K-Y BRAND LIQUIBEADS

K122061 · Johnson & Johnson Healthcare Products · NUC · Nov 8, 2013 · Obstetrics/Gynecology

Device Facts

Record IDK122061
Device NameK-Y BRAND LIQUIBEADS
ApplicantJohnson & Johnson Healthcare Products
Product CodeNUC · Obstetrics/Gynecology
Decision DateNov 8, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5300
Device ClassClass 2

Intended Use

K-Y® Brand Liquibeads® Vaginal Moisturizer is a personal lubricant for over-the-counter use, for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene, and polyurethane condoms. The product is delivered by an applicator.

Device Story

K-Y Brand Liquibeads Vaginal Moisturizer is a non-sterile, anhydrous silicone-based personal lubricant. The lubricant is contained within a gelatin-based ovule shell (gelatin NF, glycerin USP, water USP). The device is administered intravaginally using a single-use, disposable applicator. The product is intended for over-the-counter use to provide vaginal moisturization and lubrication, enhancing comfort during intimate sexual activity. It is compatible with natural rubber latex, polyisoprene, and polyurethane condoms. The device functions by delivering the silicone fill mass to the vaginal environment, where it supplements natural lubrication.

Clinical Evidence

Clinical evidence includes a human repeat insult patch test (HRIPT) demonstrating no contact sensitization. An in-home consumer use study confirmed the device provides vaginal moisturization. Non-clinical testing included biocompatibility studies (in vitro cytotoxicity, systemic toxicity in mice, contact sensitization in guinea pigs, and vaginal irritation in rabbits) conducted per ISO 10993 and GLP standards, showing the device is non-cytotoxic, non-sensitizing, and non-irritating. Condom compatibility testing confirmed compatibility with natural rubber latex, polyisoprene, and polyurethane condoms.

Technological Characteristics

Anhydrous silicone fill mass encapsulated in a gelatin shell (gelatin NF, glycerin USP, water USP). Delivered via single-use disposable applicator. Tested for biocompatibility per ISO 10993. Shelf-life stability testing performed.

Indications for Use

Indicated for over-the-counter use by individuals requiring vaginal moisturization and lubrication to enhance comfort during intimate sexual activity and supplement natural lubrication. Compatible with natural rubber latex, polyisoprene, and polyurethane condoms.

