K Number
K122146

Validate with FDA (Live)

Date Cleared
2013-04-22

(277 days)

Product Code
Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Spacelabs Healthcare Xhibit Central Station, Model 96102 intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters. Data includes physiological waveforms and calculations, cardiac arrhythmia and ST data, and patient demographic information to monitor adequacy of treatment or to exclude causes of symptoms. The Spacelabs Healthcare Xhibit Central Station, Model 96102 is a prescription device intended for use under the direct supervision of a licensed healthcare professional.

Device Description

The Spacelabs Healthcare (Spacelabs) Xhibit Central Station, Model 96102 (Xhibit), is a new version of currently marketed Spacelabs product. Xhibit offers remote surveillance of patient data for those patients connected to a Spacelabs Healthcare Ultraview, Ultraview SL bedside monitor, or telemetry system. Xhibit provides for data communication using the TCP/IP network protocol employed in the Spacelabs Patient Care Management network of hardwired and/or telemetry monitored patients. Xhibit is not the primary alarming device for the Ultraview or Ultraview SL telemetry system.

AI/ML Overview

The provided 510(k) summary (K122146) details the Spacelabs Healthcare Xhibit Central Station, Model 96102. The document describes various performance tests conducted to establish its safety and effectiveness.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based solely on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not explicitly present a table of quantitative acceptance criteria with corresponding performance metrics for the device's clinical efficacy (e.g., arrhythmia detection accuracy). Instead, the performance testing focuses on compliance with established standards and internal requirements.

Test TypeAcceptance Criteria (Implied)Reported Device Performance
Software TestingCompliance with regulatory guidance and IEC 62304:2006"Test results indicate that the Xhibit complies with its predetermined specifications and the applicable standards and guidance documents."
Electromagnetic Compatibility TestingCompliance with IEC 60601-1-2:2007"Test results indicated that Xhibit complies with the Standard."
Performance TestingCompliance with IEC 60601-1-8:2006 and IEC 62366:2007, and internal requirements"Test results indicate that Xhibit complies with its predetermined specifications and the applicable standards."

2. Sample size used for the test set and the data provenance

The document does not provide information regarding a specific "test set" in the context of clinical performance data, such as a dataset of patient physiological waveforms for evaluating alarm accuracy or arrhythmia detection. The testing described is primarily focused on engineering and software validation. Therefore, there is no information on sample size or data provenance (country of origin, retrospective/prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided. Since there's no mention of a clinical test set requiring expert ground truthing, this detail is absent.

4. Adjudication method for the test set

This information is not provided. As no clinical test set requiring ground truth is mentioned, an adjudication method is not applicable in the documented testing.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with AI assistance versus without AI assistance was not described or referenced in the provided text.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device is a "Central Station Remote Monitor" that displays physiological data and alarms. Its function is to facilitate human monitoring, not to operate as a standalone diagnostic algorithm. Therefore, a standalone algorithmic performance study, as typically understood for AI-based diagnostic devices, was not reported. The performance testing focuses on the system's compliance with safety and operational standards.

7. The type of ground truth used

For the engineering and software testing described, the "ground truth" would be the specifications and requirements defined by the manufacturer and the relevant national and international standards (e.g., IEC 62304, IEC 60601-1-2, IEC 60601-1-8, IEC 62366). There's no mention of clinical ground truth derived from expert consensus, pathology, or outcomes data for assessing diagnostic accuracy.

8. The sample size for the training set

This information is not provided. The document describes a traditional medical device submission, not one for an AI/ML-based device that would typically involve a separate training set. The "software development process" mentioned refers to conventional software engineering, not machine learning model training.

9. How the ground truth for the training set was established

This information is not provided, as no training set for an AI/ML model is mentioned.

Summary of Device and Study Focus:

The K122146 submission for the Spacelabs Healthcare Xhibit Central Station, Model 96102, primarily focuses on demonstrating substantial equivalence to a predicate device (Philips Medical Systems Intellivue M3290A). The studies cited are primarily engineering validation and verification tests ensuring compliance with established medical device safety, software development, electromagnetic compatibility, and alarm system standards. The document does not present a clinical performance study with defined acceptance criteria for diagnostic accuracy, sensitivity, or specificity, nor does it involve clinical ground truth, expert readers, or AI/ML components requiring training and test sets in the context of this summary. The "performance testing" referenced is about the system's operational adherence to standards, not a clinical trial of its diagnostic capabilities.

{0}------------------------------------------------

510(k) Summary

Submission Date:

Submitter Contact:

08 April 2013

Submitter:

Spacelabs Healthcare 5150 220th Avenue SE Issaquah, WA 98029

Spacelabs Healthcare 5150 220th Avenue SE Issaquah, WA 98029

Mr. David J. Geraghty Spacelabs Healthcare Phone: +1 (425) 657-7200, ext 5889 Fax: +1 (425) 657-7210 Email: david.geraghty@spacelabs.com

Application Correspondent: Thomas Kroenke Principal Consultant Speed To Market, Inc. PO Box 3018 Nederland, CO 80466 USA tkroenke@speedtomarket.net 303 956 4232

Manufacturing Site:

Spacelabs Healthcare 5150 220th Avenue SE Issaquah, WA 98029

Trade Name:

Central Station Remote Monitor Common Name:

