K Number
K101098

Validate with FDA (Live)

Date Cleared
2010-08-17

(119 days)

Product Code
Regulation Number
884.5300
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Replens is a personal lubricant for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex condoms and synthetic (polyurethane and polyisoprene) condoms.

Device Description

Replens Vaqinal Moisturizer is a non-sterile, water-based, white to off-white, non-irritating, non-greasy, non-staining vaginal gel delivered in a tube with a reusable applicator as a long-lasting moisturizer for vaginal dryness. The use of the reusable applicator provides less mess in application. Replens Vaginal Moisturizer is intended for ongoing use, not exclusively for use during intimate sexual activity.

Replens Vaginal Moisturizer contains ingredients commonly used in other products for vaginal use sold as medical devices and cosmetics. All ingredients are either NF, USP, or are considered "generally recognized as safe for their intended use". The quantitative formulation is considered confidential commercial information.

AI/ML Overview

The Replens Vaginal Moisturizer, a personal lubricant for vaginal application, underwent various tests and clinical studies to demonstrate its safety and effectiveness. The acceptance criteria and performance data are summarized below.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Test/ParameterAcceptance Criteria (Implicit)Reported Device Performance
BiocompatibilityCytotoxicityNo significant cytotoxic effects.Biocompatibility testing performed, implying acceptable results (specific results not detailed, but submission indicates compliance).
Acute Vaginal IrritationNo significant acute irritation.Biocompatibility testing performed, implying acceptable results.
Subacute Vaginal IrritationNo significant subacute irritation.Biocompatibility testing performed, implying acceptable results.
Subacute Vaginal Irritation with Histological ExaminationNo significant subacute irritation or adverse histological changes.Biocompatibility testing performed, implying acceptable results.
Acute Systemic ToxicityNo significant acute systemic toxicity.Biocompatibility testing performed, implying acceptable results.
HypersensitivityNo significant hypersensitivity reactions.Biocompatibility testing performed, implying acceptable results.
Acute Oral ToxicityNo significant acute oral toxicity.Biocompatibility testing performed, implying acceptable results.
Acute Dermal ToxicityNo significant acute dermal toxicity.Biocompatibility testing performed, implying acceptable results.
Dermal IrritationNo significant dermal irritation.Biocompatibility testing performed, implying acceptable results.
Eye IrritationNo significant eye irritation.Biocompatibility testing performed, implying acceptable results.
Clinical Efficacy & SafetyVaginal pHMaintain a healthy vaginal pH or normalize altered pH associated with vaginal dryness.Studies evaluated the influence of Replens on vaginal pH. One study specifically compared effects on vaginal pH in postmenopausal women. The collective conclusion was that Replens was "safe and well-tolerated."
Vaginal MucosaImprove the condition of the vaginal mucosa.Studies evaluated the influence of Replens on the vaginal mucosa. The collective conclusion was that Replens was "safe and well-tolerated."
Symptom ReliefProvide relief for symptoms of vaginal dryness.Clinical studies documented "relief of the patient's symptoms" and improvement in the "vaginal dryness index." The collective conclusion was that Replens was "safe and well-tolerated."
Safety & TolerabilityNo significant adverse events or poor tolerability.All clinical studies concluded that Replens was "safe and well-tolerated."
Product StabilityShelf LifeMaintain product integrity and effectiveness for the specified shelf life.Real-time stability data confirmed a shelf life of three (3) years.
Microbial ControlPreservative EffectivenessMeet USP <51> Antimicrobial Effectiveness Test requirements.Successfully passed the requirements of the USP <51> Antimicrobial Effectiveness Test.
Condom CompatibilityCondom Strength/IntegrityNot materially affect the strength or integrity of natural rubber latex condoms and synthetic (polyurethane and polyisoprene) condoms.Condom compatibility testing confirmed that Replens Long-Lasting Vaginal Moisturizer does not materially affect the strength or integrity of natural rubber latex or synthetic condoms (polyurethane and polyisoprene).

2. Sample Size Used for the Test Set and Data Provenance

The document refers to a series of eight clinical studies as the "test set" for Replens Long-Lasting Vaginal Moisturizer.

