K Number
K122718

Validate with FDA (Live)

Date Cleared
2013-02-06

(154 days)

Product Code
Regulation Number
866.3740
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Veritor" System for Rapid Detection of Group A Strep test is a rapid chromatographic immunoassay for the direct and qualitative detection of Group A Streptococcus antigen from throat swabs of symptomatic patients. It is intended to be used in conjunction with the BD Veritor™ System Reader as an aid in the diagnosis of Group A Strep. All negative test results should be confirmed by bacterial culture because negative results do not preclude Group A Strep infection and should not be used as the sole basis for treatment.

The BD Veritor System for rapid detection of Group A Strep test is intended for use in point-of-care or laboratory settings.

Device Description

The BD Veritor™ System for rapid detection of Group A Strep is a qualitative, lateral flow immunoassay for the detection of Strep A carbohydrate antigen in a throat swab. In this test, antibody specific to Strep A carbohydrate antigen is coated on the test line region of the Assay device. During testing, the processed throat swab specimen reacts with an antibody to Strep A that is conjugated onto detector particles. The mixture migrates up the membrane to react with the antibody to Strep A on the membrane and is captured by the line of antibody on the membrane. A positive result for Strep A is determined by the BD Veritor ™ System Reader when antigen-conjugate is deposited at the Test "T" position and the Control "C" position on the BD Veritor™ System Strep A assay device.

AI/ML Overview

Here's an analysis of the acceptance criteria and study detailed in the provided document for the BD Veritor™ System for Rapid Detection of Group A Strep:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria values for sensitivity and specificity. However, based on the Intended Use statement and the comparison to an existing predicate device, we can infer that the device aims for performance comparable to or better than current rapid chromatographic immunoassays for Group A Strep. Given the nature of a 510(k) submission, the "acceptance criteria" here are implicitly met by demonstrating substantial equivalence to the predicate and by achieving clinically acceptable performance metrics.

For the purpose of this response, I will list the observed performance metrics from the clinical study as the "reported device performance." If explicit acceptance criteria were given, they would typically be pre-specified thresholds (e.g., "Sensitivity > 85%", "Specificity > 95%").

MetricReported Device Performance (Overall)
Sensitivity96.6% (144/149)
Specificity95.5% (618/647)

Additional Performance Data (Analytical & Reproducibility):

Performance AspectDetails / Reported Performance
Analytical Sensitivity (LOD)- Strain 12384: 1 x 10^5 CFU/mL (95.0% positivity) - Strain 19615: 5 x 10^4 CFU/mL (96.7% positivity) - Strain 25663: 2 x 10^5 CFU/mL (95.0% positivity)
Analytical Specificity- Lancefield Groups A, B, C, D, F, G (at 1x10^9 CFU/mL): Negative results (no cross-reactivity with other Streptococcus groups). - Various bacteria and yeasts (e.g., Arcanobacterium haemolyticum, Candida albicans, E. coli, S. aureus) and viruses (e.g., Adenovirus, Cytomegalovirus, HSV, Influenza): No cross-reactivity observed.
Interfering Substances40+ common substances found in respiratory samples (e.g., 4-Acetamidophenol, Acetylsalicylic acid, Albuterol, various blood types, nasal sprays, throat lozenges) tested at specified concentrations: None exhibited interference with Group A positive or Group A negative samples.
Media CompatibilityFour transport media (Modified Amies, Modified Stuart's, Normal Saline, Phosphate Buffered Saline) and two agar types (Tryptic Soy Agar with 5% Sheep Blood, Selective Strep Agar) were compatible. Expected results were obtained, and acceptance criteria were met for room temperature and overnight frozen storage conditions.
Reproducibility- High negative sample: 1.1% false positive rate (1/90 samples across 3 sites). - Low positive sample: 91.1% positivity (82/90 samples across 3 sites). - Moderate positive sample: 98.9% positivity (89/90 samples across 3 sites). - Negative sample: 0% false positive rate (0/90 samples across 3 sites).

2. Sample Size and Data Provenance (Clinical Study)

  • Sample Size for Test Set: 796 prospectively collected specimens.
  • Data Provenance: The study was a multi-center clinical trial conducted at one clinical laboratory site and four Point-of-Care (POC) sites. The country of origin is not explicitly stated but is implicitly the USA, given the FDA 510(k) submission. The data was prospective.
  • Patient Demographics:
    • Gender: 51.8% female, 48.2% male.
    • Age: 39.1% <= 5 years old; 59.3% 6-21 years old; 1.6% >= 22 years old.

