K Number
K122164

Validate with FDA (Live)

Date Cleared
2012-09-07

(49 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MR VessellQ Xpress is intended to provide an optimized non-invasive application to facilitate vascular anatomy and pathology analysis from a set of DICOM 3.0 compliant 3D contrastenhanced Magnetic Resonance Angiographic (MRA) images.

MR VessellQ Xpress is a post processing application which can be used in the analysis of MRA data for the purpose of vascular disease assessment.

This software is designed to assist radiologists and other clinicians in the evaluation and assessment of vascular anatomy and disease with the capability to provide a set of tools for visualizing directional vessel tortuosity, for sizing the vessel and for measuring areas of abnormalities within a vessel.

Device Description

MR VessellQ Xpress is a post processing analysis software application designed to assist Radiologists, Cardiologists, and other clinicians in the evaluation and assessment of vascular anatomy.

MR VessellQ Xpress is a software package for the Advantage Workstation (AW) platform and AW Server platform. The MR VessellQ Xpress is an additional tool for the 2D and 3D analysis of DICOM compliant MR angiographic images/data , providing a number of display, measurements and batch filming/archive features to study user-selected vessels which include but are not limited to stenosis analysis and directional vessel tortuosity visualization.

AI/ML Overview

The GE Healthcare MR VessellQ Xpress device, a post-processing analysis software for evaluating vascular anatomy and pathology from MRA images, did not require clinical studies to support substantial equivalence. The submission states that the device was found to be as safe and effective as, and substantially equivalent to, its predicate devices (K041521 Volume Viewer Plus by GE Medical Systems and K040746 MRA-CMS by Medis Medical Imaging Systems) based on non-clinical tests and compliance with recognized standards.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Compliance with NEMA PS 3.1 - 3.18(2008) Digital Imaging and Communications in Medicine (DICOM) SetMR VessellQ Xpress and its components comply with this standard.
Compliance with ISO 13485 Quality Systems - Model for Quality Assurance in design, development, production, installation and servicing of medical deviceMR VessellQ Xpress was designed in compliance with this standard.
Compliance with ISO 14971 Medical devices - Application of Risk management to medical devicesMR VessellQ Xpress was designed in compliance with this standard.
Compliance with IEC 62304 Medical device software - Software life cycle processesMR VessellQ Xpress was designed in compliance with this standard.
Compliance with IEC 62366 - Medical devices - Application of usability engineering to medical devicesMR VessellQ Xpress was designed in compliance with this standard.
Application of Quality Assurance Measures (Risk Analysis, Requirements Reviews, Design Reviews, Unit level testing, Integration testing, Final acceptance testing, Performance testing, Safety testing)These quality assurance measures were applied to the development of the device.
No new potential safety risksThe device does not result in any new potential safety risks.
Performs as well as predicate devicesThe device performs as well as the predicate devices currently on the market.

2. Sample size used for the test set and the data provenance

The submission explicitly states: "MR VessellQ Xpress did not require clinical studies to support substantial equivalence." Therefore, no test set of patient data was used for a clinical study. The evaluation was based on non-clinical tests and compliance with standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

As no clinical studies were performed, no experts were involved in establishing ground truth for a test set of patient data.

4. Adjudication method for the test set

Not applicable, as no clinical studies with a test set were conducted.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was performed.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

The submission does not detail any standalone performance studies using clinical data. The performance claims are based on comparison to predicate devices and adherence to design/quality standards.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Not applicable in the context of a clinical test set. The ground truth for proving performance was based on adherence to engineering and quality standards, and comparison of technical specifications and intended use with predicate devices.

8. The sample size for the training set

Not applicable, as no information regarding a training set for a machine learning model is provided. The submission describes a software application that employs "the same fundamental scientific technology as its predicate devices," suggesting the device's functionality is based on established algorithms for image analysis rather than a newly trained AI model requiring a separate training set.

9. How the ground truth for the training set was established

Not applicable, as no training set is mentioned or implied for a machine learning model.

{0}------------------------------------------------

K122164
p. 1 of 3

SEP 7 2012

GE Healthcare 510(k) Premarket Notification Submission

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

July 19th, 2012 Date:

Submitter: GE Medical Systems SCS 283, Rue de la Minière 78530 Buc. France Helen Peng, RAC Primary Contact Person: Regulatory Affairs Leader GE Healthcare 3000 North Grandview Blvd. W1140 Waukesha, WI 53188, USA Tel: 262-548-5091 Fax: 262-364-2506 Email: hong.peng@ge.com

Secondary Contact Person:

Stephen Slavens , RAC Regulatory Affairs Director GE Healthcare 3000 North Grandview Blvd, W1140 Waukesha, WI 53188, USA Tel: 262-548-4992

Fax: 262-364-2506

MR VessellQ Xpress

Email: Stephen.slavens@ge.com

Class II per 21 CFR 892.2050

Ali

Advanced Vessel Analysis MR (AVA MR)

Picture, Archiving and Communication System

Device: Trade Name: Common/Usual Name: Classification Names:

Product Code:

LLZ

Predicate Device(s):

K041521 Volume Viewer Plus by GE Medical Systems

K040746 MRA-CMS by Medis Medical Imaging Systems

Device Description:

MR VessellQ Xpress is a post processing analysis software application designed to assist Radiologists, Cardiologists, and other clinicians in the evaluation and assessment of vascular anatomy.

MR VessellQ Xpress is a software package for the Advantage Workstation (AW) platform and AW Server platform. The MR VessellQ Xpress is an additional tool for the 2D and 3D analysis

GE logo

{1}------------------------------------------------

K122164
p. 2 of 3

GE Healthcare 510(k) Premarket Notification Submission

of DICOM compliant MR angiographic images/data , providing a number of display, measurements and batch filming/archive features to study user-selected vessels which include but are not limited to stenosis analysis and directional vessel tortuosity visualization.