  • Sample Sizes: The number of patients enrolled in the Replens arm across these studies varied:
    • Bachmann et al (1991): 89 patients
    • Bachmann et al (1992): 54 patients
    • Zinny and Lee (1991): 26 patients
    • Young et al (1991): 30 patients
    • Nakamura (1991): 10 patients
    • Whitehead (1991): 32 patients
    • Nachtigall (1994): 15 patients
    • Gelfand and Wendman (1994): 25 patients
  • Data Provenance: The provenance of the data is not explicitly stated in terms of country of origin but is indicated by the publication of several studies in peer-reviewed journals (e.g., Clinical Practice in Sexuality, Today's Therapeutic Trends, Fertility and Sterility, J. Women's Health). Some studies were "Unpublished." The studies appear to be prospective clinical trials given their design (e.g., double-blind, cross-over, open-label) and specific dosage regimens and durations.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The concept of "ground truth" established by experts, as typically understood in the context of AI/ML device performance evaluation, is not directly applicable to this submission. This is a 510(k) for a medical device (vaginal moisturizer), not an AI/ML algorithm.

The "ground truth" for the clinical efficacy and safety was established:

  • Clinically by the direct observation, assessment (e.g., vaginal pH, vaginal dryness index), and reported symptoms of the patients participating in the clinical trials.
  • Statistically by the analysis of these clinical outcomes by the researchers who conducted and published these studies.
  • Biologically through the results of the various biocompatibility tests.

The "experts" involved would be the clinical investigators (physicians, researchers) who designed and executed these studies, interpreted the data, and published the findings. Their qualifications are not explicitly detailed but are implied by their authorship in scientific papers.

4. Adjudication Method for the Test Set

Adjudication methods (like 2+1, 3+1) are typically used in AI/ML studies to resolve discrepancies in expert labeling or diagnoses. This concept is not applicable to this 510(k) submission as it is not for an AI/ML device. The clinical studies followed standard trial protocols for data collection and analysis, which inherently include methods for managing and interpreting patient data and outcomes. For double-blind studies, the blinding helps ensure impartial assessment.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned. This type of study comparing human readers with and without AI assistance is specific to AI/ML devices and is not relevant to this product submission.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

No standalone (algorithm only) performance study was conducted or is relevant. This device is a topical moisturizer, not an AI algorithm.

7. Type of Ground Truth Used

The "ground truth" for this device's performance was based on:

  • Clinical Endpoints: Patient-reported symptoms (e.g., relief of vaginal dryness), objective clinical measurements (e.g., vaginal pH, vaginal dryness index), and physician assessments of the vaginal mucosa.
  • Biocompatibility Standards: Pre-defined safety standards for cytotoxicity, irritation, sensitization, and toxicity.
  • Physical/Chemical Testing: Laboratory assessments for stability, preservative effectiveness, and condom compatibility.

8. Sample Size for the Training Set

The concept of a "training set" is specific to AI/ML models. This device is a traditional medical product (vaginal moisturizer), not an AI/ML algorithm. Therefore, there is no training set in the AI/ML sense. The product development and testing relied on chemical formulation, manufacturing best practices, and the clinical and non-clinical studies summarized.

9. How the Ground Truth for the Training Set Was Established

As there is no training set for an AI/ML algorithm, this question is not applicable.

{0}------------------------------------------------

K101098

Replens Vaginal Moisturizer (35g Tube)

Idnr: 2.0 510k Summary Version: 2.2 Date: August 13, 2010 Page 1 of 4

510(k) Summary

Submitter:

AUG 1 7 2010

Lil' Drug Store Products, Inc. 1201 Continental Place NE Cedar Rapids, IA 52402

Contact Person:

Tricia Miller Director of Regulatory Telephone: 319-294-3745 Facsimile: 319-393-3494 Email: tmiller@lildrugstore.com

Date:

August 13, 2010

Proprietary Name:

Replens Long-Lasting Vaginal Moisturizer (in 35g Tube with Reusable Applicator)

Common name:

Personal Lubricant

Classification name:

21 C.F.R. 884.5300 Lubricant, Patient, Vaginal, Latex Compatible Product Code: NUC Class: 2 Obstetrics/Gynecology Review Panel:

Predicate Devices:

CVS Personal Lubricant & Moisturizer Device Name: 510(k) Number: K062682 Product Code: NUC, MMS

Intended Use:

Replens is a personal lubricant for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows a logo for "Li'l Drug Store Products". The words "Li'l Drug" are stacked on top of each other in a bold, sans-serif font. The word "Store" is below "Drug" in a smaller font. The word "Products" is at the bottom of the logo in a smaller, sans-serif font.

