K Number
K122513

Validate with FDA (Live)

Date Cleared
2013-03-06

(201 days)

Product Code
Regulation Number
888.3045
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MAGNIFUSE® II Bone Graft is intended for use as a bone graft substitute in bony voids or gaps of the skeletal system (i.e., posterolateral spine and pelvis) not intrinsic to the stability of the bony structure. The voids or gaps may be surgically created defects or defects created by traumatic injury to the bone. MAGNIFUSE® II Bone Graft is resorbed/remodeled and replaced by host bone during the healing process.

Device Description

MAGNIFUSE® II Bone Graft is assembled by the clinician at the time of the procedure using the supplied human bone allograft tissue matrix mixed 1:1 with autograft tissue. The mixture is packed into a polyglycolic acid (PGA) resorbable mesh bag with the supplied injection molded plastic spatula, funnel, and plunger. This product enables clinicians to generate a construct having a particular physical form and handling property. No additional carrier is added to the allograft material. This MAGNIFUSE® II Bone Graft product was prepared from human bone tissue recovered from a cadaveric donor using aseptic surgical techniques and microbiologically tested during recovery. As a biological material, some variations in the product should be expected in both handling and appearance. The final product in packaged form was tested for sterility according to the procedures in the current U.S. Pharmacopoeia USP standard <71>.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Key Takeaway: This submission is for a bone graft substitute, and its acceptance hinges on demonstrating substantial equivalence to a predicate device (GRAFTON® II eDBM) through non-clinical testing. The "acceptance criteria" here are implicitly tied to showing comparable performance to the predicate and established safety profiles. The study primarily focuses on fusion rates and host tolerance in a validated animal model.


1. Table of Acceptance Criteria and Reported Device Performance

Given that this is a 510(k) submission for a bone graft substitute demonstrating substantial equivalence, the "acceptance criteria" are not explicitly quantitative thresholds like you might find for an AI diagnostic device (e.g., AUC > 0.9). Instead, they are performance characteristics expected to be comparable to the predicate device to establish substantial equivalence.

Acceptance Criterion (Implicit)Reported Device Performance (MAGNIFUSE® II Bone Graft)
Bone Formation Capability (Fusion Rate)Exhibited comparable fusion rates to the autograft group in a rabbit posterolateral lumbar fusion study.Context: The predicate (MAGNIFUSE® Bone Graft manual) had a fusion rate equal to autograft, and a higher radiographic fusion rate. The subject device is shown to be comparable to autograft, thus implicitly comparable to the predicate for fusion rate.
Host Tolerance/Adverse ReactionsAll animals tolerated the graft material well and exhibited remodeling of the graft site over the duration of the study.
Osteoinductivity (of allograft subcomponent)The allograft tissue subcomponent is identical in form and processing to the predicate MAGNIFUSE® device (cleared under eDBM K082165). Its processing method consistently produces demineralized bone matrix that is osteoinductive in an athymic rat assay, confirmed via ongoing testing using a five-point linear scale (0-4) at 28 days post-implantation.Note: "Bone formation in the athymic rat surrogate assay should not be interpreted as a predictor of clinical performance."
Viral Inactivation (of allograft subcomponent)Proprietary processing steps (demineralizing acid soaks, alcohol soaks, dehydration for allograft fibers; alcohol soaks, supercritical CO2 for cortical chips) have been shown and validated to inactivate viruses including HIV-1, hepatitis B virus, hepatitis C virus, CMV, and Polio virus. These are identical to methods used for GRAFTON® DBM products (K051195) and cleared in K061982.
Product Preparation Instruction (Handling)Addition of autograft to mesh bag using provided spatula, funnel, and syringe (described as a characteristic, implies it functions as intended).
Operating PrincipleIdentical to predicate device (K082615).
Basic DesignIdentical to predicate device (K082615).
Performance (Overall Equivalence to Predicate K082615)Identical to predicate device (K082615).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Non-Clinical Study: The document states "A Rabbit posterlateral lumbar fusion study was conducted..." but does not specify the exact number of rabbits used in the study. It refers to "All animals tolerated the graft material well," implying a group of subjects.
  • Data Provenance: The data comes from a prospective animal study (rabbit posterolateral lumbar fusion study). The country of origin is not explicitly stated, but Medtronic Sofamor Danek USA is based in Memphis, Tennessee, suggesting a U.S.-based study.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • The non-clinical study involves animal models (rabbits and athymic rats) where "ground truth" would likely be established through histological analysis, radiography evaluation, and macroscopic observation.
  • The document does not specify the number of experts or their qualifications (e.g., veterinary pathologists, orthopedic surgeons evaluating radiographs/histology) who assessed the outcomes of the rabbit study or the athymic rat assay. It refers to "scoring bone formation" in the rat assay, implying expert evaluation.

