Intended Use

Thermo Scientific MAS® Omni•CARDIO™ is intended for use in the clinical laboratory as an assayed control serum for monitoring assay conditions related to cardiac and associated critical marker determinations.

Device Description

Each MAS® Omni•CARDIO control kit is packaged in a plastic clamshell container with a product insert and value sheet. Each vial contains 3 mL of the control level in an amber glass bottle with black sterile caps. Kits are stored frozen at -20°C. MAS® Omni CARDIO control levels are provided in liquid form and are to be stored at -20°C until the expiration date on the label. The control levels are prepared from human serum and contain the following constituents: Beta-Human Chorionic Gonadotropin (ß-HCG), B-Type Natriuretic Peptide (BNP), Creatine Kinase-MB (CK-MB), D-Dimer, Digitoxin, Homocysteine, High Sensitivity C-Reactive Protein (hsCRP), Human Chorionic Gonadotropin (HCG), Myeloperoxidase (MPO), Myoglobin, N-Terminal Prohormone of Brain Natriuretic Peptide (NTproBNP), Procalcitonin (PCT), Total Creatine Kinase (CK), Troponin I, Troponin T.

AI/ML Overview

Here's the analysis of the acceptance criteria and study information for the MAS® Omni•CARDIO device, based on the provided 510(k) summary:

Device: Thermo Scientific MAS® Omni•CARDIO

Intended Use: Intended for use in the clinical laboratory as an assayed control serum for monitoring assay conditions related to cardiac and associated critical marker determinations.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
Shelf Life Stability36 months frozen at -20°C (equivalent to stress testing at elevated temperatures)All analytes and lots tested met the minimum number of days for all temperatures tested, supporting a 36-month shelf life claim.
Open Vial Stability (2-8°C)15 days for most analytes. Exceptions: Myoglobin and NT-ProBNP (10 days), Homocysteine (5 days).All analytes met the required claim, with the specified exceptions (Myoglobin and NT-ProBNP: 10 days, Homocysteine: 5 days).
Closed Vial Stability (2-8°C)15 days for most analytes. Exceptions: Myoglobin and NT-ProBNP (10 days), Homocysteine (5 days).All analytes met the required claim, with the specified exceptions (Myoglobin and NT-ProBNP: 10 days, Homocysteine: 5 days).
Value AssignmentValues for evaluation lots to fall within established ranges (typically +/-20% around the grand mean, expanded if needed).Values for the evaluation lots all fell within the established ranges.

2. Sample Sizes Used for the Test Set and Data Provenance

The document does not explicitly define distinct "test sets" in the way a clinical study would for a diagnostic device. Instead, the testing described relates to the manufacturing and characterization of the control material itself.

  • Stability Studies (Accelerated & Real-time): "evaluation lots" were used. The specific number of vials/samples tested per lot, per analyte, per temperature/timepoint is not provided.
  • Open and Closed Vial Testing: "evaluation lots" were used. Vials were opened/unopened daily (work days) and assayed at "multiple timepoints." Specific numbers are not provided.
  • Value Assignment Testing: "Evaluation lots were sent to multiple sites, typically 2 or 3 lots per instrument platform, and assayed over 5 days in duplicate."
    • Data Provenance: The document does not specify the country of origin of the data. It mentions "multiple sites" for value assignment, suggesting it could be multi-site data. The studies are prospective in nature, as they involve testing newly manufactured lots of the control material under defined conditions.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • This device is an assayed control serum, not a diagnostic device that interprets patient data. Therefore, the concept of "ground truth" established by experts in the context of diagnostic device validation (e.g., radiologist consensus on images) does not directly apply here.
  • The "ground truth" for this control material is its assigned values and stability properties. These are established through rigorous laboratory testing using various analytical instruments and methods, following established protocols (e.g., averaging data from multiple sites, determining failure points via linear regression).
  • The "experts" involved would be clinical laboratory professionals, assay developers, and quality control specialists, responsible for designing and executing these validation studies and analyzing the results according to scientific and regulatory standards. Specific numbers or qualifications of these individuals are not provided in this summary.

4. Adjudication Method for the Test Set

  • Adjudication methods like "2+1" or "3+1" are typically used in clinical studies for diagnostic devices where human readers interpret patient data and discrepancies need to be resolved.
  • For this control material, adjudication, if it occurred, would likely be in the form of discrepancy resolution during the laboratory testing, where any unexpected results or outliers would be investigated, re-tested, and potentially discussed among the testing team or quality control personnel to ensure data integrity and accuracy of the assigned values and stability claims. This process is implicit in good laboratory practice but not explicitly detailed as a formal adjudication method in this summary.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not done.
  • This type of study is relevant for diagnostic devices where human readers (e.g., radiologists) use the device to interpret cases, and the study compares human performance with and without AI assistance.
  • The MAS® Omni•CARDIO is a quality control material, not a diagnostic device, and therefore does not involve human readers interpreting cases for diagnosis. Its purpose is to monitor the performance of analytical instruments.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Yes, in essence, standalone performance was assessed.
  • The stability and value assignment studies for MAS® Omni•CARDIO are conducted in a "standalone" manner in the sense that they evaluate the inherent properties of the control material (its value, stability over time) independent of a "human-in-the-loop" interpretation process. The control material serves as a fixed, known standard against which analytical instruments are measured. The evaluation of its performance criteria (stability, assigned value accuracy) is a direct measurement of the product itself.

