K Number
K122971

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-10-25

(29 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Clickfine pen needle is intended for the hypodermic injection of fluids into the body when attached to an injection pen.

Device Description

The Ypsomed Clickfine Pen Needles are sterile, non-pyrogenic, single use needles designed to be used with commercially available pen-injectors for the administration of prescribed fluids. Each needle is individually packaged in a sealed protective container with a peel tab. The pen needles are used by peeling back the peel tab and snapping or screwing the hub onto the threaded end of the pen injector. The back end of the cannula punctures the rubber injection port of the drug reservoir in the peninjector. The outer protective cap is then removed. The inner protective cap will remain over the needle until the drug is ready to be injected. When the injection is needed, the inner protective cap is removed and the needle is inserted into the chosen site. The pen-injector automatically delivers the fluid through the needle. The protective cap is replaced and the needle is then removed, safely discarded and replaced with a new needle.

AI/ML Overview

The provided document describes the "Clickfine Pen Needle" and its substantial equivalence to a predicate device (K102108). The information focuses on regulatory compliance based on established international and internal standards rather than a typical clinical study with acceptance criteria often seen for AI/ML devices or diagnostic tools.

However, based on the provided text, I can extract the relevant information regarding performance and safety criteria, which serve as acceptance criteria for this device, and the "study" (referencing the assessments performed).

Here's the breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (from Standards)Reported Device Performance
Compliance with ISO 11608-2:2012 (Needle based injection systems for medical use - Requirements and test methods - Part 2: Needles)The device has met the requirements of the relevant sections of this standard.
Compliance with ISO 9626:1991/Amd.1:2001 (Stainless steel needle tubing for the manufacture of medical devices)The device has met the requirements of the relevant sections of this standard.
Compliance with ISO 7864:1993 (Sterile hypodermic needles for single use)The device has met the requirements of the relevant sections of this standard.
Overall safety and effectiveness for intended use (hypodermic injection of fluids)The verifications have shown evidence that the Clickfine Pen Needles meet the acceptance criteria of these standards. Based on the results, it can be concluded that the device performance and safety are acceptable for the product.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify a "test set" sample size in the context of a clinical trial or algorithm evaluation. The performance and safety data are derived from "relevant assessments specified in the following international and internal standards and protocols." These standards (ISO 11608-2, ISO 9626, ISO 7864) typically involve laboratory testing of device components and finished products according to defined sampling plans, but the specific sample sizes used in Ypsomed's internal testing are not disclosed in this summary.

Data provenance: Given that Ypsomed AG is located in Switzerland, it can be inferred that the internal testing and data generation likely originated from Switzerland. The studies are prospective in the sense that they are validations and verifications performed on manufactured devices to ensure they meet specified standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This type of information is not relevant or provided in the context of this device and its regulatory submission. The "ground truth" for a pen needle is its compliance with engineering and material standards, which is assessed through defined laboratory tests, not expert consensus on interpretations.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in clinical imaging studies or diagnostic evaluations, where human interpretation is involved. For a physical medical device like a pen needle, performance is assessed through objective, standardized physical and material tests.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted diagnostic or interpretation device that would involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an algorithm-based device. Its performance is inherent in its physical and material properties.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the Clickfine Pen Needle's performance and safety is its adherence to the technical specifications defined in the international standards:

  • ISO 11608-2:2012: Focuses on the performance and safety of the needle unit in injection systems.
  • ISO 9626:1991/Amd.1:2001: Specifies requirements for stainless steel tubing used in medical devices, including dimensional accuracy and material properties.
  • ISO 7864:1993: Covers requirements for sterile hypodermic needles for single use, including sterility, sharpness, and cannula integrity.

Compliance with these standards, verified through objective testing methodologies, serves as the "ground truth."

8. The sample size for the training set

Not applicable. This is not a machine learning or AI device that requires a training set.

9. How the ground truth for the training set was established

Not applicable. As there is no training set for an AI/ML model, no ground truth needed to be established for it.

{0}------------------------------------------------

7. Summary of Safety and Effectiveness

OCT 2 5 2012

"510(K) SUMMARY"

Submitted By/ Contact Person:Dr. Benjamin ReinmannVice PresidentQuality Management & Regulatory AffairsYpsomed AGBrunnmattstrasse 6CH - 3401 BurgdorfSwitzerlandTel. 0041-344244111Fax 0041-344244122E-mail: benjamin.reinmann@ypsomed.com
Date Prepared:September 24, 2012
7.1Trade/Proprietary Name:Clickfine Pen Needle
7.2Common/Usual Name:Injection Pen Needle
7.3Classification Name:Hypodermic single lumen needle
7.4Classification: FDA has classified Hypodermic single lumen needles in Cla
  • n Class II. Final Order was published in the Federal Register on October 21, 1980 after review by the General Hospital and Personal Use Devices Classification Panel.
    CFR Reference: 880.5570 Procode: FMI

  • 7.5 Purpose of Submission: To expand the product line to include smaller needles including a 4 mm, 32 gauge needle.

  • 7.6 Substantial Equivalence: The Ypsomed Clickfine Pen Needles are substantially equivalent to the Clickfine Pen Needles (K102108). The equivalence is supported by the attached documentation.

  • 7.7 Device Description

The Ypsomed Clickfine Pen Needles are sterile, non-pyrogenic, single use needles designed to be used with commercially available pen-injectors for the administration of prescribed fluids. Each needle is individually packaged in a sealed protective container with a peel tab. The pen needles are used by peeling back the peel tab and snapping or screwing the hub onto the threaded end of the pen injector. The back end of the cannula punctures the rubber injection port of the drug reservoir in the peninjector. The outer protective cap is then removed. The inner protective cap will remain over the needle until the drug is ready to be injected. When the injection is

{1}------------------------------------------------

needed, the inner protective cap is removed and the needle is inserted into the chosen site. The pen-injector automatically delivers the fluid through the needle. The protective cap is replaced and the needle is then removed, safely discarded and replaced with a new needle.

  • 7.8 Intended Use
    The intended use of the modified device remains the same as the predicate device Clickfine Pen Needles (K102108):

The Ypsomed Clickfine Pen Needles are intended for the hypodermic injection of fluids into the body when attached to an injection pen.

Technological Characteristics 7.9 The technological characteristics have not changed.

  • 7.10 Performance and Safety Data
    Y psomed has performed the relevant assessments specified in the following international and internal standards and protocols and confirmed compliance of the modified devices and equivalence to the predicate devices.

The Clickfine Pen Needles have met the requirements of the relevant sections of the following standards:

  • . ISO 11608-2:2012 Needle based injection systems for medical use -Requirements and test methods - Part 2: Needles
  • ISO 9626:1991/Amd.1:2001 Stainless steel needle tubing for the manufacture of . medical devices
  • ISO 7864:1993 Sterile hypodermic needles for single use .

The verifications have shown evidence that the Clickfine Pen Needles meet the acceptance criteria of these standards. Based on the results it can be concluded that the device performance and safety are acceptable for the product.

7.11 Conclusion

Y psomed AG concludes based on the information presented that the modified product is substantially equivalent to the current product legally marketed in the USA.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol, with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

OCT
2 5 2012

P/L Biomedical Mr. Lee Leichter 10882 Stonigton Avenue Fort Myers, Florida 33913

Re: K122971

Trade/Device Name: Clickfine Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: September 24, 2012 Received: September 26, 2012

Dear Mr. Leichter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Antim V.n

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name:

Indications For Use:

K 122971

Clickfine Pen Needle

The Clickfine pen needle is intended for the hypodermic injection of fluids into the body when attached to an injection pen.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K122971

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).