Idnr; 2.0 510k Summary Version: 2.2 Date: August 13, 2010 Page 2 of 4

with natural rubber latex condoms and synthetic (polyurethane and polyisoprene) condoms.

Description of Device

Replens Vaqinal Moisturizer is a non-sterile, water-based, white to off-white, non-irritating, non-greasy, non-staining vaginal gel delivered in a tube with a reusable applicator as a long-lasting moisturizer for vaginal dryness. The use of the reusable applicator provides less mess in application. Replens Vaginal Moisturizer is intended for ongoing use, not exclusively for use during intimate sexual activity.

Replens Vaginal Moisturizer contains ingredients commonly used in other products for vaginal use sold as medical devices and cosmetics. All ingredients are either NF, USP, or are considered "generally recognized as safe for their intended use". The quantitative formulation is considered confidential commercial information.

Technological Characteristics of the Device

Replens Vaginal Moisturizer is substantially equivalent to the identified previously cleared vaqinal lubricant predicate with respect to its design and materials, principle of operation, function, formulation, and intended use. It is also substantially equivalent to other water-based vaginal lubricants and personal lubricants being commercially marketed in the U.S.

Summary of Performance Data

Biocompatibility Testing: The following biocompatibility testing has been performed on Replens Long-Lasting Vaginal Moisturizer:

  • Cytotoxicity .
  • Acute Vaqinal Irritation .
  • Subacute Vaqinal Irritation .
  • Subacute Vaginal Irritation with Histological Examination .
  • Acute Systemic Toxicity .
  • Hypersensitivity .
  • Acute Oral Toxicity .
  • Acute Dermal Toxicity .
  • Dermal Irritation .
  • Eye Irritation. .

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the text "KI01098" at the top left corner. Below this text is an image of the logo for "Lil' Drug Store Products". The logo is in black and white and features a stylized font.

Idnr: 2.0 510k Summary Version: 2.2 Date: August 13, 2010 Page 3 of 4

Clinical Studies: The following clinical studies have been performed on Replens Long-Lasting Vaginal Moisturizer:

  • Bachmann GA, Notelovitz M, Gonzalez SJ, Thompson C, Morecraft BA. . (1991) Vaginal Dryness in Menopausal Women: Clinical Characteristics and Nonhormonal Treatment. Clinical Practice in Sexuality, 7(9): 1-8.
  • Bachmann GA, Notelovitz M, Kelly SJ, Owens A, Thompson C. (1992) ● Long Term Nonhormonal Treatment of Vaginal Dryness. Clinical Practice in Sexuality, 8(8/9): 3-8.
  • Zinny MA, Lee S. (1991) Double-Blind Study of the Comparative . Effects of Two Gels on Vaginal pH in Postmenopausal Women. Today's Therapeutic Trends, 8(4): 65-72.
  • Young R, Goldzieher J, Kaufman R. (1991) A Study of the Effects of . Col-1003 In Postmenopausal Women. Unpublished.
  • Nakamura R. (1991) Evaluation of Col-1003 in the treatment of . vaginal dryness in postmenopausal women. Unpublished.
  • Whitehead M. (1991) A Randomised Double Blind Evaluation of Col-. 1003, a bioadhesive polymer system vaginal moisturizing gel and, KY
  • Brand Lubricating Jelly in the treatment of vaginal dryness in postmenopausal women receiving concomitant oral hormone replacement therapy. Unpublished
  • Nachtigall LE. (1994) Comparative study: Replens versus local . estrogen in menopausal women. Fertility and Sterility, 61(1): 178-180.
  • Gelfand MM, Wendman E. (1994) Treating Vaginal Dryness in Breast . Cancer Patients: Results of Applying a Polycarbophil Moisturizing Gel. ]. Women's Health, 3(6): 427-433.

