K Number
K122999

Validate with FDA (Live)

Date Cleared
2012-11-30

(64 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Blue Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

in addition, these gloves were tested for use with Chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs:

Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection time inMinutes, 0.01 ug/cm2/minute
1Cisplatin, 1.0 mg/ml (1,000ppm)>240
2Cyclophosphamide (Cytoxan),20.0mg/ml (20,000ppm)>240
3Dacarbazine (DTIC), 10.0 mg/ml (10,000ppm)>240
4Doxorubicin Hydrochloride, 2.0mg/ml (2,000ppm)>240
5Etoposide (Toposar), 20.0 mg/ml (20,000)ppm>240
6Fluorouracil, 50.0 mg/ml (50,000ppm)>240
7Paclitaxel (Taxol), 6.0 mg/ml (6,000ppm)>240
8Carmustine (BCNU)3.3mg/ml(3,300ppm)0.9
9Thiotepa, 10.0mg/ml (10,000ppm)0.05

The maximum testing time is 240 minutes. Please note that the following drugs have extremely low permeation time of less than 30 minutes

Carmustine (BCNU)3.3mg/ml(3,300ppm) Thiotepa, 10.0mg/ml ( 10,000ppm)

Device Description

Not Found

AI/ML Overview

This document details the acceptance criteria and performance for "Blue Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs" in the context of their ability to prevent permeation by chemotherapy drugs.

1. Table of Acceptance Criteria and Reported Device Performance

Chemotherapy Drug and ConcentrationAcceptance Criteria (Minimum Breakthrough Detection time in Minutes, >0.01 ug/cm²/minute)Reported Device Performance (Breakthrough Detection time in Minutes, 0.01 ug/cm²/minute)
Cisplatin, 1.0 mg/ml (1,000ppm)>240>240
Cyclophosphamide (Cytoxan), 20.0mg/ml (20,000ppm)>240>240
Dacarbazine (DTIC), 10.0 mg/ml (10,000ppm)>240>240
Doxorubicin Hydrochloride, 2.0mg/ml (2,000ppm)>240>240
Etoposide (Toposar), 20.0 mg/ml (20,000ppm)>240>240
Fluorouracil, 50.0 mg/ml (50,000ppm)>240>240
Paclitaxel (Taxol), 6.0 mg/ml (6,000ppm)>240>240
Carmustine (BCNU) 3.3mg/ml (3,300ppm)Not explicitly stated as acceptance criteria (note indicates "extremely low permeation time")0.9
Thiotepa, 10.0mg/ml (10,000ppm)Not explicitly stated as acceptance criteria (note indicates "extremely low permeation time")0.05

Note: The document specifies that the maximum testing time is 240 minutes. For Carmustine and Thiotepa, while specific acceptance criteria in the ">X" format are not given, the note highlights their extremely low permeation times, which can be interpreted as a disclosure rather than a failure of an acceptance criterion that requires >240 minutes.

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not explicitly stated in the provided document. The study refers to testing in accordance with ASTM D6978-05. The sample size would be dictated by the requirements of this standard.
  • Data Provenance: Not explicitly stated. The testing was performed in accordance with ASTM D6978-05, which is an international standard.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This document describes a physical performance test for medical gloves against chemotherapy drugs, not a diagnostic or clinical AI study. Therefore, the concept of "experts establishing ground truth" in the diagnostic sense does not apply here. The "ground truth" is the objective measurement of breakthrough time, determined by standardized laboratory procedures.

4. Adjudication method for the test set

Not applicable. This is a laboratory performance test based on a standardized methodology (ASTM D6978-05), not a situation requiring expert adjudication of subjective assessments.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This document pertains to the physical performance of a medical device (gloves), not an AI-assisted diagnostic or imaging system involving human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is a physical performance test of a medical device (gloves), not an algorithm.

7. The type of ground truth used

The ground truth used is the objective measurement of chemotherapy drug permeation breakthrough time, determined by a standardized laboratory method, specifically ASTM D6978-05. This is a quantitative, empirical measurement.

8. The sample size for the training set

Not applicable. This is a physical performance test for a medical device (gloves), not an AI model that requires a training set.

9. How the ground truth for the training set was established

Not applicable. As this is not an AI model, there is no training set or ground truth for a training set in this context.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest wings or feathers.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

November 30, 2012

Mr. Eddy Rosyadie Quality Assurance Manager Top Glove Sdn. Bhd Lot 4968, Jalan Teratai, Batu 6, Off Jalan Meru Klang, Selangor Darul Ehsan Malaysia 41050

Re: K122999

Trade/Device Name: Blue Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA, LZC Dated: October 19, 2012 Received: October 22, 2012

Dear Mr. Rosyadie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

{1}------------------------------------------------

Page 2 - Mr. Rosyadie

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.goy/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony De mita

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

INDICATIONS FOR USE

510 (k) Number (if known) : K12299999

Blue Powder Free Nitrile Examination Gloves Device Name : Tested For Use With Chemotherapy Drugs

Indication For Use :

Blue Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

in addition, these gloves were tested for use with Chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs:

Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection time inMinutes, 0.01 ug/cm2/minute
1Cisplatin, 1.0 mg/ml (1,000ppm)>240
2Cyclophosphamide (Cytoxan),20.0mg/ml (20,000ppm)>240
3Dacarbazine (DTIC), 10.0 mg/ml (10,000ppm)>240
4Doxorubicin Hydrochloride, 2.0mg/ml (2,000ppm)>240
5Etoposide (Toposar), 20.0 mg/ml (20,000)ppm>240
6Fluorouracil, 50.0 mg/ml (50,000ppm)>240
7Paclitaxel (Taxol), 6.0 mg/ml (6,000ppm)>240
8Carmustine (BCNU)3.3mg/ml(3,300ppm)0.9
9Thiotepa, 10.0mg/ml (10,000ppm)0.05

The maximum testing time is 240 minutes. Please note that the following drugs have extremely low permeation time of less than 30 minutes

Carmustine (BCNU)3.3mg/ml(3,300ppm) Thiotepa, 10.0mg/ml ( 10,000ppm)

Prescription Use (part 21 CFR 801 Subpart D) AND/OR

Over-The Counter Use X ( 21 CFR 801 Subpart C )

( PLEASE DO NOT WRITE THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED )

Digitally signed by Maria Cruz-fisher Maria Cruz-fisher Mercede Colored Colored Colorenter Cou-HDS, Govenment, ou=HDS, oreFDA,
Maria Cruz-fisher Cruzer, over consideracions constitution of FDA,
Maria Cruz-fish C 0.9.2342.19200300.100.1.1=2000576169 Date: 20

(Division Sign-Off) (Division Sign~on)
Division of Anesthesiology, General Hospital Division of Ancol, Dental Devices

510(k) Number:

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.