D3578-05Meets
Freedom from Pinholes21 CFR 800.20; ASTM D5151-06Meets
Powder ResidualASTM D6124-06Meets (values below 2mg of residual powder)
Protein LevelASTM D5712-10Meets (values below 50 mcg/g of protein)
Biocompatibility - Primary Skin IrritationISO 10993-10:2010 (rabbits)Gloves are non-irritating
Biocompatibility - Dermal SensitizationISO 10993-10:2010 (guinea pig)Gloves do not display any potential for sensitization

2. Sample size used for the test set and the data provenance:

The document does not specify the exact sample sizes for each test listed (e.g., number of gloves tested for dimensions, pinholes, etc.). It only states that the device "Meets" the respective ASTM and ISO standards. These standards would inherently define the required sample sizes for testing, but the document does not explicitly report them.

Data provenance is not explicitly mentioned (e.g., retrospective or prospective, country of origin). However, given these are non-clinical (laboratory/material) tests conducted against established international standards (ASTM, ISO), they are typically conducted in a controlled testing environment, likely in the manufacturer's facility or a certified testing laboratory.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable to this type of device and study. The "ground truth" for examination gloves is established by whether they meet objective, quantifiable physical and chemical standards (ASTM, ISO, and 21 CFR). There are no human "experts" or clinical interpretation involved in establishing the performance criteria for these tests. The tests are designed to objectively measure properties like dimensions, strength, and chemical content.

4. Adjudication method for the test set:

This information is not applicable. Since the tests are objective measurements against defined standards, there is no need for an adjudication method. The results either meet the standard or they do not.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. This device is a medical glove, not an AI-assisted diagnostic or imaging tool. Therefore, MRMC studies and AI assistance are irrelevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This information is not applicable. This device is a medical glove, not an algorithm.

7. The type of ground truth used:

The ground truth used is based on established objective laboratory standards (ASTM and ISO standards) and regulatory requirements (21 CFR 800.20). These standards define specific thresholds and methodologies for evaluating the physical, chemical, and biocompatibility properties of examination gloves.

8. The sample size for the training set:

This information is not applicable as this is not a machine learning or AI-based device that would require a training set.

9. How the ground truth for the training set was established:

This information is not applicable as there is no training set for this type of device.

{0}------------------------------------------------

K123026

Section 5, 510(k) Summary

Abbreviated 510(k) Premarket Notification Latex Powder Free Examination Gloves

ട്. 510(k) SUMMARY

DATE:

OWNER:

OFFICIAL CORRESPONDENT:

February 17, 2012

Northstar Healthcare Holdings 70 Sir John Rogerson's Quay Dublin 2, Ireland

Michael Riordan Operations Manager Telephone: 00353-21-4548255 Fax: 00353-21-4548294 Email: michael.riordan@mckesson.ie

DEVICE NAME:

Trade Name: Textured, Blue, Latex, Powder Free Examination Gloves with Protein Labeling Claim (50ug/dm2 Or Less of Water Soluble Protein)

Common Name: Patient Examination Gloves

Classification: Patient Examination Gloves

Class I

Class:

LYY Product Code: Regulation: 880.6250

PREDICATE DEVICE(S):

Predicate510(k)Device NameIndicationClearanceDateCompany
K062917Latex Powder FreeExamination GlovesThe latex examination glove is adisposable device intended for medicalpurposes that is worn on the examiner'shand to prevent contamination betweenpatient and examiner.02 Nov 2006Wear Safe SDN,BHD

DEVICE DESCRIPTION: Textured, Blue, Latex, Powder Free Examination Gloves with Protein Labeling Claim (50µg/dm2 Or Less of Water Soluble Protein)

NOV 1 6 2012

Page 1 of 3

{1}------------------------------------------------

Abbreviated 510(k) Premarket Notification
Latex Powder Free Examination Gloves

K123026

Section 5, 510(k) Summary
Page 2 of 3

STATEMENT OF INTENDED USE:

The Latex Powder Free Examination Glove is a Disposable device intended for medical and dental purposes to be worn on the examiner's hand or finger to prevent contamination between patient and examiner.

TECHNOLOGICAL CHARACTERISTICS:

The Latex Powder Free Examination Glove is substantially equivalent to the predicate device with regard to physical characteristics, design, product features, and intended use. Both gloves are made with latex using similar manufacturing processes.

