(112 days)
The indications for the Trinity Acetabular System as a total hip arthroplasty include:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis o
- Rheumatoid arthritis 0
- o Correction of functional deformity
- O Developmental dysplasia of the hip (DDH) or congenital dysplasia of the hip (CDH)
The Trinity Acetabular System is intended for cementless, single use only.
The Trinity Acetabular System is a modular acetabular cup system consisting of a press fit, titanium alloy shell, ultra-high molecular weight polyethylene (UHMWPE), highly cross-linked polyethylene (HXLPE), vitamin E highly cross-linked polyethylene (ECiMa) acetabular liners in neutral offset, +4mm offset, +4mm oblique, neutral 4mm EPW, BIOLOX deltaTM ceramic and CoCr modular heads and titanium femoral stems. The acetabular shells are coated with a rough titanium plasma spray with an additional top layer of electrochemically deposited calcium phosphate (BonitTM).
The provided text is a 510(k) summary for the Corin Trinity Acetabular System with Trinity-i Acetabular Shells. It describes a medical device, specifically a hip prosthesis, and its substantial equivalence to previously marketed devices.
Based on the information provided in the text, here's a breakdown of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance/Evidence |
|---|---|---|
| Material Equivalence | The additional components of the Trinity Acetabular System are exactly the same as the predicate devices in terms of materials. | Stated as "exactly the same" as predicate devices (K093472 and K110087). |
| Intended Use Equivalence | The additional components of the Trinity Acetabular System are exactly the same as the predicate devices in terms of intended use. | Stated as "exactly the same" as predicate devices. |
| Indications for Use Equivalence | The additional components of the Trinity Acetabular System are exactly the same as the predicate devices in terms of indications for use. | Stated as "exactly the same" as predicate devices. |
| Design and Performance Equivalence | The additional components are similar in sizes, design, and performance to the predicate devices. | Stated as "similar in sizes, design and performance" to predicate devices. Specifically mentions the Trinity-i Acetabular Shells having a smaller inner diameter with the same outer diameter to accommodate a larger liner/head. |
| Shell Stiffness | Performance related to shell stiffness. | Non-clinical testing conducted, including determination of shell stiffness. Specific acceptance values are not provided in this summary. |
| Coating Process Effect on Shell Taper Form | The coating process does not adversely affect shell taper form. | Non-clinical testing conducted, including the effect of the coating process on shell taper form. Specific acceptance values are not provided. |
| Shell Deformation Effect on Liners | The shell deformation does not adversely affect HXLPE and ECiMa liners. | Non-clinical testing conducted, including the effect of shell deformation on HXLPE and ECiMa liners. Specific acceptance values are not provided. |
| Previous Bench Testing & Characterization | Bench testing for Trinity titanium shells and characterization/testing for the coating should have been completed and found acceptable in previous submissions (K093472 and K110087). | This testing was "included in previous submissions for the Corin Trinity Acetabular System (K093472 and K110087)". Specific acceptance values are not provided here. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document explicitly states: "Non-clinical testing conducted to demonstrate substantial equivalence includes: determination of shell suffness, effect of the coating process on shell taper form, and the effect of shell deformation on HXLPE and ECiMa liners."
However, no specific sample sizes for these non-clinical tests are provided. The data provenance (e.g., country of origin, retrospective or prospective) is also not mentioned in this summary. These are bench tests, so "data provenance" in the sense of clinical patient data would not apply.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This 510(k) pertains to a medical device (hip prosthesis) and its substantial equivalence based on non-clinical bench testing. Therefore, no "ground truth" derived from expert consensus on patient data or clinical scenarios is applicable here. The "truth" is established by engineering specifications and material science standards met during the bench testing.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Since this is a non-clinical bench study, no adjudication method involving human readers or experts for a "test set" in a clinical context is relevant or described. The results of the engineering tests would be assessed against pre-defined engineering and material standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was done. This 510(k) is for a hip prosthesis and its substantial equivalence through non-clinical testing. It does not involve AI or human image interpretation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
No standalone algorithm performance study was done. This device is a physical hip prosthesis, not a software algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this substantial equivalence determination is based on engineering specifications, material properties, and performance standards for the device and its predicate devices, as demonstrated through non-clinical bench testing. There is no clinical "ground truth" from pathology or outcomes data presented in this summary, as clinical testing was not deemed necessary.
8. The sample size for the training set
No training set is applicable or described. This is a physical medical device, and the evaluation is based on non-clinical bench testing and comparison to predicate devices, not on machine learning or AI models that require training data.
