K Number
K122687

Validate with FDA (Live)

Date Cleared
2013-01-10

(128 days)

Product Code
Regulation Number
862.1675
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Blood Culture Holder is used to fill blood into blood culture bottles and tubes in routine venipuncture procedures. This device is to be used by properly trained healthcare professionals only in accordance with these instructions.

The SAFETY Blood Collection Set and/or Blood Collection Set + Blood Culture Holder are used in routine venipuncture procedures. The winged needle of the SAFETY Blood Collection Set is designed with a while shield, which can be activated to cover the needle immediately following blood collection to aid in the protection against accidental needlestick injury. The Blood Culture Holder is used to fill blood into blood culture bottles and tubes.

Device Description

The Blood Culture Holder is a single-use, non-sterile, plastic holder with threads which are compatible with male luer adapters. The holder can be used with all VACUETTE® Multiple Sample Luer Adapters and Safety Blood Collection Sets for routine venipuncture into various manufacturers' evacuated blood culture bottles and tubes. It is supplied alone or with the Greiner sterile Safety Blood Collection Set with Luer Adapter.

The Blood Culture Holder does not come into contact with the blood specimen. It is an external holder for the Safety Blood Collection Set and the VACUETTE® Multiple Sample Luer Adapter. The Safety Blood Collection Set consists of tubing with a blood collection needle at one end and a female luer at the other. The VACUETTE® Multiple Sample Luer Adapter consists of a male luer at one end and a cannula covered with a sheath. The luer adapter is threaded into the Blood Culture Holder, with the male luer end on the outside and the covered cannula on the inside of the holder barrel. The blood flows through the blood collection needle and tubing of the Safety Blood Collection Set, into the male luer of the VACUETTE® Multiple Sample Luer Adapter, through the luer into the cannula, and into the blood culture tube or holder when punctured by the cannula. The fluid path consists of the Safety Blood Collection Set and the VACUETTE® Multiple Sample Luer Adapter. The Blood Culture Holder never comes into contact with the blood specimen during conditions of use.

When the Blood Culture Holder is supplied attached to the Safety Blood Collection Set and the VACUETTE® Multiple Sample Luer Adapter, it undergoes sterilization as part of the entire system because the fluid path must be sterilized. However, when the Blood Culture Holder is provided alone, it is not sterilized and the user attaches their own sterile blood collection needle and luer adapter.

The Safety Blood Collection Set with Blood Culture Holder is non-pyrogenic and tested according to ISO-10993-11 rabbit pyrogen test. The devices also are below the limit for endotoxin and tested according to ISO-10993-11 LAL Turbidimetric Method.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Blood Culture Holder:

Acceptance Criteria and Device Performance Study for Greiner Bio-One North America, Inc. Blood Culture Holder

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Fit of Blood Culture Holders to Safety Blood Collection SetsMet acceptance criteria for fit.
Leakage of Blood Culture ContainersMet acceptance criteria for leakage.
Function of Blood Culture ContainersMet acceptance criteria for function.
Leakage of Blood Collection TubesMet acceptance criteria for leakage.
Function of Blood Collection TubesMet acceptance criteria for function.
Compatibility with Blood Culture Bottles/VialsThe Blood Culture Holder can be used with blood culture bottles and blood culture vials.
Compatibility with Blood Collection Tube SizesThe Blood Culture Holder can be used with the range of blood collection tube sizes from 13x75mm (smallest diameter cap and tube size) to 16x100mm (largest diameter cap and tube size).
Non-pyrogenicTested according to ISO-10993-11 rabbit pyrogen test and found to be non-pyrogenic.
Below Endotoxin LimitTested according to ISO-10993-11 LAL Turbidimetric Method and found to be below the limit for endotoxin.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Twenty (20) Blood Culture Holders were connected to 20 Safety Blood Collection Sets with Luer Adapters.
  • Data Provenance: The study was a simulated blood draw testing conducted by the manufacturer, Greiner Bio-One North America, Inc. The country of origin of the data is not explicitly stated but can be inferred as the USA, where Greiner Bio-One North America, Inc. is located. It is a prospective study as it involved testing new devices.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • The document does not mention the use of experts to establish ground truth for the simulated blood draw testing. The evaluation of proper fit, leakage, and function appears to have been performed internally by the manufacturer's testing personnel, based on pre-defined criteria.

