(39 days)
When used as a pedicle screw fixation system, the NuVasive Precept Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the posterior thoracic, lumbar, and sacral spine:
- Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
- Degenerative spondylolisthesis with objective evidence of neurologic impairment
- Fracture
- Dislocation
- Scoliosis
- Kyphosis
- Spinal tumor and/or
- Failed previous fusion (pseudoarthrosis)
The NuVasive® Precept™ Spinal System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion.
When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the Nu Vasive Precept Spinal System is also intended for the following indications:
- Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
- Spinal stenosis
- Spondylolisthesis
- Spinal deformities
- Fracture
- Pseudoarthosis
- Tumor resection and/or
- Failed previous fusion
The NuVasive Precept Spinal System consists of a variety of polyaxial screws, reduction screws, offset connectors, rods, locking nuts, and transverse connectors. Implant components can be rigidly locked into a variety of different configurations to suit the individual pathology and anatomical conditions of the patient. This 510(k) is for the modification to a component used in the assembly of tulip to screw shanks of polyaxial screws. The system as a whole remains unchanged compared to the predicate (K102514).
The NuVasive® Precept™ Spinal System is a medical device, and the provided text describes its 510(k) summary for premarket notification. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving efficacy in a clinical study with acceptance criteria and statistical analysis as would be typical for a new drug or high-risk medical device.
Therefore, many of the requested elements for a study proving device meets acceptance criteria are not applicable in this context, as a clinical study with human subjects, ground truth, and expert adjudication was not conducted for this 510(k) submission. Instead, the device's performance was evaluated through non-clinical (mechanical) testing.
Here's a breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (from predicate or standard) | Reported Device Performance (NuVasive Precept™ Spinal System) |
|---|---|
| Mechanical Testing: | |
| Static and dynamic flexion-extension moment | Met or exceeded the performance of the predicate devices based on ASTM F1798. |
| Tulip pull-off | Met or exceeded the performance of the predicate devices. |
| Tension-torsion | Met or exceeded the performance of the predicate devices. |
| Orbital Tension | Met or exceeded the performance of the predicate devices. |
| Overall Equivalence: | |
| Substantial equivalence to predicate devices | Demonstrated through comparison in design, intended use, material composition, function, and range of sizes. System as a whole remains unchanged compared to the predicate (K102514). |
Note: The specific numerical acceptance criteria (e.g., maximum force, displacement limits) for each mechanical test are not explicitly detailed in the provided summary, only that the device "met or exceeded" the performance of predicate devices or relevant standards (ASTM F1798).
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not explicitly stated in the document. For mechanical testing, this would refer to the number of individual components or assembled systems tested.
- Data Provenance: The tests were nonclinical (mechanical) and conducted by NuVasive, Inc. at their facilities, likely in the USA (San Diego, California).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. This was a non-clinical, mechanical testing submission. There was no "ground truth" to be established by experts in the context of clinical interpretation or diagnosis. The "ground truth" for mechanical testing is typically defined by engineering standards and specifications.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable. As this was mechanical testing, no human adjudication of clinical outcomes or interpretations was performed.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. No MRMC or clinical comparative effectiveness study involving human readers or AI assistance was performed. This submission is for a physical spinal implant system, not a diagnostic imaging or AI-driven decision support tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This relates to AI/algorithm performance. The NuVasive® Precept™ Spinal System is a physical medical device (pedicle screw system), not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Mechanical Testing Standards: For the mechanical tests (static/dynamic flexion-extension, tulip pull-off, tension-torsion, orbital tension), the "ground truth" or reference for performance is established by industry standards (e.g., ASTM F1798) and the performance of the predicate devices. The goal was to show that the modified device's performance was equivalent to or better than these established benchmarks.
8. The sample size for the training set
- Not applicable. There was no "training set" as this was a non-clinical mechanical testing for a spinal implant, not a machine learning model.
9. How the ground truth for the training set was established
- Not applicable. As there was no training set, this question does not apply.
{0}------------------------------------------------
Special 510(k) Premarket Notification NuVasive® Precept™ Spinal System
510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
Submitted by: A.
Sheila Bruschi Associate Manager, Regulatory Affairs NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-1900 Fax: (858) 320-1681
Date Prepared: September 11, 2012
B. Device Name
| Trade or Proprietary Name: | NuVasive® Precept™ Spinal System |
|---|---|
| Common or Usual Name: | Pedicle Screw System |
| Classification Name: | Pedicle Screw Spinal System, Spinal Interlaminal Fixation |
| Orthosis, Spinal Intervertebral Body Fixation orthosis | |
| Device Class: | |
| Classification: | §888.3050, §888.3060, §888.3070 |
| Product Code: | NKB, KWP, MNI, MNH, KWQ |
C. Predicate Devices
The subject NuVasive Precept Spinal System is substantially equivalent to the NuVasive Precept (K102514) and SpheRx® PPS (K090981) Systems currently distributed commercially in the U.S. by NuVasive.
