K Number
K122388

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-05-02

(269 days)

Product Code
Regulation Number
870.2770
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The seca mBCA 514 measures body Weight and records body Height for calculation of Body Mass Index (BMI), Resting Energy Expenditure (REE) and Total Energy Expenditure (TEE) to monitor growth process and weight changes in individuals of all ages.

In healthy individuals aged 18 years and older the device also measures body impedance (Reactance Xc and Resistance R) for estimation of body composition. Estimation of body composition includes Total Body Water (TBW), Intra Cellular Water (ICW), Extra Cellular Water (ECW), Lean Soft Tissue (LST), Fat-Free Mass (FFM), Fat Mass (FM), Fat Mass Indices (FMI and FMMI), Skeletal Muscle Mass (SMM), Bioelectrical Impedance Vector Analysis (BIVA), Energy stored in body (Ebody) and Phase Angle (Φ).

The seca mBCA 514 can be used in conjunction with the optional PC software accessory seca analytics 115 for data management, calculations and display of information.

Device Description

The seca mBCA 514 is a scale and impedance plethysmograph body composition analyzer for estimation of the body composition of individuals based on the Bioelectrical Impedance Analysis (BIA). The device measures bioelectrical impedance values by means of multi frequency, segmental measurements (right arm, trunk, right leg, left leg, right body side and left body side).

AI/ML Overview

The seca mBCA 514 is a body composition analyzer that uses bioelectrical impedance analysis (BIA) to estimate body composition. The device's performance was evaluated through bench and clinical testing to demonstrate its safety and effectiveness compared to a predicate device, the BioSpace InBody 720 (K052646), and other established methods for specific body composition parameters.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state acceptance criteria in terms of specific thresholds (e.g., a certain percentage of agreement or a specific correlation coefficient). Instead, it states that the results of the testing "demonstrate that the seca mBCA 514 is as safe and effective as the predicate device" and that the "differences with the predicate devices and methods are sufficiently minor to establish substantial equivalence." This implies that the acceptance criteria were met by demonstrating comparable performance to established methods.

Here's a summary of the comparisons:

Parameter Measured by seca mBCA 514Comparison Method/Predicate DeviceReported Device Performance
Impedance raw data (Impedance, Resistance, Reactance, Phase Angle)Biospace InBody 720 (K052646)Compared against measurements made by the predicate device. Stated as "as safe and effective as the predicate device."
WeightBiospace InBody 720 (K052646)Compared against measurements made by the predicate device. Stated as "as safe and effective as the predicate device."
Total Body Water (TBW)Deuterium oxide dilution methodCompared against the deuterium oxide dilution method.
Extra Cellular Water (ECW)Sodium bromide dilution methodCompared against the sodium bromide dilution method.
Lean Soft Tissue (LST)GE Lunar iDXA Bone Densitometer (K052581)Compared against measurements obtained from the GE Lunar iDXA.
Fat-Free Mass (FFM)Four compartment model (incorporating GE Lunar iDXA (K052581), BOD POD Sonamet (K060848), and deuterium oxide dilution)Estimated by applying the four compartment model.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample size used for the test set in the clinical trials, nor does it specify the country of origin of the data or whether it was retrospective or prospective. It only broadly refers to "clinical testing."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document does not specify the number of experts or their qualifications for establishing the ground truth. The ground truth for different parameters was established using established scientific methods (e.g., deuterium oxide dilution, sodium bromide dilution, DXA, BOD POD), rather than solely expert consensus on interpretations of device outputs.

4. Adjudication Method for the Test Set

Not applicable. The ground truth was established by reference methods, not by expert adjudication of device outputs.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No. The study described is a comparison of the device's measurements against established methods and a predicate device, not a human reader study with or without AI assistance.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

Yes, the studies described are standalone performance evaluations of the seca mBCA 514's ability to measure or estimate various body composition parameters compared to established reference methods and a predicate device. The device's output (measurements and estimations) is directly compared.

7. The Type of Ground Truth Used

The ground truth for the device's performance was established using established scientific methods considered gold standards for body composition measurement:

  • Deuterium oxide dilution method: For Total Body Water (TBW)
  • Sodium bromide dilution method: For Extra Cellular Water (ECW)
  • GE Lunar iDXA Bone Densitometer (K052581): For Lean Soft Tissue (LST)
  • Four compartment model (incorporating GE Lunar iDXA, BOD POD Sonamet Body Composition Analyzer (K060848), and deuterium oxide dilution method): For Fat-Free Mass (FFM)
  • Biospace InBody 720 (K052646): As a predicate device for Impedance raw data and Weight.

8. The Sample Size for the Training Set

The document does not provide information about a specific training set size. Body composition analyzers typically use pre-developed algorithms and regression equations, which are often derived from large historical datasets, but no details are given here regarding the specific training set for seca's algorithms.

9. How the Ground Truth for the Training Set Was Established

The document does not provide information on how the ground truth for any potential training set was established. It only describes the methods used for the comparative clinical testing to establish substantial equivalence.