3. Number of Experts and Qualifications for Ground Truth (Clinical Study)

The document states that the performance was determined by comparison to bacterial culture. This implies that clinical laboratory professionals (e.g., microbiologists, medical technologists) conducted and interpreted the bacterial cultures, which served as the gold standard. The specific number of experts or their years of experience are not explicitly mentioned.

4. Adjudication Method for the Test Set (Clinical Study)

The document does not describe an explicit adjudication method (e.g., "2+1" or "3+1") for discrepancies between the device result and the bacterial culture. The bacterial culture itself is treated as the reference standard. Discrepancies would simply be counted as false positives or false negatives for the device.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not explicitly described in terms of human readers using the device with and without AI assistance.
  • The study compares the device's performance (interpreted by the BD Veritor™ System Reader) against bacterial culture, which is a standalone assessment of the device's accuracy. The "Detection Format" section mentions an "opto-electronic reader determines the line intensity... interprets the results using the scoring algorithm, and reports a positive, negative, or invalid result," which indicates an automated interpretation, not a human reader decision with or without AI.
  • Therefore, an effect size of how much human readers improve with AI vs. without AI assistance is not applicable/provided in this document.

6. Standalone Performance Study (Algorithm Only)

Yes, a standalone study was done. The entire clinical performance section evaluates the BD Veritor™ System (which includes the device and the associated reader/algorithm for interpretation) as a standalone diagnostic tool. The "opto-electronic reader" described in the device comparison section directly indicates algorithmic interpretation without human intervention for the final positive/negative result.

7. Type of Ground Truth Used (Clinical Study)

The type of ground truth used for the clinical study was bacterial culture (often referred to as the "gold standard" for Group A Strep diagnosis).

8. Sample Size for the Training Set

The document does not specify a separate training set or its sample size for the device's algorithm. It primarily details the clinical validation study (test set). For an IVD device like this, the "training" might involve internal development and optimization using characterized samples, but a dedicated clinical training set of patient samples is not typically reported in the same way as for AI/ML regulatory submissions.

9. How the Ground Truth for the Training Set Was Established

As no specific training set of patient samples is detailed, the method for establishing its ground truth is not provided. If such a set were used during device development, its ground truth would likely also be established through bacterial culture or other highly reliable laboratory methods, similar to the clinical test set.

{0}------------------------------------------------

BD Veritor The System for Rapid Detection of Group A Strep

K1227/8

510(k) SUMMARY

FEB 0 6 2013

BECTON, DICKINSON AND COMPANY SUBMITTED BY: 10865 Road to the Cure, Suite 200 San Diego, CA 92121 Phone: 858-795-7890 Fax: 858-812-8505 Gregory Payne

CONTACT NAME: DATE PREPARED: January 31, 2013

DEVICE TRADE NAME: BD Veritor™ System for Rapid Detection of Group A Strep

DEVICE COMMON NAME: Streptococcus spp. serological reagents

21 CFR § 866.3740 DEVICE CLASSIFICATION:

PREDICATE DEVICES: Clearview Advanced™ Strep A Test

INTENDED USE:

The BD Veritor" System for Rapid Detection of Group A Strep test is a rapid chromatographic immunoassay for the direct and qualitative detection of Group A Streptococcus antigen from throat swabs of symptomatic patients. It is intended to be used in conjunction with the BD Veritor™ System Reader as an aid in the diagnosis of Group A Strep. All negative test results should be confirmed by bacterial culture because negative results do not preclude Group A Strep infection and should not be used as the sole basis for treatment.

The BD Veritor System for rapid detection of Group A Strep test is intended for use in point-of-care or laboratory settings.

DEVICE DESCRIPTION:

The BD Veritor™ System for rapid detection of Group A Strep is a qualitative, lateral flow immunoassay for the detection of Strep A carbohydrate antigen in a throat swab. In this test, antibody specific to Strep A carbohydrate antigen is coated on the test line region of the Assay device. During testing, the processed throat swab specimen reacts with an antibody to Strep A that is conjugated onto detector particles. The mixture migrates up the membrane to react with the antibody to Strep A on the membrane and is captured by the line of antibody on the membrane. A positive result for Strep A is determined by the BD Veritor ™ System Reader when antigen-conjugate is deposited at the Test "T" position and the Control "C" position on the BD Veritor™ System Strep A assay device.

