AFFINITY FUSION OXYGENATOR WITH INTEGRATED ARTERIAL FILTER AND BALANCE BIOSURFACE MODEL BB811
K122827 · Medtronic, Inc. · DTZ · Dec 14, 2012 · Cardiovascular
Device Facts
| Record ID | K122827 |
| Device Name | AFFINITY FUSION OXYGENATOR WITH INTEGRATED ARTERIAL FILTER AND BALANCE BIOSURFACE MODEL BB811 |
| Applicant | Medtronic, Inc. |
| Product Code | DTZ · Cardiovascular |
| Decision Date | Dec 14, 2012 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.4350 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.
Device Story
Single-use, microporous, hollow-fiber gas exchange device; integrated heat exchanger; integrated arterial filter; Balance Biosurface coating (heparin-free) to reduce platelet activation/adhesion. Used in extracorporeal perfusion circuits during cardiopulmonary bypass surgery. Blood pumped into heat exchanger for temperature control via water bath; flows through fiber bundle assembly for gas transfer (O2/CO2 diffusion) across hollow fiber membranes; arterial filter removes microemboli. Operated by perfusionists/clinicians in OR. Output provides oxygenated, temperature-controlled, filtered blood to patient; supports patient physiological needs during bypass. BPA-free copolyester housing.
Clinical Evidence
Bench testing only. In vitro performance testing conducted per ISO 7199 and ISO 15675. Tests included gas transfer, heat exchanger performance, high/min flow blood trauma, time to prime, gross air handling, filtration efficiency, and mechanical integrity (cap pulls, port break, tube pull, luer port design). Balance coverage/leaching and particulate shedding also evaluated. Device passed all performance specifications.
Technological Characteristics
Microporous hollow fiber oxygenator; integrated heat exchanger; integrated arterial filter. Housing: Eastman Tritan MX731 BPA-free copolyester. Coating: Balance Biosurface (biocompatible, heparin-free). Gas transfer via diffusion across hollow fiber membranes. Temperature control via water bath heat exchanger. Single-use. Standards: ISO 7199, ISO 15675.
Indications for Use
Indicated for patients undergoing routine cardiopulmonary bypass procedures requiring extracorporeal perfusion, blood oxygenation, carbon dioxide removal, temperature regulation, and microemboli filtration for up to 6 hours.
Regulatory Classification
Identification
A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”
Predicate Devices
- Affinity NT Oxygenator 511T (K973760)
Related Devices
- K123314 — AFFINITY FUSION OXYGENATOR WITH INTEGRATED ARTERIAL FILTER AND CARMEDA BIOACTIVE SURFACE · Medtronic, Inc. · Jan 22, 2013
- K142784 — Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Affinity Fusion Oxygenator with Integrated Arterial Filter and Carmeda BioActive Surface · Medtronic, Inc. · Oct 24, 2014
- K143073 — Affinity NT Oxygenator, Affinity NT Oxygenator with Trillium Biosurface, Affinity NT Oxygenator with Carmeda Biosurface · Medtronic, Inc. · Mar 25, 2015
- K031223 — SYNTHESIS MIMESYS ADULT MEMBRANE OXYGENATOR · Dideco S.P.A. · May 2, 2003
- K140774 — TERUMO CAPIOX FX15 AND FX25 HOLLOW FIBER OXYGENATOR/RESERVOIR · Terumo Cardiovascular Systems Corporation · Jun 2, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
K122827
1 4 2012
# 510(k) Summary of Safety and Effectiveness
| Date Prepared: | September 12, 2012 |
|----------------------|----------------------------------------------------------------------------------------|
| | DEC 1 |
| Applicant: | Medtronic, Inc. |
| | Medtronic Perfusion Systems |
| | 7611 Northland Drive |
| | Brooklyn Park, MN 55428 |
| | Establishment Registration No. 2184009 |
| Contact Person: | Mary Donlin |
| | Senior Regulatory Affairs Specialist |
| | Phone: (763) 526-9172 |
| | Fax: (763) 367-8147 |
| | E-mail:mary.e.donlin@medtronic.com |
| Trade Name: | Affinity Fusion® Oxygenator with Integrated Arterial Filter and<br>Balance® Biosurface |
| Common Name: | Oxygenator |
| Classification Name: | Cardiopulmonary bypass oxygenator |
| Classification: | Class II, 21 CFR 870.4350 |
Product Code:
DTZ
Name of Predicate Device: Affinity NT Oxygenator 511T (K973760)
# Device Description:
The Affinity Fusion Hollow Fiber Oxygenator with Integrated Arterial Filter and Balance Biosurface is a single use, microporous, hollow fiber, gas exchange device with plasma-resistant fiber and integrated heat exchanger. The oxygenator is coated on its primary blood contacting surfaces with Balance Biosurface to reduce platelet activation and adhesion and preserve platelet function.
# Intended Use:
The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
{1}------------------------------------------------
The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.