Regulatory Classification

Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K - Y ® Brand Liquibeads® - 5 l0(k) K 122061: Response to Second AI Request - May 21, 2013 Johnson+Johnson CONSUMER COMPANIES, Inc. #### R COMPANIES, Inc ### 510(k) SUMMARY 2013 | Date Prepared | October 11, 2013 | |---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Applicant | Johnson & Johnson Healthcare Products<br>Division of McNeil-PPC, Inc.<br>199 Grandview Road<br>Skillman, NJ 08558 | | Contact Person | Joseph Chmielewski, R.A.C.<br>Associate Director, Regulatory Affairs<br>Johnson & Johnson Consumer Companies, Inc<br>199 Grandview Road<br>Skillman, NJ 08558 | | | Phone: 908-874-1744<br>Fax: 908-904-3712<br>E-mail: jchmiele@its.jnj.com | | Proprietary Name | K-Y <sup>®</sup> Brand Liquibeads <sup>®</sup> | | Common Name | Vaginal Moisturizer / Personal Lubricant | | Classification Name | Condom | | Classification | Class II | | Product Code | NUC | | Regulation | 21 CFR §884.5300 | | Predicate Device | K-Y <sup>®</sup> Brand Intrigue <sup>®</sup> - K062796 | | Description | K-Y <sup>®</sup> Brand Liquibeads <sup>®</sup> Vaginal Moisturizer is a non-sterile, anhydrous silicone fill mass, encapsulated with a gelatin shell (ovule). The fill mass is composed of a proprietary blend of silicones. The gelatin shell is composed of gelatin NF, glycerin USP, and water USP. The K-Y <sup>®</sup> Brand Liquibeads <sup>®</sup> Vaginal Moisturizer is a transparent ovule with a yellowish fill mass. The product is packaged into single. | 185 Tabor Road, Morris Plains, NJ 07950 {1}------------------------------------------------ : K122061 pg 2 of 3 : | | use blister packs with single use disposable applicators, which are then<br>packed into a printed carton. | |--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | K-Y® Brand Liquibeads® Vaginal Moisturizer is a personal lubricant<br>for over-the-counter use. for vaginal application, intended to<br>moisturize and lubricate, to enhance the ease and comfort of intimate<br>sexual activity and supplement the body's natural lubrication. This<br>product is compatible with natural rubber latex. polyisoprene, and<br>polyurethane condoms. The product is delivered by an applicator. | | Technological<br>Characteristics | The subject device. K-Y® Brand Liquibeads® Vaginal Moisturizer, and<br>the predicate device. K-Y® Brand Intrigue® have the same intended<br>use but have different technological characteristics as the subject<br>device encapsulates the personal lubricant in an ovule. | | Biocompatibility<br>Review & Testing | The subject device varies from the cleared device only in the fact that<br>it is encapsulated in a gelatin shell and is inserted into the vagina by<br>means of a single-use, disposable applicator. At the concentrations<br>present in the formula, all ingredients are considered safe for human<br>use under the anticipated exposure conditions. In addition, all of the<br>ingredients have been used in legally marketed devices. The use of<br>these ingredients in the subject device raises no new questions of<br>safety or effectiveness. | | | Studies of vaginal and penile irritation in rabbits were conducted using<br>the fill mass of the subject device. The results showed that the fill<br>mass is non-irritating to genital tissue. | | | Four additional biocompatibility studies were conducted on the subject<br>device. The test article evaluated in each study consisted of a mixture<br>of the Liquibeads® ovule and 0.9% sodium chloride. The endpoints<br>evaluated included in vitro cytotoxicity. systemic toxicity in mice,<br>contact sensitization in guinea pigs, and vaginal irritation in rabbits.<br>Each study was conducted in accordance with GLP requirements and<br>the applicable ISO 10093 Standard. From the results obtained, it can<br>be concluded that the subject device, K-Y® Brand Liquibeads®<br>Vaginal Moisturizer, is not cytotoxic, does not elicit systemic toxicity.<br>is not a contact sensitizer, and is not irritating to vaginal tissue. | | Performance Data<br>-Clinical | A human repeat insult patch test was conducted to evaluate the safety<br>of the fill mass of subject device. The study showed the product did<br>not induce responses indicative of contact sensitization. | | | Results from an in-home consumer use test show that K-Y® Brand<br>Liquibeads® Vaginal Moisturizer provides vaginal moisturization. | | Performance Data<br>-Non Clinical | Condom compatibility testing demonstrates that K-Y® Brand<br>Liquibeads® Vaginal Moisturizer is compatible with natural rubber<br>latex, polyisoprene, and polyurethane condoms. | | | Routine testing demonstrated that the product will remain stable for<br>the labeled shelf-life. | | Conclusion | The subject device, K-Y® Brand Liquibeads® Vaginal Moisturizer is<br>substantially equivalent to the predicate device, K-Y® Brand Intrigue®. | . . {2}------------------------------------------------ . K122061 pg 3 of 3 : . . . {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 8, 2013 Johnson & Johnson Healthcare Products Lorna-Jane Bremer, M.S., M.B.A., R.A.C. Director, Regulatory Affairs Johnson & Johnson Consumer Companies, Inc. 185 Tabor Road Morris Plains NJ 07950 Re: K122061 > Trade/Device Name: K-Y Brand Liquibeads Vaginal Moisturizer Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Procode: NUC Dated: November 4, 2013 Received: November 5, 2013 Dear Lorna-Jane Bremer, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {4}------------------------------------------------ Page 2 - Lorna-Jane Bremer, M.S., M.B.A., R.A.C. CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### INDICATIONS FOR USE STATEMENT 510(k) Number (if known): K122061 Device Name: K-Y® Brand Liquibeads® Vaginal Moisturizer Indications for Use: K-Y® Brand Liquibeads® Vaginal Moisturizer is a personal lubricant for over-the-counter use, for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene, and polyurethane condoms. The product is delivered by an applicator. #### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ## Benjamin R. Fisher -S 2013.11.08 16:20:48 -05'00' X AND/OR Over-the-Counter Use Prescription Use (21 CFR §801 Subpart C) (Per 21 CFR §801 Subpart D)
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%