K Number
K122537

Validate with FDA (Live)

Date Cleared
2013-04-12

(235 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is intended for use by a qualified physician for ultrasound evaluation of Fetal, Abdominal, Intra-operative, Intra-operative (Neuro), Laparoscopic, Pediatric, Neonatal Cephalic, Trans-rectal & vaginal, Musculo-skeletal, Gynecological, Cardiac-(Adult, Neonatal, Pediatric), TEE & Peripheral Vascular.
The device is not indicated for Ophthalmic applications.

Device Description

The Prosound Alpha 7 Ver. 6.0 Diagnostic Ultrasound System is a full feature imaging and analysis system. It consist of a mobile console that provides acquisition, processing and display capability. The user interface includes a computer type keyboard, specialized controls and a display.

AI/ML Overview

The provided document is a 510(k) summary for the Hitachi Aloka Medical Prosound Alpha 7 Ver. 6.0 Diagnostic Ultrasound System. It primarily focuses on demonstrating substantial equivalence to a predicate device and specifying the intended uses for various transducers. It does not contain acceptance criteria for device performance or a study demonstrating that the device meets such criteria.

The document states:

  • "I. Non-clinical Tests: The device and its transducers have been evaluated for acoustic output, biocompatibility, cleaning & disinfection effectiveness, electromagnetic compatibility, as well as electrical and mechanical safety, and have been found to conform with applicable medical device safety standards."
  • "2. Clinical Tests: None Required."

Given this, I cannot extract specific acceptance criteria related to clinical performance or a study proving the device meets those criteria from the provided text.

Therefore, the requested table and information cannot be fully populated as the core information (acceptance criteria and a study demonstrating meeting them) is not present in the provided 510(k) summary for clinical performance.

However, I can provide the non-clinical acceptance criteria that were met.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryReported Device Performance
Non-clinical Performance
Acoustic OutputConforms with applicable medical device safety standards.
BiocompatibilityConforms with applicable medical device safety standards.
Cleaning & Disinfection EffectivenessConforms with applicable medical device safety standards.
Electromagnetic CompatibilityConforms with applicable medical device safety standards.
Electrical SafetyConforms with applicable medical device safety standards.
Mechanical SafetyConforms with applicable medical device safety standards.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • The document states "Clinical Tests: None Required." Therefore, there is no information on a sample size for a clinical test set or data provenance from such a study. For non-clinical tests, specific sample sizes (e.g., number of units tested for acoustic output) are not detailed in this summary, but implied to be sufficient to demonstrate conformity to standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • No clinical test set was used, so no experts were involved in establishing ground truth for a test set in the context of clinical performance.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • No clinical test set was used, so no adjudication method is applicable.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC study was mentioned, nor is this device described as having AI capabilities.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • This device is a Diagnostic Ultrasound System; it does not appear to be an algorithm-only device, and no standalone performance study in that context is mentioned.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • No clinical ground truth was established as clinical tests were not required. The "ground truth" for non-clinical aspects was defined by applicable safety and performance standards.

8. The sample size for the training set

  • Not applicable. This document pertains to regulatory submission for a diagnostic ultrasound system, not typically an AI-based system that requires a training set in the machine learning sense.

9. How the ground truth for the training set was established

  • Not applicable. (See point 8)

{0}------------------------------------------------

Hitachi Aloka Medical, Ltd.

Prosound Alpha 7 Ver. 6.0

510(K) K22537

APR 1 2 2013

510(k) Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92

.