BD Diagnostic Systems Becton, Dickinson and Company

{1}------------------------------------------------

DEVICE COMPARISON:

The BD Veritor™ System for Rapid Detection of Group A Strep was compared to the Clearview Advanced™ Strep A test K091489.

Product FeatureBD Veritor™ System for Rapid Detection of Group A StrepBinax Clearview Advanced Strep A test K091489
Intended UseThe BD Veritor™ System for Rapid Detection of Group A Strep test is a rapid chromatographic immunoassay for the direct and qualitative detection of group A Streptococcus antigen from throat swab specimens of symptomatic patients to aid in diagnosis of Group A Streptococcus infection. It is intended to be used in conjunction with the BD Veritor™ System Reader as an aid in the diagnosis of Group A Streptococcal infection.The Binax Clearview Advanced™ Strep A test is a rapid chromatographic immunoassay for the qualitative detection of group A Streptococcus antigen from throat swab specimens as an aid in the diagnosis of Group A Streptococcal infection.
Specimen TypesThroat swabsThroat swabs
Assay TechnologyImmunochromatographicImmunochromatographic
Detection FormatAn opto-electronic reader determines the line intensity at each of the spatially-defined test and control line positions, interprets the results using the scoring algorithm, and reports a positive, negative, or invalid result on the LCD screen based on pre-set thresholds.Visual determination of presence or absence of pink-to-red Test Line and the appearance of a blue Procedural Control Line on the test strip indicate the presence of Group A Streptococcal antigen.
Qualitative or QuantitativeQualitativeQualitative
ProductFeatureBD Veritor™ System forRapid Detection of Group AStrepBinax ClearviewAdvanced Strep Atest K091489
Total Assay TimeApproximately 5 minutesLess than 5 minutes
Control format• Kit Strep A positive control swab• Kit Strep A negative control swab• Internal control lines• Kit Strep A positive control swab• Kit Strep A negative control swab
Detection ofGroup A StrepTest will indicate thepresence of both viable andnon-viable group AStreptococcus bacteriaTest will indicate thepresence of bothviable and non-viablegroup AStreptococcusbacteria

· ·

{2}------------------------------------------------

SUMMARY OF PERFORMANCE DATA:

Analytical Sensitivity

The limit of detection for Streptococcus pyogenes was established with the BD Veritor" System for Rapid Detection of Group A Strep test. The limit of detection (LOD) is defined as the lowest concentration that produces an approximate 95% positive reaction when tested with 60 replicates.

STRAINLOD(CFU/mL)Results%Positivity
12384$1 \times 10^5$57/60Positive95.0%
19615$5 \times 10^4$58/60Positive96.7%
25663$2 \times 10^5$57/60Positive95.0%

Analytical Specificity

The reactivity of various Streptococcal strains was determined with the BD Veritor" System for Rapid Detection of Group A Strep test. Lancefield Groups A, B, C, D, F and G were tested at 1X109 CFU/mL in triplicate and yielded negative results.

{3}------------------------------------------------

Various microorganisms (including bacteria and yeasts) that might be found in specimens
were evaluated for potential cross reactivity with the BD Veritor " System for Rapid Detection of Group A Strep test.