MR VessellQ Xpress is intended to provide an optimized non-Intended Use: invasive application to facilitate vascular anatomy and pathology analysis and vascular disease assessment from a set of DICOM 3.0 compliant Magnetic Resonance Angiographic (MRA) images.

MR VessellQ Xpress is intended to provide an optimized non-Indication for Use invasive application to facilitate vascular anatomy and pathology analysis from a set of DICOM 3.0 compliant 3D contrast-enhanced Magnetic Resonance Angiographic (MRA) images.

MR VessellQ Xpress is a post processing application which can be used in the analysis of MRA data for the purpose of vascular disease assessment.

This software is designed to assist radiologists and other clinicians in the evaluation and assessment of vascular anatomy and disease with the capability to provide a set of tools for visualizing directional vessel tortuosity, for sizing the vessel and for measuring areas of abnormalities within a vessel.

The MR VesselIQ Xpress employs the same fundamental Technology: scientific technology as its predicate devices.

Image /page/1/Picture/9 description: The image shows a blurry, black and white drawing. It is difficult to discern the exact subject of the drawing due to the low resolution and lack of detail. The image appears to be a rough sketch or a low-quality reproduction of a more detailed image.

{2}------------------------------------------------

K122164
p. 3 of 3

GE Healthcare

510(k) Premarket Notification Submission

Determination of Substantial Equivalence:

The MR VessellQ Xpress and its components comply with the following voluntary standards :

NEMA PS 3.1 - 3.18(2008) Digital Imaging and Communications in Medicine (DICOM) Set.

MR VessellQ Xpress was designed in compliance with the following Process Standards:

  • ISO 13485 Quality Systems-Model for Quality . assurance in design, development, production, installation and servicing of medical device
  • ISO 14971 Medical devices-Application of Risk . management to medical devices
  • IEC 62304 Medical device software Software life . cycle processes
  • IEC 62366. -Medical devices Application of usability . engineering to medical devices
  • .

The following quality assurance measures were applied to the development of the device:

  • . Risk Analysis
    Summary of Non-Clinical Tests:

  • Requirements Reviews .

  • Design Reviews .

  • . Unit level testing (Module Verification)

  • Integration testing (System verification) �

  • Final acceptance testing (Validation)

  • Performance testing (Verification)

  • Safety testing (Verification)

Summary of Clinical Tests:

The subject of this premarket submission, MR VessellQ Xpress did not require clinical studies to support substantial equivalence.

MR Vessell() Xpress does not result in any new potential safety Conclusion: risks and performs as well as the predicate devices currently on the market. GE Medical Systems SCS considers the MR VessellQ Xpress to be as safe, as effective as, and is substantially equivalent to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a representation of human figures.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

GE Medical Systems SCS % Ms. Helen Peng Regulatory Affairs Leader GE Medical Systems, LLC dba GE Healthcare 3000 N. Grandview, W1140 WAUKESHA WI 53188

SEP 7 2012

Re: K122164

Trade/Device Name: MR VesselIQ Xpress Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ and LNH Dated: July 19, 2012 Received: July 20, 2012

Dear Ms. Peng:

We have reviewed your Section 510(k) premarket notification of intent to market the indication we nave reviewed your becamined the device is substantially equivalent (for the indications and for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosure) to regars annual date of the Medical Device American Continuents, or to commerce provision in May 20, 1976, the encordance with the provisions of the Federal Food. Drug, uevices that have been recuire approval of a premarket approval of a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket You may, therefore, market the device, subject to the general controls provisions of the Act. The Act. The Y ou may, therefore, market the devices, ediferents for annual registration, listing of general controls provisions of the rice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such II your device is Classified (see above) into enas ffecting your device can be found in Title 21, additional controls. Existing mayor regards a 895. In addition, FDA may publish further Ode of Pederal concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I DA s issuation with the router with other requirements of the Act
that FDA has made a determination that your device complies . You must that FDA lias made a determination and regulations administered by other Federal station and light or ally Federal statutes and regulations and limited to: registration and listing (21 Compy with all the Act 3 requirements, arrants 801 and 809); medical device reporting of

{4}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice medical device-lefated adverse events) (21 CFR Pat 3-60 ), and 820). This letter requirents as set form in the quality Systems (QS) reged in your Section 510(k) prematket will allow you to ocgin marketing your actively walence of your device to a legally marketed notification. The I DFI imaing of successification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and II you desire specific advice for your device of the career Evaluation and Safety at (301) 796-607), please contact the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to under the MDK regulation (27 ℃FR Part 005), producem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Tou may other other general meeting and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a circle. There are small swirls or flourishes around the letters and the outer edge of the circle, giving it a classic and recognizable design.

GE Healthcare 510(k) Premarket Notification Submission

K122164 510(k) Number :

MR VessellQ Xpress Device Name:

Indications for Use:

MR VessellQ Xpress is intended to provide an optimized non-invasive application to facilitate vascular anatomy and pathology analysis from a set of DICOM 3.0 compliant 3D contrastenhanced Magnetic Resonance Angiographic (MRA) images.

MR VessellQ Xpress is a post processing application which can be used in the analysis of MRA data for the purpose of vascular disease assessment. ·

This software is designed to assist radiologists and other clinicians in the evaluation and assessment of vascular anatomy and disease with the capability to provide a set of tools for visualizing directional vessel tortuosity, for sizing the vessel and for measuring areas of abnormalities within a vessel.

Prescription Use_X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

(Division Sign-Off)

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) Number K122164

N/A