Section a):
1. Submitter:Hitachi Aloka Medical Co., Ltd., 10 Fairfield Boulevard, Wallingford, CT 06492
Contact Person:Angela Van Arsdale, RA/QA Mngr.,Tel: (203)269-5088 Ext. 346, Fax: 203-269-6075
Date Prepared:7/10/12
2. Device Name:Prosound Alpha 7 Ver. 6.0 Diagnostic Ultrasound SystemUltrasonic Pulsed Doppler Imaging System, 21 CFR 892.1550 , 90 IYNDiagnostic Ultrasound Transducer, 21 CFR 892.1570, 90 ITXUltrasonic Pulsed Echo Imaging System., 21 CFR 892.1560, 90 IYO
3. Substantially Equivalent devices:Aloka SSD- Alpha 7 Diagnostic Ultrasound System K072285 for probes and system. Aloka SSD-5500 (K032875), Aloka Prosound F75 (K110207), and Aloka SSD-5000 (K012080) for expanded indications.
4. Device Description:The Prosound Alpha 7 Ver. 6.0 Diagnostic Ultrasound System is a full feature imaging and analysis system. It consist of a mobile console that provides acquisition, processing and display capability. The user interface includes a computer type keyboard, specialized controls and a display.
5. Indications for Use:The device is intended for use by a qualified physician for ultrasound evaluation of Fetal, Abdominal, Intra-operative, Intra-operative (Neuro), Laparoscopic, Pediatric, Neonatal Cephalic, Trans-rectal & vaginal, Musculo-skeletal, Gynecological, Cardiac-(Adult, Neonatal, Pediatric), TEE & Peripheral Vascular.The device is not indicated for Ophthalmic applications.
6.Comparison w/ Predicate Device:The Prosound Alpha 7 Ver. 6.0 is technically comparable and substantially equivalent to the current Aloka SSD- Alpha 7 -(K072285). It has the same technological characteristics, key safety and effectiveness features, and has the same uses and basic operating modes as the predicate device.
Section b):
I. Non-clinical Tests:The device and its transducers have been evaluated for acoustic output, biocompatibility, cleaning & disinfection effectiveness, electromagnetic compatibility, as well as electrical and mechanical safety, and have been found to conform with applicable medical device safety standards.
2. Clinical Tests:None Required.
3. Conclusion:Intended uses and other key features are consistent with traditional clinical practices, FDA guidelines and established methods of patient examination. The design, development and quality process of the manufacturer conforms to 21 CFR 820, ISO 9001:2000 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effectiveness performance. Therefore, it is the opinion of Hitachi Aloka Medical, Ltd. that the Prosound Alpha 7 Ver. 6.0 Diagnostic Ultrasound System and its transducers are equivalent with respect to safety and effectiveness to its predicate and other currently cleared Hitachi Aloka Medical systems.

//

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

April 12, 2013

Hitachi Aloka Medical, Ltd. (Hitachi Aloka Medical AMERICA) % Mr. Michael S. Ogunleye 510(k) Program Manager/Medical Lead Auditor TUV Rheinland of North America 12 Commerce Road NEWTOWN CT 06470

Re: K122537

Trade/Device Name: Prosound Alpha 7 Ver. 6.0 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: January 4, 2013 Received: March 13, 2013

Dear Mr. Ogunleye:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with Prosound Alpha 7 Ver. 6.0 Diagnostic Ultrasound System, as described in your premarket notification:

Transducer Model Number

UST-533UST-534
UST-536UST-547
UST-567UST-675
UST677P (L&T)UST-678 (C&L)
UST-984-5ASU-1010

{2}------------------------------------------------

ASU-1012ASU-1013
UST-2265-2UST-2266-5
UST-5045P-3.5UST-5293-5
UST-5296UST-5411
UST-5412UST-5534T-7.5
UST-5536-7.5UST-5543
UST-5548UST-5550
UST-5712UST-5713T
UST-9101-7.5UST-9104-5
UST-9115-5UST-9118
UST-9120UST-9128
UST-9130UST-9132 I/T
UST-9133UST-9135P
UST-9136UUST-9146 I/T
UST-52101UST-52105
UST-52108UST-52109
UST-52110SUST-52114P
UST-52119SUST-52120S
UST-52121SUST-52124

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21

{3}------------------------------------------------

Page 3 – Mr. Ogunleye

CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.