Microorganism NameConcentrationTestedGroup A Strep. TestResult
Arcanobacterium haemolyticum1x109 CFU/mLNegative
Bordetella pertussis5x108 CFU/mLNegative
Candida albicans1x109 CFU/mLNegative
Corynebacterium diphtherium sp(Corynebacterium sp)1x109 CFU/mLNegative
Enterococcus faecalis1x109 CFU/mLNegative
Enterococcus faecium1x109 CFU/mLNegative
Escherichia coli1.5x109 CFU/mLNegative
Fusobacterium necrophorum1x109 CFU/mLNegative
Haemophilus influenzae1x109 CFU/mLNegative
Haemophilus parahemolyticus1.2x105 CFU/mLNegative
Haemophilus parainfluenzae1x109 CFU/mLNegative
Klebsiella pneumoniae1.5x109 CFU/mLNegative
Lactobacillus sp (Lactobacilluscasei)1x109 CFU/mLNegative
Moraxella catarrhalis1x109 CFU/mLNegative
Moraxella lacunata1x109 CFU/mLNegative
Mycobacterium tuberculosisavirulent5x106 CFU/mLNegative
Neisseria gonorrhoeae1x109 CFU/mLNegative
Neisseria lactamica1x109 CFU/mLNegative
Neisseria meningitidis1x109 CFU/mLNegative
Neisseria mucosa1x106 CFU/mLNegative
Neisseria sicca1x109 CFU/mLNegative
Neisseria subflava1x109 CFU/mLNegative
Proteus vulgaris1x109 CFU/mLNegative
Pseudomonas aeruginosa1x109 CFU/mLNegative
Serratia marcescens1x109 CFU/mLNegative
Staphylococcus aureus1x109 CFU/mLNegative
Staphylococcus epidermidis1x109 CFU/mLNegative
Staphylococcus haemolyticus1x109 CFU/mLNegative
Streptococcus anginosus1x109 CFU/mLNegative
Streptococcus mitis1x109 CFU/mLNegative
StreptococcusmutansATCC251733x109 CFU/mLNegative
Staphylococcus oralis1x109 CFU/mLNegative
Streptococcus pneumoniae1x109 CFU/mLNegative
BD VeritorTM System for Rapid Detection of Group A Strep Cross Reactivity
Study Results - Bacteria and Yeast
Microorganism NameConcentrationTestedGroup A Strep TestResult
Staphylococcus sanguis1x109 CFU/mLNegative
Streptococcus sp. Group B1x109 CFU/mLNegative
Streptococcus sp. Group C1x109 CFU/mLNegative
Streptococcus sp. (bovis II)Group D1x109 CFU/mLNegative
Streptococcus sp. Group F1x109 CFU/mLNegative
Streptococcus sp. Group G1x109 CFU/mLNegative
Yersinia enterocolitica1x109 CFU/mLNegative
Adenovirus Type 11.6x106 TCID50 /mLNegative
Adenovirus Type 72.81x105TCID50/mLNegative
Cytomegalovirus8.9x103 TCID50/mLNegative
Enterovirus (VR-28 HumanCoxsackievirus)8.9x106TCID50/mLNegative
Epstein Barr VirusN/ANegative
HSV Type 1 (HF)8.89x106TCID50/mLNegative
Human coronavirus OC432.81x104TCID50/mLNegative
Human metapneumovirus(HMPV-27 A2)2.8x106 TCID50/mLNegative
Human parainfluenza2.8x106 TCID50/mLNegative
Measles1.6x104 TCID50/mLNegative
Mumps virus1.6x105 TCID50/mLNegative
Respiratory syncytial virus VR-261.6 x 107TCID50/mLNegative
Rhinovirus2.8x106 TCID50/mLNegative

Streptococcus salivarius

Negative

1x109 CFU/mL

{4}------------------------------------------------

Of the microorganisms tested, none demonstrated cross-reactivity with the BD Veritor™ System for Rapid Detection of Group A Strep test.

Interfering Substances

Various substances were evaluated for potential interference with the BD Veritor™ System for Rapid Detection of Group A Strep test at concentrations comparable to or greater than levels that may be present in patient respiratory samples.