FeatureLatex Powder Free Examination GlovesK062917PredicateLatex Powder Free Examination Gloves(Proposed)
Intended UseIntended for medical purposes that is worn on theexaminer's hand to prevent contamination betweenpatient and examiner.Same
Indications forUse StatementThe latex examination glove is a disposable deviceintended for medical purposes that is worn on theexaminer's hand to prevent contamination betweenpatient and examiner.The latex examination glove is a disposable deviceintended for medical purposes that is worn on theexaminer's hand or finger to prevent contaminationbetween patient and examiner.
DescriptionPowder free, examination gloves made of latex. Thegloves are provided in sizes small, medium, large andextra large.Powder free, examination gloves made of latex and coloredblue. The gloves are provided in sizes small, medium,large and extra large.
PresentationNon-Sterile gloves are provided in dispenser boxes.Same
MaterialLatexSame
Single UseYesSame
DimensionsMeets ASTM D3578-05Length Small, Medium, Large, XLWidth SmallMediumLargeXLThickness FingerPalm285 mm, min.70-90mm85-105 mm101-121 mm105-125 mm0.30 mm min.0.20 mm min.
PhysicalPropertiesMeets ASTM D3578-05Before aging/after agingElongation 650% 500%Tensile Strength 18MPa 14MPa
Freedom fromPinholesMeets ASTM D5151-06Same
Residual PowderMeets ASTM D6124-06Same
Protein LevelMeets ASTM D5712-05Meets ASTM D5712-10
BiocompatibilityPasses Primary Skin Irritation in RabbitsTests Passes Guinea Pig MaximizationSame

{2}------------------------------------------------

K123026

Section 5, 510(k) Summary

Page 3 of 3

Abbreviated 510(k) Premarket Notification Latex Powder Free Examination Gloves

ASSESSMENT OF NONCLINICAL DATA:

CharacteristicStandardDevicePerformance
DimensionASTM Standard D3578-05Meets
Physical PropertiesASTM Standard D3578-05Meets
Freedom fromPinholes21 CFR 800.20; ASTMD5151-06Meets
Powder ResidualASTM Standard D6124-06MeetsResults generatedvalues below 2mgof residual powder
Protein LevelASTM Standard 5712-10MeetsResults generatedvalues below 50mcg/g of protein
BiocompatibilityPrimary Skin Irritation inrabbits(ISO 10993-10:2010)Gloves are non-irritating
Dermal Sensitization in theguinea pig (ISO 10993-10:2010)Gloves do notdisplay anypotential forsensitization

CONCLUSIONS:

The Latex Powder Free Examination Gloves meet the requirements of established standards ASTM D3578-05, ASTM D5712-10, ASTM D5151-06, ASTM D6124-06 and ISO 10993-10:2010.

Based on the comparison of intended use, design, materials and performance, the Latex Powder Free Examination Gloves are substantially equivalent to the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

November 16, 2012

Northstar Healthcare Holdings C/O Mr. Ned Devine Underwriters Laboratories, Incorporated 333 Pfingsten Road Northbrook, Illinois 60062

Re: K123026

Trade/Device Name: Textured, Blue, Latex, Powder Free Examination Gloves with Protein Labeling Claim (50μg/dm2 Or Less of Water Soluble Protein)

Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYY Dated: October 25, 2012 Received: October 31, 2012

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Devine

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to.

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

  1. Hony 12 1 12-12-

DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, cn=Anthony D. Watson. 0.9.2342.19200300.100.1.1=1300092402

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Respiratory, Infection Control and

Dental Devices

Office of Device Evaluation Center for Devices and

Radiological Health

Enclosure

0.9.2342.19200300.100.1.1=1300092402

DN: c=US, o=U.S. Government, c
ou=FDA, ou=People, cn=Anthor
Watson,
0.9.2342.19200300.100.1.1=1300

{5}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known): 143024

Device Name:

Textured, Blue, Latex, Powder Free Examination Gloves with Protein Labeling Claim (50ug/dm2) Or Less of Water Soluble Protein)

Indications for Use: The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

AND/OR

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Digitally signed by Elizabeth F. Claverie Date: 2012.11.16 1/2:42:41 -05'00'

(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital Division Control, Dental Devices

510(k) Number: K123026

Page 1 of 1

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.