Sincerely yours,

Mi

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure(s)

{4}------------------------------------------------

510(k) Number (if known): K122537

Device Name:

Prosound Alpha 7 Ver. 6.0 Diagnostic Ultrasound System

Indications for Use:

The device is intended for use by a qualified physician for ultrasound evaluation of Fetal. Abdominal. Intra-operative (Neuro). Laparoscopic. Pediatric. Neonatal Cephalic. Transrectal & vaginal. Musculoskeletal. Gynecological. Cardiac (Adult. Neonatal. Pediatric). TEE & Peripheral Vascular.

The device is not intended for ophthalmic applications.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

Page 1 of 50

Page 001 of 50

{5}------------------------------------------------

1.3.1

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0

intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follow~
Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPNote 1Note 2
AbdominalPPPPNote 1Note 2
Intra-operative (Specify)PPPPNote 1
Intra-operative (Neuro), Burr HoleNNNNNote 1
Inter-operative, LaparoscopicNNNNNote 1
PediatricNNNNNote 1
Fetal ImagingSmall Organ (Specify)PPPPNote 1Note 2
& OtherNeonatal CephalicPPPPNote 1Note 2
Adult Cephalic
Trans-rectalPPPPNote 1
Trans-vaginalPPPPNote 1
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)PPPPNote 1Note 2
Musculo-skeletal (Superfical)
Other : GynecologicalPPPPNote 1Note 2
Cardiac AdultPPPPNote 1Note 2
Cardiac Adult, TEENNNNNote 1Note 2
CardiacCardiac Pediatric, TEENNNNNote 1Note 2
Cardiac Neonatal, TEENNNNNote 1Note 2
Cardiac, PediatricNNNNNote 1Note 2
Cardiac, NeonatalNNNNNote 1Note 2
Intra-cardiac
Other (Specify)
PeripheralPeripheral vascularPPPPNote 1Note 2
VesselOther (Specify)

= new indication for system; P= Previously cleared K072285, E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

Applications:- Intra-operative- (Iiver, III GB, EroBH WD
Applications:- Intra-operative- (liver, pancreas, gall bladder, ....), Small Organ- (breast, testes, thyroid....).

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

{6}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-533

Intended Use: Diagnostic ultrasound imaging or fluid flow anaiysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track IOnly)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)EEEENote 1
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
CardiacCardiac Adult, TEE
Cardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralPeripheral vascular
VesselOther (Specify)

ation for system; E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

Applications:- intra-operative- (liver, pancreas, gall bladder, ... .),

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801 109)

{7}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-534

..

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)EEEENote 1
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralPeripheral vascular
VesselOther (Specify)

N = new indication for system; E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

Applications:- Intra-operative- (liver, pancreas, gall bladder, ....)

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Page 004 of 50

44

{8}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-536

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)EEEENote 1
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)EEEENote 1
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascularEEEENote 1
Other (Specify)

N = new indication for system; E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

Applications:- Intra-operative- (liver, pancreas, gall bladder, ....), Small Organ- (breast, testes, thyroid... . . ).

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 005 of 50

45

{9}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-547

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)EEEENote 1
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)EEEENote 1
Neonatal CephalicEEEENote 1
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

new indication for system; E = added to Alpha 7 under Appendix E,

Combined: Note I : B/M, B/PWD, M/CD, B/CD/PWD

Applications:- Intra-operative- (liver, pancreas, gall bladder, ... .), Small Organ- (breast, testes, thyroid... . .).

(BLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 006 of 50

46

{10}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-567

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

General(Track 1Only)Clinical ApplicationSpecific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)EEEENote 1
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)EEEENote 1
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascularEEEENote 1
Other (Specify)

N = new indication for system; E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD Applications:- Small Organ- (breast, testes, thyroid ... ).

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 007 of 50

47

24

.