SubstanceConcentrationTestedInterferencewith GroupA Result
4-Acetamidophenol10 mg/mLNo
Acetylsalicylic acid20 mg/mLNo
Albuterol0.083 mg/mLNo
Amantadine500 ng/mLNo
Ascorbic acid chewable tablets5% by weightNo
Beclomethasone500 ng/mLNo
Benzocaine throat spray(Cepacol)5% by volumeNo
Blood, type A2% (v/v)No
Blood, type B2% (v/v)No
Blood, type AB2% (v/v)No
Blood, type O2% (v/v)No
Budesonide500 ng/mLNo
Chlorpheniramine maleate5 mg/mLNo
Dexamethasone10 mg/mLNo
Dextromethorphan (10 mg/mL)10 mg/mLNo
Dyclonine HCI lozenges(Sucrets)5% w/vNo
Diphenhydramine HCI5 mg/mLNo
Fexofenadine500 ng/mLNo
FluMist1% v/vNo
Fluticasone500 ng/mLNo
Guaiacol Glyceryl Ether20 mg/mLNo
Ibuprofen10 mg/mLNo
Loratidine100 ng/mLNo
Menthol Throat Lozenges5% w/vNo
Mometasone500 ng/mLNo
Mouthwash (at 5% by volume)Listerine5% (v/v)No
Mouthwash Scope5% v/vNo
Mouthwash CVS5% v/vNo
Mucin, salivary protein, purified1 mg/mLNo
Nasal Spray5% v/vNo
Nasal Spray5% v/vNo
Nasal Spray5% v/vNo
Oseltamivir500 ng/mLNo
Oxymetazoline0.05 mg/mLNo
Phenol throat spray(Chloraseptic)5% v/vNo
Phenylephrine1 mg/mLNo
Pseudoephedrine HCI20 mg/mLNo
Throat drops: CVS5% w/vNo
Throat drops: Pedia Care5% w/vNo
Throat drops: Triaminic5% w/vNo
Tobramycin500 ng/mLNo
Triamcinolone500 ng/mLNo
Zanamivir1 mg/mLNo
Zicam throat spray (Zn /5% v/vNo
benzalkonium chloride)
Zinc Lozenges5% w/vNo

BD Diagnostic Systems Becton, Dickinson and Company ... . . . . . . .

{5}------------------------------------------------

.

:

{6}------------------------------------------------

BD Veritor TM System for Rapid Detection of Group A Strep

Of the substances tested in this study, none exhibited interference when either Group A positive or Group A negative samples were tested with the BD Veritor™ System for Rapid Detection of Group A Strep test.

Media Compatibility

Various types of transport media and culture plate media commonly used in Strep A testing were evaluated for compatibility with the BD Veritor™ System for Rapid Detection of Group A Strep test. The effects of frozen storage of transport media samples on the stability of the antigen were evaluated in this study. The media tested were: Modified Amies, Modified Stuart's, Normal Saline and Phosphate Buffered Saline. The agar tested were Tryptic Sov Agar with 5% Sheep Blood and Selective Strep Agar.

Of the four media tested in this evaluation, all four demonstrated the expected results and met the acceptance criteria for both room temperature and overnight frozen storage conditions. Therefore, these four media are all compatible with the BD Group A Strep test. Although the media were non-interfering, dry swab transport and storage is recommended for testing with the BD Group A Strep test. Storage and transport of Strep A specimens in liquid transport will likely dilute the antigen while streaking on solid culture media may remove some organism from the swab, thus resulting in a lower number of bacteria introduced into the extraction reagent.

No interference was seen with the agar tested.

CLINICAL STUDIES

Performance characteristics for the BD Veritor™ System for Rapid Detection of Group A Strep were established in a multi-center clinical trial conducted at one clinical laboratory site and four POC sites during the 2011-2012 respiratory season. A total of 796 prospectively collected specimens were evaluated using the BD Veritor™ System for Rapid Detection of Group A Strep and bacterial culture. Throat swabs from symptomatic patient were obtained, 51.8% from females and 48.2% were from males. Specimens from patients five years old or younger comprised 39.1% of the total, with 59.3% from patients 6 to 21 years of age and 1,6% from patients 22 years of age or older.

{7}------------------------------------------------

The performance of the BD Veritor™ System for Rapid Detection of Group A Strep was determined by comparison to bacterial culture and is presented in the table below.

Clinical Performance Data
Culture
VeritorPN
P14429173
N5618623
149647796

Specificity: 95.5% (93.6%, 96.9%)

Culture-Reference
SiteCodeVeritorPNTotal
ClinicalSiteP20222
N08282
Total2084104
Reference Method: Culture-ReferenceSensitivity: 100% (83.9%, 100%)Specificity: 97.6% (91.7%, 99.3%)
POC 1P54357
N5188193
Total59191250
Reference Method: Culture-ReferenceSensitivity: 91.5% (81.6%, 96.3%)Specificity: 98.4% (95.5%, 99.5%)
POC 2P21930
N0111111

