K Number
K122529

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-11-16

(88 days)

Product Code
Regulation Number
888.3520
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

These devices are indicated for patients with:

  • Painful and/or disabling knee joints due to osteoarthritis or traumatic arthritis.
  • Previous tibial condyle or plateau fractures with loss of anatomy or function.
  • Varus or valgus deformities.
  • Revision of previous arthroplasty procedures.
    These devices are indicated for cemented use only.
    The Zimmer Unicompartmental Knee System is designed for use when load bearing ROM is expected to be less than or equal to 155 degrees.
Device Description

The Zimmer Unicompartmental Knee System (ZUK) is a prosthesis that replaces only one compartment of the knee condyles. It is unconstrained in the anteroposterior and mediolateral directions and also allows unconstrained internal/external rotation between the femoral and tibial components. This movement is limited only by the ligaments and other soft tissues surrounding the device.

AI/ML Overview

The Zimmer Unicompartmental Knee System Vivacit-E Articular Surface is a medical device. The provided text describes the non-clinical performance and conclusions for this device, focusing on material characteristics and mechanical properties. There are no clinical performance studies described.

1. Table of Acceptance Criteria and Reported Device Performance

Property or CharacteristicAcceptance CriteriaReported Device Performance
Effect of prolonged aging on the wear performance of vitamin-E IT Liner(Implied) Wear performance should not be significantly affected by prolonged accelerated aging.Demonstrated that prolonged accelerated aging does not significantly affect wear performance of Vivacit-E material.
Delamination Resistance of Vivacit-E Ultra High Molecular Weight Polyethylene(Implied) The material should be delamination resistant.Demonstrated that Vivacit-E material is delamination resistant.
Wear Testing of Zimmer Unicompartmental Knee (ZUK) Vivacit-E UHMWPE Articular Surfaces under Load and Motion Curves from the ISO-14243 Standard(Implied) Wear characteristics should be sufficient to survive expected in vivo loading conditions.Demonstrated that the wear characteristics of the Zimmer Unicompartmental Knee Vivacit-E articular surfaces are sufficient to survive expected in vivo loading conditions.
Fatigue Evaluation of the Zimmer Unicompartmental Knee (ZUK) Vivacit-E UHMWPE Articular Surface Locking Mechanism(Implied) The locking mechanism should have adequate resistance to disassembly.Demonstrated adequate resistance of the modular articular surfaces to disassembly.
ZUK VE Posterior Edge Crush Fatigue Strength Evaluation(Implied) Posterior edge crush fatigue strength should be sufficient to survive expected in vivo loading conditions.Demonstrated posterior edge crush fatigue strength is sufficient to survive expected in vivo loading conditions.
Human Factors Comparison of Manual Assembly of the Zimmer Unicompartmental Knee (ZUK) Vivacit-E UHMWPE Articular Surfaces(Implied) Assembly force required to insert the articular surfaces should be appropriate for users (e.g., within ergonomic limits for surgeons).Demonstrated that assembly force required to insert the Zimmer Unicompartmental Knee Vivacit-E articular surfaces is appropriate for users.

2. Sample Size Used for the Test Set and Data Provenance

The provided text details non-clinical performance and conclusions based on bench testing. Therefore, there is no human "test set" or associated patient data from which to derive sample size or data provenance (country of origin, retrospective/prospective). The "test set" in this context refers to the physical samples of the device components and materials subjected to laboratory testing. The document does not specify the number of samples used for each bench test.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

As this is a non-clinical, bench-testing study, the concept of "experts" establishing ground truth in the context of clinical interpretation or diagnosis does not apply. The ground truth for these tests is based on the physical and mechanical properties of the device components as measured by standardized testing methods and engineering principles. The experts involved would be engineers, material scientists, and quality assurance personnel who design, conduct, and interpret these tests, but their number and specific qualifications are not detailed in this summary.