{11}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-675

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation
General(Track IOnly)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal ImagingSmall Organ (Specify)
& OtherNeonatal Cephalic
Adult Cephalic
Trans-rectalPPPPNote I
Trans-vaginalPPPPNote I
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralPeripheral vascular
VesselOther (Specify)

N indication for system; P= Previously cleared K072285

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 008 of 50

25

. :

{12}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST677P (L&T)

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationCombined(Specify)Other(Specify)
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDoppler
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Small Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectalEEEENote 1
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent)
Musculo-skeletal (Superfical)
Other : Gynecological
CardiacCardiac Adult
Cardiac Adult, TEE
Cardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system; E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD Applications:- Intra-operative- (liver, pancreas, gall bladder, ....)

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 009 of 50

{13}------------------------------------------------

.

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-678 (C&L)

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectalEEEENote 1
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system; E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD Applications :- Intra-operative- (liver, pancreas, gall bladder, ....)

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

.27

{14}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-984-5

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalEEEEENote 1
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalEEEENote 1
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : GynecologicalEEEENote 1
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system; E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

{15}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: ASU-1010

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther(Specify)
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)
OphthalmicOphthalmic
FetalPPPPNote 1
AbdominalPPPPNote 1
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : GynecologicalPPPPNote 1
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system; P= Previously cleared K072285,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Page 012 of 50

{16}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: ASU-1012

.

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalEEEENote 1
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalEEEENote 1
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : GynecologicalEEEENote 1
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system; E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANQTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 013 of 50

રી

{17}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: ASU-1013

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical Application
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)EEEENote I
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)EEEENote I
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascularEEEENote I
Other (Specify)

N = new indication for system; E = added to Alpha 7 under Appendix E

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

.

Applications:- Small Organ- (breast, testes, thyroid ... ).

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) . .

Page 014 of 50

{18}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-2265-2

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac AdultE
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system; E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evatuation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 015 of 50

{19}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-2266-5

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac AdultE
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascularE
Other (Specify)

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

N = new indication for system; E = added to Alpha 7 under Appendix E,

Combined: Note I: B/M, B/PWD, M/CD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 016 of 50

કુંડ

{20}------------------------------------------------

Prosound Alpha 7 Ver. 6.0

. İ

.......

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-5045P-3.5

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
AbdominalEEEENote 1
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : GynecologicalEEEENote 1
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the

N = new indication for system; E = added to Alpha 7 under Appendix E,

Combined: Note I : B/M, B/PWD, M/CD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

{21}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0

Transducer: UST-5293-5

ATTORNEY: 601-522-3

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
CardiacCardiac Adult, TEENNNNNote 1Note 2
Cardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system; P= Previously cleared-K032875

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD , Note 2: B/CWD, B/CD/CWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 018 of 50

{22}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-5296

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, NeonatalNNN.NNote 1Note 2
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system; P= Previously cleared K072285,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD , Note 2: B/CWD, B/CD/CWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 019 of 50

{23}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-5411

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human bedy as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)EEEENote 1
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)EEEENote 1
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascularEEEENote 1
Other (Specify)

E = added to Alpha 7 under Appendix E,

Combined: Note I: B/M, B/PWD, M/CD, B/CD/PWD Applications:- Small Organ- (breast, testes, thyroid ... . ).

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 020 of 50

60

{24}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-5412

T

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)PPPPPNote 1Note 2
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)PPPPPNote 1Note 2
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascularPPPPPNote 1Note 2
Other (Specify)

cviously cleared K072285,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD Applications:- Small Organ- (breast, testes, thyroid ... ).

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 021 of 50

{25}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-5534T-7.5

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)EEEENote 1
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)EEEENote 1
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascularEEEENote 1
Other (Specify)

E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

A pplications:- Intra-operative- (liver, pancreas, gall bladder, ... ), Small Organ- (breast, testes, thyroid.... ).

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 2 I CFR 801.109)

Page 022 of 50

62

{26}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-5536-7.5

.

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, LaparoscopicNNNNNote 1
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system; P= Previously cleared-K 032875

Combined: Note I : B/M, B/PWD, M/CD, B/CD/PWD

Application: Intra-operative: (liver, pancreas, gall bladder, ... )

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 023 of 50

{27}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-5543

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)EEEENote I
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)EEEENote I
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascularEEEENote I
Other (Specify)

= added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD Applications :- Small Organ- (breast, testes, thyroid ... ).