4. Adjudication Method for the Test Set

  • The document does not describe an explicit adjudication method (like 2+1 or 3+1 consensus) for assessing the outcomes of the animal studies. Assessments like "comparable fusion rates" and "tolerated the graft material well" suggest evaluation by researchers, but the specific process for resolving discrepancies is not detailed.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. An MRMC comparative effectiveness study is designed to evaluate human reader performance with and without AI assistance, typically in diagnostic imaging. This submission is for a bone graft substitute and relies on non-clinical (animal) studies to demonstrate substantial equivalence, not human reader performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Not applicable. This device is a bone graft substitute, not an algorithm or AI. Therefore, the concept of "standalone performance" for an algorithm doesn't apply. The performance evaluated is the biological performance of the graft material itself.

7. The Type of Ground Truth Used

  • Animal Model Outcomes (Histology, Radiography, Macroscopic Observation):
    • For the rabbit posterolateral lumbar fusion study, the ground truth for "fusion rates" would likely be established by a combination of radiographic assessment and histological evaluation of the spinal segments. "Tolerance" would be based on clinical observation and potentially gross pathology.
    • For the athymic rat assay, the ground truth for "osteoinductivity" is established by "scoring bone formation at 28 days post implantation" using a "five-point linear scale (0, 1, 2, 3, 4)." This is a histological assessment.

8. The Sample Size for the Training Set

  • The concept of a "training set" is relevant to machine learning algorithms. This submission is for a biological device (bone graft). Therefore, there is no "training set" in the context of AI or algorithms. The "training" for such a device is implicitly in the development and refinement of its manufacturing processes and material composition.

9. How the Ground Truth for the Training Set Was Established

  • As there is no "training set" in the AI/algorithm sense, this question is not applicable. The "ground truth" for the device's design and manufacturing is established through extensive material science, biological compatibility testing, and historical performance of similar predicate devices, which are all part of the product development and regulatory submission process.

{0}------------------------------------------------

177511 113

MAR 0 6 2013

510(K) Summary

SUBMITTER NAME AND ADDRESS: l.

Medtronic Sofamor Danek USA 1800 Pyramid Place Memphis, Tennessee 38132 (901) 396-3133 Telephone: (901) 346-9738 Fax: Establishment Registration: 1030489

Kelly Anglin Sr. Regulatory Affairs Specialist

February 14, 2013

II. PROPOSED PROPRIETARY TRADE NAME:

DEVICE CLASSIFICATION NAME:

REGULATION NUMBER: REGULATION NAME:

CONTACT PERSON:

DATE PREPARED:

REGULATORY CLASS: PRODUCT CODE:

MAGNIFUSE® II BONE GRAFT

Filler, bone void, calcium compound 21 CFR 888.3045 Resorbable calcium salt bone void filler device MQV, MBP

IDENTIFICATION OF LEGALLY MARKETED DEVICES: III. Grafton® II eDBM (K082615, SE 10/16/2008)

IV. DEVICE DESCRIPTION:

MAGNIFUSE® II Bone Graft is assembled by the clinician at the time of the procedure using the supplied human bone allograft tissue matrix mixed 1:1 with autograft tissue. The mixture is packed into a polyglycolic acid (PGA) resorbable mesh bag with the supplied injection molded plastic spatula, funnel, and plunger. This product enables clinicians to generate a construct having a particular physical form and handling property. No additional carrier is added to the allograft material.

This MAGNIFUSE® II Bone Graft product was prepared from human bone tissue recovered from a cadaveric donor using aseptic surgical techniques and microbiologically tested during recovery. As a biological material, some variations in the product should be expected in both handling and appearance. The final product in packaged form was tested for sterility according to the procedures in the current U.S. Pharmacopoeia USP standard <71>.

V. INDICATIONS FOR USE:

{1}------------------------------------------------

MAGNIFUSE® II Bone Graft is intended for use as a bone graft substitute in bony voids or gaps of the skeletal system (i.e., posterolateral spine and pelvis) not intrinsic to the stability of the bony structure. The voids or gaps may be surgically created defects or defects created by traumatic injury to the bone. MAGNIFUSE® II Bone Graft is resorbed/remodeled and replaced by host bone during the healing process.