9. How the ground truth for the training set was established
As there is no training set, the concept of establishing ground truth for it is not applicable.
{0}------------------------------------------------
K122305 (pg 1/2)
| 3. 510(K) SUMMARY | ||
|---|---|---|
| 1. Applicant/Sponsor: | Corin USA10500 University Center DriveSuite 190Tampa, Florida 33612Establishment Registration No.: 1056629 | |
| 2. Contact Person: | Diana L. MartoneRegulatory Affairs AssociateCorin USA813-977-4469diana.martone@coringroup.com | |
| 3. Proprietary Name: | Corin Trinity Acetabular System with Trinity-i Acetabular Shells | |
| 4. Common Name: | Hip Prosthesis | |
| 5. Classification Name: | Hip joint metal/ceramic/polymer semi-constrained cemented ornonporous uncemented prosthesis (21CFR 888.3353, Product Codes LZO,MEH) | |
| 6. Legally Marketed Devices to which Substantial Equivalence is claimed: | ||
| ■ | Corin Trinity Acetabular System (K093472) | |
| ■ | Corin Trinity Acetabular System with HXLPE Liners (K110087) | |
| 7. Device Description: | The Trinity Acetabular System is a modular acetabular cup system consisting of apress fit, titanium alloy shell, ultra-high molecular weight polyethylene (UHMWPE), highlycross-linked polyethylene (HXLPE), vitamin E highly cross-linked polyethylene (ECiMa)acetabular liners in neutral offset, +4mm offset, +4mm oblique, neutral 4mm EPW,BIOLOX deltaTM ceramic and CoCr modular heads and titanium femoral stems. Theacetabular shells are coated with a rough titanium plasma spray with an additional top layerof electrochemically deposited calcium phosphate (BonitTM). |
ರ್ವ inner diameter, but with the samc outer diameter, to accommodate a larger liner and corresponding head size to the Trinity Acetabular System.
The Trinity Acetabular System is intended for use in total hip arthroplasty in skeletally mature patients, to provide increased mobility and reduce pain by replacing the damaged hip
{1}------------------------------------------------
K122305 (pg 2/2)
joint articulation where there is evidence of sufficient sound bone to seat and support the components.
8. Intended Use / Indications:
The indications for the Trinity Acetabular System as a total hip arthroplasty include:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis o
- Rheumatoid arthritis 0
- o Correction of functional deformity
- O Developmental dysplasia of the hip (DDH) or congenital dysplasia of the hip (CDH)
The Trinity Acetabular System is intended for cementless, single use only.
-
- Summary of Technologies/Substantial Equivalence:
The additional components of the Trinity Acetabular System are exactly the same as the predicate devices in terms of intended use, indications for use, and materials; and are similar in sizes, design and performance. Based on these similarities, the additional components of the Trinity Acetabular System are believed to be substantially equivalent to the predicate devices.
- Summary of Technologies/Substantial Equivalence:
-
- Non-Clinical Testing:
Non-clinical testing conducted to demonstrate substantial equivalence includes: determination of shell suffness, effect of the coating process on shell taper form, and the effect of shell deformation on HXLPE and ECiMa liners. Bench testing of the Trinity titanium shells and characterization and testing for the coating was included in previous submissions for the Corin Trinity Acetabular System (K093472 and K110087).
- Non-Clinical Testing:
11. Clinical Testing:
Clinical testing was not necessary to determine substantial equivalence between the additional components of the Trinity Acetabular System and the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized abstract symbol resembling a human figure embracing a sphere or globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Letter Dated: November 21, 2012
Corin USA % Ms. Diana L. Martone Regulatory Affairs Associate 10500 University Center Drive Tampa, Florida 33612
Re: K122305
Trade/Device Name: Trinity Acetabular System Regulation Number: 21 CFR 888.3353 'Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: II Product Code: LZO, MEH Dated: November 9, 2012
Received: November 13, 2012
Dear Ms. Martone:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Diana L. Martone
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
INDICATIONS FOR USE 2.
510(k) Number (if known): K122305
Device Name: Trinity Acetabular System
Indications for Use:
The indications for the Trinity Acetabular System as a total hip arthroplasty include:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular . O necrosis
- Rheumatoid arthritis O
- Correction of functional deformity O
- Developmental dysplasia of the hip (DDH) or congenital dysplasia of the hip (CDH) O
The Trinity Acetabular System is intended for cementless use only.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Krishna Asundi (Division Sign-Off) Division of Orthopedic Devices 2012.11.20 08:12:26 -05'00'
Page 1 of _l
N/A