{0}------------------------------------------------

Image /page/0/Picture/20 description: The image shows the word "seca" in a bold, sans-serif font. The letters are connected to each other, creating a continuous, flowing design. The color of the text is black, and the background is white.

seca medical Body Composition Analyzer 514510(k) Summary - section 05
510(k) Premarket Notification - K122388page 1 of 2

2 2013 MAY

Submitter Information 1

Owner: Address:

Internet:

Contact:

E-mail:

seca gmbh & co. kg. Hammer Steindamm 9 - 25 22089 Hamburg Germany www.seca.com

Djavaneh Blalaie - Quality Management +49 40 20 00 00 - 259 +40 40 20 00 00 - 3259 djavaneh.blalaie@seca.com

Date prepared:

Telephone Number: FAX Number:

April 22, 2013

2 Device Name

Common Device Name: Trade names:

Device Class: Classification Code: Regulation number:

Body Composition Analyzer seca medical Body Composition Analyzer 514, seca mBCA 514, seca mBCA, seca 514 Class II MNW - Analyzer, Body Composition 21 CFR 870.2770

Predicate Device 0

BioSpace InBody 720

Common Device Name: Trade names: Device Class: Classification Code: Regulation number: 510(k) number:

Body fat meter Biospace Body Composition Analyzer, Model Inbody 720 Class II MNW - Analyzer, Body Composition 21 CFR 870.2770 K052646

Device Description র্ব

The seca mBCA 514 is a scale and impedance plethysmograph body composition analyzer for estimation of the body composition of individuals based on the Bioelectrical Impedance Analysis (BIA). The device measures bioelectrical impedance values by means of multi frequency,

{1}------------------------------------------------

seca medical Body Composition Analyzer 514510(k) Summary - section 05
510(k) Premarket Notification - K122388page 2 of 2

segmental measurements (right arm, trunk, right leg, left leg, right body side and left body side).

5 Intended Use

The seca mBCA 514 measures body Weight and records body Height for calculation of Body Mass Index (BMI), Resting Energy Expenditure (REE) and Total Energy Expenditure (TEE) to monitor growth process and weight changes in individuals of all ages.

In healthy individuals aged 18 years and older the device also measures body impedance (Reactance Xc and Resistance R) for estimation of body composition. Estimation of body composition includes Total Body Water (TBW), Intra Cellular Water (ICW), Extra Cellular Water (ECW), Lean Soft Tissue (LST), Fat-Free Mass (FFM), Fat Mass (FM), Fat Mass Indices (FMI and FMMI), Skeletal Muscle Mass (SMM), Bioelectrical Impedance Vector Analysis (BIVA), Energy stored in body (Ebody) and Phase Angle (Φ).

The seca mBCA 514 can be used in conjunction with the optional PC software accessory seca analytics 115 for data management, calculations and display of information.

6 Safety and Effectiveness Comparison

The results of bench and clinical testing demonstrate that the seca mBCA 514 is as safe and effective as the predicate device. Impedance raw data measurements (Impedance. Resistance, Reactance, and Phase Angle) and measurements of Weight were compared against measurements made by the Biospace InBody 720 (K052646). Estimation of Total Body Water (TBW) was compared against the deuterium oxide dilution method. Estimation of Extra Cellular Water (ECW) was compared against the sodium bromide dilution method. Estimation of Lean Soft Tissue (LST) was compared against measurements obtained from the GE Lunar iDXA Bone Densitometer (K052581). Fat-Free Mass (FFM) was estimated by applying the four compartment model, which incorporates measurements from the GE Lunar iDXA Bone Densitometer (K052581), the BOD POD Sonamet Body Composition Analyzer (K060848), and the deuterium oxide dilution method.

7 Conclusion

The differences with the predicate devices and methods are sufficiently minor to establish substantial equivalence, and do not impact the safety or the effectiveness of the device when used as labeled.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 2, 2013

seca gmbh & co. kg % Mr. Seth A. Mailhot Special Counsel Sheppard Mullin Richter & Hampton LLP 13001 Street, NW. 11th Floor East WASHINGTON DC 20005-3314

Re: K122388

Trade/Device Name: seca mBCA 514 Regulation Number: 21 CFR§ 870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: April 24, 2013 Received: April 25, 2013

Dear Mr. Mailhot:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Seth A. Mailhot

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K122388

Device Name: seca mBCA 514

Intended Use:

The seca mBCA 514 measures body Weight and records body Height for calculation of Body Mass Index (BMI), Resting Energy Expenditure (REE) and Total Energy Expenditure (TEE) to monitor growth process and weight changes in individuals of all ages.

In healthy individuals aged 18 years and older the device also measures body impedance (Reactance Xc and Resistance R) for estimation of body composition. Estimation of body composition includes Total Body Water (TBW), Intra Cellular Water (ICW), Extra Cellular Water (ECW), Lean Soft Tissue (LST), Fat-Free Mass (FFM), Fat Mass (FM), Fat Mass Indices (FMI and FMMI), Skeletal Muscle Mass (SMM), Bioelectrical Impedance Vector Analysis (BIVA), Energy stored in body (Ebody) and Phase Angle (Φ).

The seca mBCA 514 can be used in conjunction with the optional PC software accessory seca analytics 115 for data management, calculations and display of information.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Herbert - S

(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K122388 510(k) Number

§ 870.2770 Impedance plethysmograph.

(a)
Identification. An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.(b)
Classification. Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.