CharacteristicSubject Device:MAGNIFUSE® II BoneGraftPredicate DeviceGRAFTON® II eDBM
ProductPreparationInstructionAddition of autograftto mesh bag usingprovided spatula,funnel and syringeK082615(SE 10/16/2008)
OperatingPrincipleIdenticalK082615(SE 10/16/2008)
Basic DesignIdenticalK082615(SE 10/16/2008)
PerformanceIdenticalK082615(SE 10/16/2008)

SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS: VI.

SUMMARY OF NON-CLINICAL TESTING VIII.

A Rabbit posterlateral lumbar fusion study was conducted to evaluate the bone formation capability of MAGNIFUSE® II Bone Graft. The autograft group fusion rates were consistent with the literature and previous laboratory results in this model, suggesting a valid fusion test. The predicate MAGNIFUSE ® Bone Graft manual fusion rate was equal to autograft rate but the radiographic fusion rate was higher with MAGNIFUSE ® Bone Graft. The fusion rates for the subject MAGNIFUSE ® Il Bone Graft exhibited comparable fusion rates to the autograft group. All animals tolerated the graft material well and exhibited remodeling of the graft site over the duration of the study.

The allograft tissue supplied as a subcomponent of the subject MAGNIFUSE ® Il Bone Graft device is identical in form and processing to the predicate MAGNIFUSE ® device cleared under eDBM K082165 (S.E. 10/16/2008). Methods employed to ensure osteoinductivity and viral inactivation of the allograft component are described in further detail in K082615 and subsequent submissions that were cleared by the agency. The allograft tissue subcomponent will be processed via a proprietary processing method that has been shown to consistently produce demineralized bone matrix that is osteoinductive in an athymic rat assay. As the tissue processing is identical to the predicate MAGNIFUSE® device, process consistency for MAGNIFUSE ® II will be confirmed via ongoing testing of the MAGNIFUSE ® finished product for osteoinductivity in this validated athymic rat assay utilizing a five-point linear scale (0, 1, 2, 3, 4) to score

{2}------------------------------------------------

bone formation at 28 days post implantation*. Bone formation in the athymic rat surrogate assay should not be interpreted as a predictor of clinical performance. A full assessment of osteoinductivity of the MAGNIFUSE ® device can be found in K082615. Viral inactivation of MAGNIFUSE® II allograft fibers includes proprietary processing steps of demineralizing acid soaks followed by alcohol soaks and dehydration, as established for the GRAFTON® DBM products (K051195).

Viral inactivation of the cortical chips is done by alcohol soaks and by dehydration using supercritical CO2 established for viral inactivation cleared in (K061982). These processing steps have been shown and validated to inactivate viruses including; HIV-1; hepatitis B virus; hepatitis C virus, CMV, and Polio virus. These processes further reduce the risk of disease transmission via the use of this product beyond the protection provided by donor testing and screening procedures.

  • Edwards etal; Osteoinduction of Human Demineralized Bone: Characterization in a Rat Model.

Clinical Orthopaedics, December 1998, Vol 357.

  • VII. CONCLUSION:
    The design features for the subject MAGNIFUSE® II Bone Graft are substantially equivalent to the predicate. Based on the risk analysis and additional supporting documentation provided in this premarket notification, Medtronic believes the subject device demonstrates substantial equivalence to listed predicate device GRAFTON® II eDBM K082165 (S.E. 10/16/2008).

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing or supporting another, represented by flowing, abstract shapes.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 6, 2013

Medtronic Sofamor Danek USA, Incorporated % Ms. Kelly Anglin Senior Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132

Re: K122513

Trade/Device Name: MAGNIFUSE® II Bone Graft Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV, MBP Dated: February 14, 2013 Received: February 19, 2013

Dear Ms. Anglin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Ms. Kelly Anglin

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the - Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/4/Picture/6 description: The image shows the name "Erin DKeith" in a stylized font. The letters "Erin" and "eith" are in a simple, bold font. The letters "DK" are more stylized, with a circular design in the middle of the "D".

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known):

Device Name: MAGNIFUSE® II Bone Graft

INDICATIONS FOR USE:

MAGNIFUSE® II Bone Graft is intended for use as a bone graft substitute in bony voids or gaps of the skeletal system (i.e., posterolateral spine and pelvis) not intrinsic to the stability of the bony structure. The voids or gaps may be surgically created defects or defects created by traumatic injury to the bone. MAGNIFUSE® II Bone Graft is resorbed/remodeled and replaced by host bone during the healing process.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR


Over-The-Counter Use (21 CFR Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Laurence D. Coyne -A

(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K122513

§ 888.3045 Resorbable calcium salt bone void filler device.

(a)
Identification. A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.