7. Type of Ground Truth Used

  • The "ground truth" for the MAS® Omni•CARDIO control material is established through analytical testing and consensus of assay results across multiple instruments and sites.
  • For Value Assignment: The ground truth for the assigned values (e.g., BNP 32 at 56 pg/mL for Level UL) is the grand mean derived from assaying the evaluation lots over 5 days in duplicate on "multiple sites, typically 2 or 3 lots per instrument platform."
  • For Stability: The ground truth for stability claims (e.g., 36 months frozen) is derived from analytical measurements at various timepoints and conditions (accelerated and real-time testing), with "point of failure determined by linear regression analysis of the data."

8. Sample Size for the Training Set

  • This device is a quality control material, not an AI or machine learning algorithm that requires a "training set" in the conventional sense.
  • Therefore, the concept of a "training set" as described for AI devices (e.g., for model development) does not apply to the MAS® Omni•CARDIO. The data described in the summary is for product characterization and validation.

9. How the Ground Truth for the Training Set Was Established

  • As stated in point 8, the concept of a "training set" does not apply to this device.
  • The ground truth for the characterization and validation data (which some might loosely compare to a "training" or "development" phase for product features) was established as described in point 7: through analytical measurements, grand mean calculations across multiple instrument platforms and sites, and linear regression analysis for stability.

{0}------------------------------------------------

Omni Cardio 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K122291

Company / Contact Person

Karen Lee Regulatory Affairs Specialist Thermo Fisher Scientific, Clinical Diagnostics Division 46360 Fremont Blvd. Fremont, CA 94538 Phone: (510) 979-5000 x31814 Fax: (510) 979-5422 E-mail: karen.lee@thermofisher.com

Date Prepared

August 29, 2012

Regulatory Declarations

Common / Usual NameMAS® Omni•CARDIO
Trade / Proprietary NameThermo Scientific MAS® Omni•CARDIO
Classification Regulation21 CFR 862.1660
Device ClassClass I, Reserved
Device Regulation PanelClinical Chemistry
Product CodeJJY

Intended Use MAS® Omni•CARDIO

Thermo Scientific MAS® Omni•CARDIO™ is intended for use in the clinical laboratory as an assayed control serum for monitoring assay conditions related to cardiac and associated critical marker determinations.

Legally Marketed Device to Which Equivalency is Claimed

The Thermo Scientific MAS® Omni+CARDIO is substantially equivalent to the previously cleared MAS® Cardiolmmune TL (K040880) and MAS® Cardiolmmune XL (K061196).

Device Description

Each MAS® Omni•CARDIO control kit is packaged in a plastic clamshell container with a product insert and value sheet. Each vial contains 3 mL of the control level in an amber glass bottle with black sterile caps. Kits are stored frozen at -20°C.

{1}------------------------------------------------

Below are the different kit configurations:

KitConfiguration
Level UL6 x 3 mL
Level L6 x 3 mL
Level 16 x 3 mL
Level 26 x 3 mL
Level 36 x 3 mL
Multi-PackLevel 1 – 2 x 3 mLLevel 2 – 2 x 3 mLLevel 3 – 2 x 3 mL
Sample PackLevel UL – 1 x 3 mLLevel L – 1 x 3 mLLevel 1 – 1 x 3 mLLevel 2 – 1 x 3 mLLevel 3 – 1 x 3 mL

MAS® Omni CARDIO control levels are provided in liquid form and are to be stored at -20°C until the expiration date on the label. The control levels are prepared from human serum and contain the following constituents: Beta-Human Chorionic Gonadotropin (ß-HCG), B-Type Natriuretic Peptide (BNP), Creatine Kinase-MB (CK-MB), D-Dimer, Digitoxin, Homocysteine, High Sensitivity C-Reactive Protein (hsCRP), Human Chorionic Gonadotropin (HCG), Myeloperoxidase (MPO), Myoglobin, N-Terminal Prohormone of Brain Natriuretic Peptide (NTproBNP), Procalcitonin (PCT), Total Creatine Kinase (CK), Troponin I, Troponin T.

Stability Testing and Value Assignment Summary:

Accelerated testing was carried out for the evaluation lots at three elevated temperatures in order to support a shelf life storage temperature of -20°C for 36 months. For each temperature tested, the required number of days of stress without failure was determined that would be equivalent to 36 months storage at -20°C. All analytes and lots tested met the minimum number of days for all temperatures tested in support of a shelf life claim of 36 months frozen at -20℃. Real time testing at -20°C is on-going.

