(140 days)
The Eni-Eye SH Sphere (Efrofilcon A) Soft (hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons.
Eni-Eye SH Sphere (Efrofilcon A) Soft (hydrophilic) Contact Lens for Daily Wear
This document is a 510(k) clearance letter from the FDA for a contact lens. It does not contain information about acceptance criteria or performance studies as typically requested for AI/ML device evaluations. Therefore, I cannot provide the requested information based on the given text.
The document states that the FDA reviewed a premarket notification for the "Eni-Eye SH Sphere (Efrofilcon A) Soft (hydrophilic) Contact Lens for Daily Wear" and determined it to be substantially equivalent to legally marketed predicate devices. It lists the indications for use as the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons for daily wear.
However, none of the specific details about acceptance criteria, study design, sample sizes, ground truth establishment, or expert involvement for AI/ML performance testing are present in this regulatory clearance letter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
December 12, 2012
Soflex Limited c/o Kevin Walls, RAC Principal Consultant 33 Golden Eagle Lane Littleton, Colorado 80127
Re: K122213
Trade/Device Name: Eni-Eye SH Sphere (Efrofilcon A) Soft (hydrophilic) Contact Lens for Daily Wear
Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: II Product Code: LPL Dated: November 6, 2012 Received: November 7, 2012
Dear Mr. Walls:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Kevin Walls
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known): _
Device Name: Eni-Eye SH Sphere (Efrofilcon A) Soft (hydrophilic) Contact Lens for Daily Wear
Indications for Use: The Eni-Eye SH Sphere (Efrofilcon A) Soft (hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Neurological and Physical Medicine Devices K 12221 510(k) Number
N/A