.

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) .

Prescription Use (Per 21 CFR 801.109)

{28}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-5548

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)EEEENote 1
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascularEEEENote 1
Other (Specify)

E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD Applications:- Small Organ- (breast, testes, thyroid ... ).

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 025 of 50

{29}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-5550

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, LaparoscopicNNNNNote I
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system;

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD Applications:- Small Organ- (breast, testes, thyroid ... ).

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Page 026 of 50

દર્ભ

બંત્ર

{30}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer; UST-5712

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)EEEENote 1
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)EEEENote 1
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascularEEEENote 1
Other (Specify)

E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD Applications:- Small Organ- (breast, testes, thyroid ... ).

M

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 027 of 50

67

{31}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-5713T

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)EEEENote 1
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)EEEENote 1
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascularEEEENote 1
Other (Specify)

E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

Applications:- Intra-operative- (liver, pancreas, gal! bladder, .... ), Small Organ- (breast, testes, thyroid... ).

(PLEASE DO NOT WRITE BELQW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Page 028 of 50

68

નર્ડ

{32}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-9101-7.5

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
AbdominalPPPPNote 1
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
PediatricNNNNNote 1
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = acw indication for system; P= Previously cleared K072285, E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

Applications:- Intra-operative- (liver, pancreas, gall bladder, ... .), Small Organ- (breast, testes, thyroid.... ).

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 029 of 50

{33}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-9104-5

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)EEEENote 1
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal ImagingSmall Organ (Specify)EEEENote 1
& OtherNeonatal CephalicEEEENote 1
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralPeripheral vascular
VesselOther (Specify)

E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD Applications:- Intra-operative- (liver, pancreas, gall bladder, ....)

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 030 of 50

{34}------------------------------------------------

··

,

Prosound Alpha 7 Ver. 6.0

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-9115-5

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
AbdominalEEEENote 1
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric, Abdominal
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other: GynecologicalEEEENote 1
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system;

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 031 of 50

.

. .

{35}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-9118

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of OperationOther(Specify)
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)
OphthalmicOphthalmic
FetalPPPPNote I
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalPPPPNote I
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : GynecologicalPPPPNote I
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

P= Previously cleared K072285,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD Applications:- Intra-operative- (liver, pancreas, gall bladder, ... ), Small Organ- (breast, testes, thyroid... . .

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

· Page 032 of 50

4g

{36}------------------------------------------------

System; Prosound Alpha 7 Ver, 6.0 Transducer: UST-9120

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)PPPPNote I
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal CephalicPPPPNote I
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

P= Previously cleared K072285,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD Applications:- Intra-operative- (liver, pancreas, gall bladder, ... )

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 033 of 50

73

{37}------------------------------------------------

·

Г

.

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-9128

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1 Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Fetal Imaging& OtherAbdominalEEEENote 1
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric, Abdominal
Small Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
CardiacCardiac Adult
Cardiac Adult, TEE
Cardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system;

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

CLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 034 of 50

{38}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-9130

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalPPPPNote 1Note 2
AbdominalPPPPNote 1Note 2
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : GynecologicalPPPPNote 1Note 2
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

Intended Use: Diagnostic ultr d imsging or fluid Aou onalusis

P= Previously cleared K072285

Combined: Note I : B/M, B/PWD, M/CD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

.

{39}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-9132 I/T

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)EEEENote I
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)EEEENote I
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascularEEEENote 1
Other (Specify)

E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

Applications:- Intra-operative- (liver, pancieas, gall bladder, ... .), Small Organ- (breast, testes, thyroid. .. . ).

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Page 036 of 50

76

{40}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-9133

.