D. Device Description
The NuVasive Precept Spinal System consists of a variety of polyaxial screws, reduction screws, offset connectors, rods, locking nuts, and transverse connectors. Implant components can be rigidly locked into a variety of different configurations to suit the individual pathology and anatomical conditions of the patient. This 510(k) is for the modification to a component used in the assembly of tulip to screw shanks of polyaxial screws. The system as a whole remains unchanged compared to the predicate (K102514).
E. Intended Use
When used as a pedicle screw fixation system, the NuVasive Precept Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the posterior thoracic, lumbar, and sacral spine:
-
- Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
-
- Degenerative spondylolisthesis with objective evidence of neurologic impairment
-
- Fracture
-
- Dislocation
-
- Scoliosis
SEP 11 2012
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf-like shape on the left, followed by the word "NUVASIVE" in a simple, sans-serif font. Below the word "NUVASIVE" is the tagline "Speed of Innovation" in a smaller font size.
Special 510(k) Premarket Notification NuVasive® Precept™ Spinal System
-
- Kyphosis
-
- Spinal tumor and/or
-
- Failed previous fusion (pseudoarthrosis)
The NuVasive® Precept™ Spinal System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion.
When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the Nu Vasive Precept Spinal System is also intended for the following indications:
-
- Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
-
- Spinal stenosis
-
- Spondylolisthesis
-
- Spinal deformities
-
- Fracture
-
- Pseudoarthosis
-
- Tumor resection and/or
-
- Failed previous fusion
F. Technological Characteristics
As was established in this submission, the subject NuVasive Precept Spinal System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, function, and range of sizes.
G. Performance Data
Nonclinical testing was performed to demonstrate that the subject NuVasive Precept Spinal System is substantially equivalent to other predicate devices. The following component testing was performed:
- Static and dynamic flexion-extension moment testing per ASTM F1798 .
- Tulip pull-off .
- Tension-torsion .
- Orbital Tension .
The results of these studies showed that the system as a whole was unchanged compared to the predicate, and the subject NuVasive Polyaxial Spinal Screws meet or exceed the performance of the predicate devices, and the device was therefore found to be substantially equivalent.
H. Conclusions
Based on the indications for use, technological characteristics, performance testing, and comparison to predicate devices, the subject NuVasive Precept Spinal System have been shown to be substantially equivalent to legally marketed predicate devices, and safe and effective for its intended use.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Nuvasive, Incorporated % Ms. Sheila Bruschi Associate Manager, Regulatory Affairs 7475 Lusk Boulevard San Diego, California 92121
Re: K122352
Trade/Device Name: NuVasive® Precept™ Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, KWP, MNI, MNH, KWQ Dated: August 02, 2012 Received: August 03, 2012
Dear Ms. Bruschi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
SEP 11 2012
{3}------------------------------------------------
Page 2 - Ms. Sheila Bruschi
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark A. Mikkelsen
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
K122352 510(k) Number (if known):
Device Name: NuVasive® Precept™ Spinal System
Indications For Use:
When used as a pedicle screw fixation system, the NuVasive Precept Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities of the posterior thoracic, lumbar, and sacral spine:
- Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of 1. the disc confirmed by patient history and radiographic studies)
-
- Degenerative spondylolisthesis with objective evidence of neurologic impairment
-
- Fracture
-
- Dislocation
-
- Scoliosis
-
- Kyphosis
-
- Spinal tumor and/or
-
- Failed previous fusion (pseudoarthrosis)
The NuVasive Precept Spinal System is also indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebral joint in skeletally mature patients receiving fusion by autogenous bone graft, having the device fixed or attached to the lumbar and sacral spine (L3 to sacrum), with removal of the implants after attainment of a solid fusion.
When used as an anterolateral non-pedicle screw system in the thoracic and lumbar spine, the NuVasive Precept Spinal System is also intended for the following indications:
-
- Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
-
- Spinal stenosis
-
- Spondylolisthesis
-
- Spinal deformities
-
- Fracture
-
- Pseudoarthosis
-
- Tumor resection and/or
-
- Failed previous fusion
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
$
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
K122352 510(k) Number
Page 1 of 1
N/A