4. Adjudication Method for the Test Set

  • The document does not mention any adjudication method (e.g., 2+1, 3+1, none) for the test set. The evaluation seems to be a direct assessment against the defined acceptance criteria.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was done. This study focuses on the performance of a medical device (Blood Culture Holder) in a standalone, mechanical/functional context, not on the comparative effectiveness of human readers with or without AI assistance. Therefore, there is no effect size of human readers improving with AI vs. without AI assistance.

6. Standalone Performance Study

  • Yes, a standalone study was done. The "Product Performance" section describes a study of the Blood Culture Holder and Safety Blood Collection Set with Luer Adapter's performance in simulated blood draws. This study evaluates the algorithm's (in this case, the device's) function without human intervention influencing the fundamental mechanical performance being assessed.

7. Type of Ground Truth Used

  • The ground truth used was based on pre-defined engineering and functional specifications/acceptance criteria for fit, leakage, and function. This is a form of empirical ground truth where the device's physical performance is directly observed and measured against established performance standards. For pyrogenicity and endotoxin, the ground truth was based on adherence to ISO-10993-11 standards.

8. Sample Size for the Training Set

  • The document does not mention a training set as this is a physical medical device, not an AI or algorithmic system that requires machine learning training. The focus is on product performance validation, not algorithm development.

9. How the Ground Truth for the Training Set Was Established

  • As there is no training set mentioned or required for this type of device, this question is not applicable.

{0}------------------------------------------------

K122687 510(K) SUMMARY

December 7. 2012

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is:

CONTACT:

Manfred Abel Greiner Bio-One North America, Inc. P.O Box 1026 Monroe, NC 28111

NAME OF DEVICE: Trade Name:

Common Names/Descriptions: Classification Name:

Product Code:

Class:

Blood Culture Holder; Safety Blood Collection Set + Luer Adapter with preattached Blood Culture Holder

Tube Holder

21 CFR 862.1675 Blood specimen collection device

JKA

ll

PREDICATE DEVICE:

Smiths Medical ASD Inc. Saf-T Closed Blood Collection System® (K081229)

DEVICE DESCRIPTION: INTENDED USE:

Blood Culture Holder:

The Blood Culture Holder is used to fill blood into blood culture bottles and tubes in routine venipuncture procedures. This device is to be used by properly trained healthcare professionals only in accordance with these instructions.

Safety Blood Collection Set and/or Blood Collection Set + Blood Culture Holder:

The SAFETY Blood Collection Set and/or Blood Collection Set + Blood Culture Holder are used in routine venipuncture procedures. The winged needle of the SAFETY Blood Collection Set is designed with a safety shield, which can be activated to cover the needle immediately

{1}------------------------------------------------

following blood collection to aid in the protection against accidental needlestick injury. The Blood Culture Holder is used to fill blood into blood culture bottles and tubes.

PRODUCT DESCRIPTION:

The Blood Culture Holder is a single-use, non-sterile, plastic holder with threads which are compatible with male luer adapters. The holder can be used with all VACUETTE® Multiple Sample Luer Adapters and Safety Blood Collection Sets for routine venipuncture into various manufacturers' evacuated blood culture bottles and tubes. It is supplied alone or with the Greiner sterile Safety Blood Collection Set with Luer Adapter.

The Blood Culture Holder does not come into contact with the blood specimen. It is an external holder for the Safety Blood Collection Set and the VACUETTE® Multiple Sample Luer Adapter. The Safety Blood Collection Set consists of tubing with a blood collection needle at one end and a female luer at the other. The VACUETTE® Multiple Sample Luer Adapter consists of a male luer at one end and a cannula covered with a sheath. The luer adapter is threaded into the Blood Culture Holder, with the male luer end on the outside and the covered cannula on the inside of the holder barrel. The blood flows through the blood collection needle and tubing of the Safety Blood Collection Set, into the male luer of the VACUETTE® Multiple Sample Luer Adapter, through the luer into the cannula, and into the blood culture tube or holder when punctured by the cannula. The fluid path consists of the Safety Blood Collection Set and the VACUETTE® Multiple Sample Luer Adapter. The Blood Culture Holder never comes into contact with the blood specimen during conditions of use.

When the Blood Culture Holder is supplied attached to the Safety Blood Collection Set and the VACUETTE® Multiple Sample Luer Adapter, it undergoes sterilization as part of the entire system because the fluid path must be sterilized. However, when the Blood Culture Holder is provided alone, it is not sterilized and the user attaches their own sterile blood collection needle and luer adapter.

The Safety Blood Collection Set with Blood Culture Holder is non-pyrogenic and tested according to ISO-10993-11 rabbit pyrogen test. The devices also are below the limit for endotoxin and tested according to ISO-10993-11 LAL Turbidimetric Method.

Caution: US Federal law restricts this device for sale by or on the order of a physician.

{2}------------------------------------------------

SUBSTANTIAL EQUIVALENCE:

The Blood Culture Holder is substantially equivalent to the Smiths Medical ASD Inc. Saf-T Closed Blood Collection System® (K081229) in intended use and design. The similarities and differences are listed below.

Greiner GBOBlood Culture HolderSmiths Medical ASD Inc.Saf-T Closed Blood Collection System®(predicate device)
510(k) NumberBlood Culture Holder: K122687Safety Blood Collection Set withLuer Adapter: K011786K081229
Intended UseThe Blood Culture Holder is used tofill blood into blood culture bottlesand tubes in routine venipunctureprocedures. This device is to beused by properly trained healthcareprofessionals only in accordancewith these instructions.The Saf-T Closed BloodCollection System® device isintended for use as a directblood draw device into avacuum tube or to allow asyringe blood draw andtransfer to fill vacuum tubes.
DeviceDescriptionBlood culture holderBlood culture device
Device DesignCylindrical holder for vacuum bottleand tube placement : cylinder withthreads which are compatible withmale luer adaptersThe design of the holder allowscollection of blood into blood culturebottles and blood collection tubes.Cylindrical holder for vacuumbottle and tube placement :cylinder with threads whichare compatible with bloodcollection needles and maleluer adapters
CylinderPlasticPlastic

PRODUCT PERFORMANCE:

Simulated blood draw testing was performed on the Safety Blood Collection Set with Luer Adapter and the Blood Culture Holder using a variety of blood culture bottles and evacuated tubes representing the range of bottle and tube sizes. The devices used in the study were:

  • Blood collection devices: Greiner Safety Blood Collection Set with Luer . Adapter; Blood Culture Holders;
  • Blood culture bottles and vials (2 manufacturers); .
  • Blood collection tubes: VVACUETTE® EDTA K2, 4 mL, 13x75mm
    (smallest diameter cap and tube size) and VACUETTE® Z Serum .

{3}------------------------------------------------

Separator and Clot Activator, 7 mL, 16x100mm (largest diameter cap and tube size).

Twenty Blood Culture Holders were connected to 20 Safety Blood Collection . Sets with Luer Adapters. The sets were tested in simulated blood draws with blood culture bottles, blood culture vials, and the two tube types.

The results were evaluated for proper fit of the Blood Culture Holders to the Safety Blood Collection Sets, leakage and function of the blood culture containers, and leakage and function of the blood collection tubes. All Blood Culture Holders met the acceptance criteria for fit, leakage, and function when connected to the Safety Blood Collection Sets and when used for blood collection into blood culture bottles and tubes.

Conclusion: The Blood Culture Holder can be used with blood culture bottles, blood culture vials, and the range of blood collection tube sizes from 13x75mm to 16x100mm.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol associated with medicine and healthcare.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 10, 2013

Greiner Bio-One North America, Incorporated C/O Ms. Judith Smith Principal Judi Smith, Limited Liability Company P.O. Box 103 BALDWIN MD 21013

Re: K122687

Trade/Device Name: Blood Culture Holder Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: II Product Code: JKA Dated: December 17, 2012 Received: December 20, 2012

Dear Ms. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Ms. Smith

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony Lo Re

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known): K122687

Device Name: Blood Culture Holder

Indication For Use:

The Blood Culture Holder is used to fill blood into blood culture bottles and tubes in routine venipuncture procedures. This device is to be used by properly trained healthcare professionals only in accordance with these instructions.

The SAFETY Blood Collection Set and/or Blood Collection Set + Blood Culture Holder are used in routine venipuncture procedures. The winged needle of the SAFETY Blood Collection Set is designed with a while shield, which can be activated to cover the needle immediately following blood collection to aid in the protection against accidental needlestick injury. The Blood Culture Holder is used to fill blood into blood culture bottles and tubes.

Prescription Use × (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)


(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Weihong Gu -S c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, cn=Weihong Gu -S, 0.9.2342.19200300.100.1.1=2000380818 2013.01.11 11:49:54 -05'00'

Division Sign-Off Office of Device Evaluation

510(k) K122687

Page 1 of 1

§ 862.1675 Blood specimen collection device.

(a)
Identification. A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.(b)
Classification. Class II.