4. Adjudication Method for the Test Set

Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies involving interpretation of medical images or patient outcomes. Since this report describes non-clinical bench testing, no such adjudication method was used. The "results" are objective measurements against predefined engineering specifications and standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No MRMC comparative effectiveness study was done. The description explicitly states: "Clinical data and conclusions were not needed for this device." The study focuses on the physical properties and performance of the device's material and components through bench testing, not on human reader performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

This question relates to AI/algorithm performance. No AI/algorithm standalone performance study was done. The device is a physical knee prosthesis component, not an AI software or system.

7. Type of Ground Truth Used

The ground truth used for these non-clinical performance studies is derived from established engineering standards, material science principles, and functional requirements for knee prostheses. Results from bench tests (e.g., wear rate measurements, fatigue strength, delamination resistance, assembly force) are compared against these predetermined specifications and accepted industry benchmarks to determine if the device meets its performance objectives.

8. Sample Size for the Training Set

Similar to the "test set," the concept of a "training set" is not applicable here as this is a non-clinical, bench-testing study of a physical medical device. There is no machine learning model being trained.

9. How the Ground Truth for the Training Set Was Established

As there is no training set mentioned in the context of this non-clinical bench study, this question is not applicable.

{0}------------------------------------------------

K122529 ('/3)

Image /page/0/Picture/1 description: The image shows the logo for Zimmer. The logo consists of a circle with a stylized letter Z inside. Below the circle, the word "zimmer" is written in a stylized font.

NOV 1 6 2012

P.O. Box 708 Warsaw, IN 46581-0708 574 267-6131

Summary of Safety and Effectiveness

Sponsor:

Contact Person:

Date:

Trade Name:

Product Code / Device:

Regulation Number / Description:

Predicate Device:

Zimmer, Inc. P.O. Box 708 Warsaw, IN 46581-0708

Mark D. Warner Senior Specialist, Regulatory Affairs Telephone: (574)-372-4150 Fax: (574) 372-4605

August 17, 2012

Zimmer® Unicompartmental Knee System Vivacit-E® Articular Surface

HSX - Prosthesis, knee, femorotibial, nonconstrained, cemented, metal/polymer

OIY - Prosthesis, knee, patellofemorotibial, semiconstrained, cemented, polymer + additive / metal / polymer + additive

21 CFR § 888.3520 - Knee joint femorotibial metal/polymer non-constrained cemented prosthesis

21 CFR § 888.3560 - Knee joint, patellofemorotibial, polymer / metal / polymer, semi-constrained, cemented prosthesis

Zimmer Unicompartmental Knee System, manufactured by Zimmer, Inc. (K033363, cleared January 6, 2003)

Vivacit-E Vitamin E Highly Crosslinked Polyethylene Liners, manufactured by Zimmer, Inc. (K120370, cleared June 4, 2012)

{1}------------------------------------------------

K122S29(²/₃)

EITM Antioxidant Infused Technology, Manufactured by Biomet Manufacturing Corp. (K 100048, cleared March 9, 2010)

The Zimmer Unicompartmental Knee System (ZUK) is a prosthesis that replaces only one compartment of the knee condyles. It is unconstrained in the anteroposterior and mediolateral directions and also allows · unconstrained internal/external rotation between the femoral and tibial components. This movement is limited only by the ligaments and other soft tissues surrounding the device.

The Zimmer Unicompartmental Knee System is indicated for patients with:

  • Painful and/or disabling knee joints due to osteoarthritis or traumatic arthritis.
  • Previous tibial condyle or plateau fractures with loss of anatomy or function.
  • Varus or valgus deformities.
  • Revision of previous arthroplasty procedures.

These devices are indicated for cemented use only.

The Zimmer Unicompartmental Knee System is designed for use when load bearing ROM is expected to be less than or equal to 155 degrees.

The proposed Zimmer Unicompartmental Knee System components are similar or identical in intended use, materials, sterility, and performance characteristics to the predicate devices.