The format of the studies listed above is summarized in the table below:

CitationNo.PatientsEnrolled(Replens)StudyDesign*Dosage RegimenDuration
Bachmann et al89D-B, X-over2.5 g per day5 days
Bachmann et al54Open2.5 g, 3 x weekly12 months
Zinny and Lee26D-B, parallel2.5 g alternate nights4 weeks
Young et al30Open2.5 g, 3 x weekly, plusoption of additionalapplication prior tointercourse.12 months
Nakamura10Open, X-overbetweentreatmentdurations2.5 g daily1-5 days

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a logo for a business called "Li'l Drug Store Products". The words "Drug Store" are in a bold, sans-serif font, and the word "Li'l" is in a cursive font above the other words. There is a horizontal line above and below the words "Drug Store".

Idnr: 2.0 510k Summary Version: 2.2 Date: August 13, 2010 Page 4 of 4

CitationNo.PatientsEnrolled(Replens)StudyDesign*Dosage RegimenDuration
Whitehead32D-B, X-over2.5 g, 3 x weekly plusoption of additionalapplication prior tointercourse.8 weeks
Nachtigall15Open, parallel2.5 g, 3 x weekly3 months
Gelfand andWendman25Open2.5 g, 3 x weekly, plusoption of additionalapplication prior tointercourse.3 months
  • D-B = double-blind; X-over = cross-over design

The parameters employed in the studies included the influence of Replens on vaginal pH and the vaginal mucosa, the relief of the patient's symptoms, the vaginal dryness index, determination of vaginal pH at varying time intervals after single or multiple applications of the gel, PAP smears and the completion of diary cards by the patient. All of the studies concluded that Replens was safe and well-tolerated.

Stability Data: Real-time stability data confirms a shelf life of three (3) vears for Replens Long-Lasting Vaqinal Moisturizer.

Preservative Effectiveness: Replens Long-Lasting Vaginal Moisturizer has successfully passed the requirements of the USP <51> Antimicrobial Effectiveness Test.

Condom Compatibility Testing: Condom compatibility testing confirms that Replens Long-Lasting Vaginal Moisturizer does not materially affect the strength or integrity of natural rubber latex or synthetic condoms (polyurethane and polyisoprene).

Conclusion

Based on the information presented in the 510(k) notice, it is concluded that Replens Long-Lasting Vaginal Moisturizer (in 35g Tube with Reusable Applicator) for OTC use is safe and effective for its proposed indications and is substantially equivalent in intended use, formulation, safety, and technological characteristics to the identified predicate device and other similar water-based personal lubricants.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA", which is arranged in a circular pattern around the left side of the logo.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002.

Lil Drug Store Products, Inc. c/o Mr. Mark A. Job Responsible Third Party Regulatory Technology Services, Inc. 1394 25th Street, NW BUFFALO MN 55313

AUG 1 7 2010

·Re: K101098

Trade/Device Name: Replens Long-Lasting Vaginal Moisturizer (in 35 Tube with Reusable Applicator) Regulation Number: 21 CFR §884.5300 Regulation Name: Condom

Regulatory Class: II Product Code: NUC

Dated: July 30, 2010 Received: August 2, 2010

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act., The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies' You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{5}------------------------------------------------

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Duker Lemur MD

Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive. Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a logo for "Lil Drug Store Products". The words "Lil" are in a cursive font at the top of the logo. Below that, the words "Drug Store" are in a bold, sans-serif font. The word "Products" is at the bottom of the logo in a smaller, sans-serif font.

Idnr: 1.0 Indications Statement Version: 2.1 Date: May 12, 2010 Page 1 of 1

STATEMENT OF INDICATIONS FOR USE

K101098 510(k) Number:

Device Name: Replens Long-Lasting Vaginal Moisturizer (in 35g Tube with Reusable Applicator)

  • Indications for Use: Replens is a personal lubricant for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex condoms and synthetic (polyurethane and polyisoprene) condoms.
    Prescription Use (Per 21 CFR 801.109) OR

Over-the-Counter Use (Optional Format 1-2-96)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Shus L. Nolan.

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device 510(k) Number

§ 884.5300 Condom.

(a)
Identification. A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.(b)
Classification. (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.