Classification Name:

Alarms); 21 CFR §870.1025

Spacelabs Healthcare Xhibit Central Station, Model 96102

Monitor, Physiological, Patient (With Arrhythmia Detection or

Primary Classification Regulation:

Primary Product Code:

MHX

21 CFR §870.1025; 21 CFR §870.2300 : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :

: '

DSI; MLD; MSX

Secondary Classification Regulation:

Secondary Product Code:

· Page I of 4

APR 2 2 2013

{1}------------------------------------------------

510(k) Summarv

K122146

Information Center

SubstantiallyNew Spacelabs ModelPredicate510(k) NumberPredicateManufacturer / Model
Equivalent Devices:
Spacelabs HealthcareXhibit Central Station,Model 96102K050742Philips Medical SystemsIntellivue M3290A(Release G), M3155Configuration, Intellivue

Device Description:

The Spacelabs Healthcare (Spacelabs) Xhibit Central Station, Model 96102 (Xhibit), is a new version of currently marketed Spacelabs product. Xhibit offers remote surveillance of patient data for those patients connected to a Spacelabs Healthcare Ultraview, Ultraview SL bedside monitor, or telemetry system. Xhibit provides for data communication using the TCP/IP network protocol employed in the Spacelabs Patient Care Management network of hardwired and/or telemetry monitored patients.

Xhibit is not the primary alarming device for the Ultraview or Ultraview SL telemetry system.

The Spacelabs Healthcare Xhibit Central Station, Model 96102 intended use is to provide clinicians with central monitoring of adult. pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters. Data includes physiological waveforms and calculations, cardiac arrhythmia and ST data, and patient demographic information to monitor adequacy of treatment or to exclude causes of symptoms.

The Spacelabs Healthcare Xhibit Central Station, Model 96102 is a prescription device intended for use under the direct supervision of a licensed healthcare professional.

Xhibit employs the same technological characteristics as the predicate device.

CharacteristicPredicate DeviceProposed Device
Number of Patients per Display16Same
Patients per Central StationUp to 16Up to 48
Traces per DisplayUp to 32Same
Dedicated Display for DataReviewYesSame
Prioritized AlarmsYesSame

Technology Comparison:

Intended Use:

Page 2 of 4

{2}------------------------------------------------

510(k) Summary

Summary of Performance Testing:

.

Software TestingXhibit contains MAJOR level of concern software. Software wasdesigned and developed according to a robust software developmentprocess, and was rigorously verified and validated. Softwareinformation is provided in accordance with internal requirements andthe following standards and guidance documents:FDA guidance: The content of premarket submissions for software contained in medical devices, 11 May 05; FDA guidance: Off-the-shelf software use in medical devices, 09 Sep 99; FDA guidance: General principles of software validation; Final guidance for industry and FDA staff, 11 Jan 02; FDA guidance: Cybersecurity for networked medical devices containing off-the-shelf (OTS) software, 14 January 2005; and IEC 62304: 2006, Medical device software - Software life cycle processes. Test results indicate that the Xhibit complies with its predetermined specifications and the applicable standards and guidance documents.
ElectromagneticCompatibilityTestingXhibit hardware was tested for performance in accordance with the following Standard:IEC 60601-1-2: 2007, Medical electrical equipment – Part 1-2:General requirements for basic safety and essential performance -Collateral standard: Electromagnetic compatibility - Requirements and tests. Test results indicated that Xhibit complies with the Standard.
Performance TestingXhibit was tested for performance in accordance with internal requirements and the following standards:IEC 60601-1-8: 2006, Medical electrical equipment - General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems. IEC 62366: 2007, Medical devices - Application of usability engineering to medical devices. Test results indicate that Xhibit complies with its predetermined specifications and the applicable standards.

.

:

{3}------------------------------------------------

510(k) Summary

. .

. ···

,

.

Page 4 of 4

Conclusion

Verification and validation activities were conducted to establish the performance and safety characteristics of Xhibit. The results of these activities demonstrate that Xhibit is safe and effective when used in accordance with its intended use and labeling.

. .

Therefore, Xhibit is considered substantially equivalent to the predicate device. ・

..

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/4/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, with three curved lines representing its body or wings.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

April 22, 2013

Spacelabs Healthcare c/o Mr. Thomas Kroenke P.O. Box 3018 Nederland, CO 80466 US

Re: K122146 Xhibit Central Station, model 96102 Regulation Number: 21 CFR 870.1025 Regulation Name: Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms) Regulatory Class: Class II Product Code: MHX, DSI, MLD, MSX Dated: April 8, 2013 Received: July 19, 2012

Dear Mr. Kroenke:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Thomas Kroenke

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Owen-PFaris-S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known):K K122146
Device Name:Spacelabs Healthcare Xhibit Central Station, Model 96102
Indications for Use:The Spacelabs Healthcare Xhibit Central Station, Model 96102 intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters. Data includes physiological waveforms and calculations, cardiac arrhythmia and ST data, and patient demographic information to monitor adequacy of treatment or to exclude causes of symptoms. The Spacelabs Healthcare Xhibit Central Station, Model 96102 is a prescription device intended for use under the direct supervision of a licensed healthcare professional.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

. . . . . . . . . . . . . . . . . . . . . .

P. Faris -S .22 3 -04'00'

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.