Clinical Performance and Conclusions: Clinical data and conclusions were not needed for this device.

Non-Clinical Performance and Conclusions:

Vivacit-E material characteristics for the Zimmer Unicompartmental Knee System (ZUK) are identical to the predicate Zimmer Vivacit-E Acetabular Polyethylene Liners (K120370). In contrast to conventional polyethylene, the Vivacit-E material is delamination resistant and exhibits a

Device Description:

Intended Use:

Comparison to Predicate Device:

Performance Data (Nonclinical and/or Clinical):

{2}------------------------------------------------

K122529(³/₃)

reduction in wear according to knee simulator bench testing.

Bench testing outlined in table 1, below, was conducted according to FDA guidance documents:

Property orCharacteristicTest Results
Effect of prolonged aging on the wearperformance of vitamin-E IT LinerDemonstrated that prolonged accelerated aging doesnot significantly affect wear performance of VivacitE material.
Evaluation of the DelaminationResistance of Vivacit-E Ultra HighMolecular Weight PolyethyleneDemonstrated that Vivacit E material is delaminationresistant.
Wear Testing of ZimmerUnicompartmental Knee (ZUK) Vivacit-EUHMWPE Articular Surfaces underLoad and Motion Curves from the ISO-14243 Standard, ReportDemonstrated that the wear characteristics of theZimmer Unicompartmental Knee Vivacit-E articularsurfaces are sufficient to survive expected in vivoloading conditions.
Fatigue Evaluation of the ZimmerUnicompartmental Knee (ZUK) Vivacit-EUHMWPE Articular Surface LockingMechanismDemonstrated adequate resistance of the modulararticular surfaces to disassembly.
ZUK VE Posterior Edge Crush FatigueStrength EvaluationDemonstrated posterior edge crush fatigue strengthis sufficient to survive expected in vivo loadingconditions.
Human Factors Comparison of ManualAssembly of the ZimmerUnicompartmental Knee (ZUK) Vivacit-EUHMWPE Articular SurfacesDemonstrated that assembly force required to insertthe Zimmer Unicompartmental Knee Vivacit-Earticular surfaces is appropriate for users.

Table 1: Bench Performance Testing Overview

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the emblem.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Letter Dated: November 16, 2012

Zimmer, Incorporated % Mr. Mark D. Warner Senior Specialist, Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581-0708

Re: K122529

Trade/Device Name: Zimmer® Unicompartmental Knee System Vivacit-E® Articular Surface

Regulation Number: 21 CFR 888.3520

Regulation Name: Knee joint femorotibial metal/polymer non-constrained cemented prosthesis

Regulatory Class: Class II Product Code: HSX, OIY Dated: August 17, 2012 Received: August 20, 2012

Dear Mr. Warner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

Page 2 - Mr. Mark D. Warner

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. 1

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Laurence D. Coyne

  • Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{5}------------------------------------------------

Traditional 510(k) Premarket Notification

Indications for Use

510(k) Number (if known): K122529

Device Name: Zimmer Unicompartmental Knee System

Indications for Use: .

These devices are indicated for patients with:

  • Painful and/or disabling knee joints due to osteoarthritis or traumatic arthritis.

  • Previous tibial condyle or plateau fractures with loss of anatomy or function.

  • Varus or valgus deformities.

  • Revision of previous arthroplasty procedures.

These devices are indicated for cemented use only.

The Zimmer Unicompartmental Knee System is designed for use when load bearing ROM is expected to be less than or equal to 155 degrees.

(Please do not write below this line - Continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

OR

Ais

(Division Sign-Off) Division of Orthopedic Devices KI22529 510(k) Number_

Prescription Use X (Per 21 CFR 801.109) Over-The-Counter Use (Optional Format 1-2-96)

§ 888.3520 Knee joint femorotibial metal/polymer non-constrained cemented prosthesis.

(a)
Identification. A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.