## Contraindications:
Do not use this device for any purpose other than indicated.
# Comparison to the Predicate Device:
The Affinity Fusion Oxygenator Model BB811 has the same intended use and principles of operation and technology when compared to the predicate device. The design incorporated an integrated heat exchanger and an integrated arterial filter. The material used to manufacture the housing of the oxygenator is a copolvester material that is Bisphenol A-free (BPA free).
#### Intended Use:
The Affinity Fusion Oxygenator has the same intended use as the predicate Affinity NT oxygenator (K973760) with the addition of the filtering capability due to the integration of arterial filter within the fiber bundle assesmbly.
#### Design and Materials: .
The design and the materials of the Affinity Fusion Oxygenator and the predicate device are essentially the same. The design of the oxygenator device is similar in that they each contain a heat exchanger for temperature control, and a fiber bundle assembly for gas transfer. The device is manufactured with various adhesives and urethanes. The housing of the Affinity Fusion oxygenator is made of a Bisphenol A-free (BPA-free) Eastman Tritan TM Copolyester, MX731, which differs from the polycarbonate material used in the predicate devices. The Affinity Fusion oxygenator Model BB811 is provided with Balance Biosurface coating. Balance is a biocompatible surface coating that increases the thromboresistance of the blood contact surfaces. Balance is a heparin free version of Trillium coating. Trillium coating is available on the Affinity NT Hollow Fiber Oxygenator.
# • Principles of Operation and Technology:
The principles of operation of the subject device and the predicate devices are essentially identical. Blood is pumped into the heat exchanger device whereby blood temperature is controlled with the use of essentially a water bath. After the blood exits the heat exchanger, it enters the oxygenator device through the fiber bundle assembly through which the gas transfer occurs (i.e., introduction of oxygen; removal of carbon dioxide). The transfer process occurs via diffusion across the walls of the hollow fiber membranes contained within the oxygenator.
## • Performance:
In vitro testing was carried out in accordance with the relevant requirements of "Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions; Final Guidance for Industry and
{2}------------------------------------------------
FDA Staff issued on November 13, 2000, " ISO 7199 "Cardiovascular implants and artificial organs - Blood-gas exchangers (oxygenators)"; and ISO 15675 "Cardiovascular implants and artificial organs - Cardiopulmonary bypass systems - Arterial blood line filters".
In vitro testing was carried out to demonstrate both the substantial equivalence with the predicate device and also to comply with safety and effectiveness requirements.
Testing supplied in the 510(k) premarket notification includes performance tests, physical and mechanical integrity tests that demonstrate compliance with performance specifications.
The tests that were performed are listed in the following summary table. The Affinity Fusion Oxygenator passed each test mentioned in the table below.
| Test | Test Classification | Test Title |
|------|------------------------|-----------------------------------|
| 1. | Functional/Performance | Gas Transfer |
| 2. | Functional/Performance | Heat Exchanger Performance |
| 3. | Functional/Performance | High Flow Blood Trauma |
| 4. | Functional/Performance | Min Flow Blood Trauma |
| 5. | Functional/Performance | Time to Prime |
| 6. | Functional/Performance | Gross Air Handling |
| 7. | Functional/Performance | Filtration Efficiency |
| 8. | Physical/Mechanical | Cap Pulls |
| 9. | Physical/Mechanical | Integrity |
| 10. | Physical/Mechanical | Port Break and Tube Pull |
| 11. | Physical/Mechanical | Luer port design - sampling port |
| 12. | Physical/Mechanical | Luer port design - air purge line |
| 13. | Functional/Performance | Balance Coverage/Leaching |
| 14. | Functional/Performance | Particulate Shedding |
### Conclusion:
The data included in this submission is sufficient to provide reasonable assurance of the safety and effectiveness of the device and the Affinity Fusion® Oxygenator with Integrated Arterial Filter is substantially equivalent to the legally marketed predicate device, Affinity NT Oxygenator 511T (K973760).
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem that resembles an abstract caduceus or a symbol representing health and human services.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
DEC 14 2012 .
Medtronic Cardiovascular Mary E. Donlin. Senior Regulatory Affairs Specialist 8200 Coral Street NE Mailstop MVS83 Mounds View, MN 551112
Re: K122827
Trade/Device Name: Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface
Regulation Number: 21 CFR 870.4350
Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ
Dated: September 12, 2012
Received: September 18, 2012
#### Dear Ms. Donlin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Mary E. Donlin
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic ford in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you desire specific davide in your avers offices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Matthew G. Hillebrenner
for
Bram D. Zuckerman, M.D. Division Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
# 4.0 Indications for Use Statement
K122827 510(k) Number (if known):
Device Name:
Affinity Fusion® Oxygenator with Integrated Arterial Filter and Balance® Biosurface
Indications for Use:
The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|---------|
| (Division Sign-Off) Division of Cardiovascular Devices | |
| 510(k) Number | K122827 |