BD Diagnostic Systems Becton, Dickinson and Company

{8}------------------------------------------------

Clinical Performance Data - By Site
SiteCodeVeritorCulture-Reference
PNTotal
POC 3P21728
POC 3N0106106
Total21113134
Reference Method: Culture-ReferenceSensitivity: 100% (84.5%, 100%)Specificity: 93.8% (87.8%, 97.0%)
POC 4P28836
POC 4N0131131
Total28139167
Reference Method: Culture-ReferenceSensitivity: 100% (87.9%, 100%)Specificity: 94.2% (89.1%, 97.1%)

Reproducibility

The reproducibility of the BD Veritor™ System for Rapid Detection of Group A Strep was evaluated at one clinical and two POC sites. The panel was composed of 4 simulated Group A Strep samples. These included high negative samples ( i.e. samples containing a very low concentration of Group A Strep), a low positive sample (near the limit of detection), a moderate positive sample and a negative sample. The panel was tested by two operators at each site over five days. The results are summarized below.

BD Veritor™ Group A Strep Reproducibility
SampleSite 1Site 2Site 3Total
High negative3.3% (1/30)0% (0/30)0% (0/30)1.1% (1/90)
(0.6%, 16.7%)(0%, 11.3%)(0%, 11.3%)(0.2%, 6%)
Low positive96.7% (29/30)83.3% (25/30)93.3% (28/30)91.1% (82/90)
(83.3%, 99.4%)(66.4%, 92.7%)(78.7%, 98.2%)(83.4%, 95.4%)
Moderate positive100% (30/30)96.7% (29/30)100% (30/30)98.9% (89/90)
(88.6%, 100%)(83.3%, 99.4%)(88.6%, 100%)(94%, 99.8%)

{9}------------------------------------------------

BD Veritor™ System for Rapid Detection of Group A Strep

Negative.0% (0/30)0% (0/30)0% (0/30)------------------------------------------------------------------------------------------------------------------------------------------------------------------------------0% (0/90)
1 3%10%.l.J10%.11.3%)11.3%)(0%.(0%, 4.1%)

:

. . . . . . .

:

:

:

.

:

.

:

. . . . .

:

.

.

:

. . . . . . . . . . . . . .

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in bold, uppercase letters. The text is horizontally oriented and centered. To the left of the text is a stylized graphic, possibly a logo, with a series of vertical lines.

Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol. The text is in all caps and is in a sans-serif font.

February 6, 2013

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Becton, Dickinson and Company C/O Gregory Payne Director, Quality Systems and Regulatory Affairs 10865 Road to the Cure, Suite 200 San Diego, CA 92121

Re: K122718

Trade/Device Name: BD Veritor™ System for Rapid Detection of Group A Strep · Regulation Number: 21 CFR 866. 3740 Regulation Name: Streptococcus spp. Serological Reagents Regulatory Class: Class I Product Code: GTY Dated: January 17, 2013 Received: January 23, 2013

Dear Mr. Payne:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{11}------------------------------------------------

Page 2 - Gregory Payne

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Uwe Scherf for

Sally A. Hoivat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{12}------------------------------------------------

BD Veritor" System for Rapid Detection of Group A Strep CONFIDENTIAL AND PROPRIETARY Clinical Laboratory Product

510(k) Number: K122718

Device Name: BD Veritor ™ System for Rapid Detection of Group A Strep

Indications for Use:

The BD Veritor" System for Rapid Detection of Group A Strep test is a rapid chromatographic immunoassay for the direct and qualitative detection of Group A Streptococcus antigen from throat swabs of symptomatic patients. It is intended to be used in conjunction with the BD Veritor"" System Reader as an aid in the diagnosis of Group A Strep. All negative test results should be confirmed by bacterial culture because negative results do not preclude Group A Strep infection and should not be used as the sole basis for treatment.

The BD Veritor System for rapid detection of Group A Strep test is intended for use in point-of-care or laboratory settings.

Over-the-Counter Use AND/OR Prescription Use V (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

csk Raquel Peat, PhD

Division Sidn-

Office of In Vitro Diagnostics and Radiological Health

K122 718 510(k)

§ 866.3740

Streptococcus spp. serological reagents.(a)
Identification. Streptococcus spp. serological reagents are devices that consist of antigens and antisera (excluding streptococcal exoenzyme reagents made from enzymes secreted by streptococci) used in serological tests to identifyStreptococcus spp. from cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genusStreptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.