Open and closed vial testing was carried out at 5°C for all claimed analytes on the evaluation lots. Vials were thawed, mixed, and placed at 2-8°C for the study. Vials were opened briefly each work day (open vial) or remained unopened (closed vial) until time of assav. Multiple timepoints were tested and the point of failure determined by linear regression analysis of the data. All analytes met the required claim of 15 days storage for both open and closed vial testing with the exception of Myoglobin and NT-ProBNP (10 days) and Homocysteine (5 days).

Value assignment testing was carried out to determine typical values that would be seen for the product across different analyzer platforms. Evaluation lots were sent to multiple sites, typically 2 or 3 lots per instrument platform, and assaved over 5 days in duplicate. For each analyte and analyzer combination, the data was averaged to determine a grand mean. Value assignment ranges were established at +/-20% around the grand mean and were expanded if needed to

{2}------------------------------------------------

ensure the minimum and maximum values in the data set were within the established range. Values for the evaluation lots all fell within the established ranges.

AnalyteUnitsLevel ULLevel LLevel 1Level 2Level 3
BNP 32pg/mL561042166402600
CK-MBng/mL1.74.36.532.5100
D-Dimerng/mLNot ClaimedNot Claimed40013003900
Digitoxinng/mL3.98.113.529.062
Homocysteineμmol/LNot ClaimedNot Claimed6.512.031.0
hsCRPmg/L0.200.803.206.513.0
Myeloperoxidase(MPO)pmol/L204633010404550
Myoglobinng/mL11.029.057310750
NT-proBNPpg/mL45125620310015600
Procalcitoninng/mL0.180.320.653.9015.6
β-HCGmIU/mLNot ClaimedNot Claimed12.035270
Total CKIU/L4278169400780
Troponin I (Tnl)ng/mL0.0250.0800.451.5020.0
Troponin T (TnT)pg/mL10302007507800

:

Omni Cardio – Target Ranges

{3}------------------------------------------------

Comparison of Technological Characteristics

ComparisonProposed DevicePredicate1Predicate 2
Proprietary NameThermo Scientific MAS®Omni•CARDIOMAS® Cardiolmmune TLMAS® Cardiolmmune XL
510k NumberK122291K040880K061196
Intended UseThermo Scientific MAS®Omni•CARDIO™ is intendedfor use in the clinicallaboratory as an assayedcontrol serum for monitoringassay conditions related tocardiac and associatedcritical markerdeterminations.Cardiolmmune® TL isintended for use in theclinical laboratory as anassayed control serumsuitable for monitoring assayconditions in specific cardiacmarker determinations.MAS® Cardiolmmune® XL isintended for use in theclinical laboratory as anassayed control serum formonitoring assay conditionsin specific cardiac markerdeterminations.
MatrixHuman SerumHuman SerumHuman Serum
FormatFrozen LiquidFrozen LiquidFrozen Liquid
Control Levels544
Storage-20°C-20°C-20°C
Shelf Life36 months36 months36 months
Open Vial Stability15 days (2-8°C) exceptNT-ProBNP, Myoglobin (10days)Homocysteine (5 days)30 days (2-8°C) exceptNT-ProBNP (10 days)30 days (2-8°C) exceptCK-MB, Troponin-T, BNP32, NT-ProBNP (15 days)Myoglobin (5 days)
Closed Vial Stability15 days (2-8°C) exceptProBNP, Myoglobin (10days)Homocysteine (5 days)30 days (2-8°C)N/A
AnalytesBNP 32CK-MBD-DimerDigitoxinHomocysteinehsCRPMyeloperoxidase (MPO)MyoglobinNT-ProBNPProcalcitonin (PCT)bHCGTotal CKTroponin-I (Tnl)Troponin-T (TnT)CK-MBDigitoxinhsCRPMyoglobinNT-ProBNPTroponin-l (Tnl)Troponin-T (TnT)BNP 32CK-MBDigitoxinHomocysteinehsCRPMyoglobinNT-ProBNPTroponin-I (Tnl)Troponin-T (TnT)

Conclusion

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the seal is an abstract image of an eagle or bird-like figure.

10903 New Hampshire Avenue Silver Spring, MD 20993

Microgenics Corporation Thermo Fisher Scientific, Clinical Diagnostics Division c/o Karen Lee 46360 Fremont Blvd. Fremont, CA 94538

AUG 3 1 2012

Re: K122291

Trade Name: MAS® Omni·CARDIO Regulation Number: 21 CFR 8862.1660 Regulation Name: Ouality Control Material (assayed and unassayed) Regulatory Class: Class I, reserved Product Codes: JJY Dated: July 30, 2012 Received: July 31, 2012

Dear Ms. Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 PICER Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...

You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm

Sincerely vours,

Courney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use Statement

510(k) Number (if known) K122291

Device Name MAS® Omni · CARDIO

Indications For Use MAS® Omni•CARDIO

Thermo Scientific MAS® OmnivCARDIO™ is intended for use in the clinical laboratory as an assayed control serum for monitoring assay conditions related to cardiac and associated critical marker determinations.

Prescription Use

X

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

K12229 510(k)

1 Page 1 of

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.