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Fetal Imaging& OtherAbdominalEEEENote I
Intra-operative (Specify)EEEENote I
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Small Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

= added to Alpha 7 under Appendix E, ה

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD Applications :- Intra-operative- (liver, pancreas, gall bladder, ... )

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801 109)

{41}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-9135P

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalEEEENote 1
AbdominalEEEENote 1
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : GynecologicalEEEENote 1
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 038 of 50

55

{42}------------------------------------------------

.

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-9136U

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
FetalEEEENote 1
AbdominalEEEENote 1
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal ImagingSmall Organ (Specify)EEEENote 1
& OtherNeonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : GynecologicalEEEENote 1
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralPeripheral vascular
VesselOther (Specify)

E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 039 of 50

{43}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-9146 I/T

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)EEEENote 1
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD Applications:- Intra-operative- (liver, pancreas, gall bladder, ... .)

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 040 of 50

{44}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-52101

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Small Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
CardiacCardiac AdultPPPPNote 1Note 2
Cardiac Adult, TEE
Cardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

P= Previously cleared K072285

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801,109)

Page 041 of 50

81

{45}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver, 6.0 Transducer: UST-52105

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac AdultEEEEENote 1Note 2
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

E = added to Alpha 7 under Appendix E,

Combined: Note I: B/M, B/PWD, M/CD, B/CD/PWD Applications:- Intra-operative- (liver, pancreas, gall bladder, ....), Small Organ- (breast, thyroid... ).

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 042 of 50

{46}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-52108

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, PediatricNNNNNNote 1Note 2
Cardiac, NeonatalNNNNNNote 1Note 2
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

Applications:- Intra-operative- (liver, pancreas, gall bladder, .... ), Small Organ- (breast, testes, thyroid ... . . ).

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Page 043 of 50

હ્યું છે

{47}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-52109

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Mode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, LaparoscopicNNNNNote 1
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

new indication for system; P= Previously cleared-K032875

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

Applications:- Intra-operative- (liver, pancreas, gall bladder, ....)

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 044 of 50

84

{48}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-52110S

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

.

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
& OtherNeonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEENNNNNNote 1Note 2
Cardiac Neonatal, TEENNNNNNote 1Note 2
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system;

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD , Note 2: B/CWD, B/CD/CWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

{49}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-52114P

Intended Use: Diagnostic ultrasound imaging or Nuid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr HoleNNNNNote 1
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

(PLEASE DQ NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801-109)

{50}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-52119S

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEENNNNNNote 1Note 2
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system , P= Previously cleared-K110207

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD , Note 2: B/CWD, B/CD/CWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 047 of 50

{51}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-52120S

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, NeonatalNNNNNote 1Note 2
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system; P= Previously cleared K032875, E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD , Note: 2 : B/CWD, B/CD/CWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 048 of 50

{52}------------------------------------------------

DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-52121S

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track 1Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, Pediatric
Cardiac, NeonatalNNNNNote 1Note 2
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

N = new indication for system; P= Previously cleared K110207, E = added to Alpha 7 under Appendix E,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD , Note 2: B/CWD, B/CD/CWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Page 049 of 50

ים יו

{53}------------------------------------------------

System: Prosound Alpha 7 Ver. 6.0 Transducer: UST-52124

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Clinical Annlication

General(Track 1Only)Clinical ApplicationSpecific(Tracks 1 & 3)BMPWDCWDMode of Operation
ColorDopplerCombined(Specify)Other(Specify)
OphthalmicOphthalmic
Fetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro), Burr Hole
Inter-operative, Laparoscopic
Pediatric
Fetal Imaging& OtherSmall Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Convent.)
Musculo-skeletal (Superfical)
Other : Gynecological
Cardiac Adult
Cardiac Adult, TEE
CardiacCardiac Pediatric, TEE
Cardiac Neonatal, TEE
Cardiac, PediatricNNNNNote 1Note 2
Cardiac, Neonatal
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vascular
Other (Specify)

new indication for system; P= Previously cleared K110207,

Combined: Note 1: B/M, B/PWD, M/CD, B/CD/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

$\sinh(x)$

510(k) # K122537

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